Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Automated Quality Control of Pharmaceuticals using Sequential Injection Chromatography (SIC)

Paraskevas D. Tzanavaras*

Laboratory of Analytical Chemistry, Department of Chemistry, Aristotelian University of Thessaloniki, GR-54124 Thessaloniki, Greece

*Corresponding Author:
Paraskevas D. Tzanavaras
Laboratory of Analytical Chemistry, Department of Chemistry
Aristotelian University of Thessaloniki, GR-54124 Thessaloniki, Greece
Tel: 00302310997721
E-mail: [email protected]

Received date: October 19, 2012; Accepted date: October 20, 2012; Published date: October 22, 2012

Citation: Tzanavaras PD (2012) Automated Quality Control of Pharmaceuticals using Sequential Injection Chromatography (SIC). Pharmaceut Anal Acta 3:e130. doi: 10.4172/2153-2435.1000e130

Copyright: © 2012 Tzanavaras PD. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Pharmaceutica Analytica Acta

The present piece is my third Editorial contribution to the international journal Pharmaceutica Analytica Acta. The first Editorial (http://dx.doi.org/10.4172/2153-2435.1000105e) discussed some potential of automated flow injection techniques in the field of pharmaceutical analysis, focusing on chemiluminescence detection and flow optosensors [1]. The topic of the second Editorial (http://dx.doi. org/10.4172/2153-2435.1000e114) was more specialized discussing the applications of Sequential Injection Analysis (SI) to drug dissolution studies [2]. This third Editorial is also focused on automated flow techniques and will present theoretical and practical aspects of one of the newest members of the family of automated flow injection techniques, namely Sequential Injection Chromatography (SIC).

Despite the hundreds publications using Flow (FI) and Sequential Injection Analysis (SI) reporting the automated analysis of pharmaceuticals, the majority of the methods can be characterized as “single analyte assays” [3]. This is an important drawback of automated low pressure flow techniques, although there are various attempts towards this direction. These reports were in many cases ingenious, but in general were rather complicated or required strictly controlled conditions and were typically based on kinetic discrimination [4], online sample splitting [5] or usage of different detection systems [6]. On the other hand, simultaneous multi-component analysis is critical in many aspects of the quality control of pharmaceuticals. For example there are numerous two- or three-active component formulations (e.g. combinations of antihypertensive and diuretic ingredients). Additionally, purity control and stability testing of pharmaceuticals require per se the usage of analytical techniques with separation capabilities.

The breakthrough in multi-species determinations by FI or SI was made very recently by the incorporation of short monolithic columns to low pressure continuous flow manifolds, introducing the concept of Sequential Injection Chromatography (SIC) [7]. Reversed phase monolithic columns of various dimensions suitable for HPLC applications in many analytical fields were commercialized by Merck in 2000 [8] and were followed by other manufacturers (e.g. Phenomenex). The structural properties of monolithic stationary phases allow efficient separations typically comparable to 3-5 μm particulate columns at usually ten-times lower back pressures. This can be achieved by controlling independently the permeability (μm-sized throughpores) and efficiency of the analytical columns (nm-sized mesopores) [9] (Figure 1).

pharmaceutica-analytica-acta-Typical-structure

Figure 1: Typical structure of a reversed phase monolithic stationary phase.

In a typical SIC configuration (Figure 2) a short monolithic column (5-50 mm long × 4.6 mm i.d.) is placed in the flow line between the multiposition valve and the detector. Depending on the used pump (syringe pump or peristaltic) the mobile phase can either flow continuously through the holding coil to the column (peristaltic pump) or can be aspirated prior to every run (syringe pump). Defined volumes of the samples at the microliters level are injected in the usual way in the holding coil followed by propulsion to the column by selecting the appropriate port of the multiposition valve and by reversing the flow direction of the pump. Due to the pressure limitations of conventional SI setups, mobile phase flow rates are typically lower than 1.0 mL min-1. Fast separations and short analysis cycles are based on the short length of the monolithic columns, while chromatographic efficiency depends on the selection of the mobile phase and on the properties of the analytes. The interested reader can find useful information on SIC in two excellent review articles that have been published recently [10,11].

pharmaceutica-analytica-acta-selection-valve

Figure 2: Representative SIC setup: C = carrier (mobile phase); SP = syringe pump; HC = holding coil; SV = selection valve (or multiposition valve); MC = monolithic column; S = sample; W = waste; UV = detector.

Some representative pharmaceutical applications of SIC can be found in Table 1 [7,12-22]. Looking at the methods and the analytical conditions included in Table 1 it is evident that most of the so far reported and published protocols involve separation of simple two/ three component mixtures employing isocratic elution and universal UV detection. To my opinion the potentials of SIC and related techniques can be demonstrated to the scientific community by mainly two ways: (i) by developing SIC methods with gradient elution that will expand the separation capabilities of the technique to more complicated mixtures, and (ii) by exploiting and demonstrating the automated sample treatment potentials of SI combined to monolithicbased separations. Interesting work towards these directions has been carried out by some research groups. For example, Zacharis et al. [23] applied on-line filtration and dilution protocols prior to SIC separation that allowed direct analysis of samples from stress degradation studies of acyclovir (Figure 3) [23]. In 2010 Chocholous and co-workers proposed a two-column SIC setup that enables the separation of analytes with different polarities with isocratic elution [24], while the same research group one year later exploited the potentials of gradient formation in the holding coil by reproducible mixing of two mobile phases [25].

