Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Electrochemical Impedance Spectroscopy (EIS) Study of Modified Type-316L Stainless Steel (SS) as an Effective Biomaterial for Orthopedic Implant Applications

Sharif Hussein Sharif Zein*, Sharif Zein and Muhammad Faiq Abdullah

School of Chemical Engineering, Engineering Campus, Universiti Sains Malaysia, Seri Ampangan, 14300 Nibong Tebal, Seberang Perai Selatan, Pulau Pinang, Malaysia

*Corresponding Author:
Sharif Hussein Sharif Zein
School of Chemical Engineering
Engineering Campus, Universiti Sains Malaysia
Seri Ampangan, 14300 Nibong Tebal
Seberang Perai Selatan
Pulau Pinang, Malaysia
E-mail: [email protected]

Received Date: April 27, 2012; Accepted Date: April 27, 2012; Published Date: May 02, 2012

Citation: Zein SHS, Zein S, Abdullah MF (2012) Electrochemical Impedance Spectroscopy (EIS) Study of Modified Type-316L Stainless Steel (SS) as an Effective Biomaterial for Orthopedic Implant Applications. J Nanomed Nanotechol 3:e108. doi:10.4172/2157-7439.1000e108

Copyright: © 2012 Zein SHS, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Nanomedicine & Nanotechnology

For centuries, metals and alloys have been used in substitution for anatomical structures owing to their extraordinary mechanical properties [1]. Type 316L Stainless Steel (SS) is known to be widely used for implantation purposes in orthopedic surgery due to its corrosion resistance, effective cost and superior mechanical properties [1,2]. However, several incidences of failures involving type 316L SS have been occurred which released the metal ions into tissues [1] and blood stream surrounding the implants, thus causing histological changes of the local tissue by direct toxic effects or local hypersensitivity reaction [3]. It is highly probable that electrophoretically deposited Multiwall Carbon Nanotubes (MWCNTs) and Hydroxyapatite (HA) on type 316L SS could minimize this limitation, and further investigation of its corrosion behavior using in vitro electrochemical techniques [1,4] may confirm the acceptability of modified SS as an effective biomaterial to be used as orthopedic implants.

An ideal orthopedic implant should possess biocompatible, osteoconductivity [5], superior mechanical properties [6-8] and excellent corrosion resistance to physiological environments of the human body [1,9]. Different electrochemical techniques for evaluation of corrosion behavior have been studied extensively included Electrochemical Impedance Spectroscopy (EIS), cyclic potentiodynamics polarization and Open Circuit Potential (OCP) studies [2]. EIS is a useful approach to monitor in situ electrochemical changes [10] and to develop understanding of the physical process occurring at the electrode-electrolyte interface [11-13]. EIS study was carried out based on an appropriate circuit model in order to fit the experimental data to equivalent circuit model parameters, whereas each parameter would represents macroscopic physical quantity which contributing to interface impedance [11]. An equivalent circuit was selected to represent the electrochemical behavior of metals implants that have been coated with an unsealed porous film [14]. The equivalent circuit consists of solution resistance of the electrolyte (experimental human body fluid solution), Re, capacitance of non-defective coating layer, Cp, charge transfer resistance of electrolyte’s penetration through pinholes, Rp, polarization resistance of substrate, Rb and electrical double-layer capacitance at substrate/electrolyte interface, Cb [4]. This physical model of the electrochemical reactions that occurred at the electrodes is important to interpret the electrochemical behavior of the metal implants from EIS spectra [15].

Certain amount of current density is necessary to generate activity in stimulation process which resulted in interface impedance of the metal implants [4]. The equivalent model comprises an interface capacitance, shunted by charge transfer resistance and in series with solution resistance [16]. Then, a theoretical equation has been used to determine the interface capacitance impedance and solution resistance in order to confirm the results that correspond well with the fitted parameter values, hence validating the model equations [4]. Impedance behaviors of the metals implants could be expressed either in a Bode plot of logarithm of impedance modulus (|Z|) and phase angle (Φ) as a function of logarithm of frequency or in a Nyquist plot of Z” (ω) as a function of Z’(ω) [14,17]. The Bode plots of logarithm total impedance |Z| vs. frequency for uncoated SS and coated samples of SS/MWCNT-OH are shown in Figure 1(a) and for uncoated SS and coated samples of SS/MWCNT-OH/HA as in Figure 1(b). Bode plots were used to estimate the impedance parameters in equivalent circuit model, thereby indicates the corrosion behavior of modified SS after immersed in Phosphate Buffer Saline (PBS) solution. The decrease of polarization resistance and coating layer capacitance imply that the coating layers of MWCNT-OH and HA tend to protect SS by reducing the penetrations of anions from PBS solution [1,14], hence improving corrosion resistance of the modified SS [1].

nanomedicine-nanotechnology-bode-plots-logarithm

Figure 1: Bode plots of logarithm total impedence |Z| vs. frequency of (a) uncoated SS and coated samples of SS/MWCNT-OH and (b) uncoated SS and coated samples of SS/MWCNT-OH/HA.

Cyclic potentiodynamic polarization study is carried out when the susceptibility of metal implants to localized corrosion need to be addressed [18]. Pitting potential, Eb and pit protection potential, Eprot could be determined from cyclic potentiodynamic polarization curve [19] as these parameters are needed to evaluate pitting corrosion of metal implants [2]. Direct measurement of pit propagation mechanisms is proportional to the area under cyclic potentiodynamic polarization curve [2,19], as larger area of hysteresis loop indicates higher rate of pits formation [1]. OCP measurement provides simple evaluation of corrosion behavior including oxide film formation and metal implants passivation in experimental body fluid environment as a function of time [2,20]. The increasing potential in positive direction achieving steady state indicates the metal implants remains protective [17,20] while potential drops toward negative direction implies no film formation or break in the film [17] due to dissolution of carbon, ferum and oxides particles from metal implants surface when immersed in experimental body fluid solution [4]. Cyclic potentiodynamic polarization and OCP studies are able to provide useful evaluation of corrosion behavior [20] and pitting corrosion resistance [19] of metal implants, however, these electrochemical techniques provide less information of adsorption behavior and interactions of body fluid environment to metal implant surface [21], thus leaving EIS study as suitable technique to investigate the acceptability of modified SS to be used as orthopedic implants.

So, is the corrosion behavior study of the modified SS by EIS is enough to confirm its acceptability as an effective orthopedic implants? While in vitro EIS study have indeed able to show lower polarization resistance and coating layer capacitance of modified SS, more quantitative determination of corrosion products need to be performed in order to investigate their adverse effects on tissues surrounding the implants [19]. Hence, both in vitro cell-culture experiments and clinical retrieval studies [22] in conjunction with in vivo biological investigations of modified SS need to be investigated to ensure its continuous clinical applications in biomedical field.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Recommended Conferences

  • 2nd World Congress on Nano Science and Nano Technology
    August 10-11, 2018 Osaka, Japan
  • 25th Nano Congress for Future Advancements
    August 16-18, 2018 Dublin, Ireland
  • 20th Asia Pacific Nanotechnology Congress
    July 23-24, 2018 Sydney, Australia
  • 26th International Conference on Advanced Nanotechnology
    October 04-05 2018 Moscow, Russia
  • Nanotechnology & Materials Engineering
    Oct 25-26, 2018 Budapest, Hungary

Article Usage

  • Total views: 12045
  • [From(publication date):
    June-2012 - Jul 20, 2018]
  • Breakdown by view type
  • HTML page views : 8240
  • PDF downloads : 3805
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2018-19
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

+1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2018 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
Leave Your Message 24x7