alexa Development and Validation of A HPLC-UV Method for Dissolution Testing of Ciclosporin: Its Application to The Measurement of Brand and Generic Versions from Different Countries
ISSN: 0975-0851
Journal of Bioequivalence & Bioavailability
Like us on:
Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Development and Validation of A HPLC-UV Method for Dissolution Testing of Ciclosporin: Its Application to The Measurement of Brand and Generic Versions from Different Countries

Badr Aljohani1,2*, Faisal Alotaibi F1,3, Essam Ghazaly4, Jaber Al Jaber5, David Perrett1 and Atholl Johnston1

1William Harvey Research Institute, Queen Mary University of London, Charterhouse Square, London, EC1M6BQ, UK

2King Abdullah International Medical Research Centre, Ministry of National Guard Health Affairs (MNGHA), Riyadh, KSA, Saudi Arabia

3College of Pharmacy, University of Shaqra, Al Dawadmi, Saudi Arabia

4Centre for Haemato-Oncology, Barts Cancer Institute, Queen Mary University of London, Charterhouse Square, London, EC1M6BQ, UK

5Department of Pharmacology and toxicology, Faculty of Medicine, University of Kuwait, Kuwait

*Corresponding Author:
Badr Aljohani
Clinical Pharmacology
B.Pharm, M.Sc, Ph.D., MRSC
King Abdullah International Medical Research Centre
Ministry of National Guard Health Affairs (MNGHA)
Riyadh, KSA, Saudi Arabia
Tel: +44-208-793-8397
E-mail: [email protected]

Received Date: July 22, 2017; Accepted Date: September 11, 2017; Published Date: September 19, 2017

Citation: Aljohani B, Alotaibi F, Ghazaly E, Jaber JA, Perrett D, et al. (2017) Development and Validation of A HPLC-UV Method for Dissolution Testing of Ciclosporin: Its Application to The Measurement of Brand and Generic Versions from Different Countries. J Bioequiv Availab 9:509-515. doi: 10.4172/jbb.1000354

Copyright: © 2017 Aljohani B, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Bioequivalence & Bioavailability

Abstract

Ciclosporin is used as an immunosuppressant in post-organ transplantation. Recently, many questions have been raised about using generic substitutes, especially with narrow therapeutic index drugs (NTIDs). In this study, a simple high-performance liquid chromatography (HPLC) method was developed, validated and applied to detection ciclosporin in dissolution testing. Seven ciclosporin products (gelatin capsules) were included in this study, obtained from Columbia (C), Egypt (E), India (I), Jordan (J), Pakistan (P), Saudi Arabia (S), and Turkey (T). The dissolution test was done for all capsules. The Conditions were as follows: 500 ml deionized water as the medium in apparatus 2 (Pharmatest, Germany), temperature 37.5 ± 0.5°C; 50 rev/min, sampling times were 5, 10, 15, 30, 60 and 90 min with 5 ml for each sample. HPLC separation was done by a C18 column, 5 μm, (4.6 × 250 mm, ACE 5) held at 50 ± 0.3°C. Analytes were isocratically eluted at 0.7 ml/min with acetonitrile and water (70+30%) and 0.03% trifluoroacetic acid, over the 25-min run time. The intra-day and inter-day imprecision for ciclosporin across the standard range was <5% and <4%, respectively. The accuracy of the assay was within ± 13% of the true value at standard curve concentrations range from 0.1 to 2 mg/ml of ciclosporin. The lower and the upper limit of detection were 0.001 mg/ ml and 2 mg/ml of ciclosporin, respectively. All brands (S, T, P, J, E) and one generic (C) showed more than 80% of ciclosporin after 90 min (90.3, 100, 90.4, 82.7, 81.4 and 90.6%) respectively. One generic (I), showed less than the minimum percentage of labeled amount, 69.1%. Relative to the brand (T), statistical analysis showed significant differences (P<0.0001) of the mean percentage content between brand and generic. The 95% confidence interval range for the brands (E, J, P, and S) was (72.2-91.8), (73.4-93.3), (80.2-101.9), and (80.1-101.8), respectively, and (80.3-102.1), and (61.3-77.9) for the generic (C) and (I) respectively. Based on these results, we conclude that some of the ciclosporin preparations do not contain the exact mass labeled and the majority contained as yet unidentified impurities.

Keywords

Ciclosporin; Brand and generic; High performance liquid chromatography (HPLC); Dissolution testing; Narrow therapeutic index drugs (NTIDs); Immunosuppressant

Introduction

In the last few years, many clinical questions have been raised about using generic substitutes, especially those for narrow therapeutic index drugs (NTIDs) such as ciclosporin. The patient should be closely monitored [1].

Ciclosporin (cyclosporine A, ciclosporin A, cyclosporin A, CyA) is a lipophilic cyclic undecapeptide compound formed by a soil fungus called Tolypocladiuminflatum Gams [2]. It was discovered by Sandoz of Basel, Switzerland in the 1970s [3], as synthetic or semi-synthetic [4]. It has been widely used since 1978 after organ or tissue transplant to prevent rejection [5,6]. For auto-immune diseases such as severe rheumatoid arthritis and psoriasis, ciclosporin can be used (Figure 1) [7].

bioequivalence-bioavailability-structure-ciclosporin

Figure 1: The chemical structure of ciclosporin (Drawan by Aljohani, 2014, using ACD/Labs 2012, version 14.01, build 65894).

Ciclosporin can be given intravenously or orally, as a solution or soft gelatine capsule. The capsules contain several ingredients such as Corn oil-mono-di-triglycerides [8].

High-performance liquid chromatography (HPLC) is an excellent technique for measuring average drug content in tablets and capsules. It can be used for therapeutic drug monitoring (TDM) when suitable detectors such as a mass spectrometer are used. The advantages of HPLC-mass spectrometry are high sensitivity, specificity, small sample requirements, minimal sample preparation, rapid throughput, and simultaneous measurement for the drug and its impurities. However, using the current assay, it is impossible to know if unknown impurities have interfered with ciclosporine. HPLC with ultraviolet/fluorescent detection is also a good method to quantify and analyse the content of ciclosporin capsules. Like other peptides, ciclosporin can be measured by HPLC at relatively low UV absorbance wavelengths and this is considered as a gold standard for its measurement [9-11].

There have been various methods to measure the drug contents in ciclosporin capsules. Some of them used to cut ciclosporin capsules and dissolving them before subjecting it to analytical HPLC system [12]. Other suggests cutting ciclosporin capsule and obtaining the contents and dissolves a known volume in the mobile phase before quantifying using HPLC [12]. According to the United States Pharmacopeia, 2008 [13], they recommend to cut about 20 capsules and extract the contents with the aid of alcohol and make up a known solution in a volumetric flask with ethanol. Then dilute this stock to obtain a concentration of 1 mg/mL. This is used to determined ciclosporin in capsules. The other method is to obtain the contents of a capsule by aspirate, using a syringe and preparing from the obtained stock. Therefore, a dissolution method of rupturing the capsule in medium and then measuring the content of ciclosporin in the dissolution medium would give rise to more reproducible results. Several methods have been developed to measure ciclosporin in biological matrices [14,15]. However, our methos are tailored for quantification of ciclopsorin in dissolution testing.

Ciclsoporin is classified as a BCS IV because of its low solubility and poor permeability [16]. However, the dissolution test is carried out under standard conditions described in the FDA guidelines.

Material and Methods

General chemicals

Cyclosporin A (lot no. BCBD2418V, catalogue number: 30024-100 mg, 98.5% Purity) for calibrator and control samples preparation and HPLC-grades solvent were obtained from Sigma-Aldrich, Poole, UK. HPLC-grade deionised water at a purity of 18.2 MΩcm was obtained from an ultra-water system (PURELAB® ULTRA), Marlow, United Kingdom.

Equipment

PT-DT 70 dissolution tester and 20 μm Poroplast Filter Element were from PharmaTest, Hainburg, Germany. High-quality electronic instrument: Sartorius (R 160 P) Electronic Semi-Microbalance, Sartorius Stedim, Surrey, United Kingdom. Ultra-water system (PURELAB®ULTRA) for deionised water, ELGA, UK. HPLC system Jasco UV- 975 intelligent UV-VIS with Jasco AS-950 intelligent sampler injector, two Jasco PU-980 intelligent HPLC pumps and Jasco DG- 2080-53 3 lines degasser. Jones Chromatography (Column Heater) Model 7990. The HPLC experiment was carried out with a UV detection wavelength of 210 nm. The separation was carried out using a reversed phase C18 column, 5 μm, (125 x 4.6 mm, ACE 5, part number: ACE-121- 1246) Advanced Chromatography Technologies, Aberdeen, AB25 1HF, Scotland. JENWAY spectrophotometer, Ultra-Violet/Visible, Model 6715 UV/Vis. The column temperature was maintained at 50°C and the injection volume was 20 μL. The analytes were eluted by isocratic elution at a flow rate of 0.7 mL/min with acetonitrile and water (70:30%, v/v) and 0.03%, v/v trifluoroacetic acid, over 25 min. The HPLC control data acquisition was by Chromatography Data System version 1.8.6.1 from JASCO Chrom Pass.

Collection of drug samples

All brand samples manufactured by Novartis in Switzerland T (100 mg), S (100 mg), E (50 mg), J (50 mg), and P (100 mg), then repackaged in the importing country. Both generics C (100 mg) and I (100 mg) were manufactured in India. All samples were obtained from a commercial pharmacy, except one sample (S) was from a government hospital.

Extraction by dissolution

A dissolution method to analyse ciclosporin capsules were obtained from US Pharmacopeia 2008 [13], (USP-31) requirements such as rupturing time within 15 min.

Seven ciclosporin products were included in this study. All products were stored according to the conditions labelled on the package. Each capsule was weighed to check the dosage uniformity. Dissolution test carried out under the following condition: Temp: 37.5°C ± 0.5, 500 mL deionized water used as a medium, the paddle apparatus (Apparatus 2):50 rpm, sampling time at (5, 10, 15, 30, 60, and 90 min), with 5 mL volume for each sample. The samples were filtered using 20 μm filters.

HPLC experimental method development

Identification of the λmax for ciclosporin measurement: A standard ciclosporin solution of 2 mg/mL was used to detect the λmax. The spectrum was compared to an acetonitrile blank which was used as a base line. The scan range of the spectrophotometer was set from 190 nm to 400 nm to detect maximum absorbance for ciclosporin throughout this range.

Selection of columns

Various C18 columns were tested to get a good separation of ciclosporin in the HPLC method. The ACE 5 column showed the best results. All subsequent development was undertaken on this column.

Optimisation of temperature for separation of ciclosporin

Based on literature review and the US Pharmacopeia, different temperatures to heat the column were investigated, at 25, 50, and 75°C The best resolution obtained at 50°C.

Optimisation of flow rate

Different flow rates at 0.7, 1, and 1.2 mL/min used to obtain best resolution and peak shape, the best resolution was found at 0.7 mL/ min.

Preparation of stock, calibration solutions and control samples

Ciclosporin stock solution (2 mg/mL) was prepared in the mobile phase and stored at 20 ºC. The ciclosporin calibration standards were freshly prepared at the time of the experiment. Eight concentrations of ciclosporin (0.1, 0.2, 0.4, 0.6, 0.8, 1, 1.5 and 2 mg/mL) were prepared in mobile phase.

Calibration and linearity

Calibration standards (0.1-2 mg/mL) were injected into the HPLC system. The integrated peak areas of the ciclosporin were recorded. Linear regression of the standards was done using Microsoft Excel 2010, by plotting the peak area of ciclosporin against the ciclosporin concentration of the standards. The linearity was checked by calculating the r2 value using least-square linear regression analysis.

Method sensitivity

The upper limit of detection is given by the point where the detector response deviates by 1% from expected linear response.

Method specificity

The method specificity was investigated by injecting blank mobile phase before the start of each run. The blank standards were used to control for any interfering peak that elutes at the same retention time of ciclosporin. The interfering peaks might arise from the mobile phase or the extraction process.

Coefficient of variation

Each ciclosporin calibration standard peak area was divided by its concentration in order to calculate the slope of the standard curve. The average and the standard deviation of the slope of all calibration standards were then calculated. Then the intra-day coefficient of variation (CV) of this calibration line was calculated by dividing the standard deviation by the average (of at least three samples) and multiplied by 100.

Inaccuracy

The inaccuracy of the assay was determined by measuring the difference between actual (hypothetical) and the measured concentration of each ciclosporin calibration standard. The concentration difference was then divided by the actual concentration and multiplied by 100.

HPLC system carry over

The HPLC system carryover was checked by running blank samples between actual samples. The blank sample was checked for the presence of any carried over ciclosporin. Calibration standards were also randomised and the calibration curve linearity was then checked in order to detect any carryover.

Dissolution extraction recovery

The recovery of the volume was calculated by dividing the actual volume over the total volume then multiplied by concentration.

Stability

Freeze thaw stability was done for three ciclosporin calibration standards stored at -20°C, thawed and re-frozen every week for three weeks.

Statistical analysis and graphic presentation

The graphic presentations and the statistical analyses were done using Excel 2010 for Windows version, 14.0.6112.500 (32-bit) and Minitab version, 16.

Results

Optimisation of the UV detector for ciclosporin measurement

The highest absorbance was between 205 and 215 nm. Based on this result, the maximum absorbance at 210 nm was used to detect ciclosporin in all subsequent experiments.

Dissolution test

The results of the dissolution test showed that the average rupture times for ciclosporin capsules of generic C, brand T, brand J, brand E, generic I, brand P and brand S were (mean ± SD) 1.56 ± 0.02, 5.07 ± 0.02, 5.11 ± 0.01, 5.29 ± 0.02, 5.34 ± 0.03, 5.37 ± 0.02 and 5.43 ± 0.02 min respectively. The capsule rupture was determined through the dissolution vessel by visual observation of the capsule shell and/ or by the release of capsule content. A sample was tested using HPLC before putting the capsule into the vessel to ensure no ciclosporin contamination of the medium. The HPLC chromatogram shows that there was no ciclosporin peak (Figure 2). This confirms the absence of ciclosporin in the medium. Rupturing time was calculated for each capsule by using a stop watch. In this study, all the capsules met the USP requirements, rupturing within 15 min (Table 1).

bioequivalence-bioavailability-dissolution-tester

Figure 2: Chromatogram to show there is no peak before putting the ciclosporin capsule in the dissolution tester.

Ciclosporin Capsule Average Rupture Time (min) ± SD
Generic C 1.56 ± 0.02
Brand T 5.07 ± 0.02
Brand J 5.11 ± 0.01
Brand E 5.29 ± 0.02
Generic I 5.34 ± 0.03
Brand P 5.37 ± 0.02
Brand S 5.43 ± 0.02

Table 1: Average rupture time for ciclosporin capsules (n=4).

Calibration and linearity

The separation of ciclosporin was achieved using the method and a typical ciclosporin chromatogram is shown in Figures 3 and 4. A symmetrical peak shape with a retention time of 13.8 min corresponds to ciclosporin retention time. The peak areas of the ciclosporin standards were 46, 92.3, 183, 275, 367, 459, 688, 916 mVmin for 0.1, 0.2, 0.4, 0.6, 0.8, 1.0, 1.5, and 2 mg/mL respectively. The linear regression (R2) was 1.00, regression residuals less than ± 5% (Figure 4). In Figure 5 show linear regression residuals for ciclosporin standards, and Figure 6, we can see a standard curve for ciclosporin analysis using the HPLC system.

bioequivalence-bioavailability-sample-solution

Figure 3: Chromatogram obtained from a sample solution contains 200 μg/mL ciclosporin.

bioequivalence-bioavailability-minutes-sampling

Figure 4: Chromatogram obtained at 90 minutes sampling from a dissolution tester.

bioequivalence-bioavailability-regression-residuals

Figure 5: Linear regression residuals for ciclosporin standards, less than 5%.

bioequivalence-bioavailability-ciclosporin-analysis

Figure 6: Standard curve for ciclosporin analysis using the HPLC system.

Method sensitivity

On seven different days, the mean concentration measured of the lowest standard (0.1 mg/mL) was 0.09 ± 0.01 (mg/mL) with CV of 11.84%. The highest standard was 2.01 ± 0.08 (mg/mL) with a CV of 4.22% (Table 2).

Days Measured concentration of lowest standard (0.1 mg/mL) Measured concentration of highest standard (2 mg/mL)
D1 0.07 2.21
D2 0.1 1.97
D3 0.1 1.99
D4 0.1 2
D5 0.1 1.99
D6 0.1 2
D7 0.1 1.97
Average 0.09 2.01
± SD 0.01 0.08

Table 2: The concentration of low and high standards on different days.

Method specificity and carry over

The method for measuring ciclosporin was specific. Blank mobile phase samples were used before, between and after each run. No significant interfering peaks and no presence of any carried over ciclosporin at the migration time of ciclosporin samples.

Precision

Inter-day variability: Ciclosporin standards at 0.1, 0.2, 0.4, 0.6, 0.8, 1, 1.5, and 2 mg were injected into the HPLC system. Table 3 shows the hypothetical and measured concentrations by HPLC analysis for these samples over six different days along with the average, standard deviation, and the coefficient of variation percentage.

Concentration Day1 Day2 Day3 Day4 Day5 Day6
(mg/mL) (mg/mL) (mg/mL) (mg/mL) (mg/mL) (mg/mL) (mg/mL)
0.1 0.1 0.09 0.1 0.09 0.1 0.1
0.2 0.2 0.2 0.21 0.2 0.21 0.2
0.4 0.38 0.39 0.4 0.39 0.4 0.38
0.6 0.62 0.61 0.57 0.61 0.57 0.62
0.8 0.82 0.81 0.82 0.81 0.82 0.82
1 1.01 1.01 1.02 1.01 1.02 1.01
1.5 1.51 1.49 1.46 1.49 1.46 1.51
2 1.97 1.98 2 1.98 2 1.97

Table 3: The average, standard deviation (± SD) and the coefficient of variation (CV) for the inter-day variability for ciclosporin assay standards at 0.1, 0.2, 0.4, 0.6, 0.8, 1, 1.5, and 2 mg/mL.

Intra-day variability: Ciclosporin standards at 0.1, 0.2, 0.4, 0.6, 0.8, 1, 1.5, and 2 mg were injected into the HPLC system on the same day. Table 4 shows the actual and measured concentrations for these samples, average, standard deviation, and the correlation of variation percentage.

Concentration
(mg/mL)
Concentration
(mg/mL)
Concentration
(mg/mL)
Mean ± SD CV% intra-day
0.1 0.09 0.10 0.10 0.0022 2.22
0.2 0.20 0.21 0.20 0.0074 3.64
0.4 0.39 0.40 0.40 0.0092 2.30
0.6 0.61 0.57 0.59 0.0241 4.05
0.8 0.81 0.82 0.82 0.0042 0.51
1 1.01 1.02 1.02 0.0091 0.89
1.5 1.49 1.46 1.48 0.0161 1.08
2 1.98 2.00 1.99 0.0102 0.51

Table 4: The average, standard deviation (± SD) and the coefficient of variation (CV) for the intra-day variability for ciclosporin assay standard at 0.1, 0.2, 0.4, 0.6, 0.8, 1, 1.5, and 2 mg/mL.

Inaccuracy: The inaccuracy of the method was measured by calculating the difference between the actual and measured concentration of each ciclosporin calibration standard (Table 5).

Concentration Day1 Day2 Day3 Day4 Day5 Day6 Average ± SD
(mg/mL)
0.1 -32.69 0.19 2.98 0.19 2.98 1.7 -4.1 14
0.2 -29.27 0.12 5.41 0.12 5.41 1.15 -2.83 13
0.4 -23.97 0.9 2.38 0.9 2.38 4.66 -2.12 11
0.6 11.65 1.93 3.75 1.93 3.75 3.38 4.4 4
0.8 2.24 2.26 3.01 2.26 3.01 2.52 2.55 0
1 -1.96 1.64 2.94 1.64 2.94 1.96 1.52 2
1.5 -18.14 0.53 2.05 0.53 2.05 0.69 -2.04 8
2 10.62 0.61 0.11 0.61 0.11 1.41 2.24 4

Table 5: Inaccuracy of the assay was <4% of the true value at standard curve concentrations range from 0.1 to 2 mg/mL of ciclosporin.

The first day the inaccuracy of the 3 lowest concentration was above acceptable level due to unknown reason however in subsequent 6 experiments the inaccuracies were within acceptable range (less than 15%).

Drug recovery (from capsules): This was done in order to confirm that the dissolution test successfully recovered all the capsule content after 60 and 90 min. Maximal recovery (99 ± 0.4%) was after 60 min. Recovery after 90 min was 100 ± 0.5% (Table 6).

Recovery Average
Ciclosporin 60 min 90 min
Brand T 99 ± 0.04 % 100 ± 0.05 %
Brand P 91 ± 0.03 % 90 ± 0.03 %
Brand S 88 ± 0.04 % 90 ± 0.07 %
Generic C 86 ± 0.05 % 91 ± 0.01 %
Brand J 84 ± 0.02 % 83 ± 0.05 %
Brand E 76 ± 0.05 % 81 ± 0.04 %
Generic I 69 ± 0.04 % 69 ± 0.08%

Table 6: The average recovery percentage of ciclosporin mass amount after 60 and 90 min of the dissolution test (n=4).

Stability

This test was done to check the stability of ciclosporin standards in stock solutions. Ciclosporin standards were stable for the whole period of the experiment. Two different concentrations of 0.4 and 0.6 mg from day 2, 4, and 6 were measured (Table 7). The ciclosporin standards were stable for freeze-thaw cycles. Average measured concentration for 3 cycles of 0.4 and 0.6 mg of ciclosporin was 0.39 and 0.61 respectively. The CV% of 0.4 and 0.6 mg of ciclosporin standards were less than 3%. The inaccuracy of measured concentration was less than ±3%.

Ciclosporin Concentration (mg/mL) Day 2 (mg/mL) Day 4 (mg/mL) Day 6 (mg/mL) Average ± SD CV%
0.4 0.39 0.39 0.38 0.39 0.009 2.22
0.6 0.61 0.61 0.62 0.61 0.005 0.82
Inaccuracy Day 2 Day 4 Day 6 Average ± SD  
0.4 0.9 0.9 4.66 2.15 2.173  
0.6 1.93 1.93 3.38 2.41 0.839  

Table 7: The average concentration (± SD) and coefficient of variation (CV) and inaccuracy of ciclosporin stability in stock solution for day 2, 4 and 6.

Application of the assay

This method was successfully applied to measure the actual concentration in each ciclosporin product using HPLC analysis. Sampling time was done on six different intervals in order to check for ciclosporin release from capsules. All brands (T, S, E, J, P) and one generic (C) showed more than 80% of labelled amount in ciclosporin capsules after 90 min of the dissolution test giving 100 ± 0.05, 90 ± 0.07, 81 ± 0.04, 83 ± 0.05, 90 ± 0.03, and 91 ± 0.01 % (± SD) respectively. One generic (I), showed less than the minimum percentage of labelled amount 69 ± 0.08%. Relative to the brand (T), statistical analysis showed significant differences (p<0.0001) of the mean percentage content between brand and generic, the confidence interval 95% range for the brands (S, E, J, P) were (80.1-101.8), (72.2-91.8), (73.4-93.3), (80.2-101.9), respectively, and (80.3-102.1), (61.3-77.9), for the generic (C) and (I) respectively (Tables 8 and 9).

Drug Name Country of Origin Average weight (g), n=4 Sampling time (min) and Average content %, n=4
5 min 10 min 15min 30 min 60 min 90 min
Neoral®* Turkey 1.5 6.70% 55.80% 98.30% 98.60% 98.80% 100.00%
Neoral®* Saudi 1.51 0.40% 57.80% 85.00% 91.90% 88.40% 90.40%
Neoral® Jordan 1.67 3.20% 65.30% 77.60% 80.50% 84.20% 82.80%
Neoral® Egypt 1.69 3.10% 13.50% 55.70% 74.80% 76.20% 81.40%
Neoral®* Pakistan 1.5 6.90% 79.20% 93.30% 89.30% 91.30% 90.40%
Generic C* Colombia 1.32 85.00% 94.70% 88.00% 89.50% 86.10% 90.50%
Generic I* India 1.6 4.60% 47.80% 72.00% 65.70% 68.90% 69.30%

Table 8: Shows the average percentage of ciclosporin mass amount at different time intervals, reference capsule (Turkey) showed 100% mass amount after 90 min, (n=4).

Ciclosporin N % of mass amount at 90 min (mg/mL) ± SD CV (%) 95% Confidence Interval
Capsule
Generic I (100 mg) 4 69.3 7.6 11 (61.3-77.9)
Brand E (50 mg) 4 81.4 3.6 4 (72.2- 91.8)
Brand J (50 mg) 4 82.8 4.7 6 (73.4-93.3)
Brand S (100 mg) 4 90.4 6.9 8 (80.1-101.8)
Brand P (100 mg) 4 90.4 3.3 4 (80.2-101.9)
Generic C (100 mg) 4 90.5 1.2 1 (80.3-102.1)

Table 9: Average percentage content of ciclosporin in capsules.

The average percentage of ciclosporin mass amount, standard deviation, the coefficient of variation and 95% CI based on reference capsule (T) 100% mass amount, (n=4).

Discussion

Ciclosporin is an important immunosuppressant agent. In recent years, many studies have questioned the use of generic substitutes, especially when considering those that are narrow therapeutic index drugs (NTIDs).

In this work, the aim was to develop, validate and apply a dissolution test for ciclosporin in capsules using an HPLC method and then to compare brand medications to the generic or copy products. Moreover, the detection of contaminants and the source of impurities in these products are essential to understanding.

A sensitive and specific HPLC-UV method was developed to detect and subsequently quantitate the ciclosporin active base in capsules. Dissolution testing of the capsules was done as specified in the US pharmacopeia. Analytes were isocratically eluted at 0.7 mL/min with acetonitrile and water (70:30% v/v) and 0.03% v/v trifluoroacetic acid over the 25 min runtime. The hydrocarbon chain forming the hydrophobic phase is usually a hydrocarbon of eighteen (C18), eight (C8) or four (C4) carbons. Peptides need longer hydrophobic chain lengths to be resolved [17]. Various C18 columns were tested to get a good separation of ciclosporin by the HPLC method. The ACE 5 column showed the best results. All subsequent development was undertaken on this column. Although the USP recommends that the column should be heated to 80°C, at 50°C we achieved the necessary sensitivity and separation.

In addition, the dissolution profile revealed that all the ciclosporin capsules used in these investigations complied with US Pharmacopeia, rupturing within 15 min [13].

The standard curve, when fitted using least square linear regression, was linear (r2=0.9998). The separation of ciclosporin was successful at 13.8 min retention time. Detection of ciclosporin at low and high concentration was reproducible. The lower limit of quantification was 0.1 mg/mL, higher limit of quantification was 2 mg/mL and the method was sensitive with CV% of 4.2 and 11.8% at high and low quantitation limits respectively.

The within day variability was mostly less than 4%. The between day variability of the standards was within an acceptable range mostly less than 5%.

The method was accurate. The inaccuracy of the assay was within <4%. However, on the first day, the inaccuracy of the three lowest concentrations was higher than the acceptable range (15%) due to unknown reason. There was no system carry over for ciclosporin. Blank samples were run between the actual samples without detection of any carried over of ciclosporin.

Different concentrations at 0.4 and 0.6 mg used to check the stability. CV was less than 3% and the inaccuracy of measured concentration within ± 3%.

The application of the assay was successful. The Turkey brand was the best brand in terms of the content of the active ingredient with 100 ± 0.05%. The Indian brand had only 69 ± 0.08%. There were one or more impurities in all capsules. There are some discrepancies in the dissolution methods between the FDA and the USP. The FDA suggests using a L of 0.1 N HCl containing 4 mg of N, N-dimethydodecylamine- N-oxide per mL with 75 rpm. But the USP in 2008 recommends the use of 500 mL water with 50 rpm.

In the USP there is no mention of the times at which the sampling should be taken apart from the fact that they ask us to test at 90 min where there should be more than 80% of the content. In this study, we did 5, 10, 15, 30, 60 and 90 min.

The results of this study are surprised that the content is much less than 80% in some of the capsules and are not 100% in all the brand capsules either. This raise concerns over the manufacturing standards and the quality control measures employed even within the same brand manufactured in different countries. In the literature, The Danish medicine agencies suggested that the generic substitution should no longer be used for ciclosporin following a reassessment [18]. The contents of one of the ciclosporin capsules we reported in this study as an agreement with those found by other researchers [19], where they detected only 68% of the labelled concentration in one of the generic capsules.

Conclusion

This study represents a simple, rapid, specific and sensitive HPLC method for the determination of an immunosuppressive drug ciclosporin. This method was used successfully in determination of ciclosporin mass amount in brand versus generic capsules. Overall, switching among and between brands and generic ciclosporin can lead to undesirable effects.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Recommended Conferences

Article Usage

  • Total views: 289
  • [From(publication date):
    September-2017 - Dec 11, 2017]
  • Breakdown by view type
  • HTML page views : 266
  • PDF downloads : 23
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

social_politicalsc[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version