Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Extreme Twiddling: Unusual Presentation of an Old Foe

Keywords
ICD; Twiddler’s syndrome; Intracardiac twiddling
Abbreviations
ICD: Implantable Cardioverter-Defibrillator; RV: Right ventricular; HV: High voltage; SVC: Superior vena cava; PA: Postero-anterior
Case Report
A 57-year old man underwent ICD implantation for resuscitated ventricular fibrillation in the setting of ischaemic cardiomyopathy in 2006. An active fixation single coil right ventricular [RV] lead [Medtronic 6945-65] was positioned at the RV apex and connected to a Medtronic Secura VR pulse generator before being placed into a sub-pectoral pocket [R wave 14mV, threshold 1.2V, impedance 1278 ohms]. Routine annual device interrogations were uneventful until 2011 when he was found to have rising thresholds with poor sensing [R wave 1.3mV, threshold 3.75V, High Voltage [HV] impedance 42 ohms]. PA chest X-ray did not show evidence of a macroscopic lead displacement or fracture. Lead repositioning was attempted but was unsuccessful due to a complete conductor break; a stylet would not pass beyond the mid portion of the lead, nor would the helix retract. The old lead was extracted and a new dual coil RV lead [Medtronic 6947-65] placed at the RV apex [R wave 12mV, threshold 0.5V, impedance 803 ohms, HV and SVC coil impedance 42 and 61 ohms respectively].
At routine device interrogation 1 month later there was complete failure to sense or capture. Again, the chest X-ray showed satisfactory lead position with no evidence of twisting or coiling and the patient denied any manipulation of the generator. At operation, the old lead was removed and a new dual coil RV lead was positioned at the RV apex [R wave 10.5mV, threshold 0.5 V, impedance 500 ohms, HV and SVC coil impedance 44 and 61 respectively] (Figure 1a).
1 month later HV and SVC impedances, which had always been within normal limits now both measured <20 ohms [R 9.9mV, threshold 2.75V, impedance 437 ohms]. Furthermore, chest X-ray demonstrated an abnormal lead appearance with decreased distance between the proximal and distal coils. The pins and header were orientated in the opposite direction to that seen on the post implantation X-ray (Figure 1b). Lateral chest X-ray confirmed localized intracardiac twisting of the RV lead (Figure 1c). The patient subsequently admitted to manipulating the generator. At operation the RV lead was untwisted, explanted and replaced. On this occasion a non-absorbable suture was placed through the fixation hole of the generator and fixed to the floor of the subpectoral pocket to help anchor the device. No further problems have since been encountered.
Discussion
Twiddler’s Syndrome is caused by deliberate or inadvertent rotational manipulation of the pulse generator. Radiographically it is characterized by twisting of the lead with associated displacement. Twisting of the proximal portion of the lead is a commonly reported observation, whilst discrete twisting of the intracardiac portion of the lead is a less frequent occurrence. A change in orientation of the pins and header on X-ray is a poorly described entity that is clearly depicted in our case. This radiological observation is seldom reported in the literature and has previously only been described in pacemaker recipients [1].
Since Bayliss first described Twiddler’s Syndrome in 1968 [2], additional variants of the condition including Reel and Ratchet Syndrome have been described [3-4]. All are recognized causes of pacemaker, ICD and biventricular ICD malfunction. Whilst classical Twiddler’s Syndrome results from rotational manipulation of the generator about its long axis and is characterized by twisting of the leads, Reel Syndrome results from rotation of the generator about its transverse axis and causes coiling of the leads within the pocket. Ratchet Syndrome stands alone; it is not dependent on rotational manipulation of the device [4]. Rather it results due to inadequate anchoring of the lead within the fixation sleeve, such that movement of the ipsilateral arm causes retraction of the affected lead with subsequent coiling in the pocket. Though the mechanisms differ, all three result in macroscopic lead displacement.
Patient-induced ICD malfunction was initially suspected but the radiographic hallmarks of Twiddler’s Syndrome and its variants were not evident. The delay in diagnosing the cause of ICD malfunction was multifactorial. Initial radiographical findings were subtle and did not suggest classical Twiddler’s Syndrome. Discernable twisting of the RV lead within the heart was not manifest until the final X-ray. A change in orientation of the header was central to confirming that the mechanism of lead failure was due to rotational manipulation of the pulse generator. Repeated rotation of the generator first one way and then back again may have been sufficient enough to cause lead damage without resulting in overt twisting or displacement of the lead. This may explain the initial ‘concealed’ presentation. Finally, continuous uni-directional rotation of the generator resulted in twisting of the intracardiac portion of the RV lead. This caused the 2 shock coils to move closer together and subsequent drop in HV and SVC coil impedances, which until that point had always measured within normal limits. These observations all infer that ‘fulminant’ twiddling did not commence until the later in the course of events when far more convincing signs of twiddling became apparent. Subpectoral placement of the pulse generator is described as a potential remedy for Twiddler’s Syndrome. However our case demonstrates that this does not prevent a determined patient. Securing the generator with an anchoring stitch and comprehensive directives on refraining from stirring the pocket may be just as influential in reducing the incidence of this patient-induced complication.
Acknowledgements
FZA has received research funding and sponsorship to attend educational meetings from Medtronic and Boston Scientific. AZ and SA have received research funding and sponsorship to attend educational meetings from Medtronic and St. Jude Medical. No funding was obtained for this current work. Competing interests: none declared.
 
References





Visit for more related articles at Journal of Clinical & Experimental Cardiology

Abstract

Characteristic twiddling and lead macro-displacement are well-documented radiological hallmarks of Twiddler’s Syndrome. What is less well known is that the different variants of Twiddler’s Syndrome can be distinguished by assessing orientation of the pins on X-ray. Illustrations of a change in orientation of the header and pins are rarely reported in cases of pacemaker Twiddler’s Syndrome and the phenomenon has not been described in implantable cardioverter-defibrillator recipients. We highlight an important classic finding that is seldom depicted in the literature

Keywords
ICD; Twiddler’s syndrome; Intracardiac twiddling
Abbreviations
ICD: Implantable Cardioverter-Defibrillator; RV: Right ventricular; HV: High voltage; SVC: Superior vena cava; PA: Postero-anterior
Case Report
A 57-year old man underwent ICD implantation for resuscitated ventricular fibrillation in the setting of ischaemic cardiomyopathy in 2006. An active fixation single coil right ventricular [RV] lead [Medtronic 6945-65] was positioned at the RV apex and connected to a Medtronic Secura VR pulse generator before being placed into a sub-pectoral pocket [R wave 14mV, threshold 1.2V, impedance 1278 ohms]. Routine annual device interrogations were uneventful until 2011 when he was found to have rising thresholds with poor sensing [R wave 1.3mV, threshold 3.75V, High Voltage [HV] impedance 42 ohms]. PA chest X-ray did not show evidence of a macroscopic lead displacement or fracture. Lead repositioning was attempted but was unsuccessful due to a complete conductor break; a stylet would not pass beyond the mid portion of the lead, nor would the helix retract. The old lead was extracted and a new dual coil RV lead [Medtronic 6947-65] placed at the RV apex [R wave 12mV, threshold 0.5V, impedance 803 ohms, HV and SVC coil impedance 42 and 61 ohms respectively].
At routine device interrogation 1 month later there was complete failure to sense or capture. Again, the chest X-ray showed satisfactory lead position with no evidence of twisting or coiling and the patient denied any manipulation of the generator. At operation, the old lead was removed and a new dual coil RV lead was positioned at the RV apex [R wave 10.5mV, threshold 0.5 V, impedance 500 ohms, HV and SVC coil impedance 44 and 61 respectively] (Figure 1a).
1 month later HV and SVC impedances, which had always been within normal limits now both measured <20 ohms [R 9.9mV, threshold 2.75V, impedance 437 ohms]. Furthermore, chest X-ray demonstrated an abnormal lead appearance with decreased distance between the proximal and distal coils. The pins and header were orientated in the opposite direction to that seen on the post implantation X-ray (Figure 1b). Lateral chest X-ray confirmed localized intracardiac twisting of the RV lead (Figure 1c). The patient subsequently admitted to manipulating the generator. At operation the RV lead was untwisted, explanted and replaced. On this occasion a non-absorbable suture was placed through the fixation hole of the generator and fixed to the floor of the subpectoral pocket to help anchor the device. No further problems have since been encountered.
Discussion
Twiddler’s Syndrome is caused by deliberate or inadvertent rotational manipulation of the pulse generator. Radiographically it is characterized by twisting of the lead with associated displacement. Twisting of the proximal portion of the lead is a commonly reported observation, whilst discrete twisting of the intracardiac portion of the lead is a less frequent occurrence. A change in orientation of the pins and header on X-ray is a poorly described entity that is clearly depicted in our case. This radiological observation is seldom reported in the literature and has previously only been described in pacemaker recipients [1].
Since Bayliss first described Twiddler’s Syndrome in 1968 [2], additional variants of the condition including Reel and Ratchet Syndrome have been described [3-4]. All are recognized causes of pacemaker, ICD and biventricular ICD malfunction. Whilst classical Twiddler’s Syndrome results from rotational manipulation of the generator about its long axis and is characterized by twisting of the leads, Reel Syndrome results from rotation of the generator about its transverse axis and causes coiling of the leads within the pocket. Ratchet Syndrome stands alone; it is not dependent on rotational manipulation of the device [4]. Rather it results due to inadequate anchoring of the lead within the fixation sleeve, such that movement of the ipsilateral arm causes retraction of the affected lead with subsequent coiling in the pocket. Though the mechanisms differ, all three result in macroscopic lead displacement.
Patient-induced ICD malfunction was initially suspected but the radiographic hallmarks of Twiddler’s Syndrome and its variants were not evident. The delay in diagnosing the cause of ICD malfunction was multifactorial. Initial radiographical findings were subtle and did not suggest classical Twiddler’s Syndrome. Discernable twisting of the RV lead within the heart was not manifest until the final X-ray. A change in orientation of the header was central to confirming that the mechanism of lead failure was due to rotational manipulation of the pulse generator. Repeated rotation of the generator first one way and then back again may have been sufficient enough to cause lead damage without resulting in overt twisting or displacement of the lead. This may explain the initial ‘concealed’ presentation. Finally, continuous uni-directional rotation of the generator resulted in twisting of the intracardiac portion of the RV lead. This caused the 2 shock coils to move closer together and subsequent drop in HV and SVC coil impedances, which until that point had always measured within normal limits. These observations all infer that ‘fulminant’ twiddling did not commence until the later in the course of events when far more convincing signs of twiddling became apparent. Subpectoral placement of the pulse generator is described as a potential remedy for Twiddler’s Syndrome. However our case demonstrates that this does not prevent a determined patient. Securing the generator with an anchoring stitch and comprehensive directives on refraining from stirring the pocket may be just as influential in reducing the incidence of this patient-induced complication.
Acknowledgements
FZA has received research funding and sponsorship to attend educational meetings from Medtronic and Boston Scientific. AZ and SA have received research funding and sponsorship to attend educational meetings from Medtronic and St. Jude Medical. No funding was obtained for this current work. Competing interests: none declared.
 
References





Figures at a glance

Figure Figure Figure
Figure 1a Figure 1b Figure 1c
Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Article Usage

  • Total views: 11614
  • [From(publication date):
    November-2012 - Oct 21, 2017]
  • Breakdown by view type
  • HTML page views : 7858
  • PDF downloads :3756
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri, Food, Aqua and Veterinary Science Journals

Dr. Krish

[email protected]

1-702-714-7001 Extn: 9040

Clinical and Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemical Engineering and Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001 Extn: 9040

Earth & Environmental Sciences

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

General Science and Health care Journals

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics and Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001 Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Informatics Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Material Sciences Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Mathematics and Physics Journals

Jim Willison

[email protected]

1-702-714-7001 Extn: 9042

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001 Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

John Behannon

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001 Extn: 9042

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords