Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Radial Retrieval of Twisted Diagnosis Catheter Using Mother in Child Technique

Ait Mokhtar Omar*

Place du 1er Mai, CHU Mustapha, Algiers, Algeria.

Corresponding Author:
Ait Mokhtar Omar, M.D
Place du 1er Mai, CHU Mustapha, Algiers, Algeria
Tel: 00213555252845
E-mail ID: [email protected]

Received date: November 06, 2015 Accepted date: November 23, 2015 Published date: December 30, 2015

Citation: Omar AM (2015) Radial Retrieval of Twisted Diagnosis Catheter Using Mother in Child Technique. J Thrombo Cir 1:101. doi: 10.4172/JTCOA.1000101

Copyright: © 2015 Omar AM. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Thrombosis and Circulation: Open Access

Abstract

Coronary angiography and percutaneous coronary intervention (PCI) over the radial approach are becoming more frequent and are recommended by the European Society of Cardiology. The radial approach leads to reduce cost because unable ambulatory angiography and less bleeding especially in elderly who are at high risk of bleeding complication, in this population with high frequency of tortuous access routes, pronounced catheter manipulation may structurally damage the catheter shaft and facilitate twisting and bending. We present case where a twisted coronary catheter became stuck in the radial artery and describe an original percutaneous retrieval technique using mother in child technique.

Keywords

Coronary angiography; Radial access; Coronary angiography complication

Introduction

Coronary angiography and percutaneous coronary intervention (PCI) over the radial approach are becoming more frequent and are recommended by the European Society of Cardiology [1]. The radial approach leads to reduce cost because unable ambulatory angiography and less bleeding especially in elderly who are at high risk of bleeding complication [2], in this population with high frequency of tortuous access routes, pronounced catheter manipulation may structurally damage the catheter shaft and facilitate twisting and bending.

Case Description

A 76 years old man was referred to our institution for stable angina, he was smoker and diabetic. Stress echocardiography showed extended anterior ischemia. We performed coronary angiography over right radial access using 6 French sheath. We used 5 french Judkins Right catheter (JR 3.5 Cordis) to catheterize right coronary artery. After Pronounced catheter manipulation because tortuouse subclavian artery, catheter was twisted and looped at humeral level. We tried to remove twist and knot by passing 0.035 inch J guide wire (Boston Scientific) then a stiff 0.014 guide wire (Progress 200XT Abbott), both wires failed to cross knot despite use of balloon support and catheter counter clockwise torque (Figure 1).

thrombosis-circulation-twisted-catheter

Figure 1: Twisted catheter.

We hypothesized that external support may remove knot, we therefore used mother in child technique to extract catheter. We cut the catheter and used 6 French JR 3.5 guiding catheter and put it in the catheter, the problem is that the guiding catheter is longer than twisted catheter and when we pushed the guiding catheter, the twisted catheter advanced and we were not able to fixe catheter because it’s extremity was entirely into the guiding catheter, in order to fix it we introduced a balloon (2.5 X 20 mm) in catheter and inflated at 18 atmospheres, this served as extension of catheter. We pushed the guiding catheter by exerting clockwise and counter clockwise torque and fixed catheter by balloon traction, this enabled pull back of catheter in the guiding catheter and retrieval of catheter (Figures 2 and 3).

thrombosis-circulation-mother-in-child

Figure 2: Mother in child technique.

thrombosis-circulation-knot-retrieval

Figure 3: Knot retrieval.

Discussion

Catheter twist is frequent and in the most cases easily managed by introducing guide wire and torque [3], however when it is associated with looping it's difficult to retrieve the catheter over the radial sheath, because the small diameter of this artery which is smaller than loop diameter, the usual technique is to snare the catheter by femoral access and retrieve the catheter [4], the other solution is vascular surgery, these techniques are more invasive than our technique which don’t necessitate other arterial access. When radial access is used and is less invasive.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Recommended Journals

Recommended Conferences

Article Usage

  • Total views: 7832
  • [From(publication date):
    December-2015 - Jun 26, 2017]
  • Breakdown by view type
  • HTML page views : 7776
  • PDF downloads :56
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords