Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Stereotactic Radiotherapy for Localized Prostate Cancer, which is Better CyberKnife or RapidArc?

Hegazy MW1,2*, Mahmood R1, Nobah A3, Moftah B3 and Alzorkany F3

1Department of Radiation Oncology, King Faisal Specialist Hospital and Research Centre, Riyadh, Saudi Arabia

2Department of Clinical Oncology and Nuclear Medicine, Faculty of Medicine, Zagazig University, Egypt

3Department of Biomedical Physics, King Faisal Specialist Hospital and Research Centre, Riyadh, Saudi Arabia

*Corresponding Author:
Hegazy MW
Department of Clinical Oncology and Nuclear Medicine
Faculty of Medicine
Zagazig University, Egypt
Tel: +966 550742859
E-mail: [email protected]

Received date: May 09, 2016; Accepted date: July 26, 2016; Published date: August 02, 2016

Citation: Hegazy MW, Mahmood R, Nobah A, Moftah B, Alzorkany F (2016) Stereotactic Radiotherapy for Localized Prostate Cancer, which is Better CyberKnife or RapidArc? J Pros Canc 1:108.

Copyright: © 2016 Hegazy MW, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Prostate Cancer

Abstract

Aim: We aim to compare two different stereotactic body radiation therapy (SBRT) techniques, non-isocentric CyberKnife (CK) with isocentric RapidArc (RA), more widely available treatment technique for the treatment of localized prostate cancer. Methods: The study included six patients treated with CK then re-planned with the new version of flattening filter free (FFF) RA and CK. The prescription dose was 36.25 Gy in five fractions. The two SBRT techniques were compared by target coverage, normal tissue sparing, dose distribution parameters and delivery time. Results: The RA technique exhibited comparable PTV coverage and better bladder and rectum sparing at high doses. The conformity and homogeneity indices of the RA were better and statistically significant than the CK plans. Additionally, the RA resulted in statistically significant lower dose regions and faster delivery times than the CK. Conclusion: The good dosimetric distribution and shorter delivery time make the new version of RA an attractive and reasonable alternative SBRT technique for the treatment of localized prostate cancer; however, no intrafractional (real-time) target tracking is possible on the RA, which is available on the CK platform.

Keywords

Stereotactic radiotherapy; RapidArc; CyberKnife

Introduction

Over recent years, stereotactic body radiotherapy (SBRT) has been increasing in the management of low-and intermediate-risk prostate adenocarcinoma [1]. SBRT is the delivery of either a single dose or a small number of fractions of high-dose, precisely targeted, highly conformal radiation therapy [2]. SBRT improved outcomes may be due to the dose escalation alone, which is widely accepted in prostate adenocarcinoma [3,4]; or as a result of the increased dose per fraction, due to the apparent low α/β ratio of prostatic adenocarcinoma [5,6].

The SBRT, including the CyberKnife technique (CK), has achieved promising clinical results in the treatment of prostate cancer [7-10]. The CK stereotactic radiotherapy system is an accurate image-guided method with intrafractional (real-time) target tracking for delivering radiation to a precisely targeted area using multiple non-isocentric beams with steep surrounding-dose gradients [11]. RapidArc (RA) is volumetric-modulated arc radiotherapy (VMAT) technique that can deliver highly conformal, intensity-modulated radiation doses by a single or multiple rotations of the gantry of the linear accelerator [12].

One of the potential benefits of SBRT is reduced patient visits (4 to 5 fractions) rather than the standard 35 to 40 fractions. A number of small studies and single-centre series have been published showing comparable outcomes with conventional fractionation schedules [13-19]. Most of these studies have used CK treatment. These preliminary studies and case series have led to the development of the PACE study, an international, randomized, phase III study comparing SBRT with both surgery and conventionally fractionated intensity-modulated radiotherapy in two parallel arms [20].

Our study may present an alternative to the CK platform, with the additional benefits of increased availability of gantry-based volumetric systems and possibly shorter delivery times.

Materials and Methods

Patients and imaging

Six patients with localized prostate cancer who had recently received hypo-fractionated radical radiotherapy by CK were included in the study. Gold seeds are implanted to prostate by radiologists under transrectal ultrasound guidance (TRUS), and then one week after 1 mm slice thickness CT simulation in a supine position with alpha cradle with bladder and bowel preparation protocol and MR images are acquired. CT scans from the fifth lumbar vertebrae level to below the ischial tuberosities. CT and MR images were transferred to a virtual simulation program to delineate regions of interest (ROIs).

Contouring and SBRT treatment plan requirements

The same radiation oncologist delineated the target and the critical structures using CT and MR fusion. The following ROIs were defined: Clinical Tumor Volume (CTV) consisted of the prostate without margin; The CTV was increased by 5 mm to create the planning target volume (PTV) for the RapidArc and by 3 mm for PTV for the CyberKnife. Organs at risk (OARs) that were delineated included the entire rectum; the whole bladder; penile bulb and femoral heads. A treatment plan that delivered 36.25 Gy to the PTVs was attempted in both treatment planning systems. Both modalities were required to achieve these criteria as per a combination of studies [7,9,20-25]. Required planning constraints are detailed in Table 1.

PTVRectum Minimum dose received by PTV ≥34.4 Gy 95% of prescription dose
V18<50%
V28<20%
V36<1 cm3
 
Bladder V18<40%
V36<10cm3
 
Femoral heads Maximum point dose 30 Gy
Penile bulb Maximum point dose No more than 100% of prescription dose

Table 1: Dose planning constraints.

Treatment planning

RapidArc: Eclipse treatment planning system (Varian Medical Systems, Inc., Palo Alto, CA) was used along with the analytical anisotropic algorithm (AAA, Version 11.031) dose calculation algorithm. For RA optimization, progressive resolution optimizer (PRO) Version 11.031 was used. All plans generated using True Beam LINACS with 120 leaf millennium multileaf collimator and kV imaging, 2 full arcs one clockwise and the other counterclockwise, with collimator angles ± 20° from 0° and a couch angle of 0°, SRS Arc mode, 6 MV Flattening Free Filter (FFF) beam energy, and maximum dose rate of 1400 MU/min. Both arcs had the same isocenter, located at the center of the PTV.

CyberKnife: Inverse planning was carried out using the sequential optimization algorithm using MULTIPLAN v. 5.1 for delivery on a CyberKnife G4 v10.1 (only the IRIS collimators was used) which is capable of delivering 800 MUs/Min. The planning approach was to use a Prostate-path (allows 114 Nodes and 5° of pitch correction), 2 shells (i.e., 2 mm to control the conformity to the PTV and 40 mm to control the dose spillage), and the range of MUs per beam per fraction was 50- 150. The PTV was prescribed to an isodose line of (78% to 80%) and a maximum of conformity index (CI) 1.24 was achieved. The PTV was at least covered by 99.2% of dose or more. The range of beam numbers was 100-150.

Plan evaluation statistics

Target volume coverage: The percentages of the PTV that received 95% of the prescription dose, PTV mean, median and maximum were compared between the CK and RA plans.

Dosimetric parameters: The normalized conformity index (nCI) was calculated as:

CI=VRI/TV

But homogeneity index equation as follows:

HI=Imax/RI

Where, VRI: Volume of prescribed dose for PTV; TV: Total volume of PTV; Imax: Maximum dose and RI: Prescribed dose of PTV [26].

The percentages of the rectum and urinary bladder that received 18 Gy (V18), 28 Gy and the volume (cm3) that received 36 Gy (V36) of the prescription dose were compared. The dose regions (V10, V20, V30 and V36) were evaluated based on the body volume received 10, 20, 30 and 36 Gy of the prescription dose for each group.

Treatment time

The on-couch time for the patient will be added to the overall beam-on time for the RapidArc treatments, in our hospital the imaging and registration time is on average 3 min to 4 min (CBCT 1 min, reconstruction and registration in three planes 2 min to 3 min).

Statistical analysis

The dosimetric endpoints of the target volumes, normal organs, CI, HI and delivery time were analyzed using the non-parametric (small sample size) Wilcoxon signed rank test (SPSS, V19, USA), a probability value of <0.05 considered to be statistically significant (two tailed).

Results

Target volume coverage

All dose constraints regarding PTV coverage were similarly achieved by both plans except the maximal doses generated by the RA plans were statistically significant lower than those of the CK plans (p value 0.028).

Comparison of dosimetric parameters

The median of Conformity and homogeneity indices of RA plans had a statistically significant higher degree of conformity than the CK plans (p values 0.027 and 0.026) (Table 2).

  Patient 1 Patient 2 Patient 3 Patient 4 Patient 5 Patient 6
RA CK RA CK RA CK RA CK RA CK RA CK
Prostate volume(cm3) 160 95 45 79 96 63
PTV volume(cm3) 264 211 174 134 95 78 159 99 178 152 120 71
PTV34.4 (95%) 99 99.2 99.3 99.4 99.5 99.5 99.3 99.5 99.7 99.4 99.4 99.5
PTV Mean(Gy) 36.8 30.5 36.3 38.2 36.3 40.5 36.3 39.9 36.3 40.6 36.3 39.6
Mean(%) 102 84 100 105.4 100 112 100 110 100 112 100 109
PTV max(Gy) 39 45.9 38 38.2 39.4 46.5 39.3 45.9 38.3 45.9 37.9 45.3
max(%) 108 127 105 127 109 128 108 127 106 127 105 125
nCI 1.07 1.17 1.1 1.14 1.06 1.22 0.95 1.24 1.04 1.20 1.05 1.22
HI 1.08 1.27 1.05 1.27 1.09 1.28 1.08 1.27 1.06 1.27 1.04 1.25

Table 2: Target dose results.

Normal tissue sparing

We analyzed the volume dose parameters of the rectum and bladder. All requirements were comparable in both groups; however, the CK plans achieved a superior sparing of the rectum and bladder in the low dose region (V18 and 28 for rectum and V18 for bladder) but statistically insignificant (p values 0.3, 0.2 and 0.4), while the RA plans exhibited a lower percentage of rectal and bladder volumes that received 100% of the prescription dose compared with the CK plans but statistically insignificant (p values 0.17 and 0.11). The femoral heads and penile bulb received lower doses with RA. The body V35 was the same for both groups, whereas body V10, V20, V30 was all lower with RA in all cases with p values of 0.04, 0.03 and 0.06 respectively (Table 3).

  Patient 1 Patient 2 Patient 3 Patient 4 Patient 5 Patient 6
RA CK RA CK RA CK RA CK RA CK RA CK
Rectum
V18 Gy (%) 41.3 46.8 45.9 35.8 39.5 43.1 46.1 31.6 37.9 41.4 47.6 41.2
V28 Gy (%) 18.5 17.9 16.3 15.4 14.8 20.8 17.2 16.8 16.5 11.3 22.6 15.5
V36 Gy (cm3) (%) 1.4 0.70 0.01 0.30 0.05 1.7 0.01 0.81 0.01 0.34 0.03 0.40
Bladder
V18 Gy (%) 37.7 35.4 37.3 36 10.6 13.3 33.6 36 33.7 36 85.2 17.2
V36 Gy (cm3) (%) 7.2 7.3 0.3 5.2 1.5 6.88 0.8 1.37 0.01 3.6 1.05 0.20
Penile bulb
Max dose (Gy) 30.6 30.5 36.7 34.9 38.4 38.6 39.3 31.7 38 37.2 36.5 36.7
Max dose (%) 84.4 84.1 101.2 96.3 106 106.5 108.4 87.5 105 102.6 100.7 101
RFH
Max (Gy) 18.6 25 12.6 20 14.9 11.6 10.9 19.3 16.8 25.7 11.3 17.7
LFH
Max (Gy) 18.5 26.7 11.7 18.6 13.5 12.5 18.9 15.1 17.2 22 10 14.5
Body
Volume (cm3) 23469 46767 29441 32226 21826 24312
V10 Gy (%) 10.6 11.9 4.3 7.4 3.9 4.3 5.7 6 7.4 9.7 5.5 5.3
V20 Gy (%) 3 3 1.2 1.9 1 1.1 1.3 1.4 2 2.7 1.3 1.4
V30 Gy (%) 1.5 1.4 0.5 0.9 0.5 0.5 0.7 0.6 1.1 1.3 0.7 0.6
V35 Gy (%) 1.1 1 0.4 0.6 0.3 0.3 0.5 0.4 0.8 0.9 0.5 0.3
Time (min) 5.4 33 5.4 34 5.4 35 5.4 35 5.4 38 5.4 31

Table 3: Organs at risk dose and treatment time results.

Treatment time

Across all five fractions of radiation treatment, the estimated delivery time of the RA (beam on 2.4 min, CBCT 1 min, reconstruction and registration in three planes 2 min) is faster than the CK (Mean and Median 5.4, 5.4 vs 34.3, 34.5 min) with significance (p value 0.028). The details are shown in Table 4.

  RA CK P value
Mean Median Mean Median
PTV95% 99.4 99.4 99.4 99.5 0.49
PTVmax 107 106.7 127 126.6 0.028
PTVmean 100 100 105 110 0.34
nCI 1.05 1.06 1.2 1.21 0.027
HI 1.1 1.07 1.27 1.27 0.026
Rectum
V18Gy(%) 43.1 43.6 40 41.3 0.35
V28Gy(%) 17.7 16.9 16.3 16.5 0.25
V36Gy (cm3)(%) 0.25 0.02 0.71 0.54 0.173
Baldder
V18Gy(%) 39.7 35.5 27.3 30.8 0.46
V36Gy(cm3)(%) 1.8 0.93 4.1 4.4 0.116
Penile bulb max Gy(%) 101 103 96.3 99 0.173
Femur head max Gy (%)
Right 14.2 13.8 20 19.6 0.046
Left 15 15.4 18.2 16.9 0.116
Body
V10 (%) 6.23 5.6 7.4 6.7 0.046
V20 (%) 1.63 1.3 1.9 1.65 0.038
V30(%) 0.83 0.7 0.9 0.75 0.68
V36 (%) 0.6 0.5 0.6 0.5 0.78
Time 2.4 2.4 34.3 34.5 0.028

Table 4: Statistical results of both plans.

Discussion

Stereotactic body radiotherapy (SBRT) using CK has recently emerged as an alternative technique to deliver hypofractionated radiotherapy to the prostate, comparable in many respects to high dose rate (HDR) brachytherapy, but with a noninvasive approach [7,13,15,17,27-29].

Almost all the literature on prostate SBRT delivers these treatments on CyberKnife but there is increasing interest in using conventional LINACS to deliver SBRT. In our study we tried to find out a reasonable alternative to CK especially with the new versions of RA. We compared the two methods of SBRT delivery available to us at our institution- CyberKnife (using Multiplan planning software) and RapidArc (arc therapy using Varian LINACS). We used the criteria regarding to the prescription dose, dose constraints and CTV to PTV margin depending on many studies [7,9,17,20-33].

CyberKnife system incorporates near real-time tracking of the prostate which allows smaller margins to be used, as it tracks and corrects for intra-fraction motion so if intra-fraction motion could be tracked and corrected during arc-therapy on a conventional LINAC, then the dosimetry is likely to be at least as good as CK. However, without intra-fraction motion control, a margin of 2 mm to 3 mm is likely to be needed to account for intrafraction motion after initial setup to gold markers [34]. Alternatively, if new flattening filter free (FFF) LINACS can deliver these plans in much shorter times, then intrafraction motion may be less important, although transient, significant excursions of prostate position are still possible [35] and this was implicated in our study, by using RA with FFF.

In our study, the dose constraints regarding PTV coverage were achieved by both planning techniques. The conformity and homogeneity indices of the RA plans were better than the CK plans and statistically significant (p values 0.027 and 0.026). The constraints on the rectum and bladder were well achieved by both planning techniques. Additionally, the RA plans resulted in lower body dose regions (V10, V20, V30 were all lower with RA in all cases with p values of 0.04, 0.03 and 0.06 respectively but V35 was the same for both groups), lower statistically significant monitor units, and faster statistically significant delivery times (mean 2.4 min) than the CK plans (34.3 mean min).

Tree et al. [36] have shown that RA and CK can produce clinically acceptable plans where comparing the new versions of CK and RA treatments, the results showed a higher conformity and a lesser lowdose region of the RA when compared to the CK, with a similar target coverage and normal tissue sparing, even under different PTV margins. The mean estimated delivery time for treatments was 39 min for CK and 3 min for RA. These differences were the result of the different nature of the CK and RA systems.

Macdougall et al. [37] have shown that there is no discernible dosimetric advantage to choosing CyberKnife over RapidArc for SBRT delivery in prostate cancer. All dose constraints regarding PTV coverage were achieved by both planning platforms. The nCI was consistently better with RapidArc. The constraints on the rectum and bladder were well achieved by both planning platforms. The body V35 was consistently lower with CyberKnife, whereas body V10, V20, V30 were all lower with RapidArc in all cases. The mean estimated delivery time for treatments was 39 min for CyberKnife and 3 min for RapidArc.

Lin et al. [38] have shown that RA plans were shown to have a greater potential than CK plans in creating conformal dose distributions and in reducing treatment times. The RA plans consistently exhibited superior PTV coverage and better rectum sparing at low doses in the both groups. The conformity and homogeneity indices of the RA plans were better than the CK plans. Additionally, the RA plans resulted in fewer low-dose regions, lower MUs, and faster delivery times than the CK plans. Currently, RA has become one of the SBRT options for localized prostate cancer treatments at their institution (Figure 1).

prostate-cancer-isodose-lines-one-patient

Figure 1: Displayed the CK & RA isodose lines of one patient (axial, sagittal, coronal & 3D view).

Conclusion

There is no dosimetric advantage for choosing CK over RA for SBRT delivery in prostate cancer. Given the significant benefits of RA in terms of availability and delivery time, make the new version of RA reasonable alternative SBRT technique for the treatment of localized prostate cancer; however, no intrafractional (real-time) target tracking is possible on the RA, which is available on the CK platform.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Article Usage

  • Total views: 8118
  • [From(publication date):
    December-2016 - Aug 20, 2017]
  • Breakdown by view type
  • HTML page views : 8030
  • PDF downloads :88
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords