Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Endoscopic Ultrasonography Fine Needle Aspiration in Pancreatic Masses: A Study Comparing 22-gauge and 25-gauge Needle in the Same Patient

Vasseur P1,2, Puppa G3, Nguyen T1, Genevay M3 and Frossard JL1*

1Service of Gastroenterology and Hepatology, Geneva University Hospital, Rue Gabrielle-Perret-Gentil 4, 1205 Genève, Switzerland

2Service of Gastroenterology, CHU Poitiers, 2 Rue de la Milétrie, 86021 Poitiers, France

3Service of Clinical Pathology, Geneva University Hospital, Rue Gabrielle-Perret-Gentil 4, 1205 Genève, Switzerland

*Corresponding Author:
Frossard JL
Service of Gastroenterology and Hepatology
Geneva University Hospital
Rue Gabrielle-Perret-Gentil 4
1205 Geneve, Switzerland
Tel: 41223729340
Fax: 41223729366
E-mail: [email protected]

Received Date:November 18, 2015 Accepted Date: December 26, 2015 Published Date: December 28, 2015

Citation: Vasseur P, Puppa G, Nguyen T, Genevay M, Frossard JL (2016) Endoscopic Ultrasonography Fine Needle Aspiration in Pancreatic Masses: A Study Comparing 22-gauge and 25-gauge Needle in the Same Patient. J Cytol Histol 7:384. doi:10.4172/2157-7099.1000384

Copyright: © 2015 Frossard JL, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Cytology & Histology

Abstract

Background: It remains unclear whether the 22- and 25-gauge needles have equal diagnostic yields in endoscopic ultrasonography–fine needle aspiration of pancreatic masses. Methods: Patients presenting with pancreatic solid masses were prospectively included between July 2010 and June 2012. Each patient underwent ultrasonography–fine needle aspiration using two different needle sizes (22- and 25-gauge needles) during the same endoscopic session, the 25 and 22 needle sequence being randomized. Cytological preparations included smear cytology, ThinPrep and Cell Block. Specimens were analyzed for diagnosis, cellularity, amount of blood and digestive contamination. Final diagnosis was reached by the final pathological report, clinical and radiological follow-up. Results: Thirty-seven patients were included among them 34 subjects had malignant pancreatic neoplasm on final diagnosis. Pancreatic masses were located in the head of pancreas in 73% of cases. The mean size of the lesions was 34.6 ± 14.4 mm. Diagnostic yields of the 25- and 22-gauge needles were 85.3% (95%CI: 72.6-98) and 88.2% (95%CI:76.7-99.7) respectively (p > 0.05). Diagnosis was obtained with smear cytology alone in 58.8% patients with the 22-gauge needle and in 55.9% patients with the 25-gauge needle, in 64.5% patients with Cell Block for both needles and in 76.5% with ThinPrep for both needles. The quality of specimens was comparable with the two needles. No complication occurred. Conclusion: The 25- and 22-gauge needles carry the same diagnostic yield when performing ultrasonography– fine needle aspiration of solid pancreatic masses. ISRCTN69545565 Controlled trials.

Keywords

Endoscopic ultrasonography; Fine needle aspiration; Pancreatic neoplasms

Introduction

Fine-needle aspiration guided by endoscopic ultrasonography (EUS-FNA) has changed the management of solid pancreatic masses. While EUS can provide supplementary information regarding the staging of pancreatic tumors, FNA can be safely performed on pancreatic masses and/or peripancreatic lymph nodes to bring the ultimate diagnosis of malignancy. Because the differential diagnosis of solid pancreatic masses is wide including benign, premalignant and malignant conditions, sampling the mass is of great interest before sending the patient to demanding and complicated surgery. However, tumor sampling is currently recommended only when the pancreatic mass does not seem amenable to resection because of staging results or poor physical condition. In case of potentially resectable tumor, EUSFNA is usually performed in order to exclude unusual tumors that would not benefit from surgery such as lymphoma [1]. EUS-FNA has a high accuracy in diagnosing malignant pancreatic tumors. A recent meta-analysis including almost 5’000 patients who underwent EUSFNA indicates a pooled sensitivity of 85% when considering malignant cytology [2]. Of the three needle sizes available to puncture solid pancreatic masses (19-gauge, 22-gauge or 25-gauge), the 22-gauge needle is the most widely used device although there is no clear recommendation to choose among them. It is indeed unclear whether the 25-gauge needle, which is the thinner one, could provide equal or better performance in sampling the pancreas than the 22-gauge needle. It is also unclear whether the 25-gauge needle can decrease the risk of FNA-induced bleeding and gastro-intestinal contamination in the smears and cell block. In prospective studies comparing the results brought by EUS FNA using a 22-gauge and a 25-gauge needles for pancreatic masses, the 25-gauge needle appeared to have the same diagnostic yield than the 22-gauge needle [3,4]. However, the design of these studies did not compare the results obtained by the two needle types in the same patient, but compared the respective yield of each needle used in two different groups of patients randomly assigned for either a 25-gauge puncture or a 22-gauge puncture. The validity of these trials is also limited by the low number of patients enrolled [5-7]. The aim of this study was to compare, in a larger cohort, the diagnostic accuracy of a 25-gauge and a 22-gauge needles in the same patient presenting with pancreatic solid mass.

Patients and Methods

Trial design and intervention

The study is a randomized, prospective, non-inferiority study aimed at evaluating the diagnostic yield of EUS FNA using both 25- and 22- gauge needle in the same patient. The protocol was approved and cleared by our local Ethical Committee. The reporting of the trial was based on the CONSORT 2010 recommendations. The study was conducted at the Geneva university Hospital and was given number ISRCTN69545565 from Controlledtrials.com (www.controlledtrials. com).

EUS FNA was performed using two needle, a 25-gauge and a 22- gauge needle, in the same patient. The randomization sequence of the needle size was created with a 1:1 allocation using blocks of 4 without stratification. The allocation was achieved in the operating room by physicians blinded for the allocation sequence, using sequentially numbered, sealed and opaque envelopes. The corresponding envelope was opened only after the eligible participant had signed an informed consent form explaining all the modalities of the endoscopic procedure. Data collection was performed blindly by a noninterventional gastroenterologist.

EUS FNA procedure

EUS FNA was performed by three endosonographers (JLF, JMD and TNG). Patients were sedated using propofol in a left lateral position. A linear echoendoscope (EUM 160, Olympus, Hamburg, Germany) was used for FNA, doppler analysis was performed to avoid any intervened vessel in the path of the needle. Three passages without stylet in place were performed with each needle in the solid pancreatic mass, a trangastric approach being used for body and tail lesions while a transduodenal approach was used for lesions in the head and uncinate process of the pancreas. A 10 cc succion was used for each passage. The needle was rinsed with 0.5 ml of NaCl 0.9%. The collected specimen from each needle were first smeared and rinsed for Cell Block and ThinPrep preparations.

Complications associated with FNA were defined as GI bleeding by either mouth or anus, abdominal pain, acute pancreatitis and fever every item occurring within 24 hours after the procedure.

Cytopathology assessment and final diagnosis

Specimens obtained with the two needles were blindly reviewed by 2 experienced cytopathologists (MG, GP). The material that was separately evaluated included: thin-Prep, cell block and cytological smear. For each case slide labels indicated the type of needle used (22 or 25) thereby a distinct assessment was performed.

Every slide was scanned at low magnification and a semiquantitative assessment was applied allowing a 4-tiered classification, as follows:

0 = Absence; 1 = Very rare cells; 2 = Few cells aggregates and a few isolated cells;

3 = Several cells aggregates, up to covering almost all the slide.

These criteria were applied for evaluating the amount of: a) Tumor cells, b) Gastro-intestinal contamination and c) Red blood cells.

The final diagnosis was based on cytology report, surgical pathology if available, repeated radiological imaging and clinical follow-up.

Sample size, inclusions and exclusions criteria

The sample size calculation was based on previous studies comparing the two explored strategies (diagnostic yield of 25- and 22- gauge needle size) demonstrating a diagnostic yield of 85% in each arm. In order to detect the same diagnostic yield afforded by each needle size, when requiring a two-sided alpha = 0.05 and a statistical power of 0.8, a sample size of 35 patients was estimated as the minimum number as each patient was his/her own control. Thirty seven patients referred to our Hospital for EUS-FNA of solid pancreatic masses were prospectively enrolled between July 2010 and June 2012. Written informed consent was obtained before their participation in the study.

Only adults >18 years were eligible for the study. They all presented to our service with a solid pancreatic mass identified by at least two dissimilar imaging modalities (ultrasound, CT scan, MRI). Pancreatic mass was defined as abnormal nodule located within any segment of the pancreas.

Exclusion criteria were sepsis, acute pancreatitis defined as abdominal pain associated with increased serum lipase >3 the normal value), anticoagulant therapy, anti-agregant therapy other than aspirin, previous history of modified anatomy interfering with endoscopic assessment and incapacity to give informed consent.

Statistical analysis

Statistical analysis was performed using SAS version 9.3 (SAS Inc, Cary, NC, USA). Cochran-Mantel-Haenszel test was used for categorical variables (diagnostic yield), Wilcoxon signed ranks test was used for ordinal variables (contamination, amount of blood and cellularity). Results were considered statistically significant if P value <0.05. The minimum sample size was determined on the basis of a non-inferiority of the 25-gauge needle for diagnostic yield, admitting a loss of 0.4% in sensitivity.

Results

Between July 2010 and June 2012, 73 patients were assessed for eligibility. Forty-five patients did not meet the inclusion criteria. A total of 37 patients with pancreatic solid mass was prospectively enrolled. There were 21 men and 16 women with a mean age of 69.3 ± 12 years (Table 1). Pancreatic masses were mainly located (73%) in the head or in the uncinate process of the pancreas, the mean size of the lesions being 34.6 ± 14.4 mm (Table 1). Among the 37 patients included, 3 patients had a final diagnosis of non-malignant mass according to surgical reports, clinical follow-up (>18 month) and imaging studies. Inflammatory masses were related to either chronic pancreatitis in 2 patients and to acute pancreatitis in one case (Figure 1). In the 34 patients with a final diagnosis of malignant tumor, the majority of lesions (82%) were pancreatic carcinomas. Diagnosis was obtained by EUS-FNA in 32 patients (94%), whereas surgical biopsy was mandatory in the 2 remaining patients. None of the patients presented hemorrhagic or infectious complications after EUS-FNA.

cytology-histology-enrollment

Figure 1: Enrollment of patients

Characteristics of population  
Median age(y, range) 72(42-87)
Gender (males, %) 21(56.8)
Site of lesion (n, %)  
Head / uncinate 27(73)
Body / tail 10(27)
Median size of lesion (mm, range) 30.5 (17-82)

Table 1: Baseline characteristics

Cytological diagnosis

When considering the 34 patients with a final diagnosis of malignant tumor, diagnostic yield of 22-gauge and 25-gauge needles was significantly non-different (p = 0.37). Diagnosis was made in 30 patients (88.2%; 95%CI: 76.7-99.7) with the 22-gauge needle and in 29 subjects (85.3%; 95%CI: 72.6-98) with the 25-gauge needle. According to the type of cytological preparation (smear, ThinPrep or Cell Block), the diagnostic yield of both needles was not different (p > 0.05).

Diagnosis was obtained with smear cytology alone in 58.8% patients with the 22-gauge needle and in 55.9% patients with the 25-gauge needle, in 64.5% patients with Cell Block for both needles and in 76.5% with ThinPrep for both needles (Table 2).

  22-gauge 25-gauge  
Overall diagnosis     *
Yes (n, %) 30(88.2) 29(85.3)  
No (n, %) 4(11.8) 5(14.7)  
Diagnosis with smear cytology     *
Yes (n, %) 20(58.8) 19(55.9)  
No (n, %) 14(41.2) 14(44.1)  
Diagnosis with Thin-Prep     *
Yes (n, %) 26(76.5) 26(76.5)  
No (n, %) 8(23.5) 8(23.5)  
Diagnosis with Cell Block     *
Yes (n, %) 20(60.4) 20(60.4)  
No (n, %) 11(35.5) 11(35.5)  

Table 2: Diagnostic yield of the 22-gauge and 25-gauge needles

Cytological results

Smear cytology and ThinPrep were available in the 34 patients with a diagnosis a malignant tumor, Cell Block was available in 30 patients. Quality of samples did not differ between the two sizes of needles. Cellularity of samples, regardless of cytological preparation, was not inferior with 25-gauge needles. Furthermore, the contamination by digestive cells and the amount of blood was not different for the two needles (Table 3).

  Smear cytology   Thin-Prep   Cell Block  
22G 25G p 22G 25G p 22G 25G p
Amount of blood (n)     0.51     1.0     1.0
0 11 9   9 8   10 11  
1 10 9   15 16   12 11  
2 4 7   9 9   4 7  
3 9 9   1 1   4 2  
Contamination (n)     0.90     0.15     0.78
0 6 7   6 10   10 11  
1 13 12   16 15   17 15  
2 10 7   8 6   3 4  
3 5 8   4 3   0 1  
Cellularity (n)     0.52     0.12     0.98
0 11 14   7 9   10 9  
1 11 9   10 13   9 10  
2 7 6   11 9   2 8  
3 5 5   6 3   9 4  

Table 3: Cytological results of samples obtained with the 22-gauge and 25-gauge needles

Discussion

EUS-FNA has become a standard in pancreatic masses diagnosis and much work has been done to improve its diagnostic accuracy in the recent years. Although the 19-gauge needle and Trucut biopsy needles allow obtaining larger specimens, they carry a higher risk of bleeding and are less flexible in the trans-duodenal approach than thinner needles [8]. These features explain the reason why thinner needles either the 22-gauge or the 25-gauge needles are used instead for pancreatic masses. The latter may be in theory associated with a lower incidence of FNA-induced complications. In a retrospective study including 842 patients with pancreatic masses, no complication were noted in the 25-gauge needle group [9]. Moreover, this needle had a higher technical success rate and was easier to manipulate, especially in lesions of the pancreas located in the head and the uncinate process [3,5]. Nevertheless, it is still unclear whether the 25- and 22-gauge needles have the same diagnostic yield. Three randomized trials compared the diagnostic yield afforded by the two needles in patients with pancreatic masses and did not show any difference [3,4,10]. However, the full interpretation of these studies is somewhat difficult as they did not compare the accuracy of the two needles in the same lesions and furthermore in the same patient. Two pilot studies including 9 and 24 patients respectively compared the 22- and 25-gauge needles in the same patient and demonstrated both accurate diagnosis in these small groups of patients [5,6]. A third study collecting 50 patients lead to the same result with a diagnostic accuracy reaching 94% and 86% for the 25- and 22-gauge needles respectively [7]. We also show that both needles without the stylet in place have equal diagnostic yields for pancreatic masses, reaching 85% for the 25- gaugne needle and 88% for the 22-gauge needle. Depending on the type of cytological preparation, the size of the needle did not influence diagnostic accuracy. Combining the three cytological preparation increased the final accuracy, but we interestingly observed that ThinPrep showed a trend towards higher diagnostic yield than smear cytology (76% vs. 59%, p > 0.05); This characteristic is somewhat in contradiction with the findings of Leblanc et al. who showed that sensitivity of on-site smear cytology for pancreatic cancer was 98% versus 62% for ThinPrep [11]. We also show that the quality of cytological specimens was comparable with the two needles, which was consistent with previous studies [6,7]. The small number of patients included in this study may represent a limitation but could be explained by its particular design since it was a non-inferiority study. Nevertheless, to the best of our knowledge, this is the only study that studied the performance of two different needles in the same patient.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Article Usage

  • Total views: 7866
  • [From(publication date):
    January-2016 - Sep 25, 2017]
  • Breakdown by view type
  • HTML page views : 7814
  • PDF downloads :52
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri, Food, Aqua and Veterinary Science Journals

Dr. Krish

[email protected]

1-702-714-7001 Extn: 9040

Clinical and Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemical Engineering and Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001 Extn: 9040

Earth & Environmental Sciences

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

General Science and Health care Journals

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics and Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001 Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Informatics Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Material Sciences Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Mathematics and Physics Journals

Jim Willison

[email protected]

1-702-714-7001 Extn: 9042

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001 Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

John Behannon

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001 Extn: 9042

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords