Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Epidural Granuloma by Dislocated Catheter Tip Associated with Spinal Cord Compression in High-Dose Intrathecal Morphine Therapy

Koeck K1*, Grossauer S1, Trummer M1 and Kleinert R2
1Department of Neurosurgery, Medical University Graz, Austria
2Department of Pathology, Medical University Graz, Austria
Corresponding Author : Koeck K
Department of Neurosurgery
Medical University Graz, Austria
E-mail: [email protected]
Received August 20, 2013; Accepted September 10, 2013; Published September 20, 2013
Citation: Koeck K, Grossauer S, Trummer M, Kleinert R (2013) Epidural Granuloma by Dislocated Catheter Tip Associated with Spinal Cord Compression in High- Dose Intrathecal Morphine Therapy. Gen Med (Los Angel) 1:117. doi: 10.4172/2327-5146.1000117
Copyright: © 2013 Koeck K, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Related article at
DownloadPubmed DownloadScholar Google

Visit for more related articles at General Medicine: Open Access

Abstract

Intrathecal drug delivery systems are an option for selected patients with chronic pain. Complications such as intrathecal catheter-tip inflammatory masses are already reported in literature. In the present case report we describe a patient suffering from neuropathic pain located on the anterior aspects of both thighs following failed back surgery. Two years after implantation of an intrathecal morphine pump system the patient experienced symptoms of a spinal cord compression with increasing radicular pain, especially after application of morphine bolus. Within this time frame the average daily intrathecal morphine hydrochloride dose was 3.2 mg per day and the total neuraxial morphine hydrochloride dose was 2460 mg. Magnetic resonance imaging revealed spinal cord compression from an epidural mass measuring 13×11×10 millimeters, originating from an epidural dislocation of the intrathecal catheter. After surgical removal of the mass, histological examination confirmed a disseminated formation of a granuloma. What makes this case so special, is the fact that the granuloma is not located intrathecally, but in the epidural space (epidural granuloma). The formation of a purely epidural granuloma due to a dislocated catheter tip has - according to our research - not been reported about up to the present.

