Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

The Application of Portable GC-MS on the Petrochemical Wharf

Zhao Yun*

China Waterborne Transport Research Institute, China

*Corresponding Author:
Zhao Yun
China Waterborne Transport Research Institute, China
Tel: 008613810864135
E-mail: [email protected]

Received Date: June 19, 2016; Accepted Date: June 24, 2017; Published Date: June 29, 2017

Citation: Yun Z (2017) The Application of Portable GC-MS on the Petrochemical Wharf. J Chromatogr Sep Tech 8:374. doi: 10.4172/2157-7064.1000374

Copyright: © 2017 Yun Z. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Chromatography & Separation Techniques

Abstract

It is very important to quickly identify the liquid chemicals in the petrochemical wharf when they volatile in the air. Since most of the liquid chemicals transported in the petrochemical wharf are harmful to people, and they spread very fast in the air. If the exact kind of liquid chemicals is undefined or is wrongly defined, it will lead to a big mistake in detection result. The mass spectrum is very effective in determining the gas kinds, but its big size and complex sampling mode are not convenient for quick detection. To resolve this problem, a method based on portable GC_MS for VOCs determination is developed. The VOCs are directly extracted by solid phase micro extraction head for 120 seconds. After that, they are analysed on the portable GC-MS, determined by retention time and MS peaks, quantified by external standard method. It only takes 6.4 min from sample extracting to get the result. And its linear correlation is larger than 0.98, and the minimum detectable concentration is lower than half of their occupational exposure limits. In the simulation test, the relative error is lower than 4.4%. This method can directly sample in the air, rarely depends on external conditions. At the same time, it can determine and quantify multiple kinds of VOCs at a time. So, it will be very suitable for the VOCs emergency detection in the leakage scene.

Keywords

Portable GC-MS; Volatile organic compounds; Quick detection; Direct sample; Solid phase micro extraction; Petrochemical wharf

Introduction

There are many kinds of dangerous chemicals in the petrochemical wharf area. And most of them are volatile organic compounds(VOCs). The pollution accidents have occurred many times in the petrochemical wharf area [1,2]. The VOCs diffuses very fast in the air, and would pollute water and soil on a large scale. Also, there are many kinds of VOCs, and their toxicities are very complicated. Once the pollution accidents occurred, the people nearby would be in danger [3,4]. The identification of VOCs is not difficult, but there are few detecting method to quickly quantify the concentration of VOCs when they are identified. So, it is necessary to develop a quick detecting method to identify and quantify the concentration of VOCs in the petrochemical wharf area. The portable gas chromatograph-mass spectrometer (portable GC_MS) is quite small, and convenient to carry out. So, it is very suitable for quick detection on the scene. When it is used on the scene, the sampling method is very important. Solid phase micro extraction (SPME) is a sampling method without solvents. It can combine the sampling, concentration and injection [5-7]. Especially, the SPME could direct sample in the air, so it will save much time.

Experimental

Instruments and reagents

Portable GC_MS, gas distributing device (MF-3C), bladder tank (5 L), SPME sampling head, micro syringe (10 μL,100 μL), standard substance (benzene, toluene, paraxylene, aniline, trichloroethylene, tetrachloroethylene, cyclohexanone).

Analysis conditions

chromatographic condition: Injection temperature: 270°C temperature progress: hold 40°C for 10 s, then heat up at a speed of 2°C/s to 280°C, then hold for 20 s, flow speed: 0.15 mL/min.

Ionization source: EI, energy level: 70 eV, scan mode: Full scan, scan range: 45-500 m/z.

Sampling

The air sample is directly taken by the SPME sampling head for 120 s. Then it is injected into the portable GC_MS without further treatment.

Results and Discussion

The concentrations of standard series of VOCs

The most important purpose of quick detection is to evaluate whether the pollutant concentration will cause health damage and to determine the pollution scope. Here we could refer to the occupational exposure limits of each VOCs, set by national standard. When the pollutant concentrations in the field are higher than their occupational exposure limits, it should be identified as contaminated area. This requires that the minimum detectable concentration of each chemical hazardous agent must be lower than its occupational exposure limit. According to the <Occupational exposure limits for hazardous agents in the work place: Chemical hazardous agents> (GBZ 2.1-2007) [8], the occupational exposure limits of benzene, toluene, paraxylene, aniline, trichloroethylene, tetrachloro ethylene and cyclohexanone are 6, 50, 50, 3, 30, 200 and 50 mg/m3, respectively. Based on this, we set the concentrations of standard series gases as 0.5, 1, 2, 4, and 8 times of their occupational exposure limits. The specific concentrations are shown in Table 1.

Standard series Benzene (mg/m3) Toluene(mg/m3) Paraxylene (mg/m3) Aniline(mg/m3) Trichloroethylene(mg/m3) Tetrachloroethylene (mg/m3) Cyclohexanone(mg/m3)
Series 1 3 25 25 1.5 15 100 25
Series 2 6 50 50 3 30 200 50
Series 3 12 100 100 6 60 400 100
Series 4 24 200 200 12 120 800 200
Series 5 48 400 400 24 240 1600 400

Table 1: The concentration of standard series gases.

The sampling time

The more gases the SPME sample hand collected, the higher precision will be obtained. correspondingly, the sampling time needed will be longer. and if the sampling time is too long, it will not match the demand of the quick detection. So, a balance between the sampling time and detecting precision will be needed, at which, the analysis result could be obtained quickly and the analysis precision is accuracy enough. The analysis precisions at different sampling time are compared using the detection of minimum concentration in the standard series gases. When the sampling time is 120 s, the analysis precision of every gas is good, and 120 s is short enough. So, the sampling time is set as 120 s. The separating effect is shown in Figure 1.

chromatography-separation-techniques-separation-effect

Figure 1: The separation effect of 7 kinds of VOCs (1. Benzene; 2. trichloroethylene; 3. Toluene; 4. Tetrachloroethylene; 5. Paraxylene; 6. Cyclohexanone; 7. Aniline).

The standard curve

The standard mix gases of benzene, toluene, paraxylene, aniline, trichloroethylene, tetrachloroethylene and cyclohexanone are prepared with pure air. The concentrations are 48, 400, 400, 24, 240, 1600, 400 mg/ m3 respectively. And it is diluted into the series concentrations shown in Table 1. Then the series gases are analysed on portable GC_MS. All the gases are determined by the mass spectrum peak and retention time, and the standard curve are plotted through the peak area. Also, the equations of linear regression and correlation coefficients are obtained. The minimum detectable concentrations are calculated through 3 times noise signals. The equations of linear regression, correlation coefficients and the minimum detectable concentrations are shown in Table 2.

Compounds Retention Time(s) Mass Peak(m/z) Linear Equation Correlation Coefficient Minimum Detectable Concentration(mg/m3)
Benzene 43.0 77 y=4.369x-10.66 0.994 1.1
Toluene 61.4 91 y=0.891x-6.833 0.995 21
Paraxylene 68.9 104 y=2.931x-7.541 0.993 23
Aniline 83.4 91 y=4.513x-1.375 0.986 0.6
Trichloroethylene 46.1 129,131 y=10.86x-10.62 0.996 12
Tetrachloroethylene 61.2 160,162,164 y=0.099x-6.041 0.996 26
Cyclohexanone 74.0 98 y=0.262x+1.375 0.994 20

Table 2: The linear correlations and minimum detection concentrations of 7 kinds of VOCs by portable GC-MS.

Discussion

A new quick detection method for VOCs based on portable GC_MS and SPME head was developed. And a mixture of 7 kinds of VOCs was analysed by this method. This detection method took only 6.4 mins to get the analysis result. And the SPME head was used in the sampling process, and it provided convenient sampling procedure. This combination of portable GC_MS and SPME head would make the detection of VOCs more convenient and fast. So, it would be very suitable for the VOCs quick detection in the petrochemical wharf area. And it is helpful to determine the pollution scope and improve the worker’s occupational health for reduce the damage from VOCs [9-13].

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Recommended Conferences

Article Usage

  • Total views: 636
  • [From(publication date):
    August-2017 - Jul 19, 2018]
  • Breakdown by view type
  • HTML page views : 576
  • PDF downloads : 60
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2018-19
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

+1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2018 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version