Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Comparative Analysis of Dose Variations in Tumor Volumes and Organs at Risk in IMRT Plans for Head-And–Neck, Pelvis and Brain Cancers with Varying Dose Calculation Grid Sizes

Pathak P1*, Guha S2, Vitwekar Vidya3, Majumder Sourav4 and Soni Ashok5
1Chief medical Physicist, BIMR Hospitals, Gwalior, India
2Radiation Oncologist, BIMR Hospitals, Gwalior, India
3Radiotherapy Technologist, BIMR Hospitals, Gwalior, India
4Diploma in Radiotherapy Technology, Radiotherapy Technologist, India
5Service Engineer, BIMR Hospitals, Gwalior, India
Corresponding Author : Pathak P
Chief medical Physicist, BIMR Hospitals, Radiation Oncology, Surya mandir road Gwalior
Madhya Pradesh 474004, India
Tel: 07512405715
E-mail: [email protected]
Received July 25, 2014; Accepted September 23, 2014; Published September 25, 2014
Citation: Pathak P, Guha S, Vidya V, Sourav M, Ashok S (2014) Comparative Analysis of Dose Variations in Tumor Volumes and Organs at Risk in IMRT Plans for Head-And–Neck, Pelvis and Brain Cancers with Varying Dose Calculation Grid Sizes. J Cancer Sci Ther 6:394-400. doi:10.4172/1948-5956.1000298
Copyright: © 2014 Pathak P, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Related article at
DownloadPubmed DownloadScholar Google

Visit for more related articles at Journal of Cancer Science & Therapy

Abstract

Purpose: The aim of this study was to compare the plan results that were obtained by using different calculation grid sizes ranging from 3 mm to 10 mm, and the same dose calculation algorithm Pencil Beam (PB), in Intensity Modulated Radiotherapy (IMRT) for different treatment sites Head-And –Neck, Pelvis (Carcinoma Cervix) And Brain Cancers. Introduction: Ever since the advent and development of treatment planning systems, the uncertainty associated with calculation grid size has been an issue. Even to this day, with highly sophisticated 3D conformal and intensity-modulated radiation therapy (IMRT) treatment planning systems (TPS), dose uncertainty due to grid size is still a concern. Materials and methods: Twelve patients in which four patients of Head-And –Neck, Pelvis And Brain tumors respectively were considered for the study. IMRT Plans were generated for a 6,600cGy, 5,000cGy & 5,400cGy prescribed doses for Head-And –Neck, Pelvis and Brain tumors respectively using Oncentra v 4.3 TPS. For each patient, dose calculation with Pencil Beam (PB) algorithms using dose grid sizes of 3.0 mm, 5.0 mm, and 10.0 mm were performed. Results: The plans were evaluated as per the ICRU guidelines and dose constraints were maintained as per the Quantec guidelines. The dose differences for the varying grid sizes in Tumor Volumes and Organs at Risk were analyzed and tabulated. Conclusion: Overall, the effect of varying grid size on dose variation appears to be insignificant. However, 3 mm is recommended to ensure acceptable dose calculations, especially in high gradient regions.

