Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Poly(alcian blue) Modified Carbon Paste Electrode for the Determination of Catechol in Presence of Hydroquinone: A Voltammetric Study

Rekha, Swamy BEK* and Ganesh PS

Department of PG Studies and Research in Industrial Chemistry, Kuvempu University, Jnana Sahyadri, Shankaraghatta, Shimoga, Karnataka, India

*Corresponding Author:
BE Kumara Swamy
Department of PG Studies and Research in Industrial Chemistry, Kuvempu University
Jnana Sahyadri, Shankaraghatta-577451, Shimoga, Karnataka, India
Tel: +91 8282-2562251
E-mail: [email protected]

Received date: July 15, 2016; Accepted date: August 11, 2016; Published date: August 21, 2016

Citation: Rekha, Swamy BEK, Ganesh PS (2016) Poly(alcian blue) Modified Carbon Paste Electrode for the Determination of Catechol in Presence of Hydroquinone: A Voltammetric Study. J Biosens Bioelectron 7:217. doi: 10.4172/2155-6210.1000217

Copyright: © 2016 Rekha, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Biosensors & Bioelectronics

Abstract

The carbon paste electrode (CPE) was modified by electropolymerisation of alcian blue in 0.2 M phosphate buffer solution (PBS) of pH 7.4 by using cyclic voltammetric (CV) technique. The fabricated electrode was employed for the electrochemical analysis of catechol (CC) and hydroquinone (HQ). The effect of scan rate suggests the adsorption controlled electrode process. A good analytical performance was observed in terms of sensitivity, selectivity, linearity and observed detection limits. The lower limit of detection of CC and HQ was found to be 0.104μM and 0.142μM by cyclic voltammetric technique. Because of the catalytic capability of the fabricated electrode the simultaneous separation was observed in a binary mixture containing CC and HQ.

Keywords

Catechol; Hydroquinone; Poly(alcian blue); Cyclic voltammetry; Pulse voltammetry

Introduction

Phenolic compounds are formed during the biological degradation processes; they are widely used as raw materials in the production of dyes, photostabilizer, plasticizers, cosmetics, pesticides and some pharmaceuticals [1,2]. Catechol (CC) and hydroquinone (HQ) are two simple electroactive molecules, belongs to the class of dihydroxybenzene isomers. Since their wide spread applications in day to day life, they are the major cause for environmental pollutions [3,4]. HQ and CC are highly toxic and even at very low concentrations can be harmful to animals and plants [5]. Therefore, sensitive and selective analytical methods are necessary for their determination. Several methods are proposed [6-10] including chromatography [11,12], spectrophotometry [13], pH?based flow injection analysis [14], synchronous fluorescence [15], and voltammetry [16-21]. Among the various fabrication methodologies in getting the new chemically modified electrodes, electropolymerisation is a conveniently simple and powerful method by easy modification of different types of electrodes with desired matrices. The easy synthesis and deposition of desired electroactive polymers onto the conductive surface from monomer solutions and the formation rate and thickness can be precisely controlled by electrochemical input parameters. These electroactive polymers have useful properties such as electronic conductivity and ionic conductivity [3,22-25]. Most of all the bare carbon paste electrodes suffer from a fouling effect due to surface accumulation of the oxidized products, and because of the similar structure and properties the oxidation of dihydroxy benzene isomers at bare electrodes was undistinguishable. Swamy et al. developed an electrochemical sensor for the electroanalysis of CC and HQ by electropolymerising brilliant blue on the surface of carbon paste electrode (CPE) [26]. J He et al. reported the modification of the glassy carbon electrode by electropolymerised film of eosin Y and used for the voltammetric separation of CC and HQ [27]. Wang et al. studied the application of simple amino acids in the fabrication of a new modified glassy carbon electrode for the qualitative determination of CC and HQ in a binary mixture [3,22,23].

Alcian Blue is a phthalocyanine dye contains copper metal and used in the cationic form. The alcian blue has many positive charges on the molecule and it is thought to work by forming reversible electrostatic bonds between the cationic dye and the negative sites on the polysaccharide. Alcian Blue is a group of polyvalent basic water soluble compound which imparts blue color in the solution form due to the presence of copper ion in the molecule [28]. We proposed a modification of CPE by weight ratio method using alcian blue as a modifier for the determination of dopamine [29]. The present work gives a detailed work for the modification of carbon paste electrode by electropolymerising alcian blue followed by its characterization for the voltammetric determination of catechol and hydroquinone by cyclic voltammetric and differential pulse voltammetric techniques. The fabricated Poly(alcian blue) modified carbon paste electrode shows catalytic capability in the oxidation of both CC and HQ. The sensitivity, selectivity, stability and reproducibility in the result were obtained at the fabricated electrode. The structure of alcian blue was shown in the Scheme 1 (Scheme 1).

biosensors-bioelectronics-Structure-alcian-blue

Scheme 1: Structure of alcian blue.

Experimental Section

Reagents and apparatus

Hydroquinone (HQ), catechol (CC) and alcian blue were purchased from Himedia. The stock solutions of 25×10−4 M HQ, 25×10−4 M CC and 25×10−3 M alcian blue was prepared in double distilled water. Phosphate buffer solution (PBS) of same ionic strength was maintained (0.2 M) and the desired pH was obtained by mixing appropriate ratio of NaH2PO4·H2O and Na2HPO4. Graphite powder of 50 μM particle size was purchased from Merck and silicone oil from Himedia was used to prepare the carbon paste electrode (CPE). All the chemicals are of analytical grade used as received without any further purification.

All electrochemical experiments were performed using a model CHI-660c (CH Instrument-660 electrochemical work station). A conventional three electrode system was used in a electrochemical cell with a saturated calomel electrode (SCE) as a reference, a platinum wire counter electrode and bare carbon paste electrode (BCPE) or poly(alcian blue) modified carbon paste electrode as a working electrode. Oxidation potentials of all the analytes were recorded versus SCE at an ambient temperature of 25±0.5°C.

Preparation of the bare carbon paste electrode

The BCPE was prepared by homogeneously mixing graphite powder and silicone oil in a ratio of 70:30% for about 45 minutes. The paste was then packed into a PVC tube with homemade cavity of 3 mm internal diameter and the surface was smoothened on a weighing paper. Unless otherwise stated, the paste was carefully removed prior to pressing a new portion into the electrode after every measurement. The electrical contact was provided by a copper wire connected to the end of the tube.

Result and Discussion

Electropolymerisation of alcian blue on BCPE

The poly(alcian blue) MCPE was fabricated by cyclic voltammetric technique by potential sweeping the bare carbon paste electrode between the potential window of -0.8 V to + 1.8 V with scan rate 0.1 Vs-1 for 5 multiple cycles in 0.2 M PBS of pH 7.4 until a stable cyclic voltammogram was observed. During this process of recording the multiple cycles the voltammogram was slowly increased with increasing in cyclic times as shown in Figure 1a and later becomes constant [30,31]. This indicates that a thin polymer layer of alcian blue was formed and deposited on BCPE. As the extent of thickness can vary the electrocatalytic performance of the fabricated electrode, it was subjected for the oxidation of 0.2 mM CC in 0.2 M PBS of pH 7.4 with the scan rate of 0.05 Vs-1 as shown in the Figure 1b. A better catalytic performance was observed for the electrode which was modified by five consecutive cycle scans. Therefore, the Poly(alcian blue) MCPE was fabricated by considering all the above mentioned parameters (Figures 1a and 1b).

biosensors-bioelectronics-Cyclic-voltammograms-preparation

Figure 1a: Cyclic voltammograms of preparation of poly(alcian blue) MCPE. 1 mM solution in 0.2M PBS of pH 7.4 at 5 cycles with scan rate 0.1 Vs-1.

biosensors-bioelectronics-Cyclic-voltammograms-recorded

Figure 1b: Cyclic voltammograms recorded for the oxidation of 0.2 mM catechol in 0.2 M PBS of pH 7.4 at poly (alcian blue) MCPE at scan rate of 0.05 Vs-1. (A-5cycles B-10cycles C-15 cycles D-20 cycles).

Characterization of BCPE and poly(alcian blue) MCPE

The convenient redox probe of [Fe(CN)6]4-/[Fe(CN)6]3- was used for the electrochemical characterization of the fabricated electrodes. The freshly prepared stock solutions of 1 mM potassium ferrocyanide in 1M KCl as a supporting electrolyte were placed in an electrochemical cell. Figure 2 shows the cyclic voltammograms obtained for the redox behavior of [Fe(CN)6]4-/[Fe(CN)6]3- at both BCPE (dotted line) and Poly(alcian blue) MCPE (solid line) at the scan rate 0.05 Vs-1. A poor catalytic performance was observed at BCPE with low redox current response. On the other hand, Poly(alcian blue) MCPE has shown the capability of enhancing the redox peak currents with improved electrocatalytic effect. The electrochemical results obtained at Poly(alcian blue) MCPE suggests that the surface property was significantly changed. This remarkable change in the surface of the Poly(alcian blue) MCPE is a platform for the analysis of the targeted molecule. The total active surface area available for reaction of species in solution can be estimated by the Randles–Sevcik equation (1) [26,31].

biosensors-bioelectronics-potassium-ferrocyanide

Figure 2a: Cyclic voltammograms of 1 mM potassium ferrocyanide in 1M KCl at BCPE (solid line) and poly(alcian blue) MCPE (dashed line) at scan rate of 0.05 Vs-1.

Ip= 2.69× 105 n3/2 A D½ C0 υ1/2 (1)

where, Ip is the peak current in A. C0 is the concentration of the electroactive species (mol cm−3), n is the number of electron exchanged, D is the diffusion coefficient in cm2 s-1, is the scan rate (Vs-1) and A is the electroactive surface area (cm2). As calculated, the electroactive surface is maximum for Poly(alcian blue) MCPE (0.0358 cm2) as compared with BCPE (0.0279 cm2).

Scanning electron microscopy (SEM) images of BCPE and Poly(alcian blue) MCPE are shown in the Figure 2b, which reveals that the surface of BCPE is of irregular shape (A) with less available surface area. After the electropolymerisation the electrode surface is covered by a thin polymer layer of alcian blue (B), with uniform aligned surface. This morphological future is entirely different from BCPE (Figures 2a and 2b).

biosensors-bioelectronics-SEM-Images-BCPE

Figure 2b: SEM Images of BCPE (A) and poly(alcian blue) MCPE (B).

Electrochemical response of CC at Poly(alcian blue) MCPE

The Figure 3 shows the cyclic voltammograms recorded for the oxidation of 0.1 mM CC in 0.2 M PBS of pH 7.4 with the scan rate of 0.05 Vs-1 at BCPE and poly(alcian blue) MCPE. The CC shows quasireversible behavior at BCPE with a broad voltammetric response and the oxidation peak potential was located at around 0.230 V versus SCE. On the other hand, an improved current response was obtained at Poly(alcian blue) MCPE (solid line) with a slight shifting in the anodic peak potential towards the positive side comparing to BCPE. The anodic peak potential was observed at 0.176 V versus SCE. This refinement in the voltammetric response shows the catalytic capability of fabricated Poly(alcian blue) MCPE towards the oxidation of CC. The oxidation mechanism of CC and HQ was shown in Scheme 2 (Figure 3 and Scheme 2).

biosensors-bioelectronics-voltammograms-alcian-blue

Figure 3: Cyclic voltammograms of 0.1 mM CC in 0.2M PBS pH 7.4 at BCPE (dashed line) and poly(alcian blue) MCPE (solid line) at scan rate of 0.05 Vs-1.

biosensors-bioelectronics-Oxidation-mechanism

Scheme 2: Oxidation mechanism of catechol and hydroquinone.

The effect of scan rate on peak current of CC

The effect of variation of applied scan rate for 0.1 mM CC in 0.2 M PBS of pH 7.4 was examined by cyclic voltammetric technique at poly(alcian blue)MCPE as shown in Figure 4A. The experimental results obtained at Poly(alcian blue) MCPE suggests the peak current and scan rate are proportional to each. It can be observed that, there is a slight positive shift of anodic peak potential (Epa) and negative shift in the cathodic peak potential (Epc). In order to evaluate the electrode phenomenon, the graph of peak current (Ip) versus scan rate (?) was plotted and the obtained graph is a straight line with good linearity in the range from 0.01 to 0.14 Vs-1 as shown in Figure 4B with the correlation coefficient of (r2) 0.9973 and 0.9992. On other hand, the Ip versus square roots can rate (?1/2) was plotted as shown in Figure 4C with the correlation coefficient (r2) 0.9911 and 0.9949. This suggests the electrode phenomenon was an adsorption-controlled [32]. The heterogeneous rate constant (k0) values was determined from the experimental peak potential difference (ΔEp) data’s, Eq.(2) was used for the voltammograms whose ΔEp values are greater than 10 mV [26,31].

biosensors-bioelectronics-Graph-peak-current

Figure 4: (A) Cyclic voltammograms of 0.1 mM CC in 0.2 M PBS solution of pH 7.4 at different scan rate (a-l: 0.01 to 0.14 Vs-1) (B) Graph of peak current versus scan rate. (C) Graph of peak current versus square root of scan rate.

ΔEp= 201.39 log (?/k0)-301.78 (2)

From the experimental ΔEp values and corresponding k0 values for the oxidation of catechol and hydroquinone was tabulated in Table 1 (Figure 4 and Table 1).

υ/mVs-1 ΔEp/mV ko/s-1
Catechol
(CC)
Hydroquinone
(HQ)
Catechol
(CC)
Hydroquinone
(HQ)
10 0.0388 0.0403 0.3172 0.3280
20 0.0393 0.0425 0.6344 0.6212
30 0.0425 0.0445 0.9515 0.9008
40 0.0434 0.0458 1.2680 1.2242
50 0.0426 0.0462 1.5852 1.4828
60 0.0398 0.0466 1.9028 2.0012
70 0.0424 0.0468 2.2202 2.1098
80 0.0412 0.0462 2.5368 2.4880
90 0.0422 0.0458 2.8516 2.7102
100 0.0430 0.0478 3.1717 3.1882
120 0.0436 0.0460 3.8053 3.6022
140 0.0446 0.0468 4.4401 4.2202

Table 1: Voltammetric parameters gathered from Figures 4 and 8.

Effect of CC concentration

Differential pulse voltammetry (DPV) shows better defined peaks at lower concentrations than those obtained by cyclic voltammetry. The variation of concentration of CC at Poly(alcian blue) MCPE was conducted by DPV technique. The Figure 5A shows by increasing the concentration of CC from 10 to 60 μM the Ipa and Ipc goes on increasing with a small shifting in the peak potentials. The graph of Ipa versus concentration of CC was plotted as shown in the Figure 5B and it shows a straight line with good linearity. The linear regression equation is Ipa (μA) = 0.1172(C0 μM/L)+0.4275, (r2=0.9872). The limit of detection was calculated as previously reported [24,26]. The poly(alcian blue) MCPE shows a limit of detection of 0.104μM for CC by DPV technique, which is relatively lower compared to other reported as shown in Table 2 [8,33-39] (Figure 5 and Table 2).

biosensors-bioelectronics-Differential-pulse-voltammograms

Figure 5: (A) Differential pulse voltammograms of CC in 0.2 M PBS solution of pH 7.4 at poly(alcian blue) MCPE with different concentrations (a–f: 10-60 μM). (B) Graph of anodic peak current versus concentration.

Working Electrode Limit of detection in mM Technique Reference
CC HQ
Zn/Al Layered Double
Hydroxide Film MGCE
1.2 9 DPV [33]
Glassy carbon electrode
in CPB and SDBS 
3 8 DPV [34]
Silsesquioxane-MCPE 10 10 DPE [8]
Graphene oxide and
multiwall carbon nanotubes
1.8 2.6 DPV [35]
[Cu(Sal-β-Ala)(3,5-DMPz)2]
/SWCNTs/GCE
3.5 1.46 DPV [36]
Poly(calmagite) MCPE 2.55 1.70 DPV [37]
PEDOT/GO modified Electrode 1.6 1.6 DPV [38]
PASA/MWNTs/GCE 1.0 1.0 DPV [39]
Poly(alcian blue) MCPE 0.104 0.142 DPV This work

Table 2: Comparision table of limit of detection obtained at poly(alcian blue) MCPE with other electrodes.

Influence of solution pH on the determination of CC at Poly(alcian blue) MCPE

The pH of the phosphate buffer solution plays a key role in the oxidation of the electroactive molecules, and it was evaluated at Poly(alcian blue) MCPE by cyclic voltammetric technique. The effect of phosphate buffer pH value at Poly(alcian blue) MCPE for the determination of CC was evaluated in a pH range of 5.5-8.0. Figure 6A shows cyclic voltammograms recorded for 0.1 mM CC at Poly(alcian blue) MCPE. It is clear that as the pH increases the oxidation peak potential shifts to a negative potential. The linear relationship was observed between anodic peak potential (Epa) versus pH with a slope of 0.0588 V/pH as shown in inset Figure 6B. The result suggests the involvement of equal number of electrons followed by equal number of protons in the redox mechanism [24,26] (Figure 6).

biosensors-bioelectronics-anodic-peak-potential

Figure 6: (A) Cyclic voltammograms of 0.1 mM CC at poly(alcian blue) MCPE in 0.2 M PBS solution with different pH values (a–f: 5.5 to 8.0) at scan rate of 0.05 Vs-1. (B) Graph of anodic peak potential versus pH.

Electrocatalytic oxidation of HQ at Poly(alcian blue) MCPE

The Figure 7 shows the oxidation of 0.1 mM HQ at BCPE (dotted line) and Poly(alcian blue) MCPE (solid line) in 0.2 M PBS of pH 7.4 at the scan rate of 0.05 Vs-1. From the Figure 7 it is observed that the oxidation potential of HQ at BCPE was broad and poor in sensitivity, the anodic peak potential was located at around 0.126 V. However, at poly(alcian blue) MCPE the oxidation peak potential was shifted towards negative side by minimizing the over potential with enhancement in redox peak current, the anodic peak potential was situated at -0.0585 V. Therefore, by this it came to know that the fabricated poly(alcian blue) MCPE can catalyze the favorable oxidation process of HQ.

biosensors-bioelectronics-poly-alcian-blue

Figure 7: Cyclic voltammograms of 0.1 mM HQ in 0.2M PBS pH 7.4 at BCPE (dashed line) and poly(alcian blue) MCPE (solid line) at scan rate of 0.05 Vs-1.

The effect of applied potential scan rate for 0.1 mM HQ in 0.2 M PBS of pH 7.4 was examined by CV technique at poly(alcian blue) MCPE as shown in Figure 8A. The applied scan rate was proportional to the redox peak current, with a small shift in the anodic peak potential (Epa) towards more positive side and cathodic peak potential (Epc) to the less negative side. The graph Ip versus ? and Ip versus ?1/2 was plotted in the range from 0.01 to 0.14 Vs-1 as shown in Figure 8B and Figure 8C. As the obtained correlation coefficient is more linear with Ip versus ?, suggesting electrode phenomenon was an adsorption-controlled [40].

biosensors-bioelectronics-versus-square-root

Figure 8: (A) Cyclic voltammograms of 0.1 mM HQ in 0.2 M PBS solution of pH 7.4 at different scan rate (a-l: 0.01 to 0.14Vs-1) (B) Graph of peak current versus scan rate. (C) Graph of peak current versus square root of scan rate.

The electrocatalytic oxidation of HQ was conducted by varying its concentration in the range 10 to 60μM at Poly(alcian blue) MCPE as shown in the Figure 9A. A linear relationship was established between the Ipa and concentration of HQ as shown in the Figure 9B, with a linear regression equation of Ipa(μA)=0.3563(C0 μM/L)+0.55407, (r2=0.9753). The limit of detection for HQ at Poly(alcian blue) MCPE was calculated in the lower concentration range was found to be 0.142 μM by DPV technique.

biosensors-bioelectronics-peak-potential-versus-pH

Figure 9: (A) Cyclic voltammograms of 0.1 mM HQ at poly(alcian blue) MCPE in 0.2 M PBS solution with different pH values (a–f: 5.5 to 8.0) at scan rate of 0.05 V s-1 (B) Graph of anodic peak potential versus pH.

The effect of phosphate buffer pH value on the cyclic voltammetric determination of 0.1 mM HQ at poly(alcian blue) MCPE was shown in the Figure 10A. The oxidation peak potential shifts to a more negative potential with increasing pH. The Epa versus pH (Figure 10B) graph clearly indicated the depends of Epa with pH in the range of 5.5–8.0 with a slope of 0.0634V/pH (r2=0.9939). This was consistent with the reported literature [24, 26] (Figures 7-10).

biosensors-bioelectronics-PBS-solution-different-pH

Figure 10: (A) Cyclic voltammograms of 0.1 mM HQ at poly(alcian blue) MCPE in 0.2 M PBS solution with different pH values (a–f: 5.5 to 8.0) at scan rate of 0.05 V s-1. (B) Graph of anodic peak potential versus pH.

Simultaneous determination of CC and HQ

The cyclic voltammetry was employed for the simultaneous electrochemical determination of CC and HQ. The Figure 11 shows the cyclic voltammograms obtained for the equimolar (0.1 mM) mixture of CC and HQ in 0.2 M PBS of pH 7.4 at the scan rate of 0.05 Vs-1. The dotted line shows cyclic voltammogram obtained at BCPE, on the other hand the solid line for poly(alcian blue)MCPE and the oxidation potential of CC and HQ were observed at the same potential as determined individually, this voltammetric resolution of CC and HQ at poly(alcian blue) MCPE has great significance in the simultaneous analysis. Differential pulse voltammetry has some advantages like higher current sensitivity and absence of background current, The Figure 12 shows the DPV curves recorded for the equimolar mixture of CC and HQ at BCPE and poly(alcian blue) MCPE. Therefore, the discrimination of oxidation potentials of CC and HQ was achieved at Poly(alcian blue) MCPE (Figures 11 and 12).

biosensors-bioelectronics-simultaneous-determination

Figure 11: Cyclic voltammograms for simultaneous determination of 0.1 mM CC and 0.1mM HQ at BCPE (dashed line) and poly(alcian blue) MCPE (solid line) at scan rate of 0.05 Vs-1.

biosensors-bioelectronics-pulse-voltammogram-obtained

Figure 12: Differential pulse voltammogram obtained for 0.1 mM CC and 0.1 mM HQ in 0.2M PBS solution of pH 7.4 at poly(alcian blue) MCPE (A) and BCPE (B).

Conclusion

A simple and convenient method for the modification of bare carbon paste electrode was proposed by the electropolymerisation of alcian blue followed by its electrochemical and scanning electron microscopic characterization. The strong electrocatalysis in the oxidation of catechol and hydroquinone has made the fabricated electrode as a promising analytical sensor in the simultaneous analysis of both the isomers by cyclic voltammetric and differential pulse voltammetric techniques. Overall, the sensitivity, selectivity, antifouling property, reproducibility, stability was shown by the electrode.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Article Usage

  • Total views: 8533
  • [From(publication date):
    September-2016 - Feb 21, 2018]
  • Breakdown by view type
  • HTML page views : 8414
  • PDF downloads : 119
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2018-19
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2018 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version