Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Ensuring Radiation Safety to Staff in Lymphatic Tracing and Sentinel Lymph Node Biopsy Surgery – Some Recommendations

Coventry BJ1,2*, Collins PJ3, Kollias J1, Bochner M1, Rodgers N4, Gill PG1, Chatterton BE3 and Farshid G4

1Breast-Endocrine Surgical Oncology Unit, Department of Surgery, University of Adelaide, Royal Adelaide Hospital, Adelaide, South Australia

2Adelaide Melanoma Unit, Department of Surgery, University of Adelaide, Royal Adelaide Hospital, Adelaide, South Australia

3Nuclear Medicine, University of Adelaide, Royal Adelaide Hospital, Adelaide, South Australia

4Pathology, SA Pathology, Institute of Medical and Veterinary Science, Adelaide, South Australia

Corresponding Author:
Prof BJ Coventry
Breast-Endocrine Surgical Oncology Unit and Adelaide Melanoma Unit
Department of Surgery, University of Adelaide
Royal Adelaide Hospital, North Terrace, Adelaide, South Australia
E-mail: brendon.coventry@adelaide.edu.au

Received Date: June 08, 2012; Accepted Date: June 18, 2012; Published Date: June 22, 2012

Citation: Coventry BJ, Collins PJ, Kollias J, Bochner M, Rodgers N, et al. (2012) Ensuring Radiation Safety to Staff in Lymphatic Tracing and Sentinel Lymph Node Biopsy Surgery – Some Recommendations. J Nucl Med Radiat Ther S2:008. doi:10.4172/2155-9619.S2-008

Copyright: © 2012 Coventry BJ, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Nuclear Medicine & Radiation Therapy

Abstract

Introduction: Lymphatic mapping and sentinel node biopsy (LM/ SNB) techniques for melanoma and breast cancer management potentially expose staff, including operating theatre personnel, radiologists, pathologists and others, to ionising radiation.

Aims: To ascertain exposure levels in a practical setting and to establish safe work practices for staff involved in the LM/ SNB procedure pathway.

Methods: Cumulative intra-procedural extremity (hands) and whole body radiation doses were recorded separately for surgeons, pathologists and couriers during standard sentinel lymph node biopsy procedures from 13 melanoma and included also radiologists in 11 breast cancer cases.

Results: The measured extremity dose for melanoma procedures was zero for surgeons and pathologists. The extremity dose for breast cancer procedures was approximately 250 μSv for surgeons, and about 10 μSv for pathologists per breast procedure if done on the day of surgery, but is otherwise negligible; zero for the radiologist; and zero for the courier. No whole body dose was detectable for any staff member.

Conclusions: Using the international limit for skin dose some 200 breast cancer procedures could be performed per annum per surgeon (at the general public radiation limit) – and 2000 breast surgical procedures (at the radiation worker limit) based on extremity doses. Radiologists, pathologists and couriers received minimal or zero radiation doses from handling breast specimens. Melanoma procedures showed no measurable dose. Some recommendations for effective safe work practices are given.

Keywords

Radiation safety; Sentinel node biopsy; Lymphatic mapping; Melanoma; Breast cancer

Introduction

With the introduction of lymphatic mapping and sentinel node biopsy (LM/ SNB) techniques using radioactive tracers for the management of melanoma and breast cancer, operating theatre personnel, pathologists and other staff are now being exposed to ionising radiation from radiopharmaceuticals.

Lymphoscintigraphy, lymphatic tracing, sentinel node biopsy, specimen transport, specimen radiology and the subsequent pathological analysis of the tissues may result in significant external and internal radiation exposure to personnel handling the specimen or in close proximity. Several studies (refs) have been presented in abstract form and also in the published literature relating to the safety of these techniques, but recommendations for practical safe handling procedures for these agents in the clinical setting following surgery may be useful for those concerned.

To put these results into perspective, LM/ SNB techniques for breast cancer using a 99mTc colloid yields a patient effective dose of approximately 0.3 μSv. This compares with doses from 99mTc Bone Scan (3.6 μSv), 123m I thyroid scan (4.4 μSv), mammogram (0.4 μSv), chest x-ray (0.04 μSv), CT chest (8.3 μSv) or abdominal CT (7.2 μSv), return airflight London to Sydney (0.20 μSv), UK annual consumed food and drink (0.37 μSv), or one years’ residence on Denver Colorado from background radiation (0.88 μSv) [1].

The study primarily aimed to measure radiation exposure to the extremity (hands) and (whole) body of the operating surgeon during melanoma [2,3] and breast [4] sentinel node biopsy procedures. Secondary aims were to assess radiation exposure further through the entire LM/ SNB procedure pathway for radiology, pathology and courier staff. Guidelines could then be devised for safe handling and processing of radioactive tissues as part of the sentinel node biopsy procedure in order to minimise radiation exposure risk for surgical and other personnel.

Methods

Patients

Informed patient consent was obtained for all procedures. No patients were pregnant.

Lymphatic mapping and sentinel node biopsy

99mTc Antimony sulfide colloid (Lymphflo™, Adelaide, Australia) was used for all studies and 40 MBq was injected in 4 divided doses either intradermally around the primary melanoma excision site, or peri-tumourally around the primary breast cancer. The pre-operative lymphoscintigram was performed for each case prior to Patent Blue (Guerbet, Faulding, Adelaide) dye injection, according to the same protocol as that used by our group as part of the Multicenter Selective Lymphadenectomy Trial (MSLT-I) studies [2,3] and for breast cancer [4]. All patients were injected with the radiopharmaceutical on the morning of surgery and proceeded to the operating theatre later that day, within 3-5 hours of receiving the radiopharmaceutical. Personal dosimeters were worn during the operative period from the time of commencement to completion of the procedure. Measurements at the non-dominant index finger were performed using a finger badge dosimeter (Figure 1 - right), and whole body doses were measured using a waist badge (Figure 1 – left). Sterility was preserved by careful washing of the finger-badge in antiseptic solution, and then placing it between two layers of sterile gloves using a double gloving technique (Figure 2). The pathway of the sentinel node procedure was identified from initial entry into the operating theatre, transport of the specimen from theatre to specimen radiology (breast) and to the pathology department (melanoma and breast) and steps at each stage were measured for radiation exposure levels (Figure 3).

nuclear-medicine-badge-dosimeter

Figure 1: A Finger badge dosimeter (right), and a whole body waist badge dosimeter (left) were used for measurements of radiation exposure and used serially amongst surgeons performing procedures. Separate badges were used for radiology staff and for pathology staff performing wider excision and/or sentinel node diagnostic procedures.

nuclear-medicine-finger-badge

Figure 2: The finger-badge was carefully washed in antiseptic solution, and then placed between two layers of sterile gloves and worn on the non-dominant index finger, using a double gloving technique.

nuclear-medicine-sentinel-node

Figure 3: The pathway for the sentinel node procedure from the operating theatre, specimen transport from theatre to specimen radiology (breast) and to the pathology department (melanoma and breast) during which radiation exposure was measured.

Staff radiation measurement

Surgeons: Intraoperative radiation doses were recorded for surgeons during sentinel lymph node biopsy procedures. Thirty-six (36) procedures were used to estimate whole body radiation doses and 24 procedures for extremity (finger) doses. Four surgeons participated in the studies and wore the whole body (WB) and finger badge (FB) sequentially for estimation of cumulative doses. No surgeons were pregnant.

The radiation exposure to the whole body and extremity (hands) of the surgeon were evaluated using thermo-luminescent dosimeters (TLDs) worn on the waist/ chest and as a finger badge on the nondominant index finger respectively (Figure 1 and 2). The finger and whole body dosimeters were worn by different surgeons for sequential sentinel node operations for 1 and 3 months respectively, to provide cumulative doses. The TLD dosimetry was measured by ARPANSA (Melbourne Australia).

Specimen radiology: Whole body and extremity measurements using TLD badges were recorded in the radiology suite for the radiologist performing breast specimen radiographs and ultrasounds of excised breast tissue containing the primary tumour from SN procedures. This tissue included most of the injected radiopharmaceutical dose. The cumulative radiation dose to the finger and whole body of the radiologist performing the studies was measured for specimens from 28 breast cancer patients.

Pathological processing: The primary tumour specimen contained approximately 95% of the radioactivity, whereas the sentinel node only contained < 5% of the injected dose. The radiation exposure to the pathologist and technician (combined) was measured over two separate time periods using two different processing methods. In Method I the tissues (9 melanoma and breast cancer specimens) were processed the day following LM/ SNB surgery. Method II included imprint cytology which was performed intra-operatively at the time of the surgery on the margins of the resected breast tissue (21 breast cancer and no melanoma cases) containing the primary tumour and also on the bisected sentinel lymph node.

Whole body and finger dosimeters were worn by the Pathologist during processing of the tissue for routine histopathology assessment on the day after surgery (method 1), and also measured during the taking of imprint cytology of the fresh breast specimen (method 2). A total of 21 samples were analysed.

Medical couriers: The cumulative radiation exposure for courier staff transporting the breast tissues from the operating theatre, via radiology, to the pathology laboratory was measured for 19 cases.

The cumulative radiation dose was measured using a TLD whole body waist badge worn by courier staff transporting the breast tissues in plastic containers to the pathology laboratory for 19 breast specimens.

Results

Surgeons

The cumulative whole body radiation dose to the surgeons (total 36 studies) was not detectable, being below the detection sensitivity of the dosimeter (approximately 10 μSv).

Tables 1 and 2 provide the cumulative extremity dose to surgeons from 24 breast and melanoma studies over 6 separate measurementtime periods. These comprised 3 periods where radiation was measurable (A,B,C) and 3 where radiation was not measurable (D,E,F), separated for convenience into Table 1 and 2 respectively. The average extremity dose for the 3 wearing periods (A,B,C) in which a radiation dose was recordable (8 breast and 2 melanoma) (Table 1) was 250 μSv (Table 1 and 2).

Wearing period
(1 month each)
Dose (mSv) Time (min) Breast
cases
Melanoma cases
A 800 77 2 0
B 800 200 2 1
C 900 310 4 1
Totals 2500 587 8 2

Table 1: Wearing periods for finger badges in which doses for surgeons were recorded.

Wearing period
(1 month each)
Dose (mSv) Time (min) Breast
cases
Melanoma cases
D 0 178 2 3
E 0 500 1 6
F 0 200 0 2
Totals 0 878 3 11

Table 2: Wearing periods for finger badges in which doses for surgeons were not measurable.

The mean time for breast procedures was 47 minutes and for melanoma 77 minutes - these were statistically significantly different (p < 0.03). There was no correlation between extremity dose and duration of the procedure.

Radiologist

The cumulative radiation dose to whole body and extremity for the radiology staff was not detectable using the current TLD dosimeters.

Pathologists

The cumulative whole body radiation dose for the pathologist examining the breast or melanoma samples was not detectable using the current dosimeters used during either method 1 or method 2. The extremity dose for the 9 melanoma and breast specimens processed using method 1 (on the day after surgery) was below the detection limit.

However, using method 2, the extremity dose when performing imprint cytology of the margins of the fresh breast specimen was 200 μSv for the 21 procedures – approximately 10 μSv per procedure.

Medical courier

The cumulative whole body radiation dose for the couriers transporting the specimens was below the detection limit.

Discussion

Sentinel lymph node tracing and biopsy (LM/ SNB) techniques using radiopharmaceuticals have gradually gained status as a clinical staging and potential therapeutic tool. The widespread use of these techniques has meant that surgeons, pathologists and other staff are potentially exposed to radiation from the injected radioactive material. This has led to safety concerns for hospital staff involved in the sentinel node dissection and evaluation pathway. Although strict radiation handling and safety procedures almost universally apply in Nuclear Medicine Departments, the issues relating to the handling of radioactive materials outside of these areas may not be well understood by staff.

In our institution we have addressed requirements to assure safe work practices for staff performing LM and SNB studies. These have included tissue transportation, radiological handling and pathological processing. Although several studies have provided some important information concerning radiation exposure and possible suitable protective measures with LN and SNB for surgeons and pathologists [5-11], little information exists on extremity (or whole body) radiation exposure to other staff. More recent studies [12-17] have addressed parts of the LM/ SNB procedure pathway. Our studies specifically addressed the radiation exposure of staff involved in the entire LM/ SNB pathway.

Dose to the operating surgeon

A small study has demonstrated that there was no statistical difference in radiation exposure to the hands of the surgeon or first assistant between melanoma and breast cancer sentinel node biopsy procedures using 99mTc albumin sulphur colloid [5].

Our data suggest that the extremity radiation dose to the surgeon is measurably higher for breast cancer than for melanoma LM/ SNB procedures, despite a longer time being required for melanoma cases. This can be explained by (i) the shorter distance (on average) of the sentinel node(s) from the primary tumour in breast cancer studies (thus increasing the working time of the surgeon closer to the injection site) compared to melanoma procedures, and (ii) the reduced direct contact time with the primary tumour site for excision of the melanoma primary, where forceps and longer handled instruments are commonly used, compared with that for the breast primary tumour excision, where relatively more close finger manipulation of the tumour is usual.

The extremity dose (finger, hand) to surgeons from breast studies of approximately 250 μSv per study is still low compared with annual International limits (500,000 μSv for radiation workers and 50,000 μSv for general public, or non-radiation workers (ICRP 60). Therefore, at conservative, minimum radiation safety levels some 200 breast cancer procedures could be performed per annum by a single surgeon before the radiation limit for non-radiation workers is exceeded – and some 2000 surgical procedures could be performed before the limit for radiation workers is exceeded.

In addition, our data show that the whole body dose to the surgeon from LM/ SNB is negligible - not being measurable using standard personal dosimeters over a number of studies.

The dose to the radiologist

Specimen mammography and/ or ultrasound of the surgical breast specimen is now commonplace for assessment of radiological margins to predict the adequacy of surgical clearance. This has meant that imaging staff may be exposed to radiation from 99mTc tracers used for LM/ SNB during specimen radiography. However, our data show that whole body or extremity radiation exposure to radiology staff is negligible.

The dose to the pathologist

When the pathologist processed the primary melanoma specimen the day after LM/ SNB procedure, sufficient decay had occurred to reduce the radioactivity substantially, so that no whole body or extremity dose was detected. However, if the tissue was processed on the day of surgery, for example if performing breast tissue imprint cytology, the pathologist received an extremity dose of approximately 10 μSv per procedure. Imprint cytology for margin assessment is not routinely practiced in most institutions. The examination of SN imprint cytology and the handling of the breast specimen after fixation were associated with still lower exposure levels.

The dose to specimen transport and laboratory staff

Staff transporting the radioactive specimen from the operating theatre to the radiology suite and/ or pathology laboratory are potentially at risk of radiation exposure from holding the specimen container close to their bodies. This could be further increased if several radioactive specimens were transported together. Measurement of whole body doses revealed negligible radiation exposure when couriers carried radioactive materials by the handle of a standard bucket-type plastic specimen container.

The dose to nursing and ancillary staff

As radiation exposureis inversely proportional to the square of the distance from the source (inverse square law), exposure during LM/ SNB procedures for nursing staff and other staff, including anaesthetic and observing surgical staff, is considerably less than for the surgeon. Distances beyond 1-2 meters ensure that the radiation dose received by staff away from the operating table in these procedures is negligible.

The extremity dose to nursing staff was not measured in this study. However, as nursing staff routinely use long forceps to transfer radioactive specimens from the operating table into specimen containers for transport, and as the transfer time is very short, the dose would be negligible.

Dose to the nuclear medicine personnel

The major source of radiation exposure to nuclear medicine staff is from close proximity to the concentrated radionuclide prior to injection into the patient. Further potential risk occurs from leakage of radionuclide from the skin injection sites onto skin, clothing or bedlinen. Safe handling guidelines for radioactive materials are in routine use in nuclear medicine departments.

Our results demonstrate that LM/ SNB techniques are safe for surgeons, pathologists, radiologists and courier staff who are potentially at risk of radiation exposure from these sources and confirm the findings of the few studies in the literature.

Conclusions

Whole body dose to the surgeon, couriers, radiology staff and pathologist LM/ SNB is extremely low.

Some 200 (at maximum some 2000) breast cancer procedures to be performed per annum by a single surgeon before standard allowable limits are exceeded.

The extremity radiation exposure for melanoma LM/ SNB procedures appears much lower than for breast cancer LM/ SNB procedures.

Extremity doses for radiology, pathology, courier and other staff are much lower than for surgeons and are essentially negligible using safe work practices.

In conclusion, even though the radiation levels are low for LM/ SNB, good radiation handling practices should still be adopted in all cases during surgery, specimen transportation and pathological processing to ensure that the ALARA (As Low as Reasonably Achievable) Principle of radiation exposure is applied. In general terms, it is good practice for staff who are pregnant (or potentially so) to avoid any procedure involving irradiation, and exercise extra care to reduce avoidable radiation exposure risk (Table 3).

STAFF MEMBER DOSE PER CASE MAX CASE NO. *
Surgeon – melanoma case 0 mSv Effectively unlimited
Surgeon- breast case 250 mSv 200†
Radiologist 0 mSv Effectively unlimited
Pathologist (same day) 10 mSv 5000†
Pathologist (next day) 0 mSv Effectively unlimited
Medical Courier 0 mSv Effectively unlimited

Table 3: Estimated radiation safety limits for LM and SNB procedures for staff.

Some Basic Radiation Safety Recommendations for Lymphatic Mapping and Sentinel Node Biopsy

Although the radiation levels measured in this study were very low, detailed safety protocols should be in place and practiced to ensure that staff are exposed to the minimum levels of radiation in LM/ SNB procedures and tissue processing. If such procedures are followed lead aprons, which are often heavy, cumbersome and may produce skeletal strain, are not required.

Some basic information that could be incorporated into LM/ SNB protocols is given below:

Surgeon and ancillary staff

The dose to the surgeon’s hands can be effectively reduced by a factor of about 30 with the use of 20 cm forceps, artery clips, pencil electrocautery and other instruments, since radiation exposure varies inversely with the square of distance from the source.

Ensuring only a brief duration of contact or limited time spent within a 10 cm (4 inch) radius of the injection site can further reduce radiation exposure. Awareness of these points by staff can ensure the best practice during LM/ SNB procedures. If possible the primary tumour (which contains > 95% of the radioactive dose) should be removed as soon as possible after the sentinel node(s) have been excised, and radioactive materials should be placed > 2 metres away from staff.

Radiographer and radiologist

The dose to the radiologist and radiographer can be reduced by minimal direct handling and rapid processing of the specimen.

Pathologist and pathology technician

The dose to the pathologist can be substantially reduced if the pathology processing of the primary tumour is delayed for 24 hours, since radioactivity from 99mTc is halved every 6 hours (1/16 at 24 hr). Urgent specimens can be safely handled immediately using longhandled instruments and similar precautions to that used by surgeons. It is not necessary to delay processing of the sentinel node.

Specimen transport and laboratory personnel

The excised specimen should be placed in a container with a handle and with a prominently displayed radioactive substance label. The courier should carry the container using the handle, away from their body, and it should be transported expediently to the laboratory. Suitable protocols for handling of radioactive substances should be in place and actively practiced by laboratory staff to safely store and process pathological specimens.

Nuclear medicine and ward staff

Spillage and leakage of radioactive material from the injection site can be minimised by careful injection technique with the site draped and standard isotope handling protocols. Taping a swab over the injection site after completion of the isotope injection can contain any inadvertent leakage, preventing soiling of bed-linen. This technique also assists the surgeon in confirming that the agreed primary tumour site has been accurately injected.

Waste disposal

Linen, gauze swabs and other potentially low-level 99mTc contaminated material can usually be disposed of in the same way as other non-radioactive biological waste. Human tissue samples containing higher radioactivity can be processed, stored or disposed of in standard methods to that used for non-radioactive biological waste after 24-48 hours has elapsed from the initial radiopharmaceutical injection time – this time period permits adequate radioactive decay of 99mTc.

Acknowledgements

The authors thank Dr Tom Dodd, Consultant Pathologist, IMVS, Dr T Bessen, Radiology for assisting in coordination of the pathology and radiology studies respectively.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Recommended Conferences

Article Usage

  • Total views: 11711
  • [From(publication date):
    specialissue-2013 - Oct 24, 2017]
  • Breakdown by view type
  • HTML page views : 7946
  • PDF downloads :3765
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri, Food, Aqua and Veterinary Science Journals

Dr. Krish

agrifoodaquavet@omicsonline.com

1-702-714-7001 Extn: 9040

Clinical and Biochemistry Journals

Datta A

clinical_biochem@omicsonline.com

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

business@omicsonline.com

1-702-714-7001Extn: 9042

Chemical Engineering and Chemistry Journals

Gabriel Shaw

chemicaleng_chemistry@omicsonline.com

1-702-714-7001 Extn: 9040

Earth & Environmental Sciences

Katie Wilson

environmentalsci@omicsonline.com

1-702-714-7001Extn: 9042

Engineering Journals

James Franklin

engineering@omicsonline.com

1-702-714-7001Extn: 9042

General Science and Health care Journals

Andrea Jason

generalsci_healthcare@omicsonline.com

1-702-714-7001Extn: 9043

Genetics and Molecular Biology Journals

Anna Melissa

genetics_molbio@omicsonline.com

1-702-714-7001 Extn: 9006

Immunology & Microbiology Journals

David Gorantl

immuno_microbio@omicsonline.com

1-702-714-7001Extn: 9014

Informatics Journals

Stephanie Skinner

omics@omicsonline.com

1-702-714-7001Extn: 9039

Material Sciences Journals

Rachle Green

materialsci@omicsonline.com

1-702-714-7001Extn: 9039

Mathematics and Physics Journals

Jim Willison

mathematics_physics@omicsonline.com

1-702-714-7001 Extn: 9042

Medical Journals

Nimmi Anna

medical@omicsonline.com

1-702-714-7001 Extn: 9038

Neuroscience & Psychology Journals

Nathan T

neuro_psychology@omicsonline.com

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

John Behannon

pharma@omicsonline.com

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

social_politicalsci@omicsonline.com

1-702-714-7001 Extn: 9042

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords