Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

A Validated Stability?Indicating HPLC Method estimation of Clonazepam In the bulk drug and Pharmaceutical Dosage Form

Pallavi Mangesh Patil1*, Sagar Baliram Wankhede2 and Praveen Digambar Chaudhari1

1P.E.Society’s Modern College of Pharmacy Yamunanagar, Nigdi, Pune-411044 Maharashtra, India, 411044

2Padm. Dr.D.Y. Patil Institute of Pharmaceutical Sciences & Research, Pimpri, Pune-411018, Maharashtra, India

*Corresponding Author:
Pallavi Mangesh Patil
P.E.Society’s Modern College of Pharmacy
Yamunanagar, Nigdi, Pune-411044, India
Tel: +91 09823720695
E-mail: [email protected], [email protected]

Received Date: December 12, 2014; Accepted Date: January 27, 2015; Published Date: February 02, 2015

Citation: Patil PM, Wankhede SB, Chaudhari PD (2015) A Validated Stability–Indicating HPLC Method estimation of ClonazepamIn the bulk drug and Pharmaceutical Dosage Form. Pharm Anal Acta 6:332. doi: 10.4172/2153-2435.1000332

Copyright: © 2015 Patil PM, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Pharmaceutica Analytica Acta

Abstract

A rapid, accurate, linear, and sensitive RP-HPLC method has been developed and validated for estimation of Clonazepam in the bulk and Pharmaceutical Dosage Form. The chromatographic separation was performed on C18 Column (250 mm × 4.6 mm, 5 μm particle size) using a mobile phase Acetonitrile: Methanol (60:40 v/v) at flow rate of 1.0 ml/min and 30°C column temperature with the detection wavelength at 254 nm. The Clonazepam (RT 6.11 min).The linearity was performed in the concentration range of 5 to 25 μg/ml for Clonazepam R2 0.9993, Clonazepam. The percentage purity of Clonazepam was found to be 99-101%. Precision (Inertday) of the system was found to be 0.48% Clonazepam and all the method was found to be specific and found to be within the limits of the acceptance criteria. The limit of detection was 0.092 μg/ml and limit of quantification was 0.97 μg/ml. Clonazepam was found to be specific. The Proposed method has been validated for which were within the acceptance limit according to ICH guidelines. Forced degradation conditions of hydrolysis (neutral, acidic and alkaline), oxidation, photolysis and thermal stress, as suggested in the ICH guideline Q1A (R2). The drug showed instability in alkaline and oxide, while it remained stable in acid conditions.

Keywords

Clonazepam; Validation; Force degradation studies

Introduction

Clonazepam is a light yellow crystalline powder which is practically odourless. Literature survey states that Paroxetine hydrochloride and Clonazepam is official in IP [1], BP [2] with HPLC methods for the estimation individually. Clonazepam (CZ) in Figure 1 chemically, 5-(2-chlorophenyl)-7-nitro-2, 3-dihydro-1H-1, 4-benzodia zepin-2- one (Figure 1) belongs to the drug class benzodiazepines. Mechanism of action involves Allosteric interactions between central benzodiazepine receptors and gamma-minobutyric acid (GABA) receptors potentiate the effects of GABA. It’s chemical formula and molecular weight is C15H10ClN3O3 and 315.7 [3-8], HPLC [3], LCMS based methods have been reported. A UV and Spectrophotometric methods was reported for the estimation of Paroxetine and Clonazepam in combined formulations [9-10]. But no stability-indicating assay method has been reported for the simultaneous estimation of Clonazepam in the presence of their degradants using the ICH [11-13] approach of stress testing. Therefore, the present work was aimed to develop a simple, rapid, precise, and accurate isocratic reversed-phase stability indicating HPLC method for estimation of Clonazepam in the bulk drug and tablet dosage forms. The developed method was validated as per ICH Guidelines.

pharmaceutica-analytica-acta-Clonazepam

Figure 1: Chemical Structure of Clonazepam.

Materials and Methods

All the reagents as Acetonitrile and Methanol (HPLC grade), were purchased from Merck Chemicals, India. Reference standard Clonazepam was procured from Reddy Hyderabad laboratories, India as gift samples.

Instrumentation and chromatographic conditions

Different kinds of equipment like Analytical weighing balance, HPLC system (SHIMADZU-SPD 20A), Injector (Rheodyne, 20 μl), Sonicator, pH meter, vacuum filter pump, Millipore filtration kit, mobile phase reservoir. Grace C18 ACME 9000, C18 reverse phase column of 250 × 4.0 mm i.e., 5 μm dimensions Detector: UV, D2 lamp, 254 nm Column Temperature: Controlled room temperature (25°C) Injection: 20 μL sample loop was used for Clonazepam with the help of mobile phase consisted of a mixture.

Preparation of standard concentrations

Accurately weighed quantity (100 mg) of CZ respectively were transferred to 100.0 ml volumetric flask, dissolved and diluted up to the mark with mobile phase. From this solution, 5.0 ml was transferred to 50.0 ml volumetric flask and diluted to the mark with mobile phase (concentration 100 μg/ml) CZ respectively. The solution was mixed and filtered through 0.2 μ membrane filter.

Calibration standards CZ

From standard stock solution CZ (100 ug/ml) final Concentration was made in ranges 5-25 μg/ml.

Method validation

The validated HPLC method proposed according to ICH guidelines (ICH 1994, 1996). The following parameters were used for validation of the developed method.

Linearity: Linear relationship between peak area and concentration of the drugs were evaluated, making six measurements at concentration levels in the range of 5-25 μg/ml.

Accuracy: Recovery studies were carried out by spiking three different known amounts of pure drug (at 80%, 100% and 120% of label claim) to the pre-analyzed powder (standard addition method). Hence, an accurately weighed quantity of pre-analyzed tablet 10 mg CZ to the pre-analyzed.

Precision: The precision of the method was verified by repeatability and intermediate precision studies. Repeatability studies were performed by analysing the table sample six times at 100% of test concentration on the same day. The intermediate precision of the method was checked by repeating studies on three different days.

Limit of detection and quantitation: In order to estimate the limit of detection (LOD) and limit of quantitation (LOQ) linear were separately determined based on the standard deviation (σ) of the response and the slope (S) of the calibration curve and using the formula LOD=3.3 σ /S and LOQ=10 σ /S, the LOD and LOQ for CZ was estimated.

Robustness: To evaluate the robustness of the proposed method, small but deliberate variations in the optimized method parameters were done. The effect of change in flow rate and mobile phase ratio on retention time and tailing factor were studied. The solution containing 100 μg/ml of CZ was injected (in triplicate) into sample injector of HPLC three times under the varied conditions.

Stability

Stability in sample solutions

Sample solution of CZ (100 μg/ml) was prepared and was kept at room temperature (30 ± 20°C) protected from day light. The sample solution was assayed after 20 min, 1 h, 2 h, 4h, 8 h and 24 h and hence results of the remaining analysis times were compared with it. The percent label claim and the RSD values are shown in Table 1. The result of solution stability is stable for 24 hours.

Clonazepam (CZ)
Concentration(μg/mL)    Mean Peak  Area
0 0
5 344,337
10 4 52,756
15 4 226,664
20 8 518,328
25 8 79,1890

Table 1: Standard Calibration Table for Clonazepam.

System suitability

To ascertain resolution and reproducibility of proposed chromatographic system for estimation of CZ in Pharmaceutical dosage form, system suitability parameters like tailing factor (T), resolution (R) and column efficiency (number of theoretical plates, N) were studied. From stock solution D, appropriately diluted with mobile phase to obtain 100 μg/ml CZ. The diluted standard solutions were filtered through 0.2 μ membrane filter.

Force degradation studies

In order to evaluate the stability indicating property of the developed HPLC method stress studies were carried out under ICH recommended conditions. Intentional degradation was tried by exposing the tablet sample to following stress conditions: acid (0.1 N HCl at 60°C), base (0.1 N NaOH at 60°C), oxidation (3% H2O2 at 60°C), and heat (60°C), and UV light (254 nm). Ability of the proposed method to measure the analyte response in presence of its degradation products was studied.

Results and Discussion

Optimization of separation conditions

Different mobile phases containing Acetate buffer, Phosphate buffer, Methanol and Acetonitrile in different ratio, and various pH were tried and finally Acetonitrile: Methanol in the ratio (60:40 v/v) was selected as an appropriate mobile phase that resulted in good resolution and acceptable system suitability parameters for CZ was delivered at a flow rate of 1 ml/min with detection wavelength 254 nm for CZ, The injection volume was 20 μl. Analysis was performed at a temperature of 30°C, which shown in Figure 3.

Sample solution

The corresponding chromatograms were recorded and area of each peak for CZ was measured at 254.0 nm. Amount of CZ in sample (mg) was calculated by comparing the mean peak area of standard and sample solution.

Results of analysis of sample laboratory mixture are shown in Tables 2 and 3.

Level of recover Weight of tablet taken (mg) Amount of drug added (mg) CZ Amount of drug recovered (mg) CZ % Recovery CZ
8
8
8
10
10
10
12
12
12
18.01
18.21
18.00
19.11
19.13
19.17
22.02
22.10
21.97
100.01
100.27
100.02
99.91
99.95
99.97
100.02
100.10
99.95
80 % 10
10
10
100 % 10
10
10
120 % 10
10
10

Table 2: Results of Accuracy Studies.

Mean 100.18 99.94 100.04
S.D. ±0.5812 ±0.4178 ±0.219
 C.V. ±0.5810 ±0.4170 ±0.2222

Table 3: Results of Statistical Validation for Accuracy Studies.

System suitability

For system precision six replicate injections of standard solution were given, tailing factor (T), and column efficiency (number of theoretical plates, N) were studied. For CZ was recorded for each injection shown in Table 7.

Validation of the method

Linearity of response: A linear relationship was observed between peak area and concentration in the range of 5-25 μg/ml CZ respectively. The correlation coefficients for the calibration curve were found to be 0.9993, for CZ. Mean peak areas for CZ at selected wavelength are shown in Table 2 and Figure 2.

pharmaceutica-analytica-acta-chromatogram

Figure 2: Typical chromatogram of (RT 6.11 min).

Precision of the assay: Repeatability and reproducibility of the proposed method was determined by intra-day and inter-day precision studies. The Tablet was assayed three times on the same day (intra-day) and on three consecutive days (inter-day). The results of precision studies were expressed in terms of relative standard deviation (RSD) less than 2 of the percent label claim determined by developed method shown in Table 4.

Intra-day Precision
Drug % Mean* S. D. C. V.
CZ 99.98 ± 0.4189 0.4812
Inter-day Precision
CZ 99.91 ± 0.3714 0.3516

Table 4: Results of Precision.

Limit of Detection (LOD) and Quantification (LOQ): The limits of detection was 0.92 μg/ml for CZ and limit of quantification were 0.97 μg/ml for CZ respectively shown in Table 5.

Parameter CZ
Limit of Detection (μg/mL) 0.92
Limit of Quantification (μg/mL) 0.97

Table 5: LOD and LOQ of CZ.

Accuracy: The results of accuracy study are expressed in terms of percent recovery. The percent recovery at three levels (80 %, 100 % and 120 %) was found to be in the range of 98-102 % Results of recovery studies are shown in Table 2 and 3.

Robustness: To evaluate the robustness of the proposed method, small but deliberate variations in the optimized method parameters were done. The effect of change in flow rate and mobile phase ratio on retention time and tailing factor were studied. The solution containing 25 μg/ml of CZ was injected (in triplicate) into sample injector of HPLC 3 times under the varied conditions. Robustness data is given in Table 6.

  Chromatographic Changes
Factor Level Retention time Tailing factor
Flow Rate(mL/min)   CZ
6.12
6.11
6.14
6.13
± 0.2118
CZ
6.10
6.11
6.12
6.11
±0.2189
CZ
1.15
1.16
1.18
1.16
± 0.2412
CZ
1.17
1.15
1.16
1.16
±0.1446
0.9 –0.1
1.0 0
1.1 +0.1
  Mean
S.D.
Mobile Phase(±1mL changing Acetonitrile :  Methanol ) Level
59:39 – 1.0
60:40 0
61:41 +1.0
  Mean
S.D.

Table 6: Results of Robustness Studies

Stability: The stability evaluation in sample solutions (constituted with methanol) was performed up to 24 h. The sample solutions placed in the autosampler at room temperature were analysed periodically at 1, 2, 4, 8 and 24 h. The results are shown in Table 7. The peak areas of each drug were not considerably different from each other. Moreover, RSD value of each sample was <1%.

1. Tailing factor (T) 1.16
2. No. of theoretical plates (N) 7330
3. System precision (% RSD for peak area) 0.549
4. Method precision (% RSD for peak area) 0.358

Table 7: System Suitability Parameters.

The corresponding chromatogram was recorded and area of each peak for CZ at 254.0 nm. Amount of CZ in sample (mg) was calculated by comparing the mean peak area of standard and sample solution.

Assay of tablet: The resulting mixture of Tablet dosage form was centrifuged at 3500 rpm for 20 min. After centrifugation 20 μl of this mixture was injected into the chromatograph. The resulting solution was mixed and filtered through Whatman filter paper and to get approximate concentration and to obtain final concentration of 100 μg/ml CZ. The diluted solution was filtered through 0.20 μm filter.

The corresponding chromatograms were recorded and area of each peak for CZ was measured at 254.0 nm. Amount of CZ in sample (mg) was calculated by comparing the mean peak area of standard and sample solution are shown in Table 8.

Sr.no Weight of tablet taken CZ (mg) Amount of drug estimated CZ (mg) % Label Claim
1 100 99.98 99.97
2 100 99.97 99.96
3 100 100.02 100.01
4 100 100.13 100.12
5 100 99.86 99.85
6 100 99.72 99.71
  Mean 99.96
S.D. ±0.5242
C.V. ±0.5458

Table 8: Results of Analysis of Tablet.

Force degradation studies: In the forced degradation studies CZ were found to degrade under acidic (0.1 M HCL) and oxidative (3% H2O2) stress conditions employed. However it was found to be stable to the alkaline (0.1M NaOH) and unstable in photo (240 nm) degradation, thermal (60°C) and neutral conditions employed. The results for forced degradation studies are included in Tables 9 and 10. Typical chromatography obtained for CZ under different stress conditions are shown in Figures 4-8. The developed HPLC method could effectively resolve the drugs from their degradation products which confirm the stability indicating power of the developed method.

Sr. No. Stress Condition Percent assay of active substance (CZ) RT Value of degraded product
1. Acid (0.1 M HCl) 98.60 10.1
2. Alkali (0.1 M NaOH) 97.53 4.8,8.3,9.5
3. Oxide (3 % H2O2) 98.76 10.2
4. Heat (600C) 99.82  
5. UV (240nm) 99.98 --

Table 9: Results of Degradation Study.

Time Amount of drug estimated CZ % Label Claim* S.D. (±) % C.O.V.
20 min 99.98 99.88 0.376 0.452
1hr 99.78 99.67 0.311 0.389
2hr 99.04 99.89 0.378 0.392
4hr 100.00 10.00 0.408 0.455
12hr 100.10 10.07 0.455 0.478
24hr 99.98 99.91 0.377 0.394

Table 10: Results of stability.

pharmaceutica-analytica-acta-curve

Figure 3: Standard Calibration Curve for CZ.

pharmaceutica-analytica-acta-curve

Figure 4: Chromatogram of 0.1 M HCl Treated Tablet Sample.

pharmaceutica-analytica-acta-Tablet

Figure 5: Chromatogram of 0.1 M NaOH Treated Tablet Sample.

pharmaceutica-analytica-acta-Tablet

Figure 6: Chromatogram of H2O2 (3%) Treated Tablet Sample.

pharmaceutica-analytica-acta-curve

Figure 7: Chromatogram of Dry Heat Treated Tablet Sample.

pharmaceutica-analytica-acta-leaf

Figure 8: Chromatogram of UV Radiation Treated Tablet Sample.

Conclusion

The present HPLC method for determination of CZ was proved to be simple, rapid, precise, accurate and robust in their pharmaceutical dosage and validated as per ICH guidelines. Moreover, CZ was found stable in the sample solutions placed at room temperature up to 24 h. accordingly the proposed analytical procedure with detection time of 10 min can be used for reliable determination of CZ in bulk and tablet.

The Force degradation studies method was found to be simple, sensitive, selective, and suitable for determination of CZ in presence of its degradation products CZ is instable in alkaline and acidic and stable in thermal and photo light. Statistical analysis proved that the method is repeatable, reproducible, accurate and specific for the analysis of CZ. The developed HPLC method which confirms the stability indicates power of the developed method.

As the method in cost effective and less time consuming, thus, it can represent another good alternative for the already existing HPLC methods.

Acknowledgements

The authors are thankful to Reference standards of clonazepam were procured as gift samples from Dr.Reddys Laboratories, Hyderabad, India. Authors are also thankful to Modern college of Pharmacy, Nigdi Pune for carrying out some part of the analytical work.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Recommended Conferences

  • 7th International Conference on Phamaceutical Auditing
    June 08-09, 2018 Baltimore, USA
  • 4th International Conference on Drug Discovery, Designing and Development
    June 27-28, 2018 Vancouver, Canada
  • 18th World Pharma Congress
    October 18-19, 2018 Warsaw, Poland

Article Usage

  • Total views: 12550
  • [From(publication date):
    February-2015 - May 22, 2018]
  • Breakdown by view type
  • HTML page views : 8674
  • PDF downloads : 3876
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2018-19
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2018 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
Leave Your Message 24x7