alexa Reproducibility and Normative Values of the Parameters of a New Hand-Held Digital Pupillometer | OMICS International
ISSN: 2155-9570
Journal of Clinical & Experimental Ophthalmology

Like us on:

Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Reproducibility and Normative Values of the Parameters of a New Hand-Held Digital Pupillometer

Ken Asakawa* and Hitoshi Ishikawa

Department of Orthoptics and Visual Science, School of Allied Health Sciences, Kitasato University, Kanagawa, Japan

*Corresponding Author:
Ken Asakawa
Department of Orthoptics and Visual Science, School of Allied Health Sciences
Kitasato University, 1-15-1 Kitasato, Sagamihara, Minami-ku, Kanagawa 252-0373, Japan
Tel: +81-42-778-9667
Fax: +81-42-778-9684
E-mail: [email protected]

Received date: April 26, 2017; Accepted date: May 11, 2017; Published date: May 15, 2017

Citation: Asakawa K, Ishikawa H (2017) Reproducibility and Normative Values of the Parameters of a New Hand-Held Digital Pupillometer. J Clin Exp Ophthalmol 8:654. doi: 10.4172/2155-9570.1000654

Copyright: © 2017 Asakawa K, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Clinical & Experimental Ophthalmology

Hand-held digital pupillometer; Reproducibility; Normative value; Parameter

Introduction

Evaluation of pupil size and the pupil light response are important parts of the ophthalmologic examination for detecting several neurological and ocular abnormalities [1]. Our research findings suggest that colored-light pupil response can be a novel tool to differentiate between outer and inner retinal diseases [2]. In contrast, these evaluations require special devices and technical expertise; and a longtime (almost 20 minutes) is required to make the necessary measurements, due to having to adapt the eye in a dark room.

The recent development of a hand-held digital pupillometer (PLR-3000 pupillometer; NeurOptics Inc., Irvine, CA, USA) has allowed a more rapid and reproducible method to measure pupil size and pupil light response without having to rely on the dark adaptation. However, the determination of the normative values of healthy subjects remains unknown. We, therefore, evaluated the intra-examiner reproducibility and determined the normative values of pupil parameters.

Material and Methods

The PLR-3000 pupillometer is focused on the pupil, and then a white light of 180 μW intensity at the duration of 0.8 second is flashed into the pupil. In accordance with the algorithm, the pupil parameters, such as size, latency, constriction velocity, and dilatation velocity are automatically calculated (Figure 1).

clinical-experimental-ophthalmology-Waveform-parameters

Figure 1: Waveform and parameters of the PLR-3000 pupillometer.

We examined 30 eyes of healthy subjects ranging in age from 21 to 23 years. Study protocol was approved by the Institutional Ethics Committee of Kitasato University (2016-G023B). This study followed the tenets of the Declaration of Helsinki for research involving human subjects, and written informed consent was obtained.

Pupil recordings were performed from 10 AM to 2 PM in a quiet room with controlled lighting. Pupil parameters were evaluated twice at 1-minute intervals by an experienced examiner (KA). The intra-examiner reproducibility was examined by the coefficient of variation (CV) and the normal ranges were analyzed with a 95% confidence interval (CI) as the normative values of the parameters.

Results

Table 1 summarizes the mean CVs and the normative values with the normal range determined by the 95% CI of the parameters. The mean CVs for two repeated values determined by the same examiner were 6.0% for the initial pupil size (INIT)-parameter, 4.8% for the minimum pupil size at the end of light stimulus (END)-, 4.7% for the percentage of pupil constriction (DELTA)-, 9.5% for the latency time before onset of pupil constriction (LAT)-, 10.3% for the average pupil constriction velocity (ACV)-, 7.4% for the maximum pupil constriction velocity (MCV)-, 6.6% for the average pupil dilation velocity (ADV)-, 13.7% for the time 75% recovery from the minimum pupil size (T75)-parameter. The reproducibility of the T75 parameter was considerably lower than that for the other parameters.

Pupil parameters Reproducibility (CV) Normative values
Mean SD Mean 95% CI
INIT: maximum size before light stimulus (mm) 6.0 5.0 5.0 4.7-5.3
END: minimum size after light stimulus (mm) 4.8 3.6 2.9 2.8-3.0
DELTA: percentage of constriction (%) 4.7 4.0 40.1 37.6-42.6
LAT: latency time before onset of constriction (sec) 9.5 11.6 0.21 0.20-0.22
ACV: average constriction velocity (mm/sec) 10.3 9.2 3.3 3.0-3.6
MCV: maximum constriction velocity (mm/sec) 7.4 7.8 5.0 4.6-5.4
ADV: average dilation velocity (mm/sec) 6.6 8.0 1.5 1.4-1.6
T75: time at 75% recovery (sec) 13.7 10.1 1.36 1.16-1.56
CV: Coefficient of Variation; SD: Standard Deviation; CI: ConīŦdence Interval

Table 1: Reproducibility and normative values of pupil parameters.

Discussion

In the present study, the parameters had generally good reproducibility. Recently, a novel study on the usefulness of a similar device, i.e., the NPi-100 (NeurOptics Inc.) was reported by Zhao et al. [3]. They concluded that the NPi-100 has a high inter-device reliability by two examiners measuring the same patient. Our findings strongly supported this previous report.

DELTA and LAT parameters reflect the inability of the visual afferent systems. ACV, MCV and ADV, T75 derive exclusively from the parasympathetic and sympathetic nerves, respectively [4]. Intra-examiner reproducibility of the T75 parameter, which means the times to pupil recovery were lower than those in the other parameters. The problem is that the current settings of the device must not give the pupil enough time to recover 75%. The solution is to make the light stimulus less bright and shorter time so that the pupil has enough time to recover. There should be done in another settings menu or should be at least a 5-minute interval of between the light stimuli [5]. Moreover, the normative values can be used as “reference values” because the values of people in their thirties or older remain unknown. In the future, these evaluations should be included to confirm this limitation.

Consequently, for the PLR-3000 pupillometer, normative values may be useful to distinguish various neurological abnormalities, such as optic neuritis, pupil-affecting or -sparing oculomotor nerve palsies, Horner’s syndrome, and Adie’s tonic pupil in clinical practice.

Conflicts of Interest/Disclosures

The authors report no conflicts of interest and have no proprietary interest in any of the materials mentioned in this article.

Acknowledgements

We thank Mr. Robert E. Brandt of MedEd Japan for editing the manuscript. This work was supported by JSPS KAKENHI (grant number 16K21346).

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Article Usage

  • Total views: 559
  • [From(publication date):
    June-2017 - May 21, 2018]
  • Breakdown by view type
  • HTML page views : 515
  • PDF downloads : 44
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2018-19
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2018 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
Leave Your Message 24x7