Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Development and Validation of LC-MS/MS Method for Determination of Trimipramine Maleate in Human Plasma

Venkata Sairam Koganti*, Sekuri Shashidhar Reddy, Jyothigoudapura Channabasappa Thejaswini and Gurupadayya Bannimath

Department of Pharmaceutical Analysis, JSS College of Pharmacy, JSS University, Mysore 570 015, India

*Corresponding Author:
Venkata Sairam Koganti
Department of Pharmaceutical Analysis
JSS College of Pharmacy, JSS University
Shivarathreeshwara Nagar, Mysore-570 015, India
Tel: 91-9739780422
E-mail: [email protected]

Received Date: July 16, 2015; Accepted Date: August 21, 2015; Published Date: August 24, 2015

Citation: Koganti VS, Reddy SS, Thejaswini JC, Bannimath G (2015) Development and Validation of LC-MS/MS Method for Determination of Trimipramine Maleate in Human Plasma. Pharm Anal Acta 6:413. doi: 10.4172/2153-2435.1000413

Copyright: © 2015 Koganti VS, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Pharmaceutica Analytica Acta

Abstract

A simple, sensitive, Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) method was developed for the determination of Trimipramine maleate in human plasma using Opipramol dihydrochloride as internal standard (ISTD). The analyte and ISTD were isolated from plasma by protein precipitation method using methanol. The chromatographic separation was carried out on ZORBAX ECLIPSE XDB- C18 column (4.6 x 150 mm, 5 μm). The elution was done with isocratic mobile phase composed of 5mM ammonium formate with 0.1% formic acid and methanol in the ratio of 25:75(v/v) at a flow rate of 1.0 ml/min. Trimipramine and Opipramol were detected with proton adducts at m/z 295.20→100.10 and m/z 364.30→171.20 in multiple reaction monitoring (MRM) positive mode respectively. The retention time of Trimipramine and ISTD were found to be 1.67 min and 1.48 min respectively. The calibration curves were linear over concentration range of 0.1 to 100.1 ng/ml. The lower limit of quantification was found to be 0.1 ng/ml. The method was validated by EMEA (European medicines agency) guidelines for linearity, accuracy and precision, matrix factor, recovery, selectivity and stability. The validation parameters were found to be well within the acceptance limit.

Keywords

LC-MS/MS; Human plasma; Trimipramine maleate; Multiple reaction monitoring; Opipramoldihydrochloride

Abbreviations

%: Percent; Cmax: Concentration Maximum; EMEA: European Medicines Agency; LC-MS/MS: Liquid Chromatography-Mass Spectrometry/Mass Spectrometry; LLOQ: Lower Limit of Quantitation; LMQC: Lower Medium Quality Control; LOD: Limit of Detection; LOQ: Limit of Quantification; LOQ QC: Limit of Quantification Quality Control; LQC: Low Quality Control; MQC: Medium Quality Control; MRM: Multiple Reaction Monitoring; Psi: Pounds per square inch; Q1: First Quadrupole; Q3:Third Quadrupole; rpm: Rotations per minute; V: Voltage

Introduction

Trimipramine maleate is chemically dihydrobenzo [b] [1] benzazepin-11-yl)-N,N,2-trimethylpropan-1-amine (Figure 1) used as a tricyclic antidepressant (TCA) [1]. These agents act by blocking the reuptake of serotonin and norepinephrine; the brain chemicals that are related to mood. By blocking the reabsorption of these chemicals, cyclic antidepressants increase the levels of serotonin and norepinephrine which helps in elevating the mood.

pharmaceutica-analytica-acta-Structure

Figure 1: Structure of Trimipramine.

Trimipramine (marketed forms Surmontil, Rhotrimine, Stangyl) is a tricyclic antidepressant (TCA) with sedative properties. Trimipramine is primarily used in the treatment of major depressive disorder [2] especially where sedation is required due to its prominent sedative effects [3]. Generally, depression disorders are associated with an increase in the risk of suicide [4] and monitoring trimipramine plasma level concentration would be helpful in optimizing the dose for enhancement of clinical usage [5]. Quantifying the plasma drug would be an add-on for therapeutic drug monitoring and useful in pharmacokinetic studies in treatment of depression.

Several spectrometric [6,7], chromatographic techniques have been reported, including UPLC methods [8] but till date there were no perfect LC-MS/MS reported methods. But all these methods are inadequate in terms of sensitivity and employed uneconomical extraction procedures. The current method employed extraction procedure using methanol yielding good recovery results. Method development was done using opipramol dihydrochloride (Figure 2) as an internal standard, unlike deuterated standards and obtaining aqueous linearity regression r=0.9993.

pharmaceutica-analytica-acta-Opipramole

Figure 2: Structure of Opipramole.

Experimental

Chemicals and reagents

Trimipramine maleate and Opipramoldihydrochloride were obtained from RL Fine Chemicals Ltd., Bangalore. Methanol of analytical gradient was purchased from Sigma Aldrich, Ammonium Formate was purchased from Fischer Scientific and Milli-Q water was used. Blank human plasma was obtained from volunteers of human pathological unit.

Instrumentation

The LC-MS/MS system consisted of a High Performance Liquid Chromatography (Schimadzu LC-20AD, autosampler SIL-30ACMP) coupled with a Sciex Triple Quadrupole Mass Spectrometer (API 4000, MDS, Sciex, Ontario-Canada) equipped with an electrospray ionization (ESI) (Applied Biosystems). Data acquisition and processing were controlled by Applied Biosystems/MSD SCIEX Analyst software (Version 1.6.1).

Chromatographic conditions

Chromatographic separations were achieved by ZORBAX Eclipse XDB-C18 (4.6 X 100 mm; 3.5 μ) column with a flow rate of 1ml/min. Mobile phase was methanol: 5 mM ammonium formate with 0.1% formic acid in a ratio of 75:25. Injection volume was 10 μl. The run time was 3 min.

Mass spectrometry conditions

Mass parameters were set and tuned in positive ion mode for Q1 and Q3 masses. The mass parameters were listed in Table 1. The full scan and product ion spectra of trimipramine are shown in (Figure 3).

Parameter Value
Source Temperature (ºC) 500
Ion Source Gas1 (Psi) 40.00
Ion Source Gas2 (Psi) 45.00
Curtain gas (Psi) 30.00
Collision gas (Psi) 6.00
Ion Spray Voltage (V) 5500.00
Dwell time (ms) 200.00
Entrance potential (V) 10
Collision cell exit potential (V) 10
MRM Transition (amu)-Trimipramine 295.30→100.10
MRM Transition (amu)- Opipramol 364.30→171.20
Collision Energy (V) 25(analyte)/ 30(IS)
Declustering potential (V) 50(analyte)/ 80(IS)

Table 1: Mass Spectrometer Parameters.

pharmaceutica-analytica-acta-product

Figure 3: Full scan and product ion spectra of Trimipramine.

Preparation of stock and working standard solutions

Stock of both trimipramine and internal standard were prepared by weighing 5 mg of each and dissolving in 100% methanol separately giving 5 mg/ml concentrations. Subsequently analyte stock concentration is further diluted by 60% methanol to get 756 μg/ml and internal standard to 200 ng/ml.

Preparation of calibration standards and quality control samples

Calibration standards and quality control solutions were prepared based on the Cmax of the analyte [9]. Initially, all the stock solution was prepared in aqueous media. Now spiking of these aqueous stock solution into blank plasma were done to get the desired calibration standards of standards 8 to standard 1 of range 100.573, 80.458, 40.229, 20.115, 8.046, 3.218, 0.209, 0.105 ng/ml respectively. The quality control samples of HQC, MQC, LQC and LLOQC are of concentrations 75.619, 37.809, 0.284, 0.109 ng/ml.

Sample preparation

Each plasma aliquot of 100 μl was taken and to it 50 μl of Internal standard drug was added and vortexed (30 sec). Subsequently to it 400 μl of methanol was added and vortexed (30 sec). Now these are cool centrifuged at 4500 rpm for 5 min. The supernatant is collected and 150 μl is transferred to autosamplers and 10 μl is injected.

Method Validation 5.1System Suitability

To equilibrate chromatographic system, six replicate injections of reference standard solutions (MQC level concentration) have been injected.

ASCOT (Auto sampler carry over test)

To check the residual compound in standard blank from the previous injection, sequential injections of processed samples were given in the order of Diluent (mobile phase)→Standard blank→ Standard 1→Standard 8→Standard blank (re-injected).

Selectivity

To evaluate any interference at retention time of analyte and ISTD, six different individual blank plasma lots (including 1 heamolysed and 1 lipemic) were screened. Six blanks and respective lower limit of quantitation (LLOQ) samples were processed and analyzed.

Linearity

Calibration curve covering the entire range of Trimipramine from 0.1 ng/ml to 100 ng/ml was analyzed each with eight concentration levels. Calibration curves covering the entire range of the analytical method were constructed with analyte area on Y-axis and analyte concentration on X-axis with the regression model (linear 1/x2 weighting = mx+c). The correlation coefficient was found to be 0.999.

Matrix effect

It was investigated for six different individual plasma lots (4 normal, 1 haemolysed and 1 lipemic) at LQC and HQC concentration level. Six unextracted (aqueous) and six post extracted LQC and HQC samples were processed and analyzed. The matrix factor for analyte and ISTD should be within ±15% and IS normalized matrix factor should be ≤ 15% at each level.

Precision and accuracy

Three different batches of precision and accuracy were performed, intra-assay precision was performed at LLOQ (0.109 ng/ml), LQC (0.284 ng/ml), MQC (37.809 ng/ml) and HQC (75.619 ng/ml) concentration levels for a six replicates in each analytical run. Interassay precision and accuracy were calculated after the replicates in three different analytical runs.

Recovery

The extraction efficiency of the analytical procedure was evaluated. By processing and analyzing six each LQC, MQC and HQC concentration level. By comparing the responses of analyte and ISTD in unextracted and extracted samples of equivalent concentration levels.

Stability studies

Experiments were performed to evaluate the stability of analyte in plasma at different simulating conditions of processing. Freeze thaw, bench top and processed sample stability like in-injector and wet extract were performed at LQC and HQC levels in six replicates.

Results and Discussion

Method development and optimization

The chromatographic conditions were optimized with several trials while change in mobile phase composition to achieve symmetric peak shapes and good resolution. Mobile phase buffer composition was (ammonium formate and formic acid). Initially mobile phase methanol 0.1% formic acid (80:20) is used and the retention time was found to be 1.44 min for analyte and 1.52 min for ISTD, the detector response was found to be less and peak shape was not good. The mobile phase composition was altered in the next trial to Methanol:5 mm ammonium formate with 0.1% formic Acid (80:20) and retention time was found to be 1.38 min for analyte and 1.27 min for ISTD, the peak wasn’t symmetric and to this mobile phase composition further modified by changingvolume proportion (65:35) which resulted in increase of retention time of analyte to 2.66 min and for ISTD 2.14 min. Furthermore the composition proportion was changed to (75:25) methanol: buffer and retention time was found to be 1.67 min for analyte and 1.48 min for ISTD. This method was optimized as chromatographic conditions.

The mass parameters were tuned and detection was carried out in positive polarity with multiple monitoring reaction (MRM) modes. The mass parameters were tuned and optimized for Q1 (parent ion) and Q3 (daughter ion).

Method validation

The developed method was validated in terms of system suitability, auto sampler carry over test, selectivity, linearity, matrix effect, precision and accuracy, recovery and for stability.

System suitability: System suitability was performed by six replicate injections of aqueous MQC level concentration (AQS MIX). The acceptance criteria were met. The %CV of retention time limit was found to be 0 and 0 for both analyte and ISTD and % CV was found to be 2.19 for response ratio limit. Which satisfactorily passed the criteria of %CV for retention time and response ratio of analyte and ISTD NMT 5%.

(ASCOT) Autosampler carryover test: Response of interference at retention times of Analyte and ISTD in re-injected blank sample after injecting extracted Standard-1 and Standard-8 was within the limits i.e. response of interference at analyte retention time NMT 20% of LLOQ (0.109 ng/ml) response and NMT 5% for ISTD.

Selectivity: Selectivity was performed with extracted LLOQ samples and blank from each matrix were analyzed to evaluate selectivity. The % interference at the retention time of ISTD for all six biological matrices was found to be less than 20% of the mean LLOQ response (0.109 ng/ ml). The chromatograms of selectivity of Standard blank (Figure 4) of plasma lot1 and LLOQ (Figure 5) of plasma lot 1 are shown in.

pharmaceutica-analytica-acta-Plasma

Figure 4: Selectivity- Plasma Lot 1-Standard Blank.

pharmaceutica-analytica-acta-Plasma

Figure 5: Selectivity-Plasma Lot 1- LLOQ.

Linearity: Calibration curve covering the entire range of Trimipramine from 0.1 ng/ml to 100 ng/ml was analyzed each with eight concentration levels. Calibration curves covering the entire range of the analytical method were constructed with analyte area on Y-axis and analyte concentration on X-axis with the regression model (linear 1/x2 weighting = mx+c). The correlation coefficient was found to be 0.999. Calibration curve plot (Figure 6) and the chromatograms of standard-1 (Figure 7 and 8) and standard-8 are shown below.

pharmaceutica-analytica-acta-Linearity

Figure 6: Linearity Calibration Curve.

pharmaceutica-analytica-acta-Linearity

Figure 7: Linearity-Standard-1 Chromatogram.

pharmaceutica-analytica-acta-Chromatogram

Figure 8: Linearity Standard-8 Chromatogram.

Matrix effect: Processed and analyzed calibration standards in the same matrix which is to be used during validation experiment and also six different lots of plasma at LQC and HQClevels as per the procedure are described in the sample preparation section. % Nominal concentration was found to be 99. 3% (LQC) and 89.8% (HQC) for the dilutions, which passed the limit of 85–115%.

Precision and accuracy: Three batches were performed using six replicates at concentration levels LLOQ, LQC, MQC, HQC levels. The within batch % CV was 0.7% to 7.1% and % accuracy was found to be 88% to 108%. The results were found to be within ± 15%. The results of Intra batch precision and accuracy are tabulated in Table 2 and Inter batch precision accuracy is Table 3.

Quality control samples Concentration added (ng/ml) Concentration found(ng/ml) (mean ± S.D) Precision (%) CV Accuracy (%)
LLOQ QC 0.109 0.110 ± 0.005 4.50 100.91
LQC 0.284 0.270 ± 0.014 5.19 95.18
MQC 37.809 36.013 ± 0.308 0.85 95.24
HQC 75.619 72.361 ± 0.532 0.73 95.7

Table 2: Results for Intra batch Precision and accuracy.

Quality control samples Concentration added (ng/ml) Concentration found(ng/ml) (mean ± S.D) Precision (%) CV Accuracy (%)
LLOQ QC 0.109 0.110 ± 0.005 4.50 100.91
LQC 0.284 0.270 ± 0.014 5.19 95.18
MQC 37.809 36.013 ± 0.308 0.85 95.24
HQC 75.619 72.361 ± 0.532 0.73 95.7

Table 3: Results for Inter batch Precision and accuracy.

Recovery: The percentage mean recovery for trimipramine was found to be 93.2%. The mean recovery of the ISTD was found to be 100.4% at a concentration of 200 ng/ml.

Stability: Stability of analyte was assessed at all processing and sample storing conditions. The stability results are tabulated in Table 4.

Stability Experiments Sample Concentration (ng/ml) Concentration found (ng/ml) (mean ± SD)
Three Freeze thaw cycles LQC 0.283 0.2731 ± 0.0078
HQC 75.518 71.7145 ± 3.0689
Autosampler stability LQC 0.283 0.2695 ± 0.0032
HQC 75.518 71.3601 ± 0.6432
Bench top LQC 0.283 0.2925 ± 0.0089
HQC 75.518 73.9863 ± 2.6501
Wet extract LQC 0.283 0.3093 ± 0.0059
HQC 75.518 81.6491 ± 1.0609

Table 4: Stability studies of the analyte.

Freeze thaw: The stability of analyte was determined in plasma during freeze-thaw cycles by freezing at -70ºC in freezer and thawing at room temperature for three cycles (interval of 12 hours) and thawing for minimum of one hour for every cycle. The precision was found to be 2.86% (LQC) and 4.27% (HQC) and mean % bias was found to be -3.47 (LQC) and -5.03 (HQC). The results were found to be within ± 15%.

Benchtop: The samples were stored unprocessed for 6 hours on bench at room temperature. The precision was found to be 3.07% (LQC) and 3.5% (HQC) and mean % bias was found to be 3.3% (LQC) and -2.02% (HQC).

In-injector: The processed sample was stored at autosampler temperature (4ºC) for a period of 12 hours. The precision was found to be 1.8% (LQC) and 0.9% (HQC) and mean % bias was found to be -4.7% (LQC) and -5.5% (HQC).

Wet extract: The processed sample was stored at room temperature for a period of 12 hours. The precision was found to be 1.9% (LQC) and 1.2% (HQC) and mean % bias was found to be 9.3% (LQC) and 8.1% (HQC).

Conclusion

A sensitive, specific, accurate and reproducible LC-MS/MS method has been developed employing a simple protein precipitation extraction technique for the quantification of trimipramine maleate in human plasma using opipramoldihydrochlorideas an internal standard and validated. The desired sensitivity for trimipramine maleate was achieved with an LLOQ of 0.1 ng/ml. The method can be applicable for routine bioavailability/bioequivalence studies and also determination of pharmacokinetics and therapeutic drug monitoring.

Acknowledgement

Authors are thankful for Clinigene International Ltd, Bengaluru for the facilities provided and also thankful for R & L fine chemicals for providing the drug.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Article Usage

  • Total views: 12376
  • [From(publication date):
    August-2015 - Dec 10, 2018]
  • Breakdown by view type
  • HTML page views : 8525
  • PDF downloads : 3851
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2018-19
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri and Aquaculture Journals

Dr. Krish

[email protected]

+1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

immunom[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

streamtajm

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2018 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version