Keywords
Dose grid; 2D Array; Organs at risk; Intensitymodulated radiotherapy; Dose-volume changes; Head-and-neck cancers
Introduction
The benefit of intensity-modulated radiation therapy (IMRT) in the treatment of head-and-neck cancer (HNC) has been demonstrated in numerous studies [1-3]. Highly conformal radiation allows for a high dose to high-risk areas, whilst sparing adjacent organs at risk (OAR) such as the parotid glands. Clinical studies have shown that IMRT reduces grade-3 xerostomia comparison to threedimensional conformal radiotherapy (3D CRT) [4,5], for that reason IMRT has become the standard treatment in many centers. IMRT dose distributions, with steep dose gradients, are very sensitive to geometrical uncertainties, and hence, deviations between planned and delivered dose distributions have to be minimized. One way of improving the treatment accuracy is to reduce geometrical errors. Rigid errors, such as setup, have been extensively studied. Mechalakos et al. [6]. for instance evaluated the interfraction and interfraction errors in treatments of HNC and compared their results with previous studies from others authors. Margins are added to clinical volumes in order to take into account geometrical uncertainties. These planning margins are commonly calculated from measured systematic and random geometrical errors [7].
However, it is well known that many HNC patients treated with radiotherapy (RT) suffer significant anatomical changes due to tumor shrinkage or weight loss. Several scheduled rescanning studies have evaluated these volumetric changes in both target volumes and normal tissues [8-11], mostly on the parotid glands and their consequent effects on dose distribution [12-15].
The purpose of the present study was to analyze the variation on the dose distribution in Planning target volumes (PTVs) and organs at risk (OAR). The use of IMRT implies the irradiation of more OARs than conventional 3D CRT. Therefore, beside typical susceptible organs such as the eyes, optic nerves, optic chiasm, spinal cord, parotid glands, bladder, rectum and bowel we have also included additional OARs such as the brainstem, and femur head.
The IMRT technique has the potential benefit over conventional whole-pelvis irradiation of improving target dose coverage, reducing the volume of the organs at risk (OARs) that receive irradiation, and reducing the toxicity to normal tissue [16-19]. Despite the significant benefits of IMRT, there are some disadvantages. The technique usually requires multiple fixed-angle radiation beams, which can increase treatment delivery time. This has an impact on patient comfort, reproducibility of the treatment position, and intra-fraction motion. Moreover, IMRT uses a larger number of monitor units (MUs) compared with conventional conformal radiotherapy (CRT), leading to an increase in the amount of low-dose radiation received by the rest of the body. This raises the concern of secondary radiation-induced malignancy, which is of particular relevance to young patients or those with long future life expectancies [20-23].
In the past, whole-brain radiotherapy (WBRT) planning was simple. Today, new clinical and dosimetric considerations are taken into consideration when approaching such planning. It has been found that as many as 11% of patients who were treated by WBRT and survived more than 12 months developed dementia, especially with the use of a larger dose-per-fraction regimen [24]. However, regression of the lesions after WBRT was found to correlate with survival and improved neurocognitive function. Therefore, achievement of macroscopic lesion control is the mainstay of treatment. Thus, treatment-dose compromise is unjust for preserving these neurocognitive functions. Furthermore, memory functions were found to be most susceptible to early decline, even in patients with nonprogressing brain metastases [25]. These concerns became more significant as WBRT was instituted for prophylactic brain irradiation (PCI) for various neoplasms to decrease intracranial failure in patients with potential long-term survival [26].
In the case of IMRT, this is accomplished by using complex computer models to calculate the dose to a given volume. These volumes and their resolution of defined by the calculation grid, which defines the space where the dose calculation models are applied and the resolution of that space. However, the calculation grid has been generally left at a default value to minimize the amount of time that the treatment planning system needs to perform the dose calculations. The intent of the project is, therefore, to test the effect of very fine calculation grid resolutions on the accuracy of IMRT plans.
The intent of the study is, therefore, to test the effect of calculation grid sizes on the accuracy of IMRT plans.
Materials and Methods
C T acquisition and contouring
CT scans were acquired using a Somatom Power Spirit CT Simulator (Siemens) with 3–5 mm slice spacing. Patients were in the supine position and immobilized with a thermoplastic head–shoulder mask. A planning CT scan (CT) was acquired one week before RT treatment. The Oncentra version 4.3 (Nucletron) treatment planning system was used for delineation and dose distribution calculations. Target volumes and normal tissues were manually contoured by a physician on each axial slice of the CT using MRI or contrast-enhanced CT. The definition of volumes was in accordance with ICRU Reports 50-62, but dose-volume parameters were reported according to the new ICRU Report 83 IMRT recommendations. Gross tumor volume (GTV) included the primary tumor and affected lymph nodes. The GTV was expanded to include the high-risk regions (CTV).
To compensate for geometrical uncertainties such as setup and organ motion, a 5 mm margin was automatically added to CTVs to obtain the planning target volume (PTV). In order to avoid dose compensation in the build-up region, in cases with no skin infiltration, the PTVs were manually modified excluding areas where the distance to the skin was less than 3 mm. Although these modified PTVs were used during optimization process, the absorbed dose was reported over the whole PTV. Prescribed doses were 6,600cGy, 5,000cGy & 5,400cGy for HEAD-AND –NECK, PELVIS (Carcinoma Cervix), & BRAIN respectively.
The critical structures contoured were: the parotid glands, spinal cord, mandible, eyes, oral cavity, brainstem, brain, optic nerves, optic chiasm, bladder, rectum, bowel & femur heads.
Treatment planning
IMRT treatment plans were generated on the CT with nine 6 MV fields on the Oncentra treatment planning system. For each of the calculation grid sizes, three different sites; namely, Head -And- Neck, Cervix, and Brain were analyzed as shown in Figures 1(a) (b) (c), 2(a) (b) (c) and 3(a) (b) (c). The IMRT plans were optimized using an inverse planning algorithm. The final dose distribution was calculated using the Pencil Beam (PB) with heterogeneity correction and 3-10 mm grid resolution. Dose volume histograms were generated for each of the cases and statistical analysis performed included mean relative difference and Homogeneity Index for target structures. Comparison was done first by using 3mm calculation grid as a golden standard and keeping the same number of monitor units (MUs) per beam for each grid size, then the second part involved renormalizing plans to have the same target coverage (95% of the prescription dose covering at least 95% of the target volume) for each grid size used shown in Table 1 (a), (b) and (c).
Future study plans include their verification with the PTW 2D Array.
Optimization goals were as follows: 1) prescription doses (Dpres) must encompass at least 95% of target volumes; 2) near-minimum absorbed doses (D98%) of PTVs should be higher than 92% of Dpres; 3) the near-maximum absorbed dose (D2%) of the PTVs should be less than 110% of Dpres.
High priority constraints to normal critical structures were: no more than 1.0 cm3 of spinal cord could receive more than 46 Gy ; 2) no more than 1% of brainstem could receive more 54Gy; 3) the parotid gland volume receiving 26Gy should be less than 50% in at least one gland; 4) optic nerves Dmax should be less than 56Gy 5)optic chiasm Dmax should be less than 54Gy 6) Bowel 195cc should be less than 45Gy; 7) bladder Dmax should be less than 45Gy; 8) Rectum Dmax should be less than 50Gy; 9) D2% of normal tissue should be less than Dpres.
Low priority constraints that should not compromise target coverage were: 1) eyes Dmax should be less than 50 Gy.
Results
The maximum percentage of variation recorded between calculation grid sizes used was in the case of the Head and Neck treatments. For the Cervix and Brain cases there was little variation in the results based on the calculation grid size chosen. However head and neck cases with nodal involvement showed significant variation in the dosimetric results based on the grid size chosen as shown in Table 2 (a), (b) and (c). Overall results vary from case to case and also depend on the plan complexity. For larger treatment areas calculating with the grid size smaller than 3mm may be impossible as time needed for calculation rises exponentially with the field size involved. In gamma function tests, all grid sizes met the criteria of acceptability. (i.e., 95% of the region resulted in gamma index less or equal to 1 with a 3% dose difference and a 3 mm Distance to target agreement (DTA) criteria) except for deep target and 5 mm and 10 mm grid sizes where 95% of the region resulted in gamma index less or equal to 1 with a 5% dose difference and a 5 mm DTA criteria. It was observed that larger grid spacing produces higher dose gradient.
There are enduring uncertainties regarding the optimal dose grid resolution for use with pelvic intensity-modulated radiotherapy (IMRT) plans in which the adjacent organs at risk are slender and transect the field edge Table 1 (a), (b) and (c) shows target volume averaged dose parameters at CT with varying grid sizes for different sites viz. Head & Neck, Pelvis & Brain. Values are presented as a percentage of Dpres of PTV.
Conclusion
IMRT places a higher requirement on dose grid resolution than conventional radiation therapy. While 3 mm-5 mm grid was assumed adequate for conformal treatment planning, smaller dose grid is required at least in the areas of high dose. In the cases where steep dose gradients exist smaller grid size should be used while calculating and evaluating treatment plans, as the choice of the calculation grid size may in certain cases even influence clinical results. The statistical analysis showed that there were no significant differences in conformity & homogeneity except in some cases of 10 mm grid size IMRT plan as shown in Table 3 (a), (b) and (c). The CI for all the sites were 0.95 which shows that even if the grid sizes are chosen from 3 mm-10 mm the plans are conformal enough to be accepted for delivery ,but it requires little optimization for 5mm-10 mm . The average HI for Head & Neck, Pelvis & Brain are 0.135,0.1175,0.13 for 3 mm , 0.1475,0.12,0.14 for 5 mm & 0.165,0.1325,0.157 for 10 mm which shows that the plans are more homogeneous for 3 mm IMRT plans with respect to 5 & 10 mm IMRT plans .Thus 3 mm is recommended to ensure acceptable dose calculations, especially in high gradient regions.
P value and statistical significance
The two-tailed P value is less than 0.001 for target coverage.
By conventional criteria, this difference is considered to be extremely statistically significant.
References


























Tables and Figures at a glance

Figure Figure Figure
Table 1 Table 2 Table 3

 

Figures at a glance

Figure Figure Figure
Figure 1 Figure 2 Figure 3
Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Recommended Conferences

Article Usage

  • Total views: 11567
  • [From(publication date):
    October-2014 - Jun 22, 2017]
  • Breakdown by view type
  • HTML page views : 7809
  • PDF downloads :3758
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords