Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Fluoroscopy-Guided Axillary Vein Puncture, a Reliable Method to Prevent Acute Complications Associated With Pacemaker, Defibrillator, and Cardiac Resynchronization Therapy Leads Insertion

E.M. Malitha S. Hettiarachchi, Camelia Arsene*, Salah Fares, Adriss Faraj, Erik Saulitis, Salvatory Losito and Mukarram Siddiqui
Department of Medicine, Sinai-Grace Hospital / Detroit Medical Center, MI, USA
Corresponding Author : Camelia Arsene
Department of Medicine
Sinai-Grace Hospital/Detroit Medical Center
6071 West Outer Drive
Detroit Medical Center, MI 48235, USA
Tel: 313-966-9014
Fax: 313-966-1738
E-mail: [email protected]
Received November 28, 2013; Accepted December 30, 2013; Published January 06, 2014
Citation: Hettiarachchi EMMS, Arsene C, Fares S, Faraj A, Saulitis E, et al. (2014) Fluoroscopy-Guided Axillary Vein Puncture, a Reliable Method to Prevent Acute Complications Associated With Pacemaker, Defibrillator, and Cardiac Resynchronization Therapy Leads Insertion. J Cardiovasc Dis Diagn 2:136. doi: 10.4172/2329-9517.1000136
Copyright: © 2014 Hettiarachchi EMMS, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Related article at
DownloadPubmed DownloadScholar Google

Visit for more related articles at Journal of Cardiovascular Diseases & Diagnosis

Abstract

Introduction: The subclavian venous approach is a widely used method for venous access in device implantation and is associated with pneumothorax as a short term complication and lead fracture as a long term complication. The axillary vein approach is an alternate method for venous access, and this study evaluates the successfulness and immediate complications of fluoroscopy-guided axillary vein puncture compared to other venous approach methods.

Methods: This is a retrospective observational study on all patients who underwent pacemaker, defibrillator implant or lead change over 23 months. The fluoroscopy-guided modified Seldinger technique was used for axillary vein puncture and if failed, venography was performed.

Results: Out of 261 device implants or lead changes, 210 patients underwent fluoroscopy-guided axillary vein puncture. The mean age of the patients was 65.43 ± 15.7 years; 96.1% were African American; 57.6% were males. In 194 (92.3%) patients left or right axillary vein approaches were successful by either fluoroscopy or venography guidance. When anatomical abnormalities were excluded the success rate for axillary vein puncture was 97% and for fluoroscopy-guided axillary vein puncture was 94.5 %. Multiple leads were placed without any resistance and none of the patients had pneumothorax, hemothorax or hematoma as immediate complications.

Conclusion: Based on this first study conducted in a relatively large consecutive United Sates patient population, we report that fluoroscopy-guided axillary vein puncture using the first rib as a landmark, is a safe and effective method for device implantation with single or multiple leads, without patients getting exposed to intravenous contrasts.

Keywords
Fluoroscopy guided; Axillary vein; Pacemaker; Defibrillator implant; Immediate complications
Introduction
Infraclavicular subclavian approach, cephalic vein cut down, and axillary vein puncture are accepted methods for implanting endocardial pacemaker and transvenous defibrillator leads. Even though the subclavian approach is a widely used method for lead placement in cardiac pacing, different acute complications like for example a 1-3% incidence of pneumothorax or hemothorax have been noted [1-4]. Furthermore, lead fracture due to the entrapment of the lead by the costoclavicular ligament and/or the subclavius muscle, has been presented as subclavian crush and is a major concern [5,6]. As an alternative method, cephalic cut down has been shown to be associated with a lower incidence of lead fracture but the cephalic venous approach requires skills in surgical cut-down technique and the vessel’s smaller diameter makes insertion of multiple large diameter leads more technically demanding [7]. As a solution to problems associated with cephalic vein cut down, the axillary vein puncture has been tried. Experience with axillary vein approach is mainly clinical and limited, and immediate and long term complications over other methods are not well established in observational studies [8,9].
We present our single center experience in fluoroscopy-guided axillary vein puncture using the modified Seldinger technique. This study was approved by the Institutional Review Board.
Methods
Study methods
This is a retrospective observational study done on all patients who underwent a pacemaker, a defibrillator device implant or lead change for a period of 23 months in our institution. Patient information was gathered by chart review, specifically from operative/procedure notes.
Axillary vein cannulation
Clinical anatomy: The axillary vein is a large vein that is a continuation of the bacilic vein (Figure 1). It starts at the lower border of the teres major tendon and terminates immediately beneath the clavicle at the outer border of the first rib, where it becomes the subclavian vein. It is anterior and medial to the axillary artery and brachial plexus and partially overlaps. At the level of the coracoid process, the axillary vein receives the lateral cephalic vein [10].
Procedure: We access the axillary vein under fluoroscopy. An incision is made just medial to the coracoid process in the middle of the deltopectoral groove and carried inferomedially in a direction perpendicular to the deltopectoral groove. The incision is carried down to the surface of the pectoralis major muscle. The image intensifier is pulled over to the incision and the first rib is identified. An 18 gauge percutaneous needle and syringe is placed on pectoralis major muscle in the superior aspect of the incision. Using fluoroscopy the needle tip is placed in the middle of the first rib (Figure 2 and 3). The angle of the needle and syringe is advanced from 60° to 90° as needle is advanced through the pectoral muscle. Needle advancement is continued until the first rib is struck. Once the first rib is touched, the needle and syringe are slowly withdrawn under suction until the vein is entered. If the first pass is unsuccessful the needle and syringe are moved either medially or laterally along the first rib, and the maneuver repeated until the vein is entered. Once the vein is entered the guide wire is passed and the sheath applied per standard technique [10,11].
Multiple accesses are obtained if the patient requires dual chamber permanent pace maker/ defibrillator or biventricular implantable cardioverter defibrillator.
The method to fix the lead at the venous access site is the one described by Peter Belott [12].
Results
Two hundred and sixty-one patients underwent device implant or lead change during the study period (Table 1). In 210 patients fluoroscopy-guided left axillary vein approach was performed, and in 51 patients the subclavian or cephalic access was utilized.
The median fluoroscopy time was 6.8 seconds with a range of 0.1 to 58.9 seconds. Table 2 presents the fluoroscopy time data for different procedures.
The mean age of the patients who underwent axillary vein approach was 65.43 ± 15.78 years. The majority of them, 96.1%, were African American and 57.6% were males.
Out of 210 patients who had axillary vein approach attempted, in 194 patients (92.3%) left or right axillary vein approaches were successful by either fluoroscopic or venography guidance (Table 1 and Figure 4). In 16 (7.6%) patients axillary vein puncture was not successful and alternative sites were required.
The identified reasons for unsuccessful cannulation were anatomical variations/ abnormalities including persistence of left superior vena cava (then venous access was obtained in right axillary vein), venous obstruction, tortuosity of the axillary vein, and clavicular and rib abnormalities. Venous stenosis was also seen in two patients: a hemodialysis patient and an intravenous drug user. After the anatomical abnormalities were excluded, the success rate for axillary vein puncture (either fluoroscopy or venography guidance) was 97% and the success rate for fluoroscopy-guided axillary vein puncture was 94.5 % (Table 3).
Once the venous access was obtained, pacemaker leads or defibrillator leads were successfully placed via the axillary vein regardless of the number of leads, ranging from 1 to 3. Difficulty in placing introducer sheath or lead was not seen in any patients. The characteristics of the devices and the number of leads successfully placed are shown in Table 4. A total of 325 successful axillary vein punctures were performed. Lead removal and replacement were also done successfully in patients who had a preexisting device.
The patients were watched for immediate complications clinically and radiographycally: postero-anterior and lateral chests X-ray were performed and none of the patients had pneumothorax or hemothorax as immediate complications.
Discussion
Other operators have published different techniques to mark the position of the axillary vein, and for example the anterior side of the second rib was used as the guiding site for puncture site and direction (9). Our study used the first rib as landmark, not the second rib, and therefore this method prevented excessive bleeding at the puncture site and allowed for better and easier lead fixation. Although this could be considered as a technique sometimes used in the clinical practice, we have done a thorough literature review, and we found only one retrospective chart review published study conducted in a smaller patient population in Israel, that used the most lateral edge of the first rib as a fluoroscopic landmark [13].
Venous access is one of the most important steps in device implantation and the subclavian approach is a widely used method. The subclavian vein access is associated with short and long term complications: 1-3% of the incidence of pneumothorax or hemothorax as acute complications and subclavian crush phenomenon leading to lead fracture as a long term complication [5,6]. As shown in our study, axillary vein puncture is not associated with immediate complications [2-4].
Out of the small number of published data on axillary vein puncture most are non United States smaller scale studies that compare various methods, namely, the contrast venography-guided, the Dopplerguided, and the ultrasound-guided puncture methods except for a few studies done in Israel, Japan and India with fluoroscopy-guided approach [9,11,13-15]. This study is a North-American experience with fluoroscopy-guided axillary vein puncture using the first rib as a landmark; it is done in a relatively large African-American population, and discusses the immediate complications of the procedure.
Our study shows that fluoroscopy-guided axillary vein puncture is a safe method with limited immediate complications. There was an overall 97% success in axillary vein puncture (either fluoroscopy-guided or venography-guided) after excluding failures due to anatomical abnormalities. Fluoroscopy-guided axillary vein puncture alone also had a success rate of 94.5 % after excluding failures due to anatomical abnormalities.
The main advantages of fluoroscopy-guided approach are that patients are not exposed to the hazards of intravenous contrasts as in venography-guided approach, and the reduced incidence of pneumothorax or hemothorax. While some literature mentions the difficulty in measuring the depth of the vessel with fluoroscopy guidance we did not recognize it as a major barrier [14]. Most importantly, multiple access could be obtained for multiple leads to be inserted without any failure. We did not experience any immediate complications such as pneumothorax or hemothorax, and this could be correlated with the advanced experience of the principal investigator. According to the published literature, leads placed with axillary vein approach have minimum risk for lead fracture [16]. To identify the long term complications including lead fracture, we will be following these patients for 12 to 24 months prospectively.
The main limitations of this study are the fact that this is a retrospective observational study, and also there is no direct comparison between fluoroscopy-guided axillary vein puncture and the traditional methods of vascular access, like venography-guided axillary vein puncture, cephalic vein cut down or subclavian vein puncture.
Conclusion
We report that fluoroscopy-guided axillary vein puncture using the modified Seldinger technique is a safe and effective method for the implantation of pacemaker of defibrillator device with single or multiple leads, and it has limited immediate complications.
References
















Tables and Figures at a glance

image   image   image   image
Table 1   Table 2   Table 3   Table 4

Figures at a glance

image   image   image   image
Figure 1   Figure 2   Figure 3   Figure 4
Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Recommended Conferences

  • 3rd Global Summit on Heart Diseases
    November 02-04, 2017 Bangkok, Thailand
  • 22nd World Cardiology Congress
    December 11-12, 2017 Rome, Italy

Article Usage

  • Total views: 12745
  • [From(publication date):
    January-2014 - Oct 21, 2017]
  • Breakdown by view type
  • HTML page views : 8817
  • PDF downloads :3928
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri, Food, Aqua and Veterinary Science Journals

Dr. Krish

[email protected]

1-702-714-7001 Extn: 9040

Clinical and Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemical Engineering and Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001 Extn: 9040

Earth & Environmental Sciences

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

General Science and Health care Journals

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics and Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001 Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Informatics Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Material Sciences Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Mathematics and Physics Journals

Jim Willison

[email protected]

1-702-714-7001 Extn: 9042

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001 Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

John Behannon

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001 Extn: 9042

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords