Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on Medical, Pharma, Engineering, Science, Technology and Business
All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.

Calibration and Evaluation of an X-Ray Fluorescence Method for the Determination of Lead and Arsenic in Soils

Dana J DiScenza1, Alison R Keimowitz 2 and Neil Fitzgerald1*

1Department of Chemistry, Biochemistry and Physics, Marist College, Poughkeepsie, NY, USA

2Chemistry Department, Vassar College, Poughkeepsie, NY, USA

*Corresponding Author:
Neil Fitzgerald
Department of Chemistry
Biochemistry and Physics
Marist College, Poughkeepsie, NY, USA
Tel: (845) 575-3000 ext. 2491
E-mail: [email protected]

Received date: January 09, 2014; Accepted date: March 29, 2014; Published date: March 31, 2014

Citation: Di Scenza DJ, Keimowitz AR, Fitzgerald N (2014) Calibration and Evaluation of an X-Ray Fluorescence Method for the Determination of Lead and Arsenic in Soils. J Environ Anal Chem 1:103. doi: 10.4172/2380-2391.1000103

Copyright: © 2014 DiScenza DJ, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Environmental Analytical Chemistry

Abstract

A portable X-ray fluorescence method is described for the measurement of lead and arsenic in soils based on calibration with standard reference materials. Both elements exhibited good linear calibration curves. The method was used to analyze dried and ground soil samples obtained from the metropolitan New York City area. Results were compared to measurements obtained by graphite furnace atomic absorption spectrometry with microwave assisted acid leaching. No significant difference was found for lead measurements between the two methods however, arsenic values were significantly different with XRF exhibiting higher values. It is hypothesized that incomplete leaching and spectral interference are the predominant causes of this discrepancy. The XRF method provides a simple, quick, and nondestructive method for the determination of total lead and arsenic content in soils.

Keywords

X-ray fluorescence; Lead; Arsenic; Soil; Graphite furnace; Calibration

Introduction

X-Ray Fluorescence spectrometry (XRF) offers many advantages for environmental analysis of soils and sediments. XRF instruments require minimal sample preparation and are non-destructive, rapid, and portable. Numerous reports have been published demonstrating the application of XRF to the analysis of soils and sediments. A typical example was published by Chou et al. [1] who demonstrated the use of a portable XRF instrument for the assessment of lead and arsenic in the greater New Orleans area after Hurricane Katrina. Other relevant examples include the feasibility of using a portable XRF in the field for lead content in sieved soils [2], XRF analysis of arsenic contamination at an arsenic works site [3], and the application of portable XRF for heavy metal analysis of soils near abandon mines [4].

Disadvantages of XRF compared to routine laboratory instruments are the relatively high detection limits (generally in the part per million range) and inferior accuracy and precision. XRF also suffers from significant matrix effects which can make calibration problematic. Calibration methods for XRF include the use of matrix matched standards, adding an internal standard to the sample matrix, dilution to reduce matrix effects, and fusion of standards and samples with a suitable flux [5]. Calibration of modern commercial XRF instruments is often performed using calibration algorithms loaded into the instrument software. While this may be acceptable for in-situ monitoring work, it may not provide acceptable results compared to laboratory based methods. The United States EnvironmentalProtection Agency (USEPA) state precision values between 1.24 to 29.2% relative standard deviation for ground and dried soils, depending on the element, with recoveries for Standard Reference Materials between 90 and 110% [6]. Kristna et al. [7] compared XRF and Inductively Coupled Plasma Atomic Emission Spectrometry (ICP-AES) for the determination of heavy metals in soil samples. They used pressed powder pellets and empirical corrections to account for matrix effects in XRF and performed a total soil digestion with hydrofluoric acid prior to ICP-AES analysis. Using these methods good agreement was obtained. Kenna et al. [8] attempted to improve data quality of a field portable XRF by the use of soil and sediment Standard Reference Materials (SRMs) to calibrate the instrument. The reported measurements for Hudson River estuary sediment were in good agreement with independent laboratory analysis.

In this work we describe a variation of the calibration method described by Kenna et al. [8] using standard reference materials to calibrate a field portable XRF for the determination of lead and arsenic in soils obtained from a variety of locations in the New York City metropolitan area. The XRF method was compared to results obtained using Graphite Furnace Atomic Absorption (GFAA) after microwave assisted nitric acid leaching.

Materials and Methods

X-Ray fluorescence

X-Ray Fluorescence results were obtained on ground and dried soil samples obtained from the Rockaway Peninsula and Greenpoint, Queens, NY areas. The soils varied in composition from sandy to loamy. Lead and arsenic were measured using a field-portable XRF (Bruker Tracer III-V, Billerica, MA). Samples were dried at 30°C overnight, ground and placed into XRF sample cups. The XRF excitation source was set to 40 kV, at 25 μA under 5 Torr vacuum. Data was collected for 2 minutes for each sample. The L1 line at 10.55 keV was used for lead and Ka1 line at 10.54 keV for arsenic. Calibration was performed by duplicate measurements of National Institute of Standards and Technology (NIST, Washington DC) Standard Reference Materials (Table 1). Chemically pure silica was used as a blank. The XRF signal intensity was plotted against the certified value for each SRM in order to construct calibration curves.

SRM Certified Lead (mg/Kg) Certified Arsenic (mg/Kg)
1646a (Estuarine Sediment) 11.7 6.23
1944 (NY/NJ Waterway Sediment) 330 18.9
2586 (Trace Elements in Soil) 423 8.7
2711a (Montana II Soil) 1400 107
2781 (Domestic Sludge) 202.1 7.82

Table 1: NIST standard reference materials used for XRF calibration.

Graphite Furnace Atomic Absorption (GFAA)

Graphite Furnace Atomic Absorption (GFAA) spectrometry (Thermo M5, Walthan, MA) was used to compare measurements obtained using the XRF method. Instrument conditions are detailed in Table 2. Soil samples were prepared using a microwave digestion unit (CEM MARS, Charlotte, NC). Approximately 0.500 g of soil samples were combined with 10.00 mL of nitric acid (Fischer trace metal grade, Fairlawn, NJ) in microwave vessels. The microwave program involved heating to 175°C and holding at this temperature for 4 minutes 50 seconds. The vessels were left an appropriate time to cool before diluting with 18 MΩ water (Barnstead Nanopure, Lake Balboa, CA) to a volume of 100.00 mL. The determination of arsenic required a standard additions method while external standard calibration was used for lead. Arsenicstandards were prepared by appropriate dilutions of an atomic absorption standard solution (Ricca, Arlington, TX) with 10% nitric acid. Lead standards were prepared by dilution of a lead standard solution (Fisher AA standard, Fairlawn, NJ) with 10% nitric acid.

  Lead Arsenic
Wavelength (nm) 283.3 197.3
Slit Width (nm) 0.5 0.5
Matrix Modifier Ammonium Nitrate Nickel Nitrate
Ashing Temperature (°C) 800 1200
Atomization Temperature (°C) 1200 2600
Calibration Method External Standards Standard Additions

Table 2: Graphite furnace atomic absorption operating conditions.

Results and Discussion

Calibration curves for lead and arsenic using standard reference materials as calibration standards are shown in Figures 1 and 2. In both cases SRM1646a (Estuarine Sediment) produced high signals that caused significant influences on the regressions and were omitted as outliers. The Pearson correlation coefficients (r2 values) including SRM1646a were 0.9180 and 0.5162 for lead and arsenic respectively. In the absences of 1646a, they improved to 0.9764 and 0.9946 for lead and arsenic respectively.

environmental-analytical-chemistry-Calibration-curve-lead

Figure 1: Calibration curve for lead by XRF.

environmental-analytical-chemistry-Calibration-curve-arsenic

Figure 2: Calibration curve for arsenic by XRF.

In order to validate the methods, a comparison to an accepted method was performed. In our laboratory Graphite Furnace Atomic Absorption spectrometry after microwave pretreatment is the standard method for elemental analysis and generally accepted to produce accurate results. A paired t-test indicated no significance difference for lead measurements between the two techniques at 95% confidence. For arsenic, however, the paired t-test indicates a significant difference between the two techniques at 95% confidence with XRF values tending to be approximately double GFAA values. Figure 3 shows a comparison of XRF with GFAA for lead measurements. Ideally, if both techniques provide identical results, the data points would fit a linear regression line with a slope of one and a zero intercept. For lead results this is broadly true although XRF values tended to be higher on average than GFAA values. A similar trend is seen for arsenic with XRF measurements exhibiting significantly higher values than GFAA measurements in general (Figure 4). One soil value was removed as an outlier for lead measurements (r2 was 0.4731 including outlier). Significant scatter is observed in both graphs which may be an indication of inherently poor precision caused by sample inhomogeneity as well as spectral interferences on the XRF measurements. In this work duplicate measurements of samples were seen to vary by between 1.9% and 69% for lead and between 26% and 170% for arsenic highlighting precision issues. Goldstein et al. [9] also observed a positive bias for XRF arsenic measurements in soils compared to ICP data after a microwave acid leaching procedure similar to the method described here. They note that “the XRF technique is measuring total amounts of analyte, whereas standard EPA methods are measuring only the nitric acid-digestible part of the soil and not the matrix”. Akbulut et al. [10] noticed the same effect when comparing XRF measurement to ICPMS in soils. While they note that spectral interferences between arsenic and lead can affect accuracy and detection limits, they also state that incomplete leaching during the acid digestion process can lead to inaccurate results. Hassan et al. [11] studied microwave-assisted acid digestion methods and found that EPA method 3051, similar to the method used in this work, can result in recoveries as low as 47% for arsenic in reference materials. Recoveries were improved by modification of the method with the use of HCl but they note that disadvantages of using HCl may outweigh the advantages. It can be hypothesize that the higher arsenic levels in the standard reference materials used in this study to validate the GFAA method are more easily leachable by nitric acid than the arsenic strongly bound to the soil matrix in the samples. This is consistent with Goldstein’s conclusion that “XRF and EPA methods are not directly comparable for relatively pristine soils” however “in contaminated soils where most of the analyte is nitric acid-leachable, EDXRF and standard EPA methods should give more similar results”. Another explanation for the discrepancies is the well-known spectral interference of the lead line at 10.551 keV on the arsenic line at 10.543 keV. Although the spectral resolution of the instrument is sufficient to allow measurement of the arsenic in the presence of lead, spectral interference can be significant at higher lead concentrations. The calibration method used may account for interference effects to some extent by matrix matching, therefore it would be expected that spectral interference would have a greater influence on the precision than accuracy.

environmental-analytical-chemistry-Comparison-XRF-GFAA

Figure 3: Comparison of XRF to GFAA for lead.

environmental-analytical-chemistry-Comparison-XRF-GFAA

Figure 4: Comparison of XRF to GFAA for arsenic.

In conclusion, our data suggests that XRF using soil standard reference materials as calibration standards is an acceptable method for the determination of lead in dried and ground soils. XRF and acid leachable GFAA methods provide statistically similar results for lead but not for arsenic which we believe to be due to arsenic at relatively low concentrations tending to be strongly bound to the soil and not easily acid leachable. An alternative explanation is that it is an effect of uncompensated spectral interference although we believe this is more likely to affect precision than accuracy. Poor precision was observed for XRF measurements particularly for arsenic. Using soil standard reference materials offers a matrix matched calibration method and avoids needing to perform significant sample preparation or reliance on algorithms provided by the instrument manufacturer.

Acknowledgements

The authors wish to acknowledge Marist College, Vassar College, and individual contributions to a crowdfunding project on Rockethub for financial support. We wish to acknowledge Amy Mandigo (Marist College) for her help with the project.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Article Usage

  • Total views: 14087
  • [From(publication date):
    May-2014 - Jul 23, 2019]
  • Breakdown by view type
  • HTML page views : 10148
  • PDF downloads : 3939
Top