Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Design of Checking Phantom for Metrological Certification Tests of 600 L Rectangular Container Monitoring System

Alojz Slaninka1,2*, Ondrej Slávik1 and Vladimír Nečas2

1VUJE, Inc., Okružná 5, SK-91864 Trnava, Slovak Republic.

2Slovak University of Technology, Institute of Nuclear and Physical Engineering, Ilkovičova 3, SK-81219 Bratislava, Slovak Republic.

*Corresponding Author:
Slaninka A
Slovak University of Technology
Institute of Nuclear and Physical Engineering
Ilkovičova 3, SK-81219 Bratislava, Slovak Republic
Tel: +421 33 5991462
E-mail: [email protected]

Received Date: September 02, 2016; Accepted Date: August 08, 2016; Published Date:August 15, 2016

Citation: Slaninka A, Slávik O, Nečas V (2016) Design of Checking Phantom for Metrological Certification Tests of 600 L Rectangular Container Monitoring System. Int J Waste Resour 6:241. doi: 10.4172/2252-5211.1000241

Copyright: © 2016 Slaninka A, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at International Journal of Waste Resources

Abstract

New monitoring system for purpose free release of materials into the living environment with a 600 L container counting geometry was designed and developed. Considering that monitoring system is classified as a dedicated measuring instrument the metrological certification is necessary. For purpose of metrological qualification tests the checking verification container (phantom) was proposed and manufactured. The arrangement of the reference sources in the volume of checking container was designed so that it approximates evenly activity distribution considered at efficiency calibration. Description of monitoring system, measurement uncertainty assessment, design of checking container and results of metrological qualification tests are in more detail discussed. Highlights • The procedure for metrological certification of the gamma spectrometry monitoring system based on the twin HPGe detectors with 600 L rectangular container counting geometry was proposed. • The checking container (phantom) containing 24 holes for rod calibration sources evenly mounted through the container in direction of detectors axes was designed and manufactured. • Proposed arranged of holes for rod sources approximate the homogenous distribution of the activity in container volume that is expected during usual measurement.

Keywords

Container monitoring system; Gamma spectrometry; HPGe full-energy peak efficiency; Metrological certification; Massic activity; Measurement uncertainty

Introduction

Important part of decommissioning project of nuclear power plant A1 in Slovakia is free release of materials into the environment. For purpose of the monitoring capacity increasing during free release of materials, mainly soil and concrete rubble, the monitoring system of 600 L containers was recently developed and delivered. Using counting geometry of 600 L rectangular container that is equipped with self-discharger is able to increase the total monitoring capacity. The monitoring system is able to monitor a material also in standard counting geometry of 200 L drum.

According to the Slovak legislation (Metrological Act Nr. 142/200 Coll.) a metrological certification is required because the monitoring system is considered as dedicated measuring instrument.

Description of the Container Monitoring System

Monitoring system is designed to activity monitoring of radionuclides emitting gamma radiation contained in loose materials for purpose of free release. The two gamma spectrometry chains with electrically cooled HPGe detectors of type GC 3020 are the cores of the monitoring system. Detectors are posited in the reconfigurable rectangular shielding. Horizontally or vertically arrangements of the detectors is used for container or drum monitoring, respectively (Figure 1). The lead shielding of 50 mm thickness and the aperture of shielding collimator ensure that the volume of container or drum only is in view angle of detector during the measurement. Basic technical parameters of counting geometry are noted in the Table 1.

international-journal-waste-resources-Detectors-shielding-arrangement-rack

Figure 1: Detectors and shielding arrangement on rack during container measurement .

PARAMETER DRUM CONTAINER
Collimation angle 53° 53°
Lead shielding thickness 5 cm 5 cm
Distance between axes of detectors 50 cm 60 cm
Distance: detector – vessel envelop 25 cm 30 cm
Weight range 100 ÷ 400 kg 300 ÷ 1000 kg
Vessels dimension Ø60 Ã 80 cm 113.5 Ã 73.5 Ã 71 cm
(l à w à h, inside)

Table 1: Basic technical parameters of counting geometry.

All components of the monitoring system are integrated into the transportable ISO container (5680 × 2330 × 2190 mm; l × w × h) that is divided to measuring and control room. Integrated air condition unit ensures constant environmental conditions. Containers or drums are during measurement placed on the integrated electronic tensometry scale for material weight and density determination (Figure 2). Monitoring of a 600 L container is performed in two steps; firstly one side is monitored and after turning round of container the second side is monitored. Bottom discharger of the container decrease a manipulation time and increase the monitoring productivity. In case of drum monitoring the measurement post is equipped with rotating table ensuring drum rotation during the measurement.

international-journal-waste-resources-Verification-container-measuring-position

Figure 2: Verification container on the measuring position.

Control PC unit includes spectrometry, evaluation and control SW for acquired spectra evaluation, controlling particular integrated devices and protocol results report. During spectra evaluation the homogenous activity distribution is checked on the basis comparison of the responses form particular detectors.

Full-energy peak (FEP) efficiency was determined by calculation based on detector characterization and designed counting geometry by means of ISOCS calibration software [1-3]. This one is standard calibration method based on MCNP and empirically determined point responses of spherical volume up to distance 500 m from detector. During FEP efficiency calculation the sample and shielding parameters, activity distribution, attenuation coefficients of particular materials and other relevant factors are taken into the account. Basic metrology features of monitoring system are shown in the Table 2.

Designation: Measurement of massic activity [Bq/kg] of homogenously
contaminated soil or concrete debris loaded in 600 L
container or 200 L drum.
Declared
energy range of
measurement :
200 keV - 2 MeV (gamma);
RNs emitting  gamma ray: 137Cs (dominant contaminant),
134Cs, 60Co, 110mAg, 54Mn, etc. are being identified
Method of
calibration:
Full-energy peak (FEP) detection efficiency has been
calculated mathematically, by ISOCS code on the basis
ISOCS/LabSOCS characterization of used detectors and
parameters of counting geometry [1].
Density range
of monitored
material:
0.5 â 2.0 g/cm3
Full-energy peak detection efficiency polynomial curves for
various densities and both counting geometries have been
determined in advance by step of 0.025 g/cm3. Proper curve
is selected automatically on the basis determined weight
(density) of container or drum.
Acquisition
time:
10 min (rotating drum), 2 Ã 10 min (both sides of container
are monitored)
Minimum
detectable
activity:
20 Bq/kg (drum) and 10 Bq/kg (container) of 137Cs at
standard acquisition time and at radiation background  0.1
μSv/h background of 137Cs
Expected
capacity of
monitoring:
10 containers / shift or 20 drums / shift
(20 containers / 12 hour working shift)

Table 2: Basic features of monitoring system.

Measurement Uncertainty Assessment

Measurement uncertainty of the massic activity (Am) was assessed by determination of the uncertainties of particular input quantities entering into the Am calculation according to equation (1).

Am = A/m = N/(m*y*ε*t)      (1)

In compliance with standard law of uncertainty propagation the relative standard combined uncertainty of massic activity, u(Am), can be expressed by equation (2). This approach is more detail descripted in [4]. Meaning of the members of the right side of equations (1) and (2) as components of uncertainty is explained in the Table 3.

u(Am)2 = u(N)2+u(m)2+u(y)2+u(ε)2+u(t)2     (2)

Source of uncertainty, notes Counting geometry
DRUM CONTAINER
u(m) – uncertainty of weight determination, m; measurement error of scale is ± 0.2 kg 1.0% 1.0%
u(N) – uncertainty of net peak area estimation, N; results from  acquisition time and count rate on the free release level 300 Bq/kg of 137Cs 5.0% 5.0%
u(ε) – uncertainty of full-energy peak efficiency, ε; in more detail discussed in the Table 4 10.9% 13.5%
u(t) – uncertainty of the spectra acquisition time, t neglectable
u(y) – uncertainty of abundance, y 3.5% 3.5%
u(Am) – relative standard combined uncertainty of massic activity Am 11.8% 14.9%

Table 3: Relative standard combined uncertainty of massic activity and its components.

Uncertainty arising from the FEP efficiency determination, u(ε), as most significant source of uncertainty has been assessed separately. Uncertainty of the used calibration method (ISOCS [1]) as well as the uncertainties of input parameters defining the counting geometry and entering into the FEP efficiency calculation were taken into the account. Most significant components of the FEP efficiency uncertainty are provided in the Table 4.

Source of uncertainty Counting geometry
DRUM CONTAINER
uISOCS(ε) –  uncertainty of calibration method ISOCS; estimated by manufacturer during detector characterization process 6.0% 6.0%
uρ(ε) – uncertainty as a consequence of the deviation between real value of density and calibration value of density, results from decelerated energy range and sample density range according to the Table 2 5.0% 5.0%
umi(ε) – uncertainty as a consequence of the difference between the real material composition and calibration material composition; soil and concrete rubble is considered only 3.6% 3.5%
uG(ε) – geometry uncertainty as a consequence of deviation between the real measurement geometry and the calibration geometry; the change in sample position of 3 cm and 2 cm in case of the container and drum, respectively, in various directions was considered 3.5% 3.3%
uHom(ε) – uncertainty as a consequence of the inhomogeneous distribution of activity within the volume of drum or container 5.6% 9.9%
uS(ε) – uncertainty of long-time stability of efficiency 1.0% 1.0%
u(ε) – combined standard uncertainty of full-energy peak efficiency 10.9% 13.5%

Table 4: Relative standard combined uncertainty of full-energy peak efficiency and its most significant components.

Curve of relative FEP efficiency in dependence on sample density is shown on the Figure 3. It can be seen that the FEP detection efficiency of low energies is significantly lower. In additionally the sensitivity of the FEP efficiency to sample density change is most significant in low range of density. Consequently the measurable gamma energy range and sample density range was declared according to the Table 2.

international-journal-waste-resources-Curve-relative-FEP-efficiency

Figure 3: Curve of relative FEP efficiency in dependence on the sample density for gamma rays of various energies in [keV]. 1 on the y-axis corresponds to the FEP efficiency of 661.65 keV gamma rays at 1.0 g/cm3 density of sample at drum counting geometry.

Homogenous activity distribution was considered at FEP detection efficiency calculation. Therefore one of important but hard to quantify valuable source of uncertainty is potential inhomogeneous distribution of activity within the volume of container or drum, uHom(ε). It is a consequence of uneven detector response distribution from measured volume of sample because the volume of container is too large. Half of the container width (about 35 cm) presents important absorption layer. Therefore, a low inhomogeneity in sample center or close to detectors may significantly affect the measurement results. For this reason the preliminary sorting measurement and both-sides container measurement is established to decrease the uncertainty resulting from inhomogeneous activity distribution. Also container and detectors positions arrangement was proposed so that the dependency of FEP detection efficiency along the container was as even as achievable. Dependency of the FEP detection efficiency along particular directions of the container (length, height and width) is on the Figure 4.

international-journal-waste-resources-Distribution-FEP-efficiency

Figure 4: Distribution of the FEP efficiency (661.65 keV) along container length, height and width for one and pair of detectors and for one and both sides monitoring; simulated by dividing the container volume to 10 parallel layers. 1 on the y-axis referring to average FEP efficiency at homogenous distribution of activity in container at 1.0 g/cm3 density of sample.

Proposal of Container Monitor Metrological Testing Procedure

Proposal of metrological testing procedure was based on the checking phantom of the container filled by non-active material of the same kind as to be routinely monitored and with reference sources inserted into the container volume. For determination the arrangement of reference sources in the container volume the following requirements were taken into the account:

Minimization of the number of reference sources needed with regard to their price,

Simple way of insertion and removal to/from the container,

Minimization of the deviation of FEP efficiency for phantom container from the FEP efficiency for real container with homogeneous activity distribution.

Using of reference sources of rod shape for easy inserting into the container was proposed. For determination of the direction of the rod sources inserting into the container the distribution of FEP efficiency alongside particular direction of container was taken into the account. From the Figure 4 it can be seen that the change of FEP efficiency is most sensitive in direction of container width, i.e. in the direction of detectors axes due to increasing self-absorption effect. This direction was proposed for the rod sources inserting. Arrangement of the rod sources in direction perpendicular to the detectors axes (along length or height of the container) is inappropriate because small position change of a source in the container volume in the detector vicinity means significant change of FEP efficiency (response of detector. On the contrary, the presence of source in the container middle does not affect the measurement results because the detector response from this source is negligible.

Positions of the rod sources in the container volume were determined so that the container volume was divided to regular segments of the same volume with rod source in the center of all segments. By ISOCS simulation calculation has been shown that for 24 rod sources posited in the container volume (4 columns and 6 rows) according to Figure 5 and for density of material 1.2 g/cm3 the value of FEP efficiency of 662 keV differs from case of homogeneous distribution less than 5%. This model was selected to the manufacturing. For comparison, model with 60 rod sources (10 columns and 6 rows) means deviation of FEP efficiency from homogeneous distribution on level 1%, but the number of reference rod sources is 3 times higher.

international-journal-waste-resources-3D-design-verification-container

Figure 5: 3D design of the verification container phantom with 24 reference rod sources throught full width of container; container is divided to 24 regular volume segments and the rod sources are posited in the centre of each segment; dimensions of the container are in [mm].

The container phantom was manufactured by modification of the standard container. Into the rod source positions were embedded the plastic tubes (Ø22 mm) with closeable apertures for rod sources inserting. Volume of container was filled by lime stone rubble of fraction 4–8 mm. Density of this material 1.3 g/cm3 was determined by the phantom container weight measurement.

The standard calibration drums of know density with 6 apertures for reference rod sources inserting were used for metrological certification of drum counting geometry. The distances between the openings and drum axis are designed in such a way as to represent an even activity distribution in the drum volume during its rotation.

Results of Metrological Certification Tests

Metrological certification tests were performed by Slovak Institute of Metrology by help 6 rod sources of 137Cs and 152Eu. During the measurement the rod sources were placed into the apertures of one column and sequentially moved into next columns. Counting time for each sources position was the same.

In addition, tests of particular detector response were performed using standard test samples to verify soundness of ISOCS calibration method. An ampoule containing mixture of radionuclides 241Am, 85Sr, and 88Y was placed 13 cm in front of detector head in its axis. In all cases the test results were satisfactory. Results of metrological tests are provided in the Table 5.

RN Counting geometry Relative deviation of
measured results from
the reference value
152Eu Drum (ρ = 1.5 g/cm3), 6 rod sources in drum 3.3 %
152Eu Drum (ρ = 0.8 g/cm3), 6 rod sources in drum 2.5 %
137Cs Drum (ρ = 1.5 g/cm3), 48 ampules merged into 6 rod sources 11.6 %
137Cs Drum (ρ = 0.8 g/cm3), 48 ampules merged into 6 rod sources 5.6 %
152Eu Container (ρ = 1.3 g/cm3), 48 ampules merged into 6 rod sources -4.5 %
137Cs Container (ρ = 1.3 g/cm3), 48 ampules merged into 6 rod sources 8.3 %
85Sr Ampule, 1st detector 5.5 %
88Y Ampule, 1st detector -2.9 %
241Am Ampule, 1st detector -4.1 %
85Sr Ampule, 2nd detector 8.0 %
88Y Ampule, 2nd detector -4.1 %
241Am Ampule, 2nd detector -15.3 %

Table 5: Results of metrological qualification tests for particular counting geometry.

Conclusion

For the free release monitoring capacity increasing at NPP A1 in Slovakia the monitoring system using counting geometry of 600 L rectangular container equipped with bottom discharger was developed. Considering the purpose of monitoring the metrological certification was necessary. Due to non-standard counting geometry of rectangular container the special phantom container with pipe openings for reference rod sources inserting and filled up by lime stone gravel was developed and manufactured for metrological certification purpose.

The rod reference sources are posited in direction of detectors axes (along width of container) in regular grid. It has been shown that 24 rod sources is sufficient number for even activity distribution simulation at container dimensions 1135 × 710 mm (length x height). The metrological certification was successfully carried out with the 20% class of accuracy achieved.

For the free release monitoring capacity increasing at NPP A1 in Slovakia the monitoring system using counting geometry of 600 L rectangular container equipped with bottom discharger was developed. Considering the purpose of monitoring the metrological certification was necessary. Due to non-standard counting geometry of rectangular container the special phantom container with pipe openings for reference rod sources inserting and filled up by lime stone gravel was developed and manufactured for metrological certification purpose.

The rod reference sources are posited in direction of detectors axes (along width of container) in regular grid. It has been shown that 24 rod sources is sufficient number for even activity distribution simulation at container dimensions 1135 × 710 mm (length x height). The metrological certification was successfully carried out with the 20% class of accuracy achieved.

Acknowledgment

This project has been partially supported by the Slovak Grant Agency for Science through grant VEGA 1/0796/13.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Article Usage

  • Total views: 8447
  • [From(publication date):
    September-2016 - Nov 18, 2017]
  • Breakdown by view type
  • HTML page views : 8331
  • PDF downloads : 116
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords