Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Size-Exclusion Chromatography and Its Optimization for Material Science

Netopilík M* and Trhlíková O

Institute of Macromolecular Chemistry, Academy of Sciences of the Czech Republic, Heyrovský Sq. 2, 162 06 Prague 6, Czech Republic

*Corresponding Author:
Milos Netopilík
Institute of Macromolecular Chemistry
Academy of Sciences of the Czech Republic
Heyrovský Sq. 2, 162 06 Prague 6, Czech Republic
Tel: +420 296809 295
Fax: +420 296809 410
E-mail: [email protected]

Received date: June 29, 2017; Accepted date: July 05, 2017; Published date: July 15, 2017

Citation: Netopilík M, Trhlíková O (2017) Size-Exclusion Chromatography and Its Optimization for Material Science. J Material Sci Eng 6: 358. doi: 10.4172/2169-0022.1000358

Copyright: © 2017 Netopilík M, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Material Sciences & Engineering

Abstract

A theoretical analysis of improving the separation power of size exclusion chromatography by decreasing flowrate is conveyed. The variance of the elution curves is larger than expected form estimated dispersity in molecular weight, Mw/Mn. The shape of the elution curves depends strongly on experimental conditions. When the experimental conditions are approaching those to the ideal separation, judged by statistical properties of the curves, both skew and excess kurtosis of the elution curves increase near the exclusion limit in accord with theoretical prediction. In analyses of polystyrene standards by size exclusion chromatography for polymers of molecular weight about thirty thousand and higher, the longitudinal diffusion is not important even at low flow-rates.

Keywords

Separation mechanism; Size-exclusion chromatography; Band broadening function; Skew; Excess kurtosis

Introduction

The material science has ever been the driving force of the progress which has always advanced with the progress in analytical sciences. For polymer materials, the distribution of chain lengths and, therefore, molecular weight distribution (MWD) is of primary importance. MWD is characterized by its averages, Mn, Mw, Mz,... and dispersities, Mw/Mn, Mz/Mw. Material properties depend on molecular weight of polymers in different ways and they are correlated with different averages. Thus Mn may be correlative with polymer colligative properties, e.g. freezing point depression, Mw with melt viscosity and Mz with toughness [1]. Several methods for the determination of the molecular weight averages were developed, e.g., membrane osmometry, cryoscopy and vapor pressure osmometry for the determination of Mn [2], static light scattering for the determination of Mw [2], sedimentation equilibrium for the determination of Mz [3] etc. However, all methods for the determination of individual molecular-weight averages have been dwarfed by size-exclusion chromatography (SEC) [4] which is a method widely used for the determination of MWD as well as of all averages and dispersities, Mw/Mn and Mz/Mw.

SEC with single concentration detector and results evaluated according to a calibration dependence, constructed from elution curves (maxima) of reference standards, frequently polystyrene or poly(methyl methacrylate), is a relative method and for polymers of different chemical composition the error may reach up to 200% [5]. Even if the analyte is chemically identical with the calibration reference standards, the correct value of Mw/Mn is not obtained because of the effect of the band-broadening [6]. This phenomenon follows from the principle of the method [7,8] and increases the value of Mw/Mn, influencing both Mw and Mn [6]. According to the theory, MWD of polystyrene prepared by anion polymerization (reference standards) is Poissonian type with dispersity decreasing with M [9]. This leads us to the conclusion that the dispersities of the reference standards given by the producers are systematically overestimated [8,10]. However, there is some evidence that efficiency of separation increases with decreasing flow rate, rf [11,12]. and there is evidence that the values of Mw/Mn of the reference standards are much lower than commonly believed [8,10,13,14]. The knowledge of the dispersity of the analyzed samples is extremely important for our understanding of the SEC separation of polymers, because of the above discussed dependences of the material properties on MWD and its averages. The experimental elution curve is related with broadening function (BBF) and theoretical elution curve, characterized by the sample dispersity, Mw/Mn, by the Tung equation [4,6].

In the theory of chromatographic separation and band broadening, BBF is called the elution curve of an analyte uniform in molecular weight and chemical composition [15]. For its description, several mathematical models of separation were developed. The kinetic model by Giddings and Eyring assumes that the capture of molecules of the analyte (analyzed substance) is described by two kinetic constants of ingress and egress [15].

The equilibrium model is based on the concept of theoretical plate on which the equilibrium is formed between molecules of the analyte moving together with MP and those anchored on the surface by enthalpic attractive forces or penetrated into the pores by entropic process basically of Brownian diffusion into pores of the stationary phase (SP). The plate model was first proposed by Martin and Synge [16]. The spacial distribution of the analyte with respect to the longitudinal axis of the separation system, developing in time, was expressed by the binomial distribution. However, further treatments of this physical situation were approximative. The exact solution to the problem gives the probability that an experiment with given probabilities of success and failure is successful, i.e., the molecule is eluted from the column in a volume of a plate after a given number of failures [17].

The obtaining of the separation parameters and the description of elution curves by the Giddings-Eyring equation was shown for low-molecular weight substances separated in the adsorption mode by liquid chromatography (LC). This was possible because the LC separation is effective, the retention (capacity) factor [18] high and the baseline-separation of the sample-components is in many cases achieved. For polymers, separated by SEC such phenomena were not observed. In SEC compared to LC, the situation concerning BBF is much less favourable. The capacity factor [18] in SEC is low and the analyzed polymer samples are almost always disperse in molecular weight. The elution curve of such sample is then the convolution of BBF with theoretical elution curve corresponding with the sample MWD [6]. On the other hand, the elution curves of narrow-MWD are much broader than expected from their MWD [19] and therefore we expect that their shape reflects properties of BBF rather than those of narrow- MWD sample MWD. It has been known soon that the concentration detector signal gives unacceptably broad MWD and the evidence suggests that MWD of polymer standards are much narrower than believed [13,14], which supports the idea of the peak-shape formation by the BBF mechanism. According to the theory, its skew and excess kurtosis (referred to sometimes as `excess' only [7,20]) increase and variance decreases near the exclusion limit of the separation system [7].

From the plate-height mechanism [7,16] it follows strictly that the polymer occurring in MP only is not separated at all and is eluted in the volume approaching that of one plate. In this case, the variance of its BBF approaches zero. For the separation of low-molecular weight substances, characterized by high capacity factor [18], the expectation of the decrease in variance has been verified [7]. With high-molecular wight polymers, the situation is more complicated. Just below the exclusion limit the capture of macromolecules in the pores causes hindered diffusion and the egress process is expected to be slow and the BBF variance to increase. This effect has been frequently observed in times when the injected concentration was usually independent of molecular weight [21,22] and it may have been the consequence of concentration effects as discussed [7]. With concentration decreasing with molecular weight the BBF variance does not change considerably [7]. Tung and Runyon [23] report increasing variance with decreasing elution volume. This is followed by a steep decrease in the vicinity of the exclusion limit. An ample decrease throughout the whole span of elution volume cannot be expected. Such effects do not follow any consistent pattern and can hardly form a part of a theory describing the separation. Polymers eluted far after the exclusion limit show almost entirely Gaussian peaks (and therefore BBF). This can be expected from the random character.

The polymers which elute near the exclusion limit in a given separation system are those of the highest molecular weight. It is therefore necessary to increase the separation power of the system. As recognized by Giddings, two kinetic processes, viz., tortuous and obstructed flow of eluent through and around the particles of packing and the flow profile in the mobile phase whose dispersive effect is checked by transversal diffusion, may combine to reduce and control band broadening. To avoid band broadening, the individual molecule has to sample the complete range of linear flow velocities in a random way as it moves along the column; this is achieved by transverse diffusion [24]. Compared to low-molecular weight substances studied by Giddings by the gas chromatography, the rate of diffusion of polymers in solutions separated by SEC is low [25]. Therefore, by decreasing the flow-rate, an increase in separation power of polymers elution close to the elution limit is expected just by letting time enough to molecules just to compensate for the low rate of transversal diffusion as well as to reach the equilibrium of the polymer between volumes inside and outside the pores of SP in accord with experimental observation [11,12].

The aim of this paper is the demonstration of the improvement of efficiency of separation in SEC aimed at the material research with decreasing flow-rate on the basis of expected statistical parameters of BBF, derived on theoretical model [26].

Materials and Methods

The SEC analyzes were performed using a Pump Deltachrom (Watrex Ltd.) with computer-controlled piston movement, autosampler MIDAS (Spark Ltd.), one column PL gel mixed D (Polymer Laboratories), separating according to the producer in the range approximately 102M ≤ 4−5 × 105, particle size 5 μm. UV/VIS DeltaChrom UVD 200 detector (Watrex) with flow-cell volume of 8 μl, operating at wavelength λ=264 nm, the light-scattering photometer DAWN Heleos II, measuring at 18 angles of observation and Optilab T-rEX differential refractometer (both Wyatt Technology Corp.) were the detectors in the order of flow. The data were collected into the Astra 6.1 (Wyatt Technology Corp.) and Clarity (DataApex Ltd.) softwares, communicating with detectors using U-PAD2 USB acquisition device.

Results and Discussion

The dependence of a physical property P on the particular average of molecular weight Mx can often be described by

P=P(1+k/Mx)-1 (1)

where P is the value of the property at infinite molecular weight and k is a constant. As an example, Figure 1 shows the dependence of the melting point on the chain length expressed as the number of monomer units published by Flory and Vrij [27]. The dependence (Figure 1) can be described by eqn. (1) with Mx=Mn, P=418.8 and k=7.33. Hence, the determination of the correct chain-length and therefore molecular weight is of highest importance in macromolecular science.

material-sciences-engineering-melting-point-temperature

Figure 1: The dependence of the melting point temperature on the number of monomeric units in the molecule [27].

The improvement of the separation can be shown on elution curves for the standard Mp=4 × 105. With decreasing rf the curves are becoming narrower and in the near the exclusion limit they become moreover non-symmetric (Figure 2). The Figure suggests that flow-rate rf=0.5 ml·min−1as a very good compromise between separation power and the time of analysis. On the other hand, in the region 0.01 ≤ rf/ ml·min−1 ≤ 0.035 the shape of the curves changes showing the expected increase in skew, indicating thus increasingly efficient separation. The change in variance of elution curves with rf can be seen in the van Deemter plot [28]

equation (2)

where the constants a, b and c are associated, in the first approximation, with eddy diffusion (the contribution of MWD is in this case neglected), longitudinal diffusion and mass transfer [4], respectively, in logarithmic scale on the rf axis (Figure 3). Only the curve for toluene rises continuously which is apparently due to the effect of longitudinal diffusion of low molecular weight substance, expressed by the term b/ rf in eqn. (2). For the standards of higher M minima appear at rf=0.02 ml min−1.

material-sciences-engineering-elution-curves-nominal

Figure 2: Elution curves of standards of nominal Mp=4 × 105 at different rf/ml•min-1, denoted with the curves. The legends are ordered according to heights of curves in maximum.

material-sciences-engineering-van-deemter-plot

Figure 3: Van Deemter plot (eqn. (2)) for several polystyrene standards with Mp denoted with the curves and toluene.

The decrease in broadness of the elution curves can be seen also in values of the apparent dispersity in molecular weight, Mw/Mn, obtained according to calibration dependence constructed at each rf (Table 1). The values of Mw/Mn for the majority of them significantly lower than those given by the producer were obtained at very low rf. This is in contrast with relatively high values obtained at widely used rf ≈ 1 ml·min−1 (Table 1).

M × 10-3 Mw/Mn
Producer rf/ml min-1 SEC with LS
Producer LS 0.01 0.5 1
900 1084 1.1 1.027 1.087 1.196 1.026
400 381 1.06 1.029 1.092 1.019 1.028
207 206 1.05 1.033 1.066 1.068 1.021
90 84 1.04 1.024 1.068 1.064 1.014
30 31 1.06 1.023 1.072 1.071 1.012
4 4.5 1.06 1.106 1.113 1.116 1.026
Toluene            

Table 1: The molecular weights M of polystyrene standards given by the producer and obtained by the on-line light scattering (LS), dispersity Mw/Mn given by the producer and obtained by the calibration dependence constructed at a given flow-rate F and obtained from the dual light-scattering concentration detection corrected for band broadening (‘SEC with LS).

The variability of the shape with the change of experimental conditions (rf) and low dispersity justify the approximation of BBF for the studies of its statistical properties by elution curve. The values of the skew are highly scattered (Figure 4), but the lowest values are always those for the lowest flow-rate rf=0.01 ml·min−1 and are close to the theoretical line [8] calculated for parameters of the standard of Mp=4 × 105 (see below). The order of points from the left to the right is the same as the order of standards in Table 1. Near the elution limit, estimated as 4.64 ml, the curve as well as the experimental value rises abruptly together with the theoretical line and for the standard of Mp=4 × 105 at rf=0.01 ml·min−1 the value of σ2 is the highest and is close to the rising part of the theoretical curve.

material-sciences-engineering-elution-curves-analytes

Figure 4: Skew of elution curves calculated as a function of elution volume V at rf=0.2 ml.min−1. The flow-rate, rf/ml•min−1, denoted with the points and Mp in thousands with curves. The order of analytes from left to right is the same as in Table 1 from top to bottom.

The large scatter of points at lower rf suggests that operational conditions and changes of the character of the flow, characterized by irregular change in shape of elution curves with rf with rf, discussed Figure 2, rather than instrumental defects (e.g., voids in the separation system) are the cause of the skew.

The theoretical line in Figure 4 was calculated from parameters of the sample of Mp=4 × 105, which is eluted close to the exclusion limit, shows a curvature. The points at rf=0.01 ml·min−1 are for polymers of Mp < 4 × 105 the lowest and close to the theoretical dependence which was calculated for data of sample Mp=4 × 105. However, for this sample the value of γS is the highest of all, in accord with the theoretical curve. The points for the standard of Mp=9 × 105, which are in front of V0, are scattered according no rule, indicating imperfect separation, the point at rf=0.01 ml·min−1 being just in the middle.

The excess kurtosis, γE calculated shows also a strong increase of values obtained at the lowest flow-rate, rf=0.01 ml·min−1, near the exclusion limit (Figure 5). Theoretical curve obtained by the same procedure [8] as that of γS, shows again a strong increase near the exclusion limit and the curve calculated for the standard of Mp=4 × 105 fits the experimental data for this standard best. The points at rf=0.01 ml·min−1 for polymers of Mp < 4 × 105 are the lowest and close to the theoretical dependence. For data of sample Mp=4 × 105 the value of γE is the highest of all in accord with the theoretical curve. The points for the standard of Mp=9 × 105 in front of V° do not also obey any rule.

material-sciences-engineering-kurtosis-elution-curves

Figure 5: Excess kurtosis of elution curves as a function of elution volume. For the details of the points and lines calculated see legend to Figure 4 and text.

Conclusions

Due to the transversal diffusion and increases the time necessary for a molecule to sample the complete range of flow velocities which is a condition for a good separation. This is achieved in the region of high M at extreme low rf in accord with the idea of transversal diffusion as a factor decreasing band broadening, as recognized by Giddings [24].

For polymers with approximately M ≥ 3 × 104 the longitudinal diffusion ceases to play an important role in the SEC separation process. With decreasing rf the efficiency of separation increases and the separation mechanism appears, using γS and γ E as a criterion, to approach the mechanism of equilibrium on theoretical plate combined with longitudinal shift, proposed by Martin and Synge [16] and solved exactly by Netopilík [26].

For polystyrene in THF for 9 × 105 ≤ M ≤ 4 × 103, the flow-rate rf=0.5 ml·min−1 appears, rather than usual rf=1 ml·min−1, a good compromise between efficiency of separation and the time of analysis. Further decrease in rf has no observable effect down to rf ≈ 0.06 ml·min−1, then a new region of increase in separation power is for some standards of higher M observed.

Acknowledgements

The financial support of the Grant Agency of the Czech Republic (project 17-04258J) is gratefully acknowledged. The research was also supported by European Research Development Fund, project OPPK CZ.2.16/3.1.00/24504 and Egyptian Academy of Sci. & Technology for funding the joint mobility project 2016/2018 (ASRT-16-2).

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Recommended Conferences

Article Usage

  • Total views: 496
  • [From(publication date):
    August-2017 - Jul 19, 2018]
  • Breakdown by view type
  • HTML page views : 446
  • PDF downloads : 50
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2018-19
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

+1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]online.com

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2018 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
Leave Your Message 24x7