Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Preventing Extubation Failure by Monitoring Minute Ventilation with the ExSpiron 1Xi

Jaideep H Mehta*, Kayla Rohleder and David Amaro-Driedger

Department of Anesthesiology, The University of Texas Medical School at Houston, Texas, USA

*Corresponding Author:
Jaideep H Mehta
Assistant Professor of Anesthesiology
Chief of Acute Pain Medicine, Department of Anesthesiology
The University of Texas Medical School at Houston
6431 Fannin Street, MSB 5.020, Houston, Texas 77030, USA
Tel: 713-500-6200
Fax: 713-500-6201
E-mail: [email protected]

Received date: July 08, 2015 Accepted date: August 17, 2015 Published date: August 24, 2015

Citation: Mehta JH, Rohleder K, Amaro-Driedger D (2015) Preventing Extubation Failure by Monitoring Minute Ventilation with the ExSpiron 1Xi™. J Anesth Clin Res 6:555. doi: 10.4172/2155-6148.1000555

Copyright: © 2015 Mehta JH, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Anesthesia & Clinical Research

Abstract

In the continuing care of postoperative patients who have been extubated, being able to monitor tidal volume and minute ventilation can provide objective criteria to identify respiratory depression and allow for earlier intervention. We present a morbidly obese female patient that was not extubated at the end of the case due to a low tidal volume detected by the anesthesia machine and a real-time non-invasive ventilation monitor that was being trialed for the procedure. This non-invasive ventilation monitor was subsequently useful in continuing to provide continuous information on the patient’s respiratory status to provide objective criteria for extubation and after a trial of extubation. Because of its ability to provide values for minute ventilation and tidal volume, rather than subjective criteria such as chest wall rise, this device may prove useful in the context of extubation trials and beyond.

Keywords

Anesthesia; Reversal; Neostigmine; Minute ventilation; Tidal volume; Exspiron; Real-time monitor

Objectives

In current practice, oxygen saturation has been used as the standard for assessing adequate ventilation in postoperative patients [1]. However, in the setting of decreased ventilation, oxygen diffusion still takes place within the functional residual capacity of the lung, thereby consuming the oxygen present [2]. As there is no replacement of the oxygen due to the decreased minute ventilation, a patient’s oxygen saturation will decrease rapidly and with little warning. There is latency in pulse oximetry that coincides with the plateau in the sigmoid oxygen dissociation curve [3]. During this time frame, diminished minute ventilation can be detected before the pulse oximetry begins to plummet. In patients with opioid or propofol induced respiratory depression, inadequate reversal of muscle relaxation, or emergence from anesthesia, monitoring of minute ventilation in real time, can be accomplished with the ExSpiron 1Xi. This monitoring device may prove to be a critical patient safety device, by allowing more timely notification of the underlying physiology, thereby allowing for earlier intervention than current pulse oximetry [4]. In postoperative patients, deciding when to extubate can be greatly facilitated by going beyond the information provided by oxygen saturation. While patients are often times extubated in the operating room, others receive a trial extubation or require extended ventilation in the post-anesthesia care unit (PACU). It has been demonstrated that in patients with traumatic brain injury, outcomes are negatively impacted by extubation failures, highlighting the importance of having a means of monitoring ventilation parameters during a trial extubation [5]. Weissman, et al. found that “hypoxemic episodes are common in the PACU…(and) timely resolution of desaturation events outside of the operating room is challenging and protracted, potentially placing patients at increased risk” [6] (Figure 1).

anesthesia-clinical-research-Monitoring-tidal-volume

Figure 1: Monitoring tidal volume.

Methods

We present a morbidly obese 49 year old female patient with obstructive sleep apnea who was postoperatively reaccumulating insufflated CO2 and had a low tidal volume, detected by real-time noninvasive monitoring. This occurred despite the use of an opioid sparing multimodal pain regimen, appropriate reversal of muscle relaxation, and early reduction and shut off of volatile anesthetic gases. The multimodal pain regimen involved the intravenous administration of 1 g acetaminophen at the start of the case and 30 mg ketorolac once the surgeon began closing. Minimal opioids were used during the case, with a total of 100 mcg fentanyl upon induction and bolus of 50 mcg during the two and half hour procedure. The last dose of intermediate acting muscle relaxant, rocuronium, was administered one hour prior to close. A 4/4 TOF with sustained tetanus was observed prior to a dose of 70 mcg/kg neostigmine administration. Despite these measures, signs of respiratory depression were still present as measured by a low tidal volume. The borderline low minute ventilation and tidal volume detected by the ExSpiron device at the end of the case, correlated within 6-10 cc of the anesthesia machine, thus confirming the accuracy of the device. This allowed the anesthesia care team to make the decision to transport the patient to the PACU intubated, where she was later extubated having met all criteria, but now with the benefit of real time objective evidence of the adequacy of her post-extubation minute ventilation and tidal volumes. No naloxone was administered to the patient in the OR nor PACU as the total fentanyl administered was less than 1.1 mcg/kg. This highly accurate device enabled the anesthesiologists and nurses to use the objective criteria of minute ventilation throughout the post-extubation period, irrespective of the presence of anesthetics, sedatives, or secured airway, reducing the likelihood of an adverse event.

Results

The patient was safely extubated in the PACU with evidence of her ability to maintain an adequate minute ventilation and tidal volume. She was evaluated postoperatively after being extubated and found to be breathing comfortably with a respiratory rate of 18, tidal volume between 350-400, and an oxygen saturation of 100. She was alert and her minute ventilation had returned to normal, allowing her to be safely discharged from the PACU without an increased length of stay.

Discussion

Measuring a decrease in tidal volume and minute ventilation were factors crucial to delaying the extubation of this patient, which prevented her from becoming hypoxemic in the PACU. Additionally, had she received a trial of extubation, detection of this hypoxemia would likely have taken longer to have been detected by pulse oximetry than by the ExSpiron device as the patient had been extubated on 100% FiO2 [7]. When the functional residual capacity (FRC) has been filled with 100% FiO2, this masks the inadequacy of ventilation until the patient is on the steep portion of the oxygen dissociation curve. This latency of detection in pulse oximetry would likely have led to an emergent reintubation, which is an undesirable adverse event and outcome. Compounding the problem, a patient on supplemental oxygen will have a higher oxygen saturation that will not indicate the problem of low minute ventilation [8]. As put forward by Williams et al. [9] real time ventilation monitoring has previously proved useful for determining the necessity of respiratory interventions surrounding the extubation of a patient with a mass in the upper airway. Supporting this view, the Joint Commission has suggested that the monitoring of ventilation is a key safety measure in patients receiving opioid analgesia [10]. It is important to note the combination of obesity and obstructive sleep apnea in this patient are both risk factors for postoperative complications. In addition to reaching a decision regarding an appropriate time to safely extubate a patient, monitoring minute ventilation also assists in evaluating patients with sleep apnea for apnea/hypopnea frequency postoperatively [11]. Minute ventilation as measured by non-invasive real-time monitors like the ExSpiron can detect “post-operative apneic events resulting from partial or complete airway collapse due to the residual effects of narcotic administration and volatile and/or intravenous anesthetics” [11].

Conclusion

With the advent of highly accurate tidal volume and minute ventilation monitoring, it appears we are on the cusp of a significant leap in patient safety (Figure 2).

anesthesia-clinical-research-ExSpiron-matches-spirometry

Figure 2: ExSpiron matches spirometry values MV, TV, RR.

We will now be able to utilize these two objective criteria in unintubated patients to assess the adequacy of ventilation in intermediate and high-risk patients and to allow us to preemptively intervene prior to a respiratory crisis or adverse event occurring. This is especially important as the device appears sufficiently sensitive to detect respiratory depression in the setting of opioid-induced, propofol-induced and respiratory muscle weakness [12]. This high degree of sensitivity will likely drive the technology to become pervasive in the post-anesthesia care unit, intensive care units and floors, where patients may be on patient controlled analgesia devices (PCAs). The latter potentially has the greatest safety impact, as a significant number of respiratory depression events occur in patients on PCAs [13]. According to Overdyk, respiratory failure after surgery is the “third most common cause patient safety-related adverse events affecting the Medicare population in U.S. hospitals, accounting for 113 events per 1,000 at-risk patient admissions, and (results) in death or anoxic brain injury in the majority of cases” [14]. As respiratory depression is a prelude to respiratory failure, early detection of this condition is crucial to improving patient safety.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Recommended Conferences

Article Usage

  • Total views: 11768
  • [From(publication date):
    August-2015 - Nov 19, 2017]
  • Breakdown by view type
  • HTML page views : 7990
  • PDF downloads : 3778
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords