Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
 
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Optical Aberrations and Contrast Sensitivity of Spherical and Aspheric Intraocular Lenses ? A Prospective Comparative Clinical Study

Nada Jirásková*, Juraj Urminský, Vera Lorencova, Alena Feuermannova, Alexandr Stepanov, and Pavel Rozsíval
Department of Ophthalmology, University Hospital, Hradec Králové, Czech Republic
Corresponding Author : Nada Jirásková, MD, PhD
Professor, Department of Ophthalmology
University Hospital
500 05 Hradec Králové, Czech Republic
Tel: 420-495-833-392
E-mail: jirasnad@fnhk.cz
Received: October 26, 2012; Accepted: November 09, 2012; Published: November 15, 2012
Citation: Jirásková N, Urminský J, Lorencova V, Feuermannova A, Stepanov A, et al. (2012) Optical Aberrations and Contrast Sensitivity of Spherical and Aspheric Intraocular Lenses – A Prospective Comparative Clinical Study. J Clin Exp Ophthalmol 3:254. doi:10.4172/2155-9570.1000254
Copyright: © 2012 Jirásková N, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Related article at
DownloadPubmed DownloadScholar Google

Visit for more related articles at Journal of Clinical & Experimental Ophthalmology

Abstract

Purpose: To compare visual acuity, aberrometry, and contrast sensitivity in eyes with spherical intraocular lens (IOL) and aspheric IOL implanted after uncomplicated phacoemulsification.

Setting: Department of Ophthalmology, University Hospital, Hradec Králové, Czech Republic.

Methods: A prospective clinical study involving 90 patients (105 eyes) having cataract surgery, with implantation of spherical IOL AcrySof Natural (SN60AT) in 45 patients (50 eyes) with a mean age of 62.5 years, and aspheric IOL AcrySof IQ (SN60WF) in 45 patients (55 eyes) with a mean age of 64.5 years. A control group involved 22 eyes of similar age with clear lens crystallina. Postoperatively, visual acuity (ETDRS charts), aberrometry (ORK Wavefront Analyser, Schwind), and contrast sensitivity tests (CSV 1000, Contrast Sensitivity 8010 System) were evaluated. Statistical analyses were performed using Student t-test.

Results: No statistical differences were found regarding visual acuity among eyes. A lower level of higher order aberrations was achieved in the aspheric group. In the contrast sensitivity test, no statistical differences between groups under photopic conditions were noted. In mesopic conditions better performance was observed in eyes with aspheric IOLs, mainly in low spatial frequencies (statistically significant). The control group (lens crystallina) was better in all frequencies.

Conclusions: Cataract surgery outcomes cannot be measured by means of visual acuity alone. In this study, the aspheric IOLs AcrySof IQ (SN60WF) demonstrated better visual function, especially contrast sensitivity in low spatial frequences, when compared with spherical IOLs AcrySof Natural (SN60AT).

Keywords
Higher order aberrations; Spherical lens; Aspheric lens
Introduction
Both ophthalmologists and patients currently approach cataract surgery as not only a “visual rescue operation” but also as a refractive procedure. Phacoemulsification and intraocular lens (IOL) implantation is considered the gold standard for cataract treatment. Topical anesthesia, small incisions, and sophisticated microsurgery techniques and technologies, including femtosecond lasers have led to better surgical results and better visual outcomes. However, for “perfect” vision we need also a “perfect” IOL. Even though conventional spherical IOL implantation provides higher contrast sensitivity than aphakia with spectacle correction, the contrast sensitivity in pseudophakic eyes is significantly lower than in normal phakic eyes [1]. Higher order aberrations (HOA) can reduce the quality of images in the optical system. With the new generation of aspheric IOLs, the anatomical and functional success of cataract surgery can be determined by improvement in visual acuity and its effects on the optical quality of the eye [2,3].
For better understanding of the visual system, new technologies were developed. Wavefront technology can quantify low and highorder aberrations present in the optical system. The high resolution imaging in the ophthalmic optics can be affected by higher order aberrations such as coma and spherical aberration. Conventional spherical IOLs can degrade imaging quality, increasing the HOAs of the optical system, such as coma and spherical aberrations. The light rays at the peripheral zones of a positive lens are refracted with larger angles and intersect the optical axis closer to the lens than the paracentral rays, producing positive spherical aberration. Aspherical IOL design can optimize image quality by limiting ray diffraction.
The other aspect of the quality of vision is so-called “functional vision”. This term describes the effect of visual functions, especially visual acuity, on a patient’s quality of life. This includes the ability to recognize faces, drive a car, especially at night, read, work with a computer, and perform workplace and leisure activities. This functional vision cannot be entirely evaluated by means of visual acuity charts (Snellen or similar) alone [4]. Contrast sensitivity tests are efficient indicators of visual function and can detect functional vision deficiencies that standard visual acuity tests are unable to identify [5,6].
The AcrySof IQ (model SN60WF, Alcon Laboratories, Inc.) is a posterior chamber IOL with a modified posterior surface for improving contrast sensitivity. The AcrySof IQ IOL was designed using a proprietary modeling process after analyzing the aberration of more than 700 corneas. The single-piece AcrySof Natural IOL (model SN60AT) was the platform for the AcrySof IQ IOL, and filters blue light with a proprietary blue-light filtering chromophore. The AcrySof IQ IOL shares the basic design features of the AcrySof Natural IOL including a 6.0 mm optic, 13 mm overall length, and modified-L haptics.
The purpose of this study was to compare visual acuity, HOAs, and contrast sensitivity of eyes implanted with aspheric IOLs AcrySof IQ (SN60WF) with those of eyes implanted with spherical IOLs AcrySof Natural (SN60AT), and with a control group of eyes with clear crystalline lenses.
Materials and Methods
This prospective clinical trial comprised 90 patients (105 eyes) scheduled for cataract surgery at the Department of Ophthalmology, University Hospital, Hradec Králové. Patients were selected from the department queue. The study purpose, procedures, and responsibilities were explained to all potential participants, and informed consent was obtained from all selected patients. Exclusion criteria were any ocular diseases, such as corneal opacities or irregularity, dry eye, amblyopia, anisometropia, glaucoma, retinal abnormalities, and surgical complication.
All patients underwent complete ophthalmic examination, including uncorrected visual acuity (UCVA), and best corrected visual acuity (BCVA) measurement using the Early Treatment of Diabetic Retinopathy Study (ETDRS) chart (Precision in Vision) at a viewing distance of 4 m, slit-lamp biomicroscopy, applanation tonometry, fundus examination, corneal topography (Keratron SCOUT, Opticon), and biometry with IOL calculation preoperatively. All patients were targeted for emmetropia. The cataract operations were performed using the same standardized phacoemulsification techniques under topical anesthesia: clear corneal incision of 2.4 mm, two paracenteses for bimanual irrigation/aspiration (I/A), continuous curvilinear capsulorrhexis with an approximate diameter of 5.0 mm, hydrodissection, nucleus phacoemulsification using the modified “quick-chop” technique, bimanual I/A of the residual cortex, and IOL implantation in the bag with an injector.
Implantation of spherical IOL AcrySof Natural (SN60AT) was performed in 50 eyes of 45 patients with a mean age of 62.5 years (range 50 – 70 years), and the aspheric IOL AcrySof IQ (SN60WF) was implanted in 55 eyes of 45 patients with a mean age of 64.5 years (range 57 -74 years). There were no significant differences between the groups in age, corneal curvature, axial length, IOL power, or mean follow-up. The control group consisted of 22 eyes of 20 patients of similar age (mean 62.0 years, range 57- 80 years) with clear lens crystallina and BCVA 1.0 or better.
Postoperatively patients were followed-up routinely: visual acuity with the ETDRS chart, applanation tonometry, slit-lamp biomicroscopy and fundus examination. Corneal topography, wavefront analysis and contrast sensitivity testing were performed in addition to this ophthalmologic examination between six and eight weeks after surgery. Contrast sensitivity testing was performed under photopic (85 cd/m2) and mesopic (6 cd/m2) conditions. Photopic contrast sensitivity was evaluated (CSV 1000, Contrast Sensitivity 8010 System, Oculus) at spatial frequency values of 3, 6, 12, and 18 cycles per degree (cpd). Mesopic contrast sensitivity was evaluated (Contrast sensitivity 8010 System, Neuroscientific Corp, Farmingdale, USA) at 0.74, 1.97, 3.69, 14.77, and 29.55 cpd. Eyes were not dilated for the contrast sensitivity test, and therefore the patient´s pupil size was normal.
The wavefront aberration of the whole eye was measured with the ORK Wavefront Analyser (Schwind) with a pharmacologically dilated pupil. The wavefront maps were analyzed using a 6 mm pupil diameter and up to the eighth order of Zernike coefficients.
Statistical analysis was performed by means of Student t-test.
Results
The mean preoperative topographic corneal spherical aberration (SA) was 0.32 ± 0.09 μm (range 0.18 – 0.51 μm).
All eyes in both pseudophakic groups had mean postoperative BCVA 1.0 or better. Mean postoperative UCVA was 1.0 or better in 77.8% of eyes in the AcrySof SN60WF group and in 80% eyes in the AcrySof SN60AT group (Figure 1). The difference was not statistically significant.
The postoperative wavefront analysis including mean HOA root mean square (RMS) values, coma, higher order aberrations, and spherical aberration for both pseudophakic groups are demonstrated in figure 2. No statistically significant difference was found between Acrysof Natural and AcrySof IQ regarding coma and higher order aberrations. The AcrySof IQ IOL obtained statistically significant less spherical aberration when compared with the spherical AcrySof Natural IOL (IQ 0.01 ± 0.018 μm; Natural 0.28 ± 0.08 μm). Wavefront analysis using 6 mm pupil diameter also demonstrated that AcrySof IQ IOL showed statistically significant lower HOA RMS mean values (IQ 0.44 ± 0.2 μm; Natural 0.6 ± 0.48 μm).
Figures 3 and 4 show contrast sensitivity in photopic and mesopic conditions. Mean pupil diameter was similar between the groups in both conditions. There were no statistically significant differences in contrast sensitivity between the AcrySof SN60WF and AcrySof SN60AT groups in all spatial frequencies. Mesopic contrast sensitivity was higher in patients with AcrySof IQ compared to AcrySof Natural, statistically significant in low frequencies, insignificant in medium and high frequencies. The control group (clear lens crystallina) was better in all frequencies.
Discussion
In phakic eyes, the decrease in visual acuity and in contrast sensitivity that occurs with age is usually attributed to lens changes. In young subjects, the crystalline lens compensates with its negative spherical aberration for the positive spherical aberration of the cornea, resulting in a low level of spherical aberration of the entire eye. After the age of 40 years, the spherical aberration of the lens progressively turns positive, adding to the spherical aberration of the cornea to increase the total aberrations of the eye [7-9]. Conventional monofocal plane-convex or biconvex IOLs can introduce only positive spherical aberration decreasing image quality [10,11]. Some pseudophakic patients complain about glare, halos, and starburst that could be attributed to spherical aberration [12]. Based on these findings, it was postulated that an aspheric IOL that restored the conditions of the young eye with respect to spherical aberration would produce a pseuphakic eye with improved contrast sensitivity as compared to pseudophakic eyes implanted with spherical IOLs. In this study, we found statistically significant lower spherical aberration and HOA RMS mean values after implantation of the AcrySof IQ when compared with implantation of AcrySof Natural.
The most significant differences between aspheric and spherical IOLs related to contrast sensitivity occurred at mesopic levels. Mester et al. [13] found statistically significant improvement in the aspheric IOL group (Tecnis Z9000) in mesopic contrast sensitivity at low spatial frequences. These authors also found no difference between the aspheric and spherical IOL in photopic conditions. In our study we have found similar results for spherical IOL AcrySof Natural and its “aspheric variation” AcrySof IQ. There were no statistically significant differences in photopic contrast sensitivity between the AcrySof SN60WF and AcrySof SN60AT groups in all spatial frequencies. Mesopic contrast sensitivity was higher in patients with Acrysof IQ compared to Acrysof Natural, statistically significant in low frequencies, insignificant in medium and high frequencies. The control group (clear lens crystallina) was better in all frequencies.
In conclusion, the adoption of ocular wavefront technology in clinical ophthalmology has made it possible to quantify total ocular aberrations and better understand the benefits of a customized IOL to correct aberrations of the eye. The aspheric AcrySof IQ induced significantly less spherical aberration than the AcrySof Natural. It also presented better contrast sensitivity only under mesopic conditions at low frequencies.
Conflict of Interest
None of the authors has a financial or proprietary interest in any product mentioned.
References













Select your language of interest to view the total content in your interested language
 
Share This Article
   
 
   
 
Relevant Topics
Disc Acupuncture for Glaucoma
Disc Acute Glaucoma
Disc Advanced Glaucoma Therapeutics
Disc Advances in Glaucoma Diagnosis
Disc Albanism
Disc Behavioural Optometry
Disc Cataract
Disc Chronic Glaucoma
Disc Clinical Ophthalmology
Disc Clinical Optometry
Disc Congenital Glaucoma
Disc Conjuctivitis
Disc Cornea and Corneal Diseases
Disc Diamox Glaucoma
Disc Experimental Ophthalmology
Disc Experimental Optometry
Disc Eye Care Optometry
Disc Eye Design Optometry
Disc Eye diseases
Disc Eye to Eye Optometry
Disc Glaucoma
Disc Glaucoma Drugs Research
Disc Glaucoma Epidemiology
Disc Glaucoma Hereditary
Disc Glaucoma Medicine
Disc Glaucoma Pathophysiology
Disc Glaucoma Prognosis
Disc Glaucoma Screening
Disc Glaucoma and Diabetes
Disc Image Optometry
Disc Intraocular Pressure
Disc Malignant Glaucoma
Disc Modern Eye Optometry
Disc Molecular Degeneration Amsler Grid
Disc Myopia
Disc Natural Remedies for Glaucoma
Disc Neuro-Ophthalmology
Disc Occular Migraine
Disc Ocular Microbiology
Disc Ocular oncology
Disc Ophthalmic Research
Disc Ophthalmic imaging
Disc Ophthalmoscopy
Disc Opticians Communications
Disc Optometry
Disc Optometry Case Reports
Disc Optometry Equipment Market Analysis
Disc Optometry Exercises
Disc Optometry Management Software
Disc Optometry Practice
Disc Optometry Vision Science
Disc Paediatric Optometry
Disc Paediatric ophthalmology
Disc Pediatric Glaucoma
Disc Phacolytic Glaucoma
Disc Phacomorphic Glaucoma
Disc Pigmentary Glaucoma
Disc Primary & Secondary Eye care
Disc Pseudoexfoliation Glaucoma
Disc Trauma Related Glaucoma
Disc Vision Problems
Disc Vision research
Disc Visual Eyes Optometry
 
Recommended Journals
Disc Optometry Journal
Disc Clinical & Medical Imaging Journal
Disc Glaucoma Journals
Disc Ophthalmic Pathology Journal
  View More»
 
Recommended Conferences
Disc 8th Ophthalmology Meeting  
July 18-20, 2016, Chicago, USA
Disc Global Ophthalmology Meeting
July 18-20, 2016, Chicago, USA
Disc Cataract and Refractive Surgery Conference
August 04-05, 2016 Manchester, UK
Disc World Ophthalmology Conference
Sept 15-17, 2016 Berlin Germany
Disc 2nd Conference on Eye and Vision
Sept 26-28, 2016, Miami, USA
Disc 9th World Ophthalmic Conference
Oct 24-26, 2016, Istanbul, Turkey
Disc 10th Clinical and Experimental Ophthalmology Conference
Nov 28-30, 2016 Dubai, UAE
View More»
 
Article Tools
Disc Export citation
Disc Share/Blog this article
 
Article usage
  Total views: 11155
  [From(publication date):
December-2012 - May 30, 2016]
  Breakdown by view type
  HTML page views : 7429
  PDF downloads :3726
 
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

 
OMICS International Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
 
 
OMICS International Conferences 2016-17
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings
 
 

Contact Us

Agri, Food, Aqua and Veterinary Science Journals

Dr. Krish

agrifoodaquavet@omicsinc.com

1-702-714-7001 Extn: 9040

Clinical and Biochemistry Journals

Datta A

clinical_biochem@omicsinc.com

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

business@omicsinc.com

1-702-714-7001Extn: 9042

Chemical Engineering and Chemistry Journals

Gabriel Shaw

chemicaleng_chemistry@omicsinc.com

1-702-714-7001 Extn: 9040

Earth & Environmental Sciences

Katie Wilson

environmentalsci@omicsinc.com

1-702-714-7001Extn: 9042

Engineering Journals

James Franklin

engineering@omicsinc.com

1-702-714-7001Extn: 9042

General Science and Health care Journals

Andrea Jason

generalsci_healthcare@omicsinc.com

1-702-714-7001Extn: 9043

Genetics and Molecular Biology Journals

Anna Melissa

genetics_molbio@omicsinc.com

1-702-714-7001 Extn: 9006

Immunology & Microbiology Journals

David Gorantl

immuno_microbio@omicsinc.com

1-702-714-7001Extn: 9014

Informatics Journals

Stephanie Skinner

omics@omicsinc.com

1-702-714-7001Extn: 9039

Material Sciences Journals

Rachle Green

materialsci@omicsinc.com

1-702-714-7001Extn: 9039

Mathematics and Physics Journals

Jim Willison

mathematics_physics@omicsinc.com

1-702-714-7001 Extn: 9042

Medical Journals

Nimmi Anna

medical@omicsinc.com

1-702-714-7001 Extn: 9038

Neuroscience & Psychology Journals

Nathan T

neuro_psychology@omicsinc.com

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

John Behannon

pharma@omicsinc.com

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

social_politicalsci@omicsinc.com

1-702-714-7001 Extn: 9042

 
© 2008-2016 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version