Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on Medical, Pharma, Engineering, Science, Technology and Business

Development of Novel, Bioresorbable, Small-Diameter Electrospun Vascular Grafts

Jed Johnson*, Devan Ohst, Tyler Groehl, Sarah Hetterscheidt and Matthew Jones

Nanofiber Solutions, Columbus, Ohio, USA

Corresponding Author:
Jed Johnson
Nanofiber Solutions, Kinnear Road
Columbus, OH, USA
Tel: (614)-453-5877
Fax: (614)-487-3704
E-mail: [email protected]

Received date: February 12, 2015; Accepted date: March 12, 2015; Published date: March 25, 2015

Citation: Johnson J, Ohst D, Groehl T, Hetterscheidt S, Jones M (2015) Development of Novel, Bioresorbable, Small-Diameter Electrospun Vascular Grafts. J Tissue Sci Eng 6:151. doi:10.4172/2157-7552.1000151

Copyright: © 2015 Johnson J et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Tissue Science & Engineering

Abstract

This study proposes a production method capable of producing vascular grafts from fully synthetic, resorbable polymers that both meet basic minimum mechanical requirements for potential vascular grafts, and have a compliance similar to that of the intended vasculature being replaced. All of the electrospun vascular grafts in this work meet the minimum mechanical requirements for compliance, burst pressure, and suture retention strength, and could be potential candidates for off-the-shelf tissue engineered vascular grafts. Each polymer investigated in this paper has FDA approval for medical use and has been shown to be successful in various tissue engineering applications. Only recently has an electrospun small-diameter graft been fabricated with compliance and burst pressure greater than that of the human saphenous vein. We show a significant advancement in burst pressure, compliance, and suture retention strength in the novel electrospun grafts presented in this work which demonstrates the potential use of these tissue engineered vascular grafts for coronary artery bypass graft and other smalldiameter graft indications.

Keywords

Nanofibers; Vascular graft; Conduit; Shunt; Electrospinning; Tissue engineering

Introduction

Cardiovascular disease is one of the leading causes of death in western countries. According to the American Heart Association, almost 3 million procedures associated with the heart or blood vessels are performed in the United States each year, with over $25 billion spent on 400,000 coronary artery bypass graft (CABG) procedures [1-4]. Success has been found in creating vascular grafts with diameters greater than 6 mm because their massive blood flow allows for them to overcome native thrombogenicity in the graft material [5]. However, there is still a great demand for small diameter (<6 mm) vascular grafts for vessels such as the coronary artery. Autologous vein, specifically the saphenous vein, remains the gold standard of vascular replacement, but its use is restricted in up to 20% of patients due to comorbidities [6] and the replicability of grafts from older donors is limited [7,8].

The alternative to an autologous vein is the use of synthetic grafts made from either expanded polytetrafluoroethylene (ePTFE) or polyethylene terephthalate (PET), also known as Gore-Tex™ (PTFE) or Dacron™ (PET), respectively. In circumstances where autologous sources are in limited supply, such as combat wound scenarios [9], these synthetic grafts are appealing because they are readily available and can be used directly off-the-shelf. The use of these synthetic grafts has a long history, but their problems are well known. Infection, thrombosis, compliance mismatch, and the inability to match biomechanical properties of the native vasculature are problems with these synthetics [8,10-12]. Therefore, research is being done on new alternatives to ePTFE and PET for small caliber vascular replacements.

Since the greatest clinical need for vascular replacements is for CABG, mechanical properties of the vascular graft such as compliance and burst pressure are critical. The compliance mismatch between the graft and the native vessel disrupts blood flow resulting in zones of recirculation, flow separation, and low wall shear stress [13]. Low wall shear stress initiates the release of vasoactive substances, gene activation, protein expression, and cytoskeletal rearrangement that stimulate neointimal hyperplasia [14]. This is seen as a cause of failure in both autologous and prosthetic bypass grafts, though saphenous vein grafts also suffer from stenosis at sites away from the anastomosis [15].

