Management and Treatment of Type 1 And 2 Diabetes: State of Art | OMICS International
ISSN: 2327-5146
General Medicine: Open Access
Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.

Management and Treatment of Type 1 And 2 Diabetes: State of Art

Tagliente I1,2,6*, Digilio G2, Ullmann N3, Solvoll T5, Trieste L8, Murgia F1, Bella S7, Bini F2 and Schiaffini R4

1Scientific Direction, Health Care Innovations and Management, Bambino Gesù Children's Hospital, IRCCS, Rome, Italy

2Department of Information, Electronic and Telecommunication, University of Sapienza

3Respiratory Unit, Bambino Gesù Children's Hospital, IRCCS, Rome, Italy

4Unit of Endocrinology and Diabetes, Bambino Gesù Children's Hospital, Rome, Italy

5Norwegian Centre for Integrated Care and Telemedicine, University hospital of North Norway, Tromsø, Norway

6Department of Human Social and Health Sciences, University of Cassino and Southern Lazio, Cassino, Italy

7Department of Pediatric Medicine, High Specialization in Continuity of Care in Chronic Diseases, Pediatric Hospital Bambino Gesù, Rome, Italy

8Institute of Management, Scuola Superiore Sant’Anna, Pisa, Italy

Corresponding Author:
Irene Tagliente
Respiratory Unit, Bambino Gesù Children's Hospital
Bambino Gesu Via Ferdinando Baldelli 41 00146, Rome, Italy
Tel: +39 066859 4760
Fax: +39 066859 4760
E-mail: [email protected]

Received Date: May 23, 2016; Accepted Date: June 28, 2016; Published Date: June 30, 2016

Citation: Tagliente I, Digilio G, Ullmann N, Solvoll T, Trieste L, et al. (2016) Management and Treatment of Type 1 And 2 Diabetes: State of Art. Gen Med (Los Angeles) 4:259. doi:10.4172/2327-5146.1000259

Copyright: © 2016 Tagliente I, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at General Medicine: Open Access

Abstract

Background: Diabetes is a hormonal disorder that affects an estimated three hundred million people worldwide. In this review (meta-analysis) of the literature, we carefully analyzed the limits and methods of the management of diabetes in adolescents and adults. Objective: The study focused on the identification of technologies and therapeutic procedures currently used and tried to identify possible factors for future improvement.

Methods: The literature review was carried out on search engines like: PubMed, ADA, Google Scholar and Google Patents. According to our search criteria, the selected articles were analyzed regarding evidence and critical issues reported by patients and clinicians. The results were then compared with the recalls of the U.S. Food and Drug Administration (FDA).

Results: The currently in use drugs and the insulin infusion systems vary a lot and simultaneous develop of new technology adds to existing chaos. This study revealed possible areas for improvements both from a technological and a medical point of view. For instance: the multiple causes related to the different timing of insulin absorption according to the insertion site, the design of new algorithms focused on simplification of practical activities, and the identification of new methods for training patients and families on the different device’s functions. Moreover, some critical points were emerged related to the current limitations and lack of interchangeability between sensors for monitoring vital parameters. This condition restricts the correct sensor selection according to the patient’s personal preference, age and needs. However, the lack of interchangeability between different devices limits the choice of the most appropriate monitoring software and data storage.

Conclusions: The collaboration between medical doctors and biomedical engineers is of great importance when analyzing available medical devices for the purpose, identifies possible weaknesses, technical problems, and areas for improvements. Moreover, multi-specialist collaboration is important to conceive new methods of training for patients and families.

Keywords

Diabetes type 1; Infusion pump; Technical problems; European norms; Medical devices; Self-management; CGM; Stylus injectors; Glucometers; Software management

Abbreviations

CGM: Glucose Monitoring In Continuous; CG-EGA: Continuous Glucose-Error Grid Analysis; CSII: Continuous Subcutaneous Insulin Infusion; DM1: Diabetes type 1; DM2: Diabetes type 2; EGA: Error Grid Analysis; IFCC: International Federation of Clinical Chemistry and Laboratory Medicine; MARD: Median Absolute Relative Difference; NIST: National Institute of Standards and Technology

Introduction

Diabetes is a hormonal disorder that affects around three hundred million people worldwide [1]. Several therapies have been proposed and a good compliance to treatment offers patients a good quality of life. Diabetes is a disease characterized by a group of metabolic disorders caused by defects of insulin’s secretion and / or activity. The first condition is represented by a pancreas inability to produce insulin and the latter is characterized by incorrect use of normal secreted insulin (insulin resistance). Produced in the beta cells from the Langerhans islets, insulin has a dual function: to transport glucose from the bloodstream to the body's cells, where it is converted into energy and it is involved in the anabolic metabolism of glycogen. Glucose, produced by the digestion of food, is transported by the blood to all body cells. The rate of glucose in the blood is called glycemia. Insulin is essential to allow the cells to use glucose as an energy source. Diabetes is divided into mellitus TYPE1 and TYPE2:

Diabetes type 1 (DM1) is characterized by immune-mediated destruction of pancreatic beta cells with absolute insulin deficiency. This condition needs to be treated with insulin injections or administered through a pump. DM1 appears more frequently in children and adolescents and accounts for about 5% -10% of all cases of diabetes in the world.

Diabetes type 2 (DM2) can be characterized by two conditions: inadequate insulin’s secretions or peripheral insulin resistance. This is by far the most common form of this disease, with 90% - 95% of cases. DM2 usually can be controlled with diet, physical activity and oral drugs but, in a high percentage of patients, insulin administration via syringes or infusion pumps are needed. Several clinical trials show that obesity and being overweight can cause the onset of type 2 diabetes.

In this paper, we present a review of the literature about the limits and methods of management of diabetes in adolescents and adults. The goal of the study was identify technologies and therapeutic procedures currently used for diabetes’ treatment and monitoring and subsequently, highlighting possible areas for improvement. The aim was analyzing challenges and limits of currently available medical devices for the management of diabetes. The focus has been extended to stylus injectors, insulin pumps, sensors for Glucose Monitoring in Continuous (CGM), glucometers and finally the artificial pancreas. Therefore, in this paper we collected the European regulations for manufactures. The study highlighted shows possible ideas for both technological and scientific improvement.

Methods

Research setting

The study was conducted at Bambino Gesu’ Children’s Hospital, Rome Italy. The Hospital encourages healthcare providers to establish an open and friendly communication relationship with patients and families. Moreover, healthcare providers are recommended to provide accurate information about medical conditions and treatments, that can help patients and families deal with the illness in a more informed manner.

Research method

The literature review was carried out via the major search engines (PubMed, ADA, Google Scholar, Google Patents, CDA) considering the following keywords: diabetes, insulin pump, glucometer, biosensor, pen injector, oral hypoglycemic agents, insulin therapy, pancreas artificial, closed-loop, CGM, continuous Subcutaneous Insulin Infusion (CSII). From the first analysis (n ° 2000 Articles) were selected abstracts and relevant full text of interest. The research led to the selection of the 15 items listed in the references. From the articles selected were analyzed evidence and critical issues reported by patients and clinicians, exposing the current technological advances aimed at overcoming those limits (bio-artificial pancreas, nano-biosensors, closed loop system). The selection of publications was performed according to: number of citations, year of publication (2004-2013), IF of the journal, significance of the results and published data the authority of the authors (we chose to consider only items written by diabetologists who strongly contributed to the scientific community taking part of National or International commission and or authors of guideline or medical books) and, finally, the importance of the conference if the article was presented. Finally, we defined the role that biomedical engineer must have to introduce these improvements in therapy for the treatment of diabetes.

Emergence was compared with the recall of the U.S. Food and Drug Administration (FDA) to the manufacturers of medical devices as glucometer or insulin pumps and deepened through the information made available by manufacturers.

Therefore in this paper we collected the European regulations for manufactures.

Data Analysis

Stylus injectors

The pen, or pen needle, is very practical and convenient for easy to control diabetes. They can be recharged or disposable and are visually similar to a pen. In the case of rechargeable Stylus injectors, a reservoir containing a vial of insulin (vial format pencil), whose apex is present with a wheel which can be set and modified from time to time, representing the doses needed for the patient. On one end there is a button to release insulin while at the opposite side (tip) there is the needle.

The advantages found by clinicians and users are: the ability to set the amount of insulin with extreme simplicity (it is set to a numeric selector), the needle is disposable and the dimensions of the stylus injectors are reduced in order to make discreet its use. The main disadvantage is that the units are not divisible, therefore it’s not possible to use fractional units and this seriously limits treatment’s adaptation for patients. In addition, when different types of insulin are needed, more stylus injectors would be also needed [2-5].

Insulin pumps

The insulin pump is a small precision electro-mechanical device that releases insulin into the body via a thin plastic tube (or infusion set).

The infuser is a long thin plastic tube (thin plastic tubing), which connects the pump to a small needle or flexible cannula, inserted through the skin at the point of infusion (infusion site), usually in the abdominal area in contact with the interstitial fluid. The infuser can be held under the skin for about three days, after which it must be replaced to avoid problems of infection. Unfortunately, the pump is not able to self-define the correct insulin dosage based on the blood glucose measurements. The pump is a precise instrument controlled by a small programmable computer, which releases insulin in previously set dosage (CSII) according to diabetologist prescription [6]. The pump can be programmed to release the drug at any time. However, the patient must check the blood glucose levels several times a day, a factor that, in parallel, requires much more attention by the patient compared to traditional systems of injection [6]. Some Insulin pumps are selflock in case of hypoglycemia thanks to special sensors [7].

Sensors for glucose monitoring in continuous (CGM)

A system for continuous measurement of blood glucose consists of a subcutaneous sensor which detects the concentration of glucose in the interstitial fluid. This system is connected to a transmitter that sends data to a receiver handheld which processes all information in order to adjust the insulin infusion (by the patient). Technological advances have increased precision and accuracy of analytical biosensors. The sensors currently available can reach a maximum MARD (Median Absolute Relative Difference), including measurement of blood glucose and blood glucose sensor of reference, between 11% to 14% [7]. Crucial indicator for the prescription of an effective therapy, but nevertheless it cannot be considered as an absolute indicator [8]. The MARD, as the Continuous Glucose-Error Grid Analysis (CG-EGA), is a Windows-based, statistical program designed to be used by the diabetes technology industry for the evaluation of continuous glucose monitoring devices. It is a logical extension of the original error grid analysis (EGA), which was developed to assess the clinical accuracy of patient-determined blood glucose values using either estimation or self-monitoring blood glucose systems. CG-EGA is based on the idea that information, generated by a monitoring system, should be reliable enough to result in clinically accurate decision. The CG-EGA technology was originally developed by Drs. William Clarke and Boris Kovatchev at the University of Virginia in Charlottesville, VA, USA.

Glucometers: A portable medical device is able to estimate with good approximation the glucose concentration in the blood. To estimate the blood glucose, the glucometer analyzes a small drop (1μL) of blood - usually taken from fingertip - on a special test strip. In most of the models, the glucose content in the blood sample activates an enzymatic reaction oxidizes by the presence of the enzyme glucose oxidase in the test strip; this oxidation produces a color change or an electrical current (depending on the model) convertible to glucose value [8].

Software Management

The main goal is to facilitate the control of the variance as a function of glycemic carbohydrate ingested and insulin units needed. The currently available devices, interfaced with PC, have their software for the storage and consultation of recorded data, allowing the visualization of the parameters recorded for more than 24 hours [9].

European Norms

The European regulations that manufacturers must follow are (from the most recent to the most dated):

UNI EN ISO 15197:2013: Systems of in vitro diagnostic assay - Requirements for systems for self-testing of blood glucose monitoring in the treatment of diabetes mellitus.

Directive EN 61326-1:2006: Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements.

Directive EEC / EAEC / EC n° 79 of 27/10/1998: 98/79/EC: Directive of the European Parliament and of the Council of 27 October 1998 on in vitro diagnostic medical devices.

Directive 1999/5/EC of the European Parliament and of the Council of 9 March 1999 on radio equipment and telecommunications terminal equipment and the mutual recognition of their conformity.

Directive 93/42 EEC for Medical Devices (abbreviated as DDM 93/42).

Reference Methods for the execution of medical devices validation tests:

Standard of the National Institute of Standards and Technology (NIST).

Standards and Recommendations of the ‘International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)’.

Recall

After this first phase we checked the major recall of FDA to the manufacturer of devices (Figure 1), numbers of recalled (Figure 2).

general-medicine-manufacturer

Figure 1: Number of recall for each manufacturer.

general-medicine-FDA

Figure 2: Way of recalls by FDA.

Therefore, we analyzed the why of the recalled to understand if there were some common factors to be studied, to perform the management of diabetes with devices (Figure 3).

general-medicine-hardware-involved

Figure 3: Kind of hardware involved 3a) for Glucometer 3b) for Insulin Pump.

The analysis was principally focused on glucometer and insulin pump. We studied the way of recalls, the motivation (e.g., was analyzed also the hardware that caused the error) and the risk for the patients as showed by diagrams on (Figure 4).

general-medicine-Alert-percentage

Figure 4: Alert percentage divided by type of device’s defects or user’s error.

Results

The intent of this study was to identify the state of art of the limits and methods of management of diabetes type 1 and 2 through the literature analysis.

Various are the drugs and the insulin infusion systems currently in use and under development [10].

Great is the industry attention to miniaturize all devices, with the use of nanotechnology, as well as to study biomaterials that are more suitable for insulin infusion.

Among many topics covered in the literature (Table 1), there are three issues considered of greater impact in the short-medium term.

Factors Desired improvements
Insulin administration Identification of new techniques to increase the rate of absorption of insulin (ultra rapid wording, dosing in other compartment, warmth the infusion site)
Training Identify new methods and training for patients and their families to facilitate the use of all the functions of the various devices that can be used for the monitoring and treatment of diabetes type I and II
Management software and communication protocols Example: 1) Communication between glucose sensor and insulin pump: make the connections more reliable; 2) Interoperability of devices and interfacing software: standardization of the communication protocol
Improving the performance of the algorithms Example: 1) Recognition of glucose: Reduce the false positive alarms of hypoglycemia 2) Identify new algorithms for insulin dose in closed loop configuration

Table 1: Objectives to be achieved to improve performance.

Lack of know how about the time of absorption of insulin, that is currently not less than 15 min [11,12]. Currently there are no studies that define the optimal measurement site, but it is known that there are different concentration of glucose in the blood compare to the interstitial fluid [13]. It is well shown the 'importance of a constant glucose concentration in the blood and prevention of hypoglycemia or hyperglycemia, a result of improper use of insulin pump (CSII). In addition, the insertion of interstitial or intra-cutaneous sensors [14] is often problematic due to poor training, as well as to the currently available intrinsic technological limits of (materials, biological component and hardware) [15-17].

As there is no standardized protocol for communication between devices, each manufacturer adopts its own software, confusing in the user, and limiting the use and interpretation of the data (useful purpose in a proper and continuous monitoring of clinical parameters of diabetic subjects).

The frenetic introduction into the market of new devices by the manufacturers does not allow the development of algorithms able to minimize detection errors of the data [18,19].

The major leaders in the field of biomedical instrumentation for diabetics have proven to be: Abbot, AgaMatrix, Animas, Arkay, Bayer, Bayer, BIONIME, Diabestic Supply of Suncoast, Diagnostic Devices, Dexcom, Lifiscan, Medtronic, Menarini, Nova Biomedical, Roche, Sanofi Aventis. Topic of scientific discussion now turns out to be the finalization of various systems "closed-loop" pointing to the best technological solution that takes into account all the variables that describe the biological system. The main difficulties in the development are related to the difficulties in reproduction of the physiology of insulin and glucose in the data due to the complexity of the human body.

The analyses of recalls showed:

For glucose meters, 59% of the problems of the devices are data from 'hardware rather than the software (41%).

77% of the insulin pumps considered encountered problems of hardware rather than software (23%) and this is a critical issue that could lead to a bad dose of insulin in patients with consequent effects on health.

In 22% of cases, the user induces problems in the management of the device, because of bad training by the staff, or inadequate interface to the user type.

In addition, the subjectivity of the disease, which is not a negligible element since each patient typically receives a different treatment, affects the therapy by artificial pancreas that cannot be standardized [20-23].

Discussions

The hospital and industrial research centers should endeavour to improve existing technologies in line with the daily needs of patients. Moreover, it would be important to identify new:.

Hardware, customized on the patients target;

Systems that can reduce the problems of interoperability between devices;

Semi-automatic calibrations to reduce possible operating errors [24].

Bambino Gesu' Children's Hospital, by the way of great result on a telemedicine case study [25,26], is developing a new Network to help the adolescent patients and their family. By a new App for Android Tablet a Team of Engineers and Doctors are looking to have a standard interface to analyze the diabetic patient data on Telemedicine follow up.

Evaluating the processes of absorption of insulin would facilitate the identification of the more appropriate dosage and reduce the absorption (with inevitable positive impact on the levels of glycosylated hemoglobin), a more appropriate treatment that also means a higher quality of life and reduced direct (health and nonhealth) and indirect costs. The synergy between Clinical, Biomedical Engineers and Computer Services is of great importance to problems resolution and to identify new methods for training for patients and their families [26,27]. The regulations currently in force have not been updated the last five years despite the strong development and innovative technologies.

Conclusion and Future Work

The study highlighted possible ideas for both technological and scientific improvement, presenting possible new studies to be undertaken, for example about: evaluation of different times depending on the absorption of insertion sites, development for algorithms focus on simplification of business routine and identification for new training techniques for patients and their families on the use of all functions that can be set on different devices. Moreover, critical issues related to the current limits for interoperability between sensors emerged. Insulin pumps and blood glucose meters, should be chosen according to patient’s needs, age and personal perceived convenience of use. In addition, the lack of interoperability between devices connected to computers, causes limitations in the choice of software, monitoring and archiving parameters, more 'suitable and intuitive by different users.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Article Usage

  • Total views: 9508
  • [From(publication date):
    June-2016 - Mar 23, 2019]
  • Breakdown by view type
  • HTML page views : 9378
  • PDF downloads : 130

Review summary

  1. Denisa Pinto
    Posted on Oct 13 2016 at 4:48 pm
    The authors have reviewed the current technology for the measurement of blood glucose and subsequent release/administration of insulin. The user based complaint and issues were recorded and compared to the recalls done by the FDA on glucose monitoring technology. Based on the data, the authors advocate improvement in glucose monitoring technology, better understanding of insulin metabolism in the body dependent on the different sites of insulin injection, better training of users and clinical practitioners and common software for multiple technologies available from different biomedical instrumentation companies. The article stressed that there is a need for greater collaboration between biomedical engineers and medical doctors for the standardization and advancement of the field.