Analytes Analytical Column Detection Mobile Phase Flow rate (mL min-1) Reference
Salicylic acid/
Methyl salicylate
Chromolith® SpeedRod
(50 × 4.6 mm i.d.)
UV at 240 nm ACN-water  (36/65) pH=2.5 0.6
[7]
Methylparaben/
Propylparaben/ 
triamcinolone acetonide
Chromolith®FlashRod
(25 × 4.6 mm i.d.)
UV at 243 nm ACN-MeOH-water  (36/65) +0.05% nonylamine, pH=2.5 0.6 [12]
Methylparaben/
Propylparaben/  diclofenac
Chromolith® FlashRod
(25 ×4.6 mm i.d.)
UV at 275 nm ACN-water  (40/70) +0.05% triethylamine, pH=2.5 0.48-1.2 [13]
Ambroxol hydrochloride / deoxycycline Chromolith® FlashRod
(25 × 4.6 mm i.d.)
UV at 213 nm ACN-water  (20/90), pH=2.5 0.48 [14]
Ambroxol hydrochloride / methylparaben /
benzoic acid
Chromolith® FlashRod
(25 × 4.6 mm i.d.)
UV at 245 nm ACN-THF-0.05 M CH3COOH (10/10/90), pH=3.75 0.48 [15]
Lidocaine/prilocaine Chromolith®FlashRod
(25 × 4.6 mm i.d.)
UV at 212 nm ACN-0.05 M Phosphate (40/80) + 0.01% triethylamine 0.6 [16]
Naphazoline nitrate / methylparaben Chromolith® FlashRod
(25 × 4.6 mm i.d.)
UV at 220 nm and 256 nm MeOH-water  (40/65), pH=5.2 0.9 [17]
Betamethasone/
Chloramphenicol
Chromolith®FlashRod
(25 × 4.6 mm i.d.)
UV at 241 nm and 278 nm ACN-water  (30/80) 0.48 [18]
Triamcinolone acetonide / Salicylic acid Onyx®C18
(50 × 4.6 mm i.d.)
UV at 240 nm ACN-water  (35/65) 0.9 [19]
sildenafil Chromolith® FlashRod
(25 × 4.6 mm i.d.)
UV at 240 nm Ammonium acetate (0.3 mol L-1)/ Acetonitrile (50:50) 2.4 [20]
Diclofenac Chromolith® FlashRod
(25 × 4.6 mm i.d.)
UV at 248 nm MeOH-water  (50/50) 2.4 [21]
Amoxicillin and
clavulanic acid
Chromolith® FlashRod
(25 × 4.6 mm i.d.)
UV at 228 nm Phosphate (25 mmol L-1, pH=3.0) / MeOH (85/15) 2.4 [22]

Table 1: Representative applications of sequential injection chromatogarphy (SIC) to pharmaceutical analysis.

pharmaceutica-analytica-acta-impurity-guanine

Figure 3: SIC chromatogram of acyclovir and its major impurity guanine (from [23]).

The common denominator of all the SIC procedures mentioned and discussed so far in this editorial was the usage of monolithic stationary phases. Very recently Chocholous et al. [19] reported, for the first time, an interesting alternative that could expand the potentials of SIC [26]. This alternative was the incorporation of short HPLC columns based on the novel “fused core” or “core shell” technology. This technology is based on sol–gel techniques to achieve a homogeneous porous layer on a solid core of silica. It provided higher efficiencies due to the smaller diffusion distance and improved mass transfer, lower internal porosity and narrower particle size distribution (Figure 4) [27].

pharmaceutica-analytica-acta-shell-particles

Figure 4: Typical structure of core shell particles (from Phenomenex, Kinetex).

To conclude this editorial SIC is definitely a viable technique that offers enhanced potentials to chemical analysis. To my opinion SIC researchers should not focused on “competing” conventional HPLC to straightforward applications. The real strength of SIC is - as already To conclude this editorial SIC is definitely a viable technique that offers enhanced potentials to chemical analysis. To my opinion SIC researchers should not focused on “competing” conventional HPLC to straightforward applications. The real strength of SIC is - as already

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Recommended Conferences

  • 4th International Conference on Drug Discovery, Designing and Development
    June 27-28, 2018 Vancouver, Canada
  • 18th World Pharma Congress
    October 18-19, 2018 Warsaw, Poland

Article Usage

  • Total views: 11759
  • [From(publication date):
    October-2012 - Jun 21, 2018]
  • Breakdown by view type
  • HTML page views : 7990
  • PDF downloads : 3769
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2018-19
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

+1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2018 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
Leave Your Message 24x7