Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Hemodynamic Consequences of Laparoscopy for Patients on Mechanical Circulatory Support

Heidi Reich1,2, Danny Ramzy1, Alagappan Annamalai2*, Lawrence Czer1, Fardad Esmailian1, Jaime Moriguchi1, Kai Ihnken1, Taizoon Yusufali3 Nicola D'Attellis3 and Francisco Arabia1

1Cedars-Sinai Heart Institute, Cedars-Sinai Medical Center, Los Angeles, USA

2Department of Surgery, Cedars-Sinai Medical Center, Los Angeles, USA

3Department of Anesthesiology, Cedars-Sinai Medical Center, Los Angeles, USA

*Corresponding Author:
Alagappan Annamalai
Comprehensive Transplant Center
8900 Beverly Blvd, Suite 267
Los Angeles, CA 90048
Tel: 310-423-2975
Fax: 310-423-0234
E-mail: [email protected]

Received date: March 26, 2015; Accepted date: April 28, 2014; Published date: May 02, 2015

Citation: Reich H, Ramzy D, Annamalai A, Czer L, Esmailian F, et al. (2015) Hemodynamic Consequences of Laparoscopy for Patients on Mechanical Circulatory Support. J Anesth Clin Res 6:526. doi: 10.4172/2155-6148.1000526

Copyright: © 2015 Reich H. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Anesthesia & Clinical Research

Abstract

Background: Technologic advances and superior survival with mechanical circulatory support (MCS) led to an expanding population that develops intra-abdominal conditions requiring intervention. Whether laparoscopy can be performed without detrimental effects on hemodynamics and device function is not well-described.

Methods: Effects of laparoscopy performed on MCS were retrospectively assessed. Intraoperative hemodynamics and device function were compared to the same time interval 24h prior to surgery using intrapatient paired t-tests. Outcomes included survival, transfusion, thromboembolic events, and infection.

Results: Twelve patients with ventricular assist devices or total artificial hearts underwent laparoscopy from 2012-2014. Median follow-up was 116 days. Operations included cholecystectomy, diagnostic laparoscopy, gastrojejunostomy, and gastrostomy. There were no differences between preoperative and intraoperative mean arterial pressure, heart rate, or inotrope or vasopressor requirements (p>0.05). Device fill volume, flow, rate, and power were unchanged (p>0.05), while pulsatility index decreased by 0.2, 95% CI [0.03, 0.36], with laparoscopy (p=0.03). All intraoperative fluctuations in hemodynamics and device function improved with reduction of pneumoperitoneum, adjusting device speed, or pharmacologic support. There were no operative mortalities. 30-day survival and survival to discharge were 75% and 50%. Despite antiplatelet therapy and preoperative INR of 2.2 ± 0.9, there were no re-operations for bleeding and 50% did not require transfusion. Two patients with recent cardiac surgery had thromboembolic events: 1 stroke, 1 device thrombus. None had postoperative bacteremia or drive-line infection.

Conclusions: Laparoscopy can be performed on mechanical circulatory support with low morbidity and mortality and minimal perturbations in hemodynamics and device function.

Keywords

Circulatory assist devices (LVAD, RVAD, BVAD, TAH); Circulatory; Hemodynamics; Minimally invasive surgery (incl port access, minithoracotomy); Outcomes (incl mortality, morbidity, survival, etc.)

Abbreviations

MCS: Mechanical Circulatory Support; INR: International Normalized Ratio; LVAD: Left Ventricular Assist Device; RVAD: Right Ventricular Assist Device; Bivad: Biventricular Assist Device; TAH: Total Artificial Heart; REMATCH: Randomized Evaluation of Mechanical Assistance for the Treatment of Congestive Heart Failure; Mmhg: Millimeters of Mercury; Mm: Millimeters; CT: Computed Tomography; MAP: Mean Arterial Pressure; HR: Heart Rate; N: Number; M: Male; F: Female; BMI: Body Mass Index; ICM: Ischemic Cardiomyopathy; NICM: Nonischemic Cardiomyopathy; INTERMACS: Interagency Registry for Mechanically Assisted Circulatory Support; BTT: Bridge to Transplantation; L: Left; R: Right; Bpm: Beats per Minute; Mcg: Micrograms; Min: Minute; Kg: Kilogram; Rpm: Revolutions per Minute; Ml: Milliliters; Lpm: Liters per Minute

Introduction

Over the last 15 years, advances in mechanical circulatory support (MCS) have dramatically changed mortality, morbidity, and quality of life for patients with advanced heart failure. Over 5.8 million U.S. adults have heart failure [1], which includes an estimated 250-500,000 with end-stage heart failure refractory to medical management [2]. Current indications for durable MCS include acute cardiogenic shock with ventricular function that is unrecoverable or unlikely to recover without a long-term device support, inability to maintain normal hemodynamics and vital organ function with temporary devices or inability to wean from temporary devices or inotropic support, capacity for meaningful recovery of end-organ function and quality of life, and absence of irreversible end-organ damage [3]. In addition, patients with inotrope-dependence or patients with heart failure and high predicted one-year mortality should be evaluated for MCS [3].

In the United States alone, over 2,500 durable mechanical circulatory support devices, which include left and/or right ventricular assist devices (LVADs, RVADs, BiVADs) and total artificial hearts (TAHs), are implanted annually [4]. The increasing rate of device implantations reflects improved understanding of the benefit devices offer patients with advanced heart failure that began with the REMATCH study [5]. One and two year survival with continuous flow LVADs are now 80% and 70% and over 40% of devices are placed as destination therapy [4] for patients who are ineligible for heart transplantation, often because of advanced age or comorbidities.

As the rate of device implantation and duration of support increase, there is an expanding population of MCS patients at risk for developing intra-abdominal conditions that warrant evaluation by a general surgeon. Hemorrhage, thromboembolism, infection, and malperfusion can lead to intra-abdominal pathology that may require operative intervention. Abdominal operations performed most frequently in MCS patients include cholecystectomy, appendectomy, exploratory laparotomy, hernia repairs, intestinal resections, and weight loss operations [6,7]. Outcomes for LVAD patients undergoing non-cardiac surgery have been favorable with equivalent one-year survival when compared to LVAD patients not requiring non-cardiac surgery and conflicting findings on survival to heart transplantation [8,9]. Laparoscopy is an effective approach to the diagnosis and treatment of intra-abdominal pathology. As improvements in laparoscopic equipment and technical skills have facilitated safe applications of laparoscopic surgery to treat increasingly broad conditions and patient populations, many historical contraindications to laparoscopy are no longer considered absolute. Current dogma identifies 4 absolute contraindications to laparoscopy, inability to tolerate laparotomy, hypovolemic shock, lack of proper surgeon training and/or experience, and lack of appropriate institutional support, as well as 5 relative contraindications, inability to tolerate general anesthesia, long-standing peritonitis, large abdominal or pelvic mass, massive incarcerated ventral and inguinal hernias, and severe cardiopulmonary disease [10]. Carbon dioxide pneumoperitoneum can have pronounced effects on the cardiovascular, pulmonary, and renal systems (Figure 1). The physiologic consequences of laparoscopy, including changes in systemic vascular resistance, increased preload and afterload, reduced cardiac output, cardiac arrhythmias, hypercarbia, cephalad displacement of the diaphragm, reduced renal blood flow, release of antidiuretic hormone, and oliguria becoming an unreliable indicator of hypovolemia, were established in healthy patients not on MCS [10,11]. Whether pneumoperitoneum has similar effects on the physiology of patients on MCS remains uninvestigated.

anesthesia-clinical-research-carbon-dioxide

Figure 1: Physiologic consequences of carbon dioxide pneumoperitoneum. The physiologic changes during laparoscopic surgery are multifactorial, but are largely attributed to the combined influences of peritoneal absorption of carbon dioxide, elevated intraabdominal pressures, and patient positioning. The effects on various organ systems were established in healthy patients not on mechanical circulatory support devices. Concerns that patients with advanced heart disease may be more susceptible to these changes and less able to mount appropriate compensatory responses remain are counterbalanced by a small, but growing, number of reports that laparoscopy can be safely performed in patients with mechanical circulatory support devices.

To explore our hypothesis that laparoscopy can be performed safely on MCS patients, we reviewed our experience with a focus on intraoperative hemodynamics, device parameters, and complications such as death, bleeding, thrombosis, or infection.

