Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

End-tidal CO2 Should not be a Parameter for Ventilatory Adjustment during Low Cardiac Output States Like Off-pump Coronary Artery Bypass Grafting

Manender Kumar Singla1*, Kanwalpreet Sodhi2, Anupam Shrivastava3, Manpreet Singh Salooja4, Kishore C. Mukherjee4
and Sonia Saini1
1Department of Cardiac Anesthesia, SPS Apollo Hospitals, Ludhiana, India
2Department of Critical Care, SPS Apollo Hospitals, Ludhiana, India
3Department of Anesthesia and Critical Care, SPS Apollo Hospitals, Ludhiana, India
4Department of Cardiothoracic and Vascular Surgery, SPS Apollo Hospitals, Ludhiana, India
Corresponding Author : Dr. Manender Kumar Singla, MD
Consultant, Department of Cardiac Anesthesia
SPS Apollo Hospitals, Ludhiana, India- 141003
Tel: +919465430748
E-mail: [email protected]
Received April 23, 2014; Accepted April 26, 2014; Published April 29, 2014
Citation: Singla MK, Sodhi K, Shrivastava A, Salooja MS, Mukherjee KC (2014) End-tidal CO2 Should not be a Parameter for Ventilatory Adjustment during Low Cardiac Output States Like Off-pump Coronary Artery Bypass Grafting. J Gen Pract 2:159. doi: 10.4172/2329-9126.1000159
Copyright: © 2014 Singla, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Related article at
DownloadPubmed DownloadScholar Google

Visit for more related articles at Journal of General Practice

Abstract

Background: Capnography is widely used as a non-invasive monitoring technique for intra-operative and postoperative mechanically ventilated patients. Aim: To study whether the EtCO2 values should be used to adjust mechanical ventilation during low cardiac output states such as off pump cardiac bypass surgery (OPCABG). Setting and design: Prospective study in 50 patients undergoing OPCABG. Material and Methods: The values of cardiac index (CI), EtCO2, PaCO2 and P(a-Et)CO2 gradient were compared at two time points during OPCABG: immediately before skin incision and during grafting of obtuse marginal artery, as usually OM grafting is considered to be associated with low cardiac output. Statistics: Data was analysed using student’s t-test. Results: There was a significant decrease in the value of CI from point A to point B (2.33 vs. 1.75 L/min/m2, p<0.001), significant decrease in EtCO2 from 25.28 to 21.88mmHg (p<0.001) and statistically significant increase in P(a-Et)CO2 gradient 4.39 at point A to 9.18 mmHg at point B (p<0.003). Conclusion: The decrease in EtCO2 had a positive correlation with the decrease in cardiac index. So during the low flow periods as found in OPCABG, EtCO2 cannot reliably estimate the adequacy of ventilation.

Abstract
Background: Capnography is widely used as a non-invasive monitoring technique for intra-operative and postoperative mechanically ventilated patients.
Aim: To study whether the EtCO2 values should be used to adjust mechanical ventilation during low cardiac output states such as off pump cardiac bypass surgery (OPCABG).
Setting and design: Prospective study in 50 patients undergoing OPCABG.
Material and Methods: The values of cardiac index (CI), EtCO2, PaCO2 and P(a-Et)CO2 gradient were compared at two time points during OPCABG: immediately before skin incision and during grafting of obtuse marginal artery, as usually OM grafting is considered to be associated with low cardiac output.
Statistics: Data was analysed using student’s t-test.
Results: There was a significant decrease in the value of CI from point A to point B (2.33 vs. 1.75 L/min/m2, p<0.001), significant decrease in EtCO2 from 25.28 to 21.88mmHg (p<0.001) and statistically significant increase in P(a-Et)CO2 gradient 4.39 at point A to 9.18 mmHg at point B (p<0.003).
Conclusion: The decrease in EtCO2 had a positive correlation with the decrease in cardiac index. So during the low flow periods as found in OPCABG, EtCO2 cannot reliably estimate the adequacy of ventilation.
Keywords
End-tidal CO2; Capnography; PaCO2; Off-pump coronary artery bypass grafting; Mechanical ventilation
Introduction
Capnography is routinely used in monitoring of intra-operative and postoperative mechanically ventilated patients [1]. It is a non-invasive technique used for continuous measurement of exhaled carbon dioxide (CO2) throughout the respiratory cycle, commonly referred to as end-tidal CO2 (PEtCO2 or EtCO2) [2]. EtCO2 is a measure of alveolar partial pressure of CO2 (PACO2), which in turn is a measure of arterial CO2 (PaCO2). It is commonly used to assess the correct placement of endotracheal tube, to monitor the integrity of mechanical ventilation equipment, weaning from mechanical ventilation and monitoring quality of resuscitation [2-4]. PEtCO2 is governed by metabolism, ventilation and circulation [2,5]. When the first two factors are controlled, PEtCO2 reflects the pulmonary blood flow and therefore cardiac output (CO) [3]. Attempts have been made to use PEtCO2 measurement as an alternative to thermo-dilution technique for determining CO in various cardiac surgical settings [3,6,7]. The normal gradient between PaCO2 and EtCO2 [P(a-Et)CO2] is approximately 4-5 mmHg which represents the normal dead space ventilation [2]. But as PEtCO2 is governed by pulmonary blood flow, it may not be true in conditions with acutely low CO [1,8]. In these situations, there is an increase ventilation-perfusion ratio (V/Q), which decreases PEtCO2 and increases P(a-Et)CO2 [1]. Applying this principal in clinical scenario means that the ventilatory settings should not be adjusted taking into account the low values of EtCO2 in different low output states.
Off-pump coronary artery bypass grafting (OPCABG) is a complex procedure with abrupt hemodynamic alterations and periods of low CO as a result of surgical manipulations, unprotected myocardial ischemia and use of stabilisers on the beating heart [9,10]. We wanted to study the changes in EtCO2 values during the cardiac output period in OPCABG and whether the EtCO2 can be relied upon as a marker of adequacy of ventilation during such low cardiac output states. For this, we assessed and compared the values of cardiac index (CI), EtCO2, PaCO2 and P(a-Et)CO2 gradient immediately before skin incision and during grafting of obtuse marginal artery (OM), as OM grafting is usually associated with a period of low CO [9].
Material and Methods
The study was approved by the hospital ethics committee. 50 patients planned for OPCABG with OM grafting were enrolled in the study. Written informed consent was taken from all patients preoperatively. On arrival in the operating room, monitoring was done in the form of continuous ECG with 5-electrodes and pulse oximetry. Under local anaesthetic infiltration, a 16-G cannula was inserted in antecubital vein and right radial artery was cannulated with 20-G cannula that was connected to a pressure transducer for arterial blood pressure (ABP) monitoring. After preoxygenation, general anaesthesia was induced with midazolam (0.03-0.05 mg/kg BW), fentanyl (3-5 mcg/kg BW) and propofol (1.5-2 mg/kg BW). Endotracheal intubation was facilitated with rocuronium (0.6 mg/kg BW) and mechanical ventilation was initiated after confirming the ET placement using 5-point auscultation and EtCO2 using side-stream capnography. All the patients were ventilated with a tidal volume of 6 mL/kg and a respiratory rate of 15/minute. After induction of anaesthesia, 7.5 Fr thermodilution pulmonary artery catheter (PAC) was inserted into right internal jugular vein. Other intra-operative monitoring included nasopharyngeal temperature and urine output. Anaesthesia was maintained with isoflurane in air-oxygen mixture, fentanyl and rocuronium. Normothermia was maintained using warm intravenous fluids and circulating warm water mattress under the patients. Medtronic octopus stabiliser was used for positioning of heart during beating heart surgery.
For study purpose, CI, EtCO2, PaCO2 and P(a-Et)CO2 values were recorded at two time points: immediately before skin incision (A) and during OM grafting (B). CI was calculated using thermodilution method by PAC (10 mL of saline at room temperature, at end expiration in triplicate). EtCO2 value was directly recorded from the side-stream capnograph monitor. Arterial blood gas sample from radial artery was withdrawn for estimation of PaCO2 values. During OM grafting, these parameters were recorded after arteriotomy and placement of shunt. P(a-Et)CO2 gradient was calculated as difference between PaCO2 and EtCO2.
Statistical analysis
The collected data is expressed as mean ± standard deviation. All data was analysed using SPSS 17.0 software. Student’s t-test was applied for statistical analysis. A p-value of <0.05 was considered significant.
Results
Out of the 50 study patients, there were 40 males and 10 females. Mean age of the study patients was 63.54 ± 9.27 years. Mean left ventricular ejection fraction was 47.76 ± 9.67%. The average grafting time of OM artery was 10.72 ± 1.88 mins. The demographic profile of the patients is shown in Table 1.
Table 2 shows the mean values of the parameters studied at two time points and their statistical relationship. There was a significant decrease in the value of CI from point A to point B (2.33 vs. 1.75 L/min/m2, p<0.001). A significant decrease was also found in the values of EtCO2 from 25.28 to 21.88 mmHg at both these time points (p<0.001). PaCO2 values increased significantly from 29.68 to 31.06 mmHg (p<0.003). A statistically significant difference was also noted in P(a-Et)CO2 gradient. The gradient changed from 4.39 at point A to 9.18 at point B, showing an increase of more than 100% (p<0.000). Thus as the CI decreased from point A to point B, there was a significant reciprocal increase in P(a-Et)CO2 gradient.
Discussion
Arterial blood gas sampling is conventionally used for estimation of PaCO2 and optimizing ventilatory strategies in mechanically ventilated patients. However the appeal for this technique is blunted clinically because it is invasive, labor intensive and expensive and it generally offers only intermittent measure of PaCO2 [11]. Time based capnography, best known as EtCO2 monitoring, provides valuable qualitative information on the waveforms associated with mechanical ventilation and a quantitative indirect estimation of PaCO2 [12]. CO2 is a product of organic cell metabolism which is uptaken and transported by venous circulation to the lungs where it is excreted by the alveolo-capillary membrane and thus, in normal ventilation/perfusion conditions, EtCO2 is an approximate measure of PaCO2 [5].
Various researchers have studied the efficacy of EtCO2 as a monitoring tool for adjustment in mechanical ventilation. Eun found a positive correlation between EtCO2 and PaCO2 in 24 critically ill neurological patients [2]. Similar findings were observed by McDonald in 129 children receiving mechanical ventilation who concluded that EtCO2 reliably estimates ventilation [12]. Russell in 59 postoperative cardiac surgery patients also observed a good correlation between PaCO2 and EtCO2 throughout the period from ICU admission to extubation [1]. On the contrary, a meta-analysis by Wahba concluded that changes in EtCO2 do not always indicate the direction and extent of change in PaCO2 [13]. Belpomme also found an inconsistent relationship between EtCO2 and PaCO2 during prehospital transport of 100 mechanically ventilated patients [14]. Prause also could not establish any correlation in 27 emergency patients and concluded that ventilation of emergency patients can only be correctly adjusted according to values from arterial blood gas [15]. Khan in 50 patients for craniotomy has also reported similar findings [16].
When metabolism and ventilation are constant, EtCO2 is largely determined by pulmonary blood flow and hence the CO. Many studies have been published that demonstrate the correlation between EtCO2 and CO in different clinical settings. Jin studied the relationship between cardiac index (CI) and EtCO2 in 16 pigs by inducing low flow states in the form of hemorrhagic, septic and cardiogenic shock and found EtCO2 as a good non-invasive alternative for continuous measurement of CO [17]. Feng, in 1994 had proposed that there exists a good correlation between pulmonary blood flow and EtCO2 while coming off cardiopulmonary bypass (CPB) in cardiac surgical patients.3 In another study by Baraka in 32 patients for CABG on CPB, a positive correlation was observed between adequacy of CO at EtCO2 values>30 mmHg [18]. Maslow also found that EtCO2 is a useful index of pulmonary arterial blood flow during separation from CPB [19]. Isserles, by inflating venacaval balloon to impede venous return in pentobarbital anaesthetised dogs, found that percentage decrease in EtCO2 is directly correlated to percentage decrease in CO [6]. However sustained decrease in CO led to accumulation of CO2 in peripheral tissues, leading to return of EtCO2 values towards baseline [3,6].
Hemodynamic changes are an important challenge during off-bypass procedures [9,20,21]. These may be secondary to the heart positioning or stabilization techniques. In order to work on the posterior or lateral walls, the heart must be lifted and tilted out of the pericardial cradle which results in a significant increase in atrial pressures, and a marked decrease in cardiac output [9]. Saleh observed in OPCABG surgery that during grafting on lateral wall, there is a decrease in EtCO2 values usually preceding major hemodynamic upset and dysrhythmias [7].
We aimed to study the appropriateness of EtCO2 monitoring during OPCABG surgery. Thus, we compared CI, EtCO2, PaCO2 and P(a-Et)CO2 gradients at two time points, before skin incision and during OM grafting. We observed a statistically significant correlation in CI, EtCO2, PaCO2 values and the P(a-Et)CO2 gradients at both these time points. Thus, we observe that similar to the findings in many studies, there exists a positive correlation between the values of EtCO2 and CO in our study. So whenever there is a decrease in CO, there is concomitant decrease in EtCO2 values, but there is an increase in the P(a-Et)CO2 gradient. Considering the OM grafting period to be a low cardiac output state, we can extrapolate our results to low cardiac output conditions. Thus we propose that the changes in EtCO2 values and P(a-Et)CO2 gradient is related to the low CO and no attempts should be made to alter the ventilatory settings during this period to normalize the EtCO2. Instead attempts should be made to increase the cardiac output.
Limitations
Several limitations of our study warrant further investigations. Firstly, the small sample size of our study limited our generalization of results. Secondly, use of side-stream capnometer might have led to under-estimation of the PEtCO2values as with this, a sample of exhaled air drawn through a sampling tube was sent to an analyzer placed at a distance away.
References





















Tables and Figures at a glance

image   image
Table 1   Table 2
Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Article Usage

  • Total views: 11726
  • [From(publication date):
    May-2014 - Dec 16, 2017]
  • Breakdown by view type
  • HTML page views : 7922
  • PDF downloads : 3804
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version