Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Determination of Drugs and Metabolites in Raw Wastewater Using Liquid Chromatography-Mass Spectrometry

Klein DM*, Pandey G, Blackwell B and Sperry K

Department of Environmental Toxicology, Institute of Environmental and Human Health, Texas Tech University, USA

*Corresponding Author:
David Klein, Ph.D.
HCLD, Associate Professor
Department of Environmental Toxicology
Institute of Environmental and Human
Health, Texas Tech University
1207 South Gilbert Drive, USA
Tel: 806-834-6044
E-mail: [email protected]

Received date: December 27, 2014; Accepted date: January 20, 2015; Published date: January 25, 2015

Citation: Klein DM, Pandey G, Blackwell B, Sperry K (2015) Determination of Drugs and Metabolites in Raw Wastewater Using Liquid Chromatography-Mass Spectrometry. J Forensic Res 6:268. doi: 10.4172/2157-7145.1000268

Copyright: © 2015 Klein DM, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Forensic Research

Abstract

A previous study in Lubbock, TX detected the cocaine metabolite benzoylecgonine in wastewater entering the municipal wastewater treatment plant. This work was conducted using chemical derivatization followed by gas chromatography-mass spectrometry (GC-MS) determination. This is a report using liquid chromatography-mass spectrometry (LC-MS) for determination of drug metabolites isolated from wastewater in Lubbock, TX by solid phase extraction (SPE). Using the LC-MS method, chemical derivatization is eliminated and multiple analytes can be determined in a single run. The results obtained showed that several compounds from different drug classes were detected. Compounds observed included cocaine, heroin, tetrahydrocannabinol, amphetamine, and benzodiazepine derivatives.

Keywords

Sewage epidemiology; Liquid chromatography-mass spectrometry; Drugs; Pharmaceuticals; Metabolites; Wastewater.

Introduction

The abuse of drugs, both legal and illegal, is a worldwide problem [1]. Several different approaches to characterizing drug use have been proposed. These include population surveys, checks and updates of criminal records, and consumer interviews. One method that has been reported to help track drugs and their metabolic products is the analysis of wastewater [2].

Zuccato et al. has termed this approach ‘sewage epidemiology.’ Using this method, it is possible to gain information about the range of substances that might be a problem in a community. It is also possible to develop some temporal information about when drugs are taken by sampling wastewater at particular intervals. Sewage epidemiology has been used in several countries [3-5].

A previous study was undertaken to examine the sewage epidemiology approach in Lubbock, TX [6]. The cocaine metabolite, benzoylecgonine was determined by gas chromatography – mass spectrometry (GC-MS). However, in this study only a single compound was monitored because it had to be isolated from the wastewater, purified, and chemically derivatized in order to be amenable to GCMS determination. Thus, GC-MS limits the number and type of compounds that can be included in a study.

The purpose of this study was to demonstrate the use of an ion trap based liquid chromatography-mass spectrometer (LC-MS) for the determination of drugs and drug metabolites in wastewater from Lubbock (TX). Unlike GC-MS, LC-MS does not require chemical derivatization of drug compounds. The use of LC-MS for the determination of pharmaceuticals products in wastewater has been previously reported [7]. In that case, several different classes of compound were determined including cocaine, opiates, cannabinoids and amphetamine compounds. Another advantage of LC-MS is that both parent compounds and metabolites can be determined in the wastewater samples.

This study employed an ion trap LC-MS. This is one of the least expensive configurations to use for this investigation. While other types of mass spectrometer may have greater sensitivity, the cost of these systems is prohibitive for most law enforcement groups. Further, the information derived from this study was used in forensic study, not the treatment and removal of the compounds where water may be further purified for reuse as drinking water. This allowed the work to be done on the least expensive platform and still have adequate sensitivity to detect compounds of interest.

In Lubbock (TX) the LC-MS was method was used to determine several classes of drugs and metabolites found in wastewater with a single isolation and purification scheme that does not require chemical derivatization [8]. By working with the local forensic science community, it was possible to select several different classes of compounds that were of interest [9]. This study is an investigation of representative compounds to determine which if any were present in the Lubbock (TX) wastewater. From this work we are expanding the concept to use in other law enforcement situations such as prisons where monitoring of wastewater would provide rapid information about the use of illegal drugs in this situation.

The use of LC-MS to monitor drugs of abuse in wastewater provides a clear way to have forensic evidence obtained. The method requires only minimal sample clean-up and rapid determination of compounds of interest.

Materials and Methods

All drug standards and reference compounds (Table 1) were purchased from Cerilliant (Austin, TX). Standards were stored at -20°C. Methanol (LC-MS grade) was purchased from Honeywell Burdick & Jackson (Morristown, NJ) and acetic acid was purchased from Fisher Scientific (Fair Lawn, NJ). 18 MΩ water was purified using a Barnstead system. A Kinetex® 5 μC-18 100 Å column (150 mm × 4.6 mm) was used for separation of analytes. Strata X® 33 μ polymeric reversed phase (200 mg/6 mL) solid phase extraction (SPE) cartridges were used to separate the analytes from the sample matrix. The column and SPE cartridges were obtained from Phenomenex (Torrance, CA).

Analayte Structures Metabolites and deuteratedstandards Mol. Wt. (g/mol) CAS number
Amphetamines image   135 300-62-9
THC image THC 314 1972-08-3
THC-OH (11-Hydroxy-Δ9-THC) 331 34675-49-5
THC-OH-d3 (11-Hydroxy-Δ9-THC-d3) 333 362044-74-4
Benzodiazepine image Benzodiazepine 144 12794-10-4
diazepam 285 439-14-5
Lorazepam 321 846-49-1
Flurazepam 388 17617-23-1
Alprazolam 309 28981-97-7
Methamphetamine image Methamphetamine 149 4846-07-5
MDMA 193 42542-10-09
diacetylmorphine (Heroin) image diacetylmorphine (Heroin) 369 561-27-3
Morphine D3 288 67293-88-3
Cocaine image Cocaine 303 50-36-2
Benzoylecognine 289 519-09-5

Table 1: Drug standards and reference compounds, Note. Mol. Wt.=Molecular weight; CAS number=Chemical Abstracts Service registry number.

Samples were obtained from the Lubbock Wastewater Treatment Plant (LWWTP). This is the only site in the city of Lubbock for wastewater treatment. Both weekday and weekend samples were collected and all samples were obtained in the morning using amber glass collection bottles. The collected samples were adjusted to pH 2 using 37% Hydrochloric acid, and were stored in dark at 4°C until they were processed. It was found in a previous study that pH adjustment was necessary to prevent degradation of the analytes during storage; pH adjusted samples were found to be stable for one month [6].

The wastewater samples were held stable in order to allow particulate matter to settle to the bottom of the container. The water was then decanted and 500 mL samples were filtered using 1.2 μ glass filters (Whatman). SPE was conducted using cartridges conditioned with 6 mL methanol followed by 6 mL water. The filtered samples were applied to the cartridges and the cartridges were washed with ultrapure water and subsequently dried with N2 for 30 min. Elution of the compounds was performed with 6 mL of methanol. The eluate was evaporated to near dryness under a gentle stream of N2. The residue was re-dissolved in 1 mL of H2O/Methanol (9/1; v/v) containing 0.1% acetic acid. 100 ng of deuterated standard (hydroxy THC-d3) was added to each sample as a surrogate.

LC-MS was conducted using a Surveyor LC, Thermoelectron, USA) interfaced to an ion trap mass spectrometer (Thermo Finnigan, LCQ Advantage, USA) equipped with an electrospray ionization source. The chromatographic conditions are shown in Table 2. Tandem mass spectrometry (MS/MS) was used to identify the compounds in this study. A full scan of the compounds of interest was run to determine the retention times of all the compounds shown in Table 3. This table provides information about the compounds selected for this study. It should be noted that this is just a representative group of compounds on interest. The number of compounds could be expanded greatly for application to other areas or concerns.

Column: PhenomenexKintex  C-18  
Dimensions: 5 µ, 100 A, 150 mm x 4.6 mm  
Mobile Phase:  0.1% acetic acid (v/v) in H2O (Eluent A)  
  0.1% acetic acit (v/v) in MeOH (Eluent B)  
Flow: 1 mL/min  
Gradient: 0% B to 10% B 0-8 min
  10% B to 60% B 8-15 min
  60% B to 70% B 15-20 min
  70% B to 80% B 20-25 min
  80% B 25-30 min
  80% B to 5% B 30-35 min
Injection Vol. 20 μL  
Column Temp: 35°C  
Sample Tray: 4°C  

Table 2: Chromatographic conditions.

Sample (Reference) Retention Time Precursor Reaction Ions
Ion 1 Ion 2
Amphetamine 8.75 136.1 119 NA
Methamphetamine 10.07 150.2 119 91.2
MDMA 11.39 194.1 163 NA
Heroin 13.73 370.4 328.2 268.3
Cocaine 13.74 304.1 182.1 150.3
Benzoylecognine 14.3 290.2 272.1 168.1
Flurazepam 15.75 388.1 315.3 288.3
Alprazolam 19.08 309.3 281.2 274.3
Lorezepam 19.09 321.2 275.5 302.9
Diazepam 20.85 285.3 257.2 228.2
THC-OH 27.51 331.2 313.1 NA
THC-OH-d3 27.45 334.2 316.2 NA

Table 3: The retention times of all the compounds, Note: The information was obtained by running the standard mix of the compounds and tuning the compounds in the LC-MS instrument. NA represents that there was only 1 reaction ion for the compound.

The selected reaction monitoring (SRM) MS/MS used the positive ionization mode. The capillary temperature was maintained at 300°C. Auxiliary gas was N2 and Ar was applied as collision gas. The intensity of these gases and the collision energy varied based on an optimized tuning conducted prior to analysis for each compound. The run was divided into segments and the compounds each had individual scan events. The SRM was setup with the 1st segment to identify amphetamine, methamphetamine, and 3,4-methylenedioxymethamphetamine (MDMA), the 2nd segment had heroin, the 3rd segment had flurazepam, alprazolam, lorazepam, and diazepam with the last segment containing THC-OH (11-Hydroxy-Δ9-THC) and THC-OH-d3 (internal standard surrogate compound). Cocaine and benzoylecgonine (BE) were run in a separate scan which had only 1 segment and 2 scan events for these two compounds.

It was not possible to obtain stable isotope-labeled standards for all of the compounds of interest to provide specific quantification for each component. The stable isotope-labeled hydroxy-THC-d3 was used as a surrogate to measure ionization effects in the electro-spray LC-MS analysis.

Results and Discussion

Other researchers have reviewed the use of LC-MS for application in sewage epidemiology [10,11]. The current study was a survey of compounds of local interest to determine if ion trap LC-MS could detect the compounds of interest in Lubbock, Texas wastewater. Detections were observed from each of the classes of compounds of interest as seen in Figure 1. These results demonstrate measureable quantities of both drugs and metabolites in the wastewater inflow.

forensic-research-Drugs-metabolites-identified

Figure 1: Drugs and metabolites identified in raw wastewater using liquid chromatography-mass spectrometry.

In this study, several compounds were detected in the samples by LC-MS (Table 4). While metabolic compounds were found in higher levels than the parent compounds, it was observed that parent drug of abuse compounds were present as well. This was a limited study and with specific interest and greater funding, more standards could be obtained to expand the number of compounds identified in the wastewater.

Drug Compound   Weekday Sample   Weekend Sample    
  Collection Presence Response Collection Presence Response
Amphetamine 1st - NA 1st - NA
  2nd + 816 2nd - NA
Methamphetamine 1st - NA 1st + 4770
  2nd - NA 2nd + 4720
MDMA 1st + 370 1st - NA
  2nd + 673 2nd - NA
Heroin 1st + 1380 1st - NA
  2nd - NA 2nd - NA
Cocaine 1st - NA 1st + 331000
  2nd + 187000 2nd + 411000
Benzoylecognine 1st + 389000 1st + 834000
  2nd + 421000 2nd + 1000000
Flurazepam 1st - NA 1st - NA
  2nd - NA 2nd - NA
Alprazolam 1st + 2550 1st - NA
  2nd + 2650 2nd + 7900
Lorezepam 1st + 44800 1st + 44600
  2nd + 30500 2nd + 30800
Diazepam 1st - NA 1st - NA
  2nd - NA 2nd - NA
THC-COOH 1st + 2630 1st + 1910
  2nd + 17400 2nd + 14000
THC-OH d3 1st + 16300 1st + 32900
  2nd + 35500 2nd + 64500

Table 4: Compounds detected in the samples by LC-MS, Note: ’+“=Present; ’-“=Absent, Weekday sample 1st=Tuesday, 2nd=Thursday , Weekend samples are all Staken on Saturday.

The results indicate that further investigation into sewage epidemiology is warranted in Lubbock, TX wastewater. One application of this method has been proposed using the method of sewage epidemiology to indicate if a prison has significant problems with drug use at their facility [12]. Texas has a large prison system and sewage epidemiology would be a very cost effective way for the state of Texas to monitor substance abuse in the prison system [13].

In order to more fully utilize these data, future work will include obtaining more metabolic compound standards and stable isotope internal standards. A review of the surrogate deuterium labeled hydroxyl-THC shows a considerable range of responses. In the laboratory, this work was repeated several times but the amount of material dissolved in the raw wastewater varies widely. Having stable isotopic internal standards for all compounds would be required for specific quantitation of each analyte.

Finally, the parent compounds are excreted in some cases. However, clearly benzoylecgonine occurs at some of the highest concentrations. Future studies will focus on metabolic break-down products which should be present at much higher concentrations than the parent compounds of interest.

Conclusions

The determination of several drugs of abuse, pharmaceuticals and associated metabolites was achieved with a simple extraction and sample clean up. Ion trap mass spectrometry was used successfully to determine these compounds in wastewater samples. The number of compounds of interest could readily be increased and determined if there is a need for this in other applications. The LC-MS has proven superior to GC MS in ease of sample preparation and the number of compounds that can be determined simultaneously. Overall, this method is appropriate for use in forensic applications without requiring the more expensive instrumentation that is employed for residual analysis in water that is going to be routed for reuse and drinking water.

Acknowledgement

The authors would like to thank the Lubbock Water Reclamation Plant for access and assistance in sample collection as well as Prof. Todd Anderson who reviewed an earlier version of this manuscript.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Recommended Conferences

Article Usage

  • Total views: 12960
  • [From(publication date):
    January-2015 - May 21, 2018]
  • Breakdown by view type
  • HTML page views : 9057
  • PDF downloads : 3903
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2018-19
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2018 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
Leave Your Message 24x7