Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Mass Resolution Study in the Cylindrical Ion Trap by Using Three-Point One Block Method

Sarkhosh Seddighi Chaharborj1,2* and Yousof Gheisari2

1School of Mathematics and Statistics, Carleton University, Ottawa, K1S 5B6, Canada

2Department of Mathematics, Islamic Azad University, Bushehr Branch, Bushehr, Iran

*Corresponding Author:
Sarkhosh Seddighi Chaharborj
School of Mathematics and Statistics Carleton
University, Ottawa, K1S 5B6, Canada
Tel: +16139816230
E-mail: [email protected]

Received date: May 14, 2017; Accepted date: May 23, 2017; Published date: May 29, 2016

Citation: Chaharborj SS, Gheisari Y (2017) Mass Resolution Study in the Cylindrical Ion Trap by Using Three-Point One Block Method. Mass Spectrom Purif Tech 3:117. doi:10.4172/2469-9861.1000117

Copyright: © 2017 Chaharborj SS, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Mass Spectrometry & Purification Techniques

Abstract

Motion equations in cylindrical ion trap (CIT) coupled in u and v (respective r and z), and thus, can only be treated as a rough approximation. Hence, studies on cylindrical ion trap (CIT) equations are more complicated and involved. Therefore, a three point one block method (3POBM) of Adams Moulton type is presented to study a cylindrical ion trap (CIT) motion equations. The advantage of the three point one block method (3POBM) is to estimate the approximate solutions directly at three points simultaneously. Numerical results from three point one block method (3POBM) will compare with fifth order Runge-Kutta method (RKM5). The proposed three points one block method has a potential application to solve complicated linear and nonlinear equations of the charge particle confinement in the cylindrical field especially in fine tuning accelerators, and generally speaking, in physics of high energy. The physical properties of the confined ions in the r and z axises are illustrated and the fractional mass resolutions m/Δm of the confined ions in the first stability region was analyzed by the fifth order Runge-Kutta method (RKM5) and three points one block method (3POBM).

Keywords

Cylindrical ion trap; Three point one block method; Fifth order Runge-Kutta method; Lagrange interpolating polynomial; Ion trajectories

Introduction

An ion trap mass spectrometer may incorporate a Penning trap [1], Paul trap [2] or the Kingdon trap [3]. The Orbitrap, introduced in 2005, is based on the Kingdon trap [4]. The two most common types of ion traps are the Penning trap and the Paul trap (Quadrupole ion trap) [5-9]. Other types of mass spectrometers may also use a linear Quadrupole ion trap as a selective mass filter.

A Quadrupole ion trap mass analyzer with simplified geometry, the cylindrical ion trap (CIT), was shown to be well-suited to use in miniature mass spectrometers and even mass spectrometer arrays. Experiments with a single miniature CIT showed acceptable resolution and sensitivity, limited by the ion trapping capacity of the miniature device. The CIT has received much attention of a number of research groups because of several merits. The CIT is easier to fabricate than the Paul ion trap which has hyperbolic surfaces [10]. And the relative simplicity and small size of the CIT make it an ideal candidate for miniaturization. With these interests, many groups in, such as Purdue University [11] and Oak Ridge National Laboratory [12] have researched on the applications of the CIT to a miniaturized mass spectrometry.

Formulation of Three Point One Block Method

Numerical solutions for ODEs have great importance in scientific computation as they were widely used to model in representing the real world problems such as Quadrupol ion trap with damping force. Many of these problems are of higher order and cannot be solved analytically and hence the use of numerical methods is advocated. There are several methods that can be used to solve the higher order ODEs numerically [13]. Block method for numerical solution of higher order ODEs directly had been proposed by several researchers [13-17]. The general theory of block method for solving first order ODE is given in Fatunla [15]. A two point block methods for solving higher order ODEs directly are described by Majid [14]. In this paper, we shall consider two point one block method for the numerical solution of solving second order ODEs in the form

equation (1)

subject to an initial conditions

equation

in the interval [a, b]. The proposed two point block method will approximate the solutions at two points simultaneously.

In this section, we have divided the interval [a, b] into a series of blocks with each block containing three points. Three approximate values that are yn+1, yn+2 and yn+3 are simultaneously found using the same back values.

The first point of the method will be obtained by integrating Equation (1) once and twice over the interval [xn, xn+1]. By integrating Equation (1) once gives:

equation

equation (2)

Integrating Equation (1) twice gives:

equation (3)

The second point of the method obtain by integrating (1) over the interval [xn, xn+2] once and twice will respectively give,

equation (4)

equation (5)

The same process is applied to derive the third point of the method, integrate (1) over the interval [xn, xn+3] once and twice will respectively give,

equation (6)

equation (7)

We can rewrite the (2), (3), (4), (5), (6) and (7) into r point k step Adams Moulton:

equation (8)

equation (9)

Figure 1 showed the seven mesh points i.e., (xn−3, fn−3), (xn−2, fn−2), (xn−1, fn−1), (xn, fn), (xn+1, fn+1), (xn+2, fn+2) and (xn+3, fn+3) involved in the three point one block method. Therefore, we substitute k=6 into Equation (8) and (9) will obtain,

mass-spectrometry-purification-techniques-block-method

Figure 1: Three point one block method.

equation (10)

equation

with r=1, 2, 3. The function f(x, y, y0) in (10) will be replaced by Lagrange interpolating polynomial. Lagrange Interpolating is a method of finding an unique polynomial, Pn which passes through a specified set of points. Define x0, x1, x2, ..., xnare (n+1) distinct numbers obtained from a function f, then exists a polynomial as follows:

equation (11)

equation

equation

For the general case when there are n+1 points, the polynomial, Pn in Equation (11) are nth-order. Seven interpolating points involved in three point one block method, we will obtain the lagrange interpolating polynomial, P6 from Equation (11).

equation

equation

equation (12)

Evaluate these integrals in (12) using MAPLE. The value of the coefficients, βi and γi, i=0, 1, 2, ..., 6 can be obtained. The corrector formulae of three point one block method as follows:

equation

equation (13)

with r=1, 2, 3. Where the coefficient βi and γi, i=0, 1, 2, 3, 4, 5, 6 shown in Table 1.

r i 0 1 2 3 4 5 6


1
  βi equation equation equation equation equation equation equation
γi equation equation equation equation equation equation equation
 
2

βi
equation equation equation equation equation equation equation

γi
equation equation equation equation equation equation equation
 
3

βi
equation equation equation equation equation equation equation

γi
equation equation equation equation equation equation equation

Table 1: The coefficients for the corrector formulae of three point one block method.

Apply the same process to find the predictor formulae of the two point direct method. The method is the combination of predictor of one order less than the corrector. The Euler method will be used only once at the beginning of the code to find the additional points for the starting initial points. The Euler method also will solve the problem directly. Then, the predictor and corrector direct method can be applied until the end of the interval. The sequence of computation involved was PECE where P and C indicate the application of the predictor and corrector formula respectively and E indicate the evaluation of the function f. The convergence test involved the corrector formulae and it will be iterated to convergence. The iterates are said to have converged when an iterate equation satisfies as follows:

equation (14)

where equation is come from predictor formula and “TOL” mean Tolerance. For numerical results has been used Tol =10−10.

Electric Field Inside Cylindrical Ion Trap

In cylindrical ion trap (CIT), the hyperbolic ring electrode [18], as in Paul ion trap, is replaced by a simple cylinder and the two hyperbolic end-cap electrodes are replaced by two planar end-plate electrodes [19]. If opposite pairs of the electrodes have steady potentials Ψ0 and −Ψ0 [20], the potential difference applied to the electrodes [18,21] is:

equation (15)

With

equation (16)

Where Udc is a direct potential, Uac is the zero to peak amplitude of the RF voltage, and Ω is RF angular frequency [20]. The electric field in a cylindrical coordinate (r, z, θ) inside the CIT can be written as follows:

equation (17)

here, ∇ is gradient. From Equation (17), the following is retrieved:

equation (18)

The equation of the motions [8,18,19,22-24] of the ion of mass m and charge e can be written as

d2uJ1iu) ch(λiv)

equation (19)

equation (20)

and the following

equation

Where J0 and J1 are the Bessel functions of the first kind of order 0 and order 1, respectively, whereas ch is the hyperbolic cosine function, mir is the roots of equation J0(mir)=0, Udc, whereas Vac are the amplitudes and the radio frequency (rf) drive frequency. Equations (19) and (20) are coupled in u and v (respective r and z), and thus, can only be treated as a rough approximation [19,23,25-27]. Therefore, studies on CIT equations are more difficult and complex.

Implementation of Block Method in Cylindrical Ion Trap Motion Equations

In this section implementation of block method has been done for cylindrical ion trap motion equations. From Equation (13) with Equation (19) and Equation (20), respectively we have, the corrector formulae of the three point one block method as follow:

equation (21)

equation(22)

equation (23)

equation (24)

equation (25)

equation (26)

here y=u or y=v with

equation

equation

and

equation

Numerical Results

Figures 2a and 2b show the ion trajectories in real time for cylindrical ion trap with α=0 and χ=0.4, z1=0.05 and equation red color (solid line): RKM5 method using h=0.01 steps, (b): black color (dash line): three-point one block method using h=0.05 steps. Figure 3 shows the ion trajectories in u −v plan for cylindrical ion trap with α=0 and χ=0.4, z1=0.05 and equation red color (solid line): RKM5 method using h=0.01 steps, (b): black color (dash line): three-point one block method using h=0.1 steps. Figure 4a and 4b show the ion trajectories for cylindrical ion trap in z-z? plan for α=0 and χ=0.4, z1=0.05 and equation red color (solid line): RKM5 method using h=0.01 steps, (b): black color (dash line): three point one block method using h=0.05 steps.

mass-spectrometry-purification-techniques-cylindrical-ion-trap

Figure 2: The ion trajectories in real time for cylindrical ion trap with α=0 and χ=0.4, z1=0.05 and equation red color (solid line): RKM5 method using h=0.01 steps, (b): blue color (dash line): three-point one block method using h=0.05 steps (Color figure just online).

mass-spectrometry-purification-techniques-ion-trajectories

Figure 3: The ion trajectories in u-v plan for cylindrical ion trap with α=0 and χ=0.4, z1 =0.05 and equation red color (solid line): RKM5 method using h=0.01 steps, (b): black color (dash line): three-point one block method using h=0.1 steps (color figure just online).

mass-spectrometry-purification-techniques-three-point-one

Figure 4: The ion trajectories for cylindrical ion trap in z-z? plan for α=0 and χ=0.4, z1 =0.05 and equation red color (solid line): RKM5 method using h=0.01 steps, (b): blue color (dash line): three-point one block method using h=0.05 steps (Color figure just online).

Numerical examples has been tested for t ∈ [0,60] and TOL=10−10. The results obtained from three point one block method and fifth order Runge-Kutta method are comparable for h=0.05 and h=0.01, respectively. The running steps taken by three-point one block method using h=0.05 and RKM5 using h=0.01 are 601 and 6000, respectively. Therefore, to obtain the comparable results from both method the total number of running steps for RKM5 is around 10 times more than three-point one block method. Hence, the execution time taken by three-point one block method using h=0.05 is less than RKM5 method using h=0.01.

The effect of three-point one block method on the mass resolution

The resolution of a cylindrical ion trap mass spectrometry [22] in general, is a function of the mechanical accuracy of the hyperboloid of the CIT Δr1, and the stability performances of the electronics device such as, variations in voltage amplitude ΔV, the rf frequency ΔΩ [22,28], which tell us, how accurate is the form of the voltage signal. Here we study the resolution of a cylindrical ion trap with RKM5 and three-point one block methods.

Table 2 shows the values of χmax and Vmax for the cylindrical ion trap in the first stability region when α=0 with using RKM5 and three-point one block method when h=0.01 and h=0.05, respectively. The values of Vmax has been obtained for 131Xe with Ω=2Π × 1.05 × 106 rad/s, U=0 V and z0=0.82 cm in the first stability region when α=0. To derive a useful theoretical formula for the fractional resolution, one has to recall the stability parameters of the impulse excitation for the CIT with RKM5 and three-point one block methods, respectively as follows,

Methods ↓ χmax Vmax
RKM5 0.64 5982.9425
3POBM 0.63 6077.9099

Table 2: The values of χmax and Vmax for the cylindrical ion trap in the first stability region when α=0 with using RKM5 and three-point one block method (3POBM) when h=0.01 and h=0.05, respectively.

equation (27)

equation (28)

Here “RKM5” and “3POBM” mean fifth order Runge-Kutta method and three-point one block method, respectively. By taking the partial derivatives with respect to the variables of the stability parameters χRKM5 for Equation (27) and χ3POBM for Equation (28), then the expression of the resolution Δm of the CIT with RKM5 and 3POBM are as follows,

equation (29)

equation (30)

Now to find the fractional resolution we have,

equation (31)

equation (32)

here Equation (31) and Equation (32) are the fractional resolutions for CIT with RKM5 and 3POBM, respectively. For the fractional mass resolution we have used the following uncertainties for the voltage, rf frequency and the geometry; ΔV/V=10−5, ΔΩ/ Ω=10−7, Δr1/r1=3 × 10−4. The fractional resolutions obtained are mm=1638.806949;1638.380841 for RKM5 and three-point one block methods with using 0.01 and 0.05 steps, respectively. When three-point one block method is applied, the rf only limited voltage is increased by the factor of approximately 1.01587, therefore, we have taken the voltage uncertainties as ΔV3POBM/V3POBM =1.01587 × 10−5. When these fractional resolutions are considered for the tritium isotope mass m=3.20, then, we have Δm=0.001952640 and 0.1953148 for RKM5 and three-point one block methods with using 0.01 and 0.05 steps, respectively. This means that, as the value of m/Δm is decreased, the power of resolution is increased due to increment in Δm. Experimentally, this means that the width of the mass signal spectra is better separated (Figure 5).

mass-spectrometry-purification-techniques-function-ion-mass

Figure 5: The resolution of Δm as function of ion mass m, dash line (red line): three-point one block method with using 0.05 steps and dash point line (blue line): RKM5 with using 0.01 steps.

Discussion and Conclusion

We have demonstrated the three points one block method to solve some complicated differential equations concerned the cylindrical ion trap. We can observe to obtain the comparable results from both methods the total number of running steps for RKM5 is around ten times more than three point one block method. When these fractional resolutions are considered for the tritium isotope mass m=3.20, then, we have Δm=0.001952640 and 0.1953148 for RKM5 and three point one block methods with using 0.01 and 0.05 steps, respectively. This indicates that as the value of m/Δm decreased, the power of resolution is increased due to increment in Δm. However, it is clear that, at least in the lower mass range the impulse voltage with using the three points one block method is quit suitable than using the RKM5 for the CIT with higher in mass resolution. Apart from numerical result, the method showed a potential application to solve complicated linear and nonlinear equations of the charged particle confinement in the cylindrical field especially in fine tuning accelerators, and generally speaking, in physics of high energy.

Acknowledgements

The authors thank the referees for valuable comments and suggestions which improved the presentation of this manuscript.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Recommended Conferences

Article Usage

  • Total views: 119
  • [From(publication date):
    June-2017 - Sep 23, 2017]
  • Breakdown by view type
  • HTML page views : 97
  • PDF downloads :22
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri, Food, Aqua and Veterinary Science Journals

Dr. Krish

[email protected]

1-702-714-7001 Extn: 9040

Clinical and Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemical Engineering and Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001 Extn: 9040

Earth & Environmental Sciences

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

General Science and Health care Journals

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics and Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001 Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Informatics Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Material Sciences Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Mathematics and Physics Journals

Jim Willison

[email protected]

1-702-714-7001 Extn: 9042

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001 Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

John Behannon

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001 Extn: 9042

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords