Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Preparation of the Silicone-Modified Antimicrobial Polyethylene Endotracheal Tube (PE ETT)

Xuhong J1,2*, Bin LV1,2, Xinmin W3 and Qianhong S3

1The First Clinical Medical College of Zhejiang Chinese Medical University, Hangzhou, China

2Emergency Department, The First Affiliated Hospital of Zhejiang Chinese Medical university, Hangzhou, China

3Zhejiang-California International NanoSystems Institute, Zhejiang University, Hangzhou, China

*Corresponding Author:
Jiang Xuhong
Shangcheng Huansha road 54
Hangzhou city, Zhejiang, China
Tel: +8615988890828
E-mail: [email protected]

Received date: June 25, 2014; Accepted date: September 23, 2014; Published date: September 25, 2014

Citation: Xuhong J, Bin LV, Xinmin W, Qianhong S (2014) Preparation of the Silicone-Modified Antimicrobial Polyethylene Endotracheal Tube (PE ETT). J Med Microb Diagn 3:163. doi: 10.4172/2161-0703.1000163

Copyright: © 2014 Xuhong J, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Medical Microbiology & Diagnosis

Abstract

Polyethylene endotracheal tube (PE ETT) antimicrobial coating has been proved to be a more effective method to prevent endoluminal biofilm formation. A transparent silicone-modified antimicrobial PE ETT was obtained by coating SiO2/KH570/MTES/Ag-SiO2 solution prepared by chemical mixing Ag-SiO2 and SiO2/KH570/MTES solution by means of dip-coating method, followed by drying. All the films were characterized by various techniques including Pencil Hardness Tester, IR, SEM, UV-vis, and ICP-MS. The results indicated that the TEOS/KH570/MTES/Ag-SiO2 (15:6:1:0.6~1.0) films, which exhibit the simple solution-processable film formation on PE ETT, show homogeneous morphology, and have the high transmittance of above 87%, high hardness of 5H, strong adhesion. Furthermore, the silicone-modified antimicrobial PE ETTs show excellent antimicrobial property with the sterilization rate up to 93.5%, it demonstrates that the antimicrobial films have excellent biocompatibility and potent antimicrobial activity against protein adhesion. Pyrogen test, hemolysis test and oral mucous irritation test results are fit for biosecurity use requirement for biological materials. So, silicone-modified antimicrobial PE ETT is hoped to be used in the field of medical materials and can be used as an efficient and economical method to extend the service life of the PE ETT.

Keywords

Endotracheal tube; Antimicrobial film; Silicone; Ag-SiO2

Introduction

Ventilator-associated pneumonia (VAP) is a common complication for mechanical ventilation patients in ICU and trachea cannula, repeated cannula and mechanical ventilation way and mechanical ventilation time are one VAP vulnerable factor [1]. Endotracheal tube (ETT) is artificial biomaterial and it does not have antibacterial property. Once inserted into human body, it becomes the potential source of infection because pathogenic bacteria is absorbed on its surface for growing and breeding to form bacterial biofilm which can adhere dissociated bacteria and avoid bacteria from killing and phagocytosing under the effect of antibiotics and body’s immune system [2]. In order to prevent infection after biomaterial placement, the antibacterial coatings are painted on its surface, such as nanoZnO [3], TiO2 photocatalytic materials [4], Ag-loaded coating [5], etc to inhibit bacteria from adhering on the surface of cannula to form bacterial biofilm.

Among that, there is more research for Ag-loaded coatings in antibacterial materials that Ag+ ion can destroy cell division and mutual respiratory function of bacteria and virus, having stronger bactericidal ability and inhibiting pathogen activity [6,7]. However, there is less research for Ag-loaded antibacterial coating in biomedical materials. The majority research is concentrated on modified or composited antibacterial materials and it is rearly verified for moldability and biocompatibility of antibacterial coatings. Furthermore, Ag+ion is sedimented on the surface of biomaterial or in the meso pore through ion exchange and physical absorption [8]. The poor binding force with base material leads to antibacterial coating shedding, influencing on its antibacterial property and safety. So people prepare antibacterial coatings by recombinating polymer and Ag+ion to enhance binding force between antibacterial coating and base material and improve compactibility, such as polyvinylpyrrolidone [9], polyurethane [10], chitosan [11], etc. However, there is less research for silicone materials. Silicone materials contain many kinds of active group and present excellent cohesiveness and film-processing property [12]. The hydrophobic groups of Si-CH3 and -CH3 could reduce coating surface energy and improve chemical inertness. Its polymer is compatible with tissue and blood of human body and has certain anticoagulation in effect so it is widely applied in biomaterials [13].

In this study, silicon compound of SiO2/KH570 and SiO2/KH570/ MTES prepared through the sol-gel process with TEOS as inorganic silicon source, KH570 and MTES as the organicsilicon source, were used to prepare series silicone-modified antimicrobial films on conventional PE ETT by mixing silver antibacterial agent. Effects of dif??ferent alkoxysilanes of KH570 and MTES content on the properties of silicon compound solutions and their films were investigated. We also study the influence of different Ag-SiO2 content on the stability of the silicone-modified antimicrobial solution and the sterilizing rate of their films. A transparent silicone-modified antimicrobial PE ETT and their properties were introduced in this paper. Its biosecurity is verified in pyrogen test, hemolysis test and oral mucous irritation test. Such silicone-modified antibacterial coatings are hoped to be used in surface coating field of medical materials or medical apparatus and instruments to improve infection situation in the hospital.

Materials and Methods

Regents and materials

Tetraethyl orthosilicate (TEOS), γ-Methacryloxy propyl trimethoxylsilane (KH570), methacryltriethoxysilane (MTES), silver antibacterial agent (Ag-SiO2), ethanol, distilled water, acetic acid and potassium oxalate were analytically grade. 0.9% sodium chloride injection, Zhejiang Shapuaisi Pharmaceutical Co. Ltd. Normal PE tracheal cannula (blank ETT), Covidien.

The silicone-modified antimicrobial solution were synthesized by sol-gel method and mechanical mixing as described in Figure 1. In the preparation of silicon compound solutions of SiO2/KH570 and SiO2/ KH570/MTES, TEOS were first diluted in ethanol with subsequent hydrolysis in water by catalyst under vigorous stirring at the room temperature, then mixed with different amounts of KH570 or the mixture of KH570 and MTES at different ratio were added under stirring until completely hydrolyzing and crossing linked with silica solution. The films of SiO2/KH570 and SiO2/KH570/MTES were obtained by dip-coating on PE ETT. The silicone-modified antimicrobial solution of TEOS/KH570/MTES/Ag-SiO2 was prepared by shear mixing Ag- SiO2 with the silicon compound solutions of SiO2/KH570/MTES.

medical-microbiology-diagnosis-silicone-modified

Figure 1: Flow diagram of the synthesis of silicone-modified antimicrobial solutions.

Figure 2 shows the flow diagram of the preparation of the siliconemodified antimicrobial PE ETT. The PE ETT were treated by corona discharging, then dipped into the solution of TEOS/KH570/MTES/ Ag-SiO2, finally baked at 80~100°C for 1 h to obtain the antimicrobial PE ETT with high transmittance and silicone-modified antimicrobial PE ETT

medical-microbiology-diagnosis-preparation-silicone

Figure 2: Flow diagram of the preparation of silicone-modified antimicrobial PE ETT.

Characterization of the materials

The hardness and adhesion of all the films were measured by pencil hardness tester and Squaring-off, respectively. Infrared spectrums (IR) of two gels of TEOS/KH570/MTES and TEOS/KH570/MTES/ Ag-SiO2 were performed on IR spectrometer. The micrograph and transmittance of the resulting films were measured by scanning electronic microscope (SEM) and a UV spectrometer. The antimicrobial property was evaluated by the method described as GBT 21866-2008 Test method and effect for antibacterial capability of paints film, and the precipitation of Ag+ ion was measured by inductively coupled plasma mass spectrometry (ICP-MS). The biosecurity used in the field of medical materials was evaluated by Pyrogen test, hemolysis test and oral mucous irritation test.

Results

The hardness and adhesion of the films and the stability of the solutions

Organosilicone with various active groups like Si-OH, R1-COO-R, Si-O-Si, etc, is especially preferred to enhance the adhesion of hybrid films on many substrates. KH570 with long chain is an excellent organic modified silicone, which can improve the flexibility of films, and MTES with two stiff groups of CH3 gives the film an improved hardness and is considered to have influence on reducing the interface energy to prevent protein adhesion. So we investigated the ratio of TEOS, KH570 and MTES in order to make better performance films and solutions. The hardness and adhesion of the obtained films including TEOS/KH570, TEOS/KH570/MTES and TEOS/KH570/ MTES/Ag-SiO2, and the storage stability of their solutions are showed in Table 1. The result silicon solution of TEOS/KH570 was flocculated and their film has worse adhesion when the ratio of TEOS and KH570 is 18:3. With increasing the KH570 content, the TEOS/KH570 films exhibit strong adhesion, and their solutions have good storage stability at room temperature for 72 h, but the hardness reduce from 7H to 1H. The films with TEOS:KH570 ratio of 15:6 and 13:8 exhibited desirable hardness of 3H and 5H and excellent adhesion, were investigated further to prepare a higher-quality basic film for the antimicrobial film on PE ETT. MTES used to prepared TEOS/KH570/MTES films can exactly enhance the hardness of the films but sacrifice the adhesion property. So it is especially desirable to keep TEOS/KH570/MTES at the ratios of 15:6:1 and 13:8:1 to prepared the silicone-modified antimicrobial solution by mixing different content of Ag-SiO2. The TEOS:KH570:MTES:Ag-SiO2 films with lower Ag-SiO2 content performed high hardness and strong adhesion, and their solutions are stable when the ratio of TEOS:KH570:MTES:Ag-SiO2 is 15:6:1:0.3, 15:6:1:0.8, 13:8:1:0.3 and 13:8:1:0.8, the solution become unstable and sediment can be observed with too much more Ag-SiO2 because of the agglomerating of nanoparticles.

Samples Ratio(ml/ml) Ag(ml%) Hardness(H) Adhesion Stability
TEOS/KH570 18:3 0 7H 1 Gel
15:6 0 5H 0
13:8 0 3H 0
11:10 0 2H 0
8:13 0 1H 0
TEOS/KH570/MTES 15:6:1 0 5H 0
15:6:3 0 6H 4
15:6:5 0 6H 5
13:8:1 0 3H 0
13:8:3 0 4H 3
13:8:5 0 4H 5
TEOS/KH570/MTES/Ag-SiO2 15:6:1:0.3 1.3 5H 0
15:6:1:0.8 3.5 5H 0
15:6:1:1.5 6.4 5H 0 Deposit
13:8:1:0.3 1.3 3H 0
13:8:1:0.8 3.5 3H 0
  13:8:1:1.5 6.4 3H 0 Deposit

Table 1: The hardness and adhesion of the films and the storage stability of the solutions.

Constituent phase of the films

The IR spectrum of resulting films of TEOS/KH570/MTES (15:6:1) and TEOS/KH570/MTES/Ag-SiO2 (15:6:1:0.8) are shown in Figure 3(a) and 3(b), respectively. Obviously, the IR spectrums are similar and the analyses are gave in Table 2. The broad peak at 3440 cm-1 was assigned to O-H bonds, the strength of that peak in (b) is stronger, because of much more Si-OH groups on the surface of Ag-SiO2. The presence of a strong peak between 1000 cm-1 and 1110 cm-1 was attributed to The IR spectrum of resulting films of TEOS/KH570/MTES (15:6:1) and TEOS/KH570/MTES/Ag-SiO2 (15:6:1:0.8) are shown in Figure 3(a) and 3(b), respectively. Obviously, the IR spectrums are similar and the analyses are gave in Table 2. The broad peak at 3440 cm-1 was assigned to O-H bonds, the strength of that peak in (b) is stronger, because of much more Si-OH groups on the surface of Ag-SiO2. The presence of a strong peak between 1000 cm-1 and 1110 cm-1 was attributed to the stretching vibration of Si-OH or Si-O-Si group, which indicated that TEOS, KH570 and MTES had been reacted sufficiently. The C-H stretching peak at 2850?3000 cm-1 in the spectrum can be an indication of Si-OC2H5 and-CH3 groups mainly from KH570 and MTES. The IR spectrums reflect that TEOS/KH570/MTES and Ag-SiO2 have well compatibility, which is consistent with the SEM analysis.

Samples Wavenumber,cm-1 groups
Fig. ?a? 3446 Si-OH
2850?3000 Si-OC2H5,-CH3,C-H
1740 -C?=O?-O-R
1400 -CH3
1110?1000 Si-O-Si
Fig. ?b? 3446 Si-OH
2850?3000 Si-OC2H5,-CH3,C-H
1740 -C?=O?-O-R
1400 -CH3
1110?1000 Si-O-Si

Table 2: Analyses of the infrared spectrum.

medical-microbiology-diagnosis-IR-TEOS

Figure 3: IR of (a) the TEOS/KH570/MTES (15:6:1); (b) the TEOS/KH570/ MTES/Ag-SiO2 (15:6:1:0.8)

Surface morphology of the films

The surface of coatings is uniform and flat in Figure 4(a) without granule and phase separation. Meanwhile, Ag antibacterial agent content is little in Figure 4(b) coatings imbedded by silicone network polymer, it is not found Ag antibacterial agent particles and the surface is flat. It shows SEM micrographs of the films of TEOS/KH570/MTES and TEOS/KH570/MTES/Ag-SiO2, which were prepared by dipcoating the solution on ETT and baked at 80°C for 1 h. The films is smooth and crack-free, the SEM results show that the films are uniform and amorphous, there is no significant inorganic domain size shown in the two series of films. It indicates that there is a crosslinking of silica sol with alkoxysilanes, and Ag-SiO2 has been homogeneously dispersed in TEOS/KH570/MTES/Ag-SiO2 film.

medical-microbiology-diagnosis-SEM-films

Figure 4: SEM of the films of (a) TEOS/KH570/MTES (15:6:1); (b) TEOS/ KH570/MTES/Ag-SiO2 (15:6:1:0.8).

Transmittance of the films

Figure 5 shows the optical transmittance spectra of ETT (a), ETT coated with TEOS/KH570/MTES film (b) and ETT with TEOS/KH570/ MTES/Ag-SiO2 film (c). The PE ETTs before or after coating with the solution maintained the high transmittance, which are 87.55%, 87.45% and 87.39%, respectively. There is no absorption can be observed in the figure, which indicates that nanoparticles have been homogeneously dispersed in the films, all the films on ETT’s surfaces are homogeneous, and the condensation reaction between the hydroxyl groups of the hydrolysates of the alkoxysilanes and the hydroxyl groups on the surface of the colloidal silica from SiO2 sol and Ag-SiO2 have occurred. It means that antibacterial coating can increase transmittance which keeps the original appearance of ETT and improves its view definition in the application process, convenient for clinical observation.

medical-microbiology-diagnosis-PVE-ETT

Figure 5: The transmittance of PVE ETT coated with (a) blank; (b) TEOS/ KH570/MTES; (c) TEOS/KH570/MTES/Ag-SiO2

Antimicrobial property

The antimicrobial property of TEOS/KH570/MTES/Ag-SiO2 films on PE ETT surface with different Ag-SiO2 content can be researched and tested by the method described as GB/T 21866-2008, and the results are showed in the Table 3. The PE ET T and PE ETT coated with TEOS/KH570/MTES without Ag-SiO2 have no antimicrobial effect. The sterilizing rate increased non-linear with the increasing of Ag-SiO2 content. When the ratio of Ag-SiO2 is 2.6%, the sterilizing rate of antimicrobial PE ETT were up to 93.5%, but there is not obvious change for the sterilizing rate by increasing Ag-SiO2 content. The results indicated that the content of silver is an important factor which influences the antimicrobial properties of the obtained films and it is desirable to contain at least 2.6% by volume or more.

Samples Ratio?ml:ml? Ag(ml%) Sterilizing Rate
PE ETT 0 0
TEOS/KH570/MTES 15:6:1 0 0
TEOS/KH570/MTES/Ag-SiO2 15:6:1:0.2 0.9 71.3%
  15:6:1:0.4 1.8 71%
  15:6:1:0.6 2.6 93.5%
  15:6:1:0.8 2.5 93.7%
  15:6:1:1.0 4.3 93.3%

Table 3: The antimicrobial property of the silicone-modified antimicrobial PE ETT with or without Ag-SiO2.

Analysis for the stabilities of Ag+ ion

In order to research the stability of silver, the antimicrobial film was dipped in hot water for 24 h, then the extraction of Ag+ ion was measured by ICP-MS, and the results are showed in the Table 4. The Ag+ ion concentration of the deionized water and the antimicrobial solution of TEOS/KH570/MTES/Ag-SiO2 are 0.002 μg/L and 0.025×1000 μg/L, respectively. We can see that silver in the antimicrobial solution is about 10000 times more than deionized water. But, in the sample A and sample B, the Ag+ ion concentration are 0.005 μg/L and 0.009 μg/L, respectively, only a little higher than deionized water. That affirms that the Ag-SiO2 has been homogeneously dispersed in the antimicrobial film and firmly chemically bonded with the silicon compound solution. It means that the firm chemical bond is established between Ag antibacterial agent and composite silicone sol and the cohesiveness between antibacteraial coating and ETT base material surface is strong. It can keep antibacterial property for a long time to extend ETT’s use time to certain extent and decrease drawing times thus to reduce inflammation occurance probability.

Treating Condition Samples Ag (µg/L)
Deionized water Blank 0.002
The film dip in 40? hot-water for 24h A 0.005
The film dip in 80? hot-water for 24h B 0.009
The solution diluted 1000 times with deionized water C 0.025

Table 4: The extraction of Ag+ ion of the silicone-modified antimicrobial PE ETT.

Biosecurity

Pyrogen test, hemolysis test and oral mucous irritation test are conducted for the blank ETT and the silicone-modified antimicrobial PE ETT with ratio of 2.6% Ag-SiO2 in order to investigate security of the silicone-modified antimicrobial PE ETT with ratio of 2.6% Ag- SiO2. The results of pyrogen test are shown as data in Table 5. The total rose temperature is 0.7°C for blank ETT and 0.03°C for the siliconemodified antimicrobial PE ETT with ratio of 2.6% Ag-SiO2, which are fit for pyrogen inspection range regulated in the standard of Chinese Pharmacopoeia (Version 2010). It is indicated no pyrogenic effect caused by materials for the silicone-modified antimicrobial PE ETT with ratio of 2.6% Ag-SiO2 and better cohesion between antibacterial coatings and ETT which is not easy for dipping and precipitation. The dipping liquid injecting into rabbit blood is guaranteed not to cause temperature rise reaction and keeps rabbit’s temperature.

Sample Blank ETT silicone-modified antimicrobial PE ETT with ratio of 2.6% Ag-SiO2
B1 B2 B3 C1 C2 C3
mWeight / Kg 2.3 2.3 2.2 2. 2 2.2 2.2
Tbefore /? 0 h 38.69 38.62 38.8 38.69 38.94 38.75
0.5h 38.63 38.5 38.86 38.69 38.87 38.69
equation 38.66 38.56 38.83 38.69 38.9 38.72
Tafter/? 0.5h 38.57 38.56 38.81 38.37 38.62 38.44
1.0h 38.75 38.56 39.25 38.37 38.69 38.56
1.5h 38.75 38.69 39.31 38.25 38.69 38.63
2.0h 38.75 38.62 39.31 38.31 38.69 38.69
2.5h 38.75 38.56 39.25 38.25 38.69 38.63
3.0h 38.69 38.56 39.25 38.37 38.81 38.75
?T /°C
?T=Tafter-equation, Tafter>equation
0.09 0.13 0.48 0 0 0.03
Σ?T/°C 0.7 0.03

Table 5: The results of pyrogen test of blank ETT and silicone-modified antimicrobial PE ETT with ratio of 2.6% Ag-SiO2.

The results of hemolysis test are shown as data in Table 6. Absorbance A in the table is the average absorbance of three same samples. The average absorbance of negative control liquid and positive control liquid is 0.0110 and 0.8097 respectively. The test complies with requirements and the results are valid. The hemolysis ratio of the blank ETT (H1) and the silicone-modified antimicrobial PE ETT with ratio of 2.6% Ag-SiO2 (H2) is 0.5885% and 0.7512% respectively. The hemolysis ratio should be less than 5% according to the standard of Biological Evaluation of Medical Devices: GB/T16886.4-200 Temolysis Test Inspection. It is found that hemolysis test of the silicone-modified antimicrobial PE ETT with ratio of 2.6% Ag-SiO2 is qualified and it is testified that antibacterial ETT is safe to red cell of human body.

Sample H1 H2 H3 H4
A 0.0157 0.0170 0.0110 0.8097
HR / % 0.5885 0.7512

Table 6: The hemolytic ratio of blank ETT and silicone-modified antimicrobial PE ETT with ratio of 2.6% Ag-SiO2.

According to GB/T16886.10-2005 standard, the results of the golden hamster oral mucous irritation test of the blank ETT and the siliconemodified antimicrobial PE ETT with ratio of 2.6% Ag-SiO2 are shown in Figure 6. The extracts of the blank ETT and the silicone-modified antimicrobial PE ETT with ratio of 2.6% Ag-SiO2 showed no irritation reactions to golden hamsters oral mucous. Both the right and left sides of oral mucosals were epithelial integrity, they had no proliferation of epithelial cells, flattened, degeneration or necrosis, submucosal had no congestion, edema, inflammatory infiltration and necrosis and or other pathological changes. The two groups’ oral mucosal tissue reactions of acute exposure were scored 0, showing no irritation.

medical-microbiology-diagnosis-golden-hamsters

Figure 6: The golden hamsters oral mucous irritation test of the blank ETT and the silicone-modified antimicrobial PE ETT with ratio of 2.6% Ag-SiO2.

Above all, the silicone-modified antimicrobial PE ETT with ratio of 2.6% Ag-SiO2 is fit for biosecurity use requirement for biological materials.

Discussion

In this paper, the silicon compound of SiO2/KH570 and SiO2 KH570/MTES prepared through the sol-gel process with TEOS as inorganic silicon source, KH570 and MTES as the organicsilicon source, were used to prepare series silicone-modified antimicrobial films on conventional PE ETT. By researching and analysising the characterization of antimicrobial PE ETT, we determine the proportion of the TEOS/KH570/MTES/Ag-SiO2 (15:6:1:0.6~1.0) films with ratio at least 2.6% to prepare silicone-modified antimicrobial PE ETT. The silicone-modified antimicrobial PE ETT coated by TEOS/ KH570/MTES/Ag-SiO2 (15:6:1:0.6~1.0) films with ratio at least 2.6% has the high transmittance of above 87%, high hardness of 5H, and strong adhesion, exhibits excellent antimicrobial property with the sterilization rate up to 93.5%. What’s more, it is proven in the pyrogen test, hemolysis test and oral mucous irritation test that the siliconemodified antimicrobial PE ETT has excellent biocompatibility and biosecurity of requirement for biological materials and equipped with excellent physical and biological properties. It is hoped to be used in the field of coated materials for medical materials or medical equipment to improve infection situation in the hospital.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Recommended Conferences

  • 10th World Congress on Alzheimers Disease & Dementia
    May 30-31, 2018 Osaka, Japan
  • 4th World Congress and Exhibition on Antibiotics and Antibiotic Resistance
    June 14-15, 2018 Barcelona, Spain
  • 4th Annual Congress on Infectious Diseases  
    September 17-18, 2018 San Diego, USA
  • 13th World Congress on Virology
    September 26-27, 2018 Montreal, Canada

Article Usage

  • Total views: 11882
  • [From(publication date):
    specialissue-2015 - May 22, 2018]
  • Breakdown by view type
  • HTML page views : 8097
  • PDF downloads : 3785
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2018-19
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2018 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
Leave Your Message 24x7