Tissue engineering offers an alternative approach to existing vascular repair methods. Tissue engineering was first proposed in the 1980’s, with the general idea that biodegradable scaffolds could be used for cell attachment and proliferation. These scaffolds, when implanted into the body, can help to facilitate the regeneration of the native tissue. The formation of neotissue is critical to the success of a tissue engineered vascular graft (TEVG), the most crucial being a fully formed endothelium. The ideal TEVG has high biocompatibility, matching compliance to the native vessel, high suture retention strength and burst pressure, non-thrombogenic, resistance to neointimal hyperplasia, resistance to infection, easy to suture, and easily manufactured [16]. Matching the native vessel compliance is arguably the most critical attribute and requires the ability to finely tune the mechanical properties of the graft to that of the native vasculature.

There are several tissue engineering approaches being taken to address this challenge. One approach taken by Cytograft™ is a cell sheet method, which bypasses the use of biodegradable polymer scaffolds. In this approach, fibroblasts are cultured in a culture flask with a high sodium ascorbate environment to promote the formation of an extracellular matrix. This cell sheet is then matured and wrapped around a Teflon-coated, stainless-steel temporary support tube and continued to mature for at least 10 weeks [2]. Although the Cytograft™ cell sheets have found some success in clinical trials [17], the manufacturing time of the cell sheet vascular grafts can be upwards of 36-40 weeks. This raises concerns about the cost to produce these grafts and the ability to be available off-the-shelf. Although the Cytograft™ graft has be shown to match compliance of the saphenous vein, there is compliance mismatch when compared to the coronary artery [2].

Another approach, used by Humacyte™, is to grow vascular smooth muscle cells on a biodegradable polyglycolic acid (PGA) scaffold in vitro then decellularize the cultured blood vessel to produce a cell-free vascular graft [18]. Ideally, this leaves behind a new scaffold comprised entirely of the extracellular matrix left behind by the decellularized vessel presumably without risk of an immunogenic response [19]. The inherent risk is that the cell-based material is not completely removed from the extracellular matrix leaving the possibility of an immunogenic response. Additionally, the manufacturing time of the Humacyte™ vascular graft is at least 10 weeks, raising concerns about the cost to produce these grafts and ability to be available off-the-shelf [18].

Another widely used and successful approach to tissue engineering is the use of electrospun nanofiber scaffolds. These fibers mimic the native extracellular matrix architecture and have been shown to be useful in producing small caliber vascular graft replacements [12,20-22]. These grafts can be created from various polymers and can be co-spun with natural polymers and peptides [21,23-25]. These grafts retain the off-the-shelf appeal of the traditional synthetic grafts, but are designed to support the formation of neotissue and thus avoid the problems of ePTFE and PET because of their improved biocompatibility. Additionally, the burst pressures of these electrospun grafts can be customized by adjusting the electrospinning parameters and have been shown to be comparable to that of the native vasculature [26]. It is also possible to produce grafts with various fiber sizes and pore sizes by altering the polymer/solvent combination and the concentration of the polymer solute. By altering these electrospinning parameters, the fibers can be given many different biomechanical and biochemical properties which could help with matching the native vasculature compliance and in the formation of neotissue. Co-spinning, co-spraying, coaxial setups, micro-patterning, and use of hydrogels can also be implemented in order to increase the biocompatibility of these scaffolds [27,28].

This paper will focus on the biomechanical properties of various types of synthetic, fully resorbable polymers that have been electrospun into scaffolds for use in TEVGs. Each of these polymers has FDA approval for medical use and has been shown to be successful in various tissue engineering applications. Only recently has an electrospun small-diameter graft been fabricated with compliance and burst pressure greater than that of saphenous vein [29]. We show a significant advancement in burst pressure, compliance, and suture retention strength (SRS) in the novel electrospun grafts presented in this work. In an attempt to establish a baseline characterization and comparison in electrospun grafts to commercially available grafts, we also show the mechanical properties from a variety of natural and synthetic grafts. As the research community continues to build a critical mass of knowledge in the design and use of electrospun TEVGs, we hope to translate this promising technology to the clinic and improve patient care across the world.

Materials and Methods

Polymer Solution Preparation

Several different resorbable polymers were used to fabricate the TEVGs as shown in Table 1. Polymer solutions were prepared for electrospinning by dissolving the polymer in hexafluro-Isopropanol (HFIP) (Oakwood Chemicals). PCL/CS was prepared differently using a blend of HFIP and Acetic Acid (AA) (Sigma-Aldrich) at a weight ratio of 9:1. Each solution was mixed for at least 24 hours at room temperature before electrospinning to ensure homogeneity.

Polymer Abbreviation Manufacturer Ratio of polymers wt% of polymer in solution
Polycaprolactone PCL Sigma-Aldrich - 5
Polycaprolactone and Chitosan blend PCL/CS Sigma-Aldrich 20:1 3.15
Polydioxanone PDO Evonik - 7
Polyglycolic acid PGA Foryou - 10
Poly (lactic-co-caprolactone) PLCL Purac 70:30 5
Poly (lactic-co-glycolic acid) PLGA8218 Purac 82:18 6
Poly (lactic-co-glycolic acid) PLGA1090 Purac 10:90 8
Poly-L-lactide PLLA Good fellow - 8

Table 1: Polymers used in each solution and their solution weight percent.

Electrospinning

A 20cc syringe (Becton-Dickenson, USA), with a 20-gauge needle tip, was filled with the polymer solution. The 20cc syringe was secured in a syringe pump (Harvard Apparatus, USA), and positioned to achieve a tip to collector distance of 20 cm. The solution was ejected at 5 mL/hr with an applied voltage between +9 kV and +20 kV on the needle and between -1 kV and -4 kV on the collector, such that a stable Taylor Cone was established. Each polymer solution was electrospun onto a cylindrical mandrel rotating at 15 rpm with an outer diameter of 6 mm, until the fiber deposition generated a vascular graft with a wall thickness of 650 μm ± 65 μm. The vascular grafts were soaked in 100% ethanol and removed from the mandrel for characterization.

Mechanical Testing

Data for the compliance, burst pressure, and SRS were acquired using a universal testing machine (MTS Systems Corporation) fitted with a 50 lb load cell with a force resolution of 10-4 pounds and a linear displacement resolution of 10-8 inches. Compliance testing was performed using a displacement velocity of 1.5 mm per minute and acquisition rate of 4 data points per second utilizing Laplace’s Law [30,31] to correlate linear force and displacement to compliance. Burst pressure testing was performed using a displacement velocity of 50 mm per minute and acquisition rate of 4 data points per second utilizing Laplace’s Law [31,32] to correlate linear force and displacement to burst pressure.

Linear Tensile Mechanical Testing

Using a scalpel, the vascular graft was cut perpendicular to the longitudinal axis to form a cylindrical sample with a length of no less than the internal diameter of the graft. The cylindrical sample was then placed around two steel rods, one rod was attached to the load frame base of the universal testing machine and the other to the load cell. The sample was strained perpendicular to the longitudinal axis. With this test setup and assumptions [30,31], we assume that the circumference (C) is increased by twice the displacement (δ). This is displayed in equation 1 with DO being the original diameter.

πDO +2δ =Ca            (1)

Assuming that a fractional change in circumference is equal to a fractional change in the diameter, we can use equation 2. This equation relates systolic circumference (CS) and diastolic circumference (CD) to systolic diameter (DS) and diastolic diameter (DD).

equation            (2)

equation 3 substitutes the systolic and diastolic forms of equation 1 into equation 2.

equation            (3)

The calculation for compliance (C%) (equation 4) can be simplified where PS represents the systolic pressure and PD represents the diastolic pressure.

equation            (4)

Substituting equation 3 into equation 4 will give us equation 5.

equation            (5)

The pressure within the vascular graft can be calculated using the data from a universal testing machine and equation 6 where P is pressure, F is the force perpendicular to the longitudinal axis, and l is the length of the graft. To represent native blood pressures, systolic and diastolic pressures of 120 mmHg and 80 mmHg, respectively, were used.

equation            (6)

Substituting the systolic and diastolic forms of equation 6 into equation 5 yields equation 7. With this equation and the data received from the universal testing machine the compliance was calculated.

equation            (7)

Using the universal testing machine, the vascular graft was cut perpendicular to the longitudinal axis to form a cylindrical sample with a length of no less than the internal diameter of the graft. The cylindrical sample was then placed around two steel rods, one rod was attached to the load frame base of the universal testing machine and the other to the load cell. The sample was strained perpendicular to the longitudinal axis until failure. With this test setup and assumptions [31,32], the burst pressure (PB) was determined using wall thickness (TW), ultimate tensile strength (UTS) (σts), and the initial internal diameter (DO). (equation 8)

equation            (8)

Suture Retention Strength

SRS data was collected using a universal testing machine setup according to ISO 7198. A 2-0 catgut suture was pierced through the wall of the scaffold 2.0 mm from the edge and tied to form a loop. The non-sutured end of the scaffold and the suture loop were each attached to separate clamps in the universal testing machine and the suture pulled at 50 mm per minute with an acquisition rate of 4 data points per second until the thread was pulled through the scaffold.

Catheter Balloon Mechanical Testing

An alternative method to measure the burst pressure and compliance of the vascular grafts was also used to characterize the electrospun grafts in this paper and outlined in ISO 7198:1998. A 60cc syringe (Becton-Dickenson, USA) was filled with water and placed in a syringe pump (KD Scientific, USA). The syringe pump was set to a constant flow rate of 5 mL/min. Surgical tubing was connected to the 60 mL syringe, and passed through a pressure transducer (Honeywell). The end of the surgical tubing was connected to a FR18 pediatric Foley catheter (Coloplast, Denmark).

A 2.5 cm long section of the vascular graft was positioned directly over the catheter balloon. The vascular graft section was centered in the field of view of a High Accuracy CCD Micrometer (Keyence, USA). Pressure and scaffold diameter readings were taken using Labview 2010 software and recorded four times per second. Testing was stopped at the point of failure of the scaffold or when the pressure reached 30 psi due to physical constraints of the catheter, tubing connections, and syringe pump.

Compliance (C%) for this test was calculated using the compliance equation (equation 4) where PS is the systolic pressure, PD is the diastolic pressure, DS is the diameter at the systolic pressure, and DD is the diameter at the diastolic pressure.

Results

Compliance

Figure 1 provides a visual comparison for the compliance values of the TEVGs to commercially available grafts, the carotid artery and saphenous vein. The PLCL graft exhibited the highest value for compliance at 8.2%/mmHg, while the PDO graft held a value of 5.7%/ mmHg. Both of these electrospun polymer grafts displayed a higher compliance value than the human carotid artery, with an average compliance of 5.4%/mmHg. The PGA graft exhibited a slightly lower compliance than the carotid artery, at 4.8%/mmHg. The human coronary artery, PLLA, silk fibroin, Cytograft™, PCL, and PCL/CS grafts all displayed similar compliance values ranging from 3.2-3.8%/ mmHg. The bovine heterograft and PLGA8218 graft exhibited even lower compliances at 2.6 and 2.5%/mmHg, respectively. The electrospun polymer graft that demonstrated the lowest compliance was PLGA1090 at 1.9%/mmHg. The Dacron knitted graft and the IMPRA ePTFE graft were even less compliant at 1.9 and 1.6%/mmHg, respectively. The human saphenous vein had the lowest compliance of all at 1.1%/mmHg.

tissue-science-engineering-electrospun-vascular-grafts

Figure 1: Compliance values for the electrospun vascular grafts, commercially available synthetic grafts, bovine heterograft, silk fibroin graft, CytograftTM, human carotid artery, human coronary artery, and human saphenous vein [1-3]. Error bars represent standard deviation when available.

Burst Pressure

Burst pressure values for the electrospun grafts, commercially available grafts, human carotid artery and human saphenous vein are displayed in Figure 2. The PLGA1090 graft had the highest burst pressure at 3.3 MPa followed closely by the Dacron knitted graft with a burst pressure of 3.2 MPa. The PLCL graft had a burst pressure strength of 2.5 MPa, while the PLGA8218 and PCL/CS grafts displayed lower burst pressure strengths of 2.2 and 1.9 MPa, respectively. Subsequently, the PLLA graft exhibited a burst pressure strength of 1.8 MPa and the PGA graft was 1.5 MPa. The electrospun grafts with the lowest burst pressure strengths were PDO and PCL, at 0.9 MPa and 0.8 MPa. The human carotid artery, human saphenous vein, Cytograft™, IMPRA ePTFE graft and silk fibroin graft all displayed significantly lower burst pressure strengths when compared to our electrospun grafts, with values ranging from 0.1 MPa to 0.5 MPa.

tissue-science-engineering-synthetic-grafts

Figure 2: Burst pressure values for the electrospun vascular grafts, commercially available synthetic grafts, silk fibroin graft, CytograftTM, human carotid artery, and human saphenous vein [1,2]. Error bars represent standard deviation when available.

Suture Retention Strength

SRS values, measured in grams, are displayed in Figure 3. The Dacron knitted graft exhibited the highest SRS value when compared to the electrospun TEVGs, commercially available grafts, human carotid artery and human saphenous vein, with a value of 3,270 g. The electrospun graft with the highest SRS was PLLA at 1,022 g followed by the PLGA1090 and PLGA8218 grafts with values of 997 g and 995 g, respectively. The PGA, PDO, and PCL grafts exhibited values in the middle of the data set, ranging from 674 g to 748 g.

tissue-science-engineering-retention-strength

Figure 3: Suture retention strength values for the electrospun vascular grafts, commercially available synthetic grafts, CytograftTM, human carotid artery, and human saphenous vein [1,2]. Error bars represent standard deviation when available.

The PLCL graft followed next with a SRS of 521 g. The IMPRA ePTFE graft had a lower SRS of 451 g and the PCL/CS graft displayed the lowest SRS for the electrospun grafts at 355 g. The human carotid artery, human saphenous vein, and Cytograft™ exhibited the lowest SRS values of the data set ranging from 152 g to 200 g.

Compliance and Burst Pressure as a Function of Wall Thickness

To investigate the potential effects of sidewall thickness on the graft compliance and burst pressure, we electrospun PCL grafts of varying sidewall thicknesses and tested their mechanical properties as described previously. As shown in Figure 4A, compliance remains relatively constant once the graft sidewall thickness reaches a thickness of approximately 400 μm. For sidewall thicknesses of 400 μm to 1,000 μm, the compliance remains steady within a range of 2-4%/mmHg. However, it is visible that at smaller sidewall thickness values, the PCL grafts exhibit a large variance in compliance measurements ranging from 2-11%/mmHg.

tissue-science-engineering-PCL-grafts

Figure 4: Compliance (A) and Burst Pressure (B) values for electrospun PCL grafts versus wall thickness.

Figure 4B illustrates that the burst pressure increases as the graft sidewall thickness increases. For the PCL grafts with a sidewall thickness ranging from 200-1,000 μm, the burst pressure values increase in a linear fashion from 0.6 MPa to 2.9 MPa. This linear relationship is displayed by the best fit line applied to the data points in a scatter plot.

Burst Pressure Method Comparison

To investigate any potential difference in the burst pressure values obtained utilizing the linear tensile approach with equation 8 versus those values obtained with the catheter balloon method, we tested five electrospun PCL samples with the linear tensile method and five electrospun PCL samples with the catheter balloon method of identical wall thicknesses. The results are shown in Figure 5, where the linear tensile method averaged a 30% higher burst pressure compared to the catheter balloon method.

tissue-science-engineering-electrospun-PCL

Figure 5: Comparison of burst pressure values of electrospun PCL grafts for the linear tensile method and the catheter balloon method.

Discussion

Of the data found on the compliance of commercially available vascular grafts, none showed a compliance that matched that of the human carotid artery. The average compliance of native human carotid artery (5.4%/mmHg) was approximately 2x greater than the most compliant commercially available vascular graft (bovine heterograft 2.6%/mmHg) included in this study (Figure 1). Similarly, the average compliance of human coronary artery (3.8%/mmHg) was approximately 1.5x of the most compliant commercially available graft. This compliance mismatch may be a major contributing factor in the poor clinical performance of these grafts. In this study, multiple types of biodegradable, synthetic electrospun grafts were fabricated that had compliance values near that of the human carotid and coronary arteries. The PDO and PGA grafts had a compliance of 5.7 ± 2.2%/mmHg and 4.9 ± 1.0%/mmHg, respectively, similar to that of the human carotid artery with a compliance of 5.4 ± 0.9%/mmHg (Figure 1). The PLLA graft had a compliance of 3.8 ± 0.9%/mmHg, similar to that of the human coronary artery with a compliance of 3.9 ± 0.6%/ mmHg.

Other polymers investigated in this paper may match the compliance of the desired vasculature by manipulating the wall thickness of the graft. For example, the wall thickness of the PLGA1090 graft could be decreased to increase the compliance since the burst pressure value is relatively high. While most of the concern in optimizing the mechanical properties of vascular grafts is focused on the choice of the polymer and the method of production, this work illustrates the significant importance of the wall thickness on the mechanical properties of the graft. The differences in the mechanical properties based on dimensional changes have been shown to be greater than the differences in mechanical properties between polymers. This leads to the idea that when characterizing the mechanical properties of potential vascular grafts, the dimensions of the grafts being implanted in vivo must match the dimensions of the grafts for mechanical characterization, otherwise there is a potential for significant differences in the mechanical properties that could lead to complications in vivo .

All of the burst pressures of the tested electrospun TEVG compositions are above those seen in native human saphenous veins and carotid arteries (Figure 2). This further supports the notion of adjusting the sidewall thickness to achieve the desired graft compliance, as long as the minimum burst pressure is achieved.

All of the electrospun TEVGs in this study surpassed the 2 N or 204 g standard minimum SRS recommended for potential vascular prosthetics [33] (Figure 3). While the SRS may not be as critical for clinical performance as the burst pressure and compliance, the duration of the surgery and risk of complications to the patient may increase if the sutures tear through the vascular graft during anastomosis in vivo .

Investigating the effect of wall thickness on the compliance of the vascular graft reveals a significant change in the compliance values when the wall thickness was below 400 μm as compared to grafts with a wall thickness greater than 400 μm (Figure 4A). When the graft wall thickness was less than 400 μm, there was a significant increase in the variability of compliance values and an overall average increase in the compliance values. This abrupt change in mechanical behavior is indicative of a change in how the nanofibers within the graft react to the applied stresses. The thin walled electrospun grafts have a higher percentage of fibers exposed to the outer surface (i.e. no surrounding fibers) than the thicker grafts, allowing for increased fiber mobility within the sidewall which allows the fibers to re-orient in response to the applied stresses. After a certain thickness threshold is reached, in this case approximately 400 μm, the percentage of fibers exposed to the outer surface (i.e. no surrounding fibers) becomes negligible and the fiber motion is restricted which results in a less compliant vascular graft.

The relationship between wall thickness and burst pressure of the vascular graft is evident (Figure 4B). As expected, a strong linear correlation can be seen between the burst pressure of the graft and the wall thickness of the graft. As the wall thickness increases, more fibers are present which results in more material present to bear the applied pressure. The greater number of fibers also increases the chance for the fibers to re-orient themselves in the direction of the loading [34,35].

Comparison of the two methods, the linear tensile method using equation 8 and the catheter balloon method, for determining the burst pressure of a small diameter vascular graft revealed a significant difference in the values of the grafts (Figure 5). The linear tensile method was shown to have average burst pressures approximately 30% higher than similar grafts tested using the catheter balloon based method. The difference in burst pressure is likely due to differences in the mobility of the individual fibers in the nanofiber mesh. Previous studies have shown that balloon based methods of burst pressure determination result in artificially lower burst pressures due to the pinning of fibers, which prevents the fibers from translating and reorienting in the direction of loading [29,34,35]. Visual observation of the failure of the electrospun grafts during the catheter balloon burst pressure testing method demonstrated that the pressurized balloon squeezes and compresses the sidewall of the nanofiber mesh, preventing the nanofibers from re-orienting as the diameter increases and initiates failure of the graft in a very localized region. In contrast, visual observation of the failure of the grafts during the linear tensile method showed no signs of localized failure. Although the catheter balloon method of mechanical testing may be more representative of the pressures experienced by the graft in vivo, the limitations in the sensitivity of the pressure transducer at lower pressures and the limitation of the strength of the catheter balloon itself at maximum pressure limits the usefulness of the catheter balloon method in the determination of the compliance and burst pressure of these small diameter vascular grafts.

Conclusions

All of the electrospun vascular grafts in this work meet the minimum mechanical requirements for compliance, burst pressure and SRS, and could be potential candidates for off-the-shelf tissue engineered vascular grafts. Each of the polymers investigated in this paper has FDA approval for medical use and has been shown to be successful in various tissue engineering applications. Only recently has an electrospun small diameter graft been fabricated with compliance and burst pressure greater than that of a human saphenous vein. We show a significant advancement in burst pressure, compliance and suture retention strength in the novel electrospun tissue engineered grafts presented in this work. Further in vivo testing will be necessary to determine the degradation rate and remodeling profile for each of these TEVGs before they can be translated to the clinic.