Keywords
Granuloma; Intrathecal drug delivery systems; Morphine; Catheter
Introduction
Intrathecal opioid administration in the treatment of intractable chronic non-malignant pain and cancer pain is well documented in the neurosurgical literature. Indications for intrathecal drug delivery systems in patients with non-malignant pain include failed back surgery syndrome, neuropathic pain, complex regional pain syndrome, diffuse pain, axial spinal pain, brachial plexitis, central pain, failed spinal cord stimulation therapy, arachnitis, post stroke pain, spinal cord injury pain and peripheral neuropathy [1]. Serious side-effects of intrathecal morphine pumps, such as nausea, vomiting, pruritus, urinary retention, constipation, sexual dysfunction and peripheral edema are rare and mostly dose-dependent [2-8]. Those complications are usually mediated by opioid receptors. Intrathecal granulomas developing around the tip of the intrathecal catheter are a less common complication leading to symptoms of spinal cord compression and reduced therapeutic effects. Drug concentration and dose are supposed to be potential causative factors in the formation of catheter-associated masses [9,10]. A strong evidence that inflammatory masses at the catheter tips are caused by highly concentrated morphine infusate was identified by Yaksh et al. [11] in animal models.
The incidence and prevalence of these complications are inconsistent in literature and range from approximately one intrathecal mass per 1000 patients [12] to three intrathecal masses out of a group of 60 patients [13]. The incidence of inflammatory mass lesions after 1 year of therapy has been estimated to be 0.04% but increases to 1.15% after 6 years [11].
According to our research there has been no report about the formation of a purely epidural granuloma due to a dislocated catheter tip to the present.
Case Report
We present the case of a 47-year-old-woman with a 5 years history of lumbar spine pain. The most important side diagnoses were obesity (height: 165 cm, weight: 86 kg), a history of cigarette smoking (18 pack years) and non-insulin dependent diabetes mellitus. She has undergone three lumbar laminectomies and an AXIAL- LIF at L5/S1-without pain relieving outcome. Consequently a spinal cord stimulator (SCS) was placed but the pain-reduction was not satisfactory. At that time the patient suffered from neuropathic pain and paraesthesia located at the anterior aspects of both thighs without any further neurologic deficit. Ultimately, an intrathecal morphine pump was placed in 2006. Operation technique: After percutaneous puncture of the dural sac with a 17 Gauge Tuohy needle at L2-L3, the intrathecal needle tip location was verified by cerebrospinal fluid flow at 5.5 cm puncture depth. A spinal catheter (Spinal Catheter Set 4000®) with a closed catheter tip and 6 separate holes at the end of the catheter was advanced 14.5 cm under fluoroscopic control up to the level of T10. Again the catheter was checked for cerebrospinal fluid flow. Then the Tuohy needle was removed and a small skin incision was made directly below the skin entry point of the catheter. The catheter was fixed with two stitches on the supra spinal ligament using a silicone anchor. Contrast agent was administered through the catheter to verify the intradural position using fluoroscopy. Finally the catheter was tunneled subcutaneously to be connected with the port and the pump. Her post-operative response was good, her pain was reduced and the patient was satisfied. Two years later, she experienced increasing back pain with radiation into both legs and abdomen, especially when she was applicating bolus of morphine. Her morphine dose escalated rapidly up to 35 mg per day. Within this time frame the average daily intrathecal morphine hydrochloride dose was 3.2 mg per day and the total neuraxial morphine hydrochloride dose was 2460 mg.
Because of increasing radicular leg pain and constrained walking abilities magnetic resonance imaging was arranged and revealed an epidural expansion measuring 13×11×10 millimeters located on the tip of the catheter, which was obviously dislocated from intrathecal to epidural space (Figures 1-3). The epidural catheter tip was encased by the mass at level T11-12, compressing the spinal cord. The patient underwent surgery and had the mass resected via a hemilaminectomy at the left side of T11. The expansion was immediately apparent underneath the lamina and the yellow ligament; it was carefully dissected from the dura. The granuloma was incised and decompressed. The tip was encountered in surgery and removed during the preparation of the granuloma mass. The granuloma-wall adherent to the dura was left since it could not be freed from the dura. During the whole surgical procedure there was no sign of dural laceration or cerebrospinal fluid leakage at the level of surgery.
Histological analysis of the mass revealed an inflammatory reaction in the dura corresponding to a granuloma (Figure 4 and 5). A new intrathecal catheter was placed and the dosage was adjusted again to the patient´s daily morphine demands. The sensations from her abdomen to the lower extremities improved.
Discussion
The phenomenon of inflammatory masses at the tips of intrathecal catheters was first reported in 1991 [14]. Hypotheses that could explain the formation of catheter tip inflammatory masses, such as foreignbody reaction to silicone and other catheter material, infection, catheter-tip design, and the infusate-especially opioid analgetics, have been proposed [9,11].
In a prospective analysis 3% of 208 patients developed intrathecal granuloma-80 percent of the patients were asymptomatic [15]. It seems therefore considerable, that patients receiving long-term intrathecal analgesia should undergo periodic radiographic surveillance [16]. However, due to the low incidence of mass formation developing at the tip of intrathecal catheters, routine imaging to identify granulomas is not indicated [15]. Only if correlated with the symptoms mentioned above, especially by bolus application. In the presented case we hypnotize that the catheter was initially placed intradurally which has been proven by fluoroscopy after intrathecal application of contrast agent through the catheter, at the time of initial catheter placement. Also the fact that the analgetic effect was sufficient over months supports this hypothesis. The catheter tip must have migrated through the dura into the epidural space at the level where the epidural granuloma mass has been found together with the encased catheter tip at the time of surgical revision. This led to a loss of therapeutic effect and a need for increase of opioid dosage. It is not clear which pathomechanism led to dural laceration and transdural migration of the catheter tip. Also the place of dural laceration could not be visualized during surgery, probably because it was covered by remnant granuloma wall adherent to the dura, which has been left intentionally. The literature concerning the pathomechanisms of catheter dislocation is scarce. Extensive patient movements especially inclination and reclination [17], or muscle spasms [18] are thought to be causing factors.
In conclusion, a case is presented with a patient suffering from spinal cord compression from an inflammatory lesion. What makes this case so special, is the fact that the granuloma is not located intrathecally, but in the epidural space.
At this time, no reports about epidural granulomas caused by intrathecal drug administration are described in literature. Therefore, this seems to be an even less frequent complication than the before mentioned intrathecal granulomas. Physicians should be vigilant to the early symptoms and signs of spinal cord compression, such as new neurological deficits, myelopathy or radiculopathy.
References


















Figures at a glance

image   image   image   image   image
Figure 1   Figure 2   Figure 3   Figure 4   Figure 5
Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Recommended Conferences

Article Usage

  • Total views: 12120
  • [From(publication date):
    December-2013 - Dec 16, 2017]
  • Breakdown by view type
  • HTML page views : 8311
  • PDF downloads : 3809

Review summary

  1. MaryAnn Cottrell
    Posted on Sep 14 2016 at 5:42 am
    I had a catheter tip granuloma and it ended up paralyzing me cause my dr didn't pay attention to the signs of the complaints I was telling him. I had all the same symptoms as stated above. I had the pump and catheter removed but the granuloma couldn't be completely removed. I am a low income female at the age of 46. I have been paralyzed for the past 2 years and I hate it every single day. What options do I have so I can regain feeling again and possibly walk again and how can i get help financially to help pay for anything I could possibly try. I don't want to spend my life in a wheelchair and be a burden to my family or anyone else it would mean so much to me to find someone to help me please I'm so desperate. I hate living like this. It has caused so many other health issues for me and I don't know what to do. Please let me know who can help me with any procedure and also funding. Thank you.
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version