Methods

Study design

All patients with left ventricular assist devices, biventricular assist devices, or total artificial hearts undergoing laparoscopic operations performed by one surgeon at Cedars-Sinai Medical Center from August 2012 to April 2014 were included for retrospective analysis to characterize the effects of laparoscopy on MCS. MCS patients undergoing planned laparotomy were excluded. To account for inter-patient variation in preoperative clinical status, each patient served as his/her own control: intraoperative hemodynamics and device function were compared to the same time interval 24 hours prior to surgery using intra-patient paired t-tests. This study was reviewed and approved by the Cedars-Sinai Institutional Review Board.

Operative approach and anesthetic technique

All operations were performed with cardiac anesthesia and perfusion teams. Anticoagulation (Warfarin or Heparin) and antiplatelet (Aspirin) therapies were not routinely held; anticoagulation was reversed with fresh frozen plasma selectively. All patients underwent general anesthesia with endotracheal intubation and were positioned supine with reverse Trendelenberg. All patients underwent arterial line placement; central venous and pulmonary artery catheters were used selectively. Intravenous fluids were minimized, with 3/12 patients receiving no crystalloid or colloid and 4/12 receiving less than 500 cc (mean volume of fluids administered: 550 ± 541 cc). Drivelines were prepped and draped within the operative field. Preoperative antibiotics were administered prior to skin incision. The most frequent antibiotic selections were Piperacillin-Tazobactam or Cefazolin for cholecystectomy or gastrotomy, respectively. Carbon dioxide pneumoperitoneum was delivered via a Veress needle to achieve pressures of 12-15 mmHg. The position of trocar sites were modified based on the location of drivelines (Figure 2). Fascia for all trocar sites ≥ 10 mm was closed and skin was closed with absorbable monofilament suture.

anesthesia-clinical-research-Standard-trocar

Figure 2: Standard trocar locations for a laparopscopic cholecystectomy include supraumbilical, subxiphoid, medial right subcostal, and lateral subcostal sites (A) and modified trocar placement relative to driveline locations for patients with left ventricular assist device (B), biventricular assist device (C), and total artificial hearts (D) undergoing laparoscopic cholecystectomy. Initial trocar placement was via supraumbilical for LVAD and TAH and infraumbilical for BiVAD. The subxiphoid trocar site was placed more caudally to avoid LVAD and BiVAD drivelines and more caudally and more rightward to avoid TAH drivelines. Avoidance of the BiVAD drivelines also required shifting all trocar sites caudally.

Outcomes

Hemodynamic and device parameter data recorded by anesthesiologists, perfusion technologists, and cardiac surgery intensive care nurses were collected. Postoperative outcomes and major adverse events, including patient survival, reoperation for bleeding, need for transfusion, thromboembolic events or infections, were also analyzed. Thromboembolic events included stroke, defined as the presence of any neurologic deficit and/or CT findings consistent with intracranial hemorrhage or ischemic stroke, and device-associated thrombus identified by echocardiography. Infections were defined as clinically-suspected or culture-proven infection requiring antibiotic treatment or opening of wound and were categorized as bacteremia, urinary tract infection, pneumonia, abdominal surgical site infection, or drive line infection.

Statistical Analysis

Continuous variables are expressed as the mean ± standard deviation or median (range) and categorical variables are presented as count (%). Differences between preoperative and intraoperative parameters were assessed by paired student’s t-tests with significance defined as two-tailed p-values<0.05. The statistical analysis was performed using GraphPad Prism 5 for Windows (GraphPad Software, San Diego CA).

Results

Twelve patients with ventricular assist devices or total artificial hearts underwent laparoscopy. Median follow-up was 116 (3-371) days. The types of device and operations performed are summarized in Table 1 and preoperative patient characteristics in Table 2. Median procedure time was 2 hours 23 minutes (53 minutes–4 hours 12 minutes). There was a single conversion to open.

Device Operation N Indication
LVAD laparoscopic cholecystectomy 4 acute cholecystitis (3), gallstone pancreatitis
LVAD laparoscopic gastrojejunostomy 1 malnutrition
BiVAD diagnostic laparoscopy 1 possible acalculouscholecystitis
BiVAD laparoscopic gastrostomy 1 dysphagia
TAH laparoscopic cholecystectomy 3 acute cholecystitis (2), biliary dyskinesia
TAH laparoscopic gastrojejunostomy 1 failure to thrive
TAH diagnostic laparoscopy 1 sepsis, possible intestinal ischemia

Table 1: Mechanical circulatory support device and operation performed.

Age 63.5 (35-81) years
M:F 7:5
BMI 23.2 (18.8-39.9) kg/m2
ICM:NICM 7:5
Intermacs profile 1 or 2 7 (58%)
Device Strategy: BTT 8 (75%)
Duration of MCS 50 (1-863) days
Mechanical Ventilation 5 (42%)
Inotropes, Vasopressors 5 (42%)
Hemodialysis 3 (25%)
Prior abdominopelvic surgery 5 (42%)
INR 2.2±0.9
Hematocrit 25.8±3.57 mg/dL
Platelet Count 261(166-470) thousand
Aspirin 11 (92%)
Red blood cell transfusion 2 (20%)
Fresh frozen plasma transfusion 2 (20%)
Platelet transfusion 0

Table 2: Preoperative patient characteristics

At baseline, average heart rate (HR) was 83.2 ± 18.04 bpm, mean arterial pressure (MAP) was 77.02 ± 10.89 mmHg, inotrope requirement was 2.5 ± 5.00 mcg/min, and vasopressor requirement was 0.01 ± 0.02 U/min. There were no differences between preoperative and intraoperative HR, MAP, or inotrope or vasopressor requirements (Table 3, p-values 0.23-0.45). Device fill volume, flow, rate, and power were unchanged (p-values 0.20-0.93), while pulsatility index decreased by 0.2, 95% CI [.03, 0.36], with laparoscopy (Table 3, p=0.03). All intraoperative fluctuations in hemodynamics and device function improved with reduction of pneumoperitoneum, adjusting device speed, or pharmacologic support. There were no device power interruptions due to electrocautery interference. Of 5 patients on inotropic or pressor support prior to surgery, 1 (20%) required a brief increase in support (epinephrine raised from 1.5 to 2.5 mcg/min), 1 (20%) tolerated intraoperative reduction of support, and 3 (60%) were stable without adjustments of pharmacologic support. 2 of seven patients not on inotropes or pressors were given intraoperative inotropes.

HD Parameters # pairs Mean Difference [95% CI] p-value
HR (bpm) 7 -5.71 [-23.08, 11.65] 0.45
MAP (mmHg) 12 -4.47 [-12.28, 3.34] 0.23
Inotrope (mcg/min, mcg/kg/min) 12 0.92 [-1.31, 3.14] 0.38
Pressor (units/min) 12 0.00 [-0.00, 0.00] 0.34
       
Device Parameters      
Device Rate (rpm, bpm) 9 0.85 [-0.63, 2.33] 0.22
L Fill Volume (mL) 6 -0.49 [-7.96, 6.98] 0.87
R Fill Volume (mL) 6 0.21 [-5.77, 6.18] 0.93
L Flow (lpm) 12 0.24 [-0.15, 0.64] 0.2
R Flow (lpm) 6 0.10 [-0.38, 0.58] 0.61
Pulsatility Index 4 0.20 [0.33, 0.36] 0.03
Power (Watts) 5 1.39 [-1.73, 4.52] 0.28
Mean difference >0 corresponds to a decrease in the parameter with laparoscopy
Bpm, beats per minute; mmHg, millimeters of mercury; mcg/min, micrograms per minute, kg, kilograms; rpm, revolutions per minute; mL, milliliters; lpm, liters per minute

Table 3: Difference between baseline and intraoperative hemodynamics and device parameters

There were no operative mortalities. 30-day survival and survival to discharge were 75% and 50%. Despite antiplatelet therapy and preoperative INR of 2.2 ± 0.9, there were no re-operations for bleeding and 50% did not require transfusion 48 hours before, during, or after surgery. Two patients with recent cardiac surgery had thromboembolic events: 1 stroke, 1 device thrombus. No patient had postoperative bacteremia, urinary tract infection, pneumonia, or drive line infection. Two patients had superficial surgical site infections: one at the umbilical trocar site that resolved after opening of the incision and one at the gastrostomy tube site that resolved with local wound care.

Discussion

In the largest study to date describing the experience of laparoscopy performed on MCS patients, we found that laparoscopy was both feasible and safe to perform in MCS patients. Previous reports of applications of laparoscopy to MCS patients were similarly encouraging, but limited to single patient case reports and very small case series (Table 4). Although prior reports have stated that laparoscopy was well-tolerated, our study is the first to report and quantitatively analyze whether hemodynamics and device parameters change with laparoscopy. Our sample represents a diverse cohort of MCS patients with multiple types of devices and inclusion of patients who were clinically stable or unstable prior to surgery.

Previous Reports of Laparoscopy on MCS Device Operation
Caceres (2013, n=1)12 HMII LVAD Roux-en-Y gastric bypass
Nayak (2013, n=1)13 HMII LVAD Left nephroureterectomy
Kartha (2008, n=1)14 HMII LVAD Cholecystectomy
Sathishkumar (2012, n=1)15 HMII LVAD Ileocolectomy
Eck (2014, n=2)16 HMII LVAD Cholecystectomy
Samoukovic (2011, n=1)17 HMII LVAD Splenectomy
Nissen (2005, n=1)18 ParacorporealBiVAD Cholecystectomy
Brown (2009, n=3)9 HMII LVAD Cholecystectomy, appendectomy (2)
Prendergast (1996, n=1)19 Abiomed BVS 5000 BiVAD Diagnostic laparoscopy
Votapka (1994, n=1)20 Thoratec LVAD Cholecystectomy

Table 4: Reports of laparoscopy performed in patients with mechanical circulatory support.

We identified a small but statistically significant decrease in the pulsatility index associated with pneumoperitoneum. The pulsatility index is calculated as (flowmax−flowmin)/flowmean × 10 and ranges from 1-10. It represents the magnitude of the flow pulse generated by an LVAD through each cardiac cycle and reflects the balance between ventricular contractility and the degree of unloading. The pulsatility index is dependent on LVAD speed and preload. The magnitude of the reduction in pulsatility index we observed, 0.2, was less than our prospectively assigned threshold for clinical relevance of 0.5. Although the reduction was small in magnitude, this change was consistent with the effects of pneumoperitoneum described in non-MCS patients, which include preload reduction. If the pulsatility index were to decrease during laparoscopy in association with hypotension or reduced flow, appropriate responses include looking for bleeding, reducing insufflation pressures, leveling the bed if in reverse trendelenberg, giving volume, and obtaining an echocardiogram. In an otherwise hemodynamically stable patient with a decrease in pulsatility index, transesophageal echocardiography is a useful adjunct to assess for evidence of hypovolemia or right ventricular depression to further guide management.

We believe that a well-coordinated team approach to perioperative care is an important element in ensuring the best outcomes for MCS patients in need of abdominal operations. Our team approach included close coordination of perioperative care with the MCS team, performing all operations with the assistance of cardiac anesthesiologists in a cardiac operating room with a perfusionist present and the cardiac surgeon immediately available if questions about device function or driveline positions arose. Modifications in traditional laparoscopic port placement are sometimes necessary to avoid the drivelines.

Decisions concerning perioperative management of anticoagulation and antiplatelet therapy should be based on the input of the MCS team and the operating surgeon. The 2013 International Society for Heart and Lung Transplantation guidelines for MCS patient requiring non-cardiac surgery recommend continuation of anticoagulation and antiplatelet therapies if bleeding risk is low, holding with or without a heparin bridge if warranted by risk of bleeding, reversal of Coumadin with fresh frozen plasma or vitamin K if needed for emergency procedures, and resuming anticoagulation and antiplatelet therapy when risk of surgical bleeding is acceptable.3 Historically, coagulopathy was considered a contraindication to laparoscopy, but this has largely fallen out of favor as meticulous surgical technique and advanced instruments have facilitated the safe application of laparoscopy patients with cirrhosis and hematologic disorders.21 Previous studies, in which open approaches were more frequent, described bleeding as the most common complication [8,9]. While the risk of bleeding could be a major limitation to performing laparoscopy on MCS, we did not find this to be a problem. In our series, we found that meticulous attention to hemostasis facilitated avoidance of withholding or reversing anticoagulation or antiplatelet therapy without leading to any reoperations for bleeding. Because MCS patients are at elevated risk of thromboembolic events, withholding or reversing anticoagulation and antiplatelet therapy could lead devastating complications, such as pump thrombosis or stroke.

Conclusion

Laparoscopy can be performed on mechanical circulatory support with low morbidity and mortality and minimal perturbations in hemodynamics and device function. This is the first study to quantify the effects of pneumoperitoneum for patients on mechanical circulatory support.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Recommended Conferences

Article Usage

  • Total views: 11825
  • [From(publication date):
    specialissue-2015 - Jun 23, 2018]
  • Breakdown by view type
  • HTML page views : 8050
  • PDF downloads : 3775
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2018-19
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

+1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

neurop[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2018 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
Leave Your Message 24x7