Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

UPLC Analytical Method Development and Validation for the Simultaneous Estimation of Paracetamol and Caffeine Capsules Dosages Form

Bikash Ranjan Jena1, Sitty Manohar Babu1, Debi Prasad Pradhan2, Suryakanta Swain1*

1College of Pharmacy, Southern Institute of Medical Sciences, Guntur, Andhra Pradesh, India

2College of Pharmacy, GITAM University, Visakhapatnam, Andhra Pradesh, India

*Corresponding Author:
Suryakanta Swain
College of Pharmacy
Southern Institute of Medical Sciences
Mangaldas Nagar, Vijyawada Road
Guntur-522001, Andhra Pradesh, India
Tel: +917702247710
E-mail: [email protected]

Received Date: April 11, 2017; Accepted Date: April 12, 2017; Published Date: April 20, 2017

Citation: Jena BR, Babu SM, Pradhan DP, Swain S (2017) UPLC Analytical Method Development and Validation for the Simultaneous Estimation of Paracetamol and Caffeine Capsules Dosages Form. Pharm Regul Aff 6:186. doi: 10.4172/2167-7689.1000186

Copyright: © 2017 Jena BR, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Pharmaceutical Regulatory Affairs: Open Access

Abstract

In the present study, a rapid, an accurate and precise Ultra Performance Liquid Chromatography (UPLC) method was developed and validated for simultaneous estimation of paracetamol and caffeine in its capsule dosage form (325 mg and 30 mg) by selecting chromatographic parameters. The UPLC method was developed using 2.1 × 50 mm, reverse phase C18 column (Acquity UPLC ethylene bridge hybrid (BEH) C18 1.7 μm) with mobile phases containing 0.1% w/v H3PO4 and 100% v/v buffer as mobile phase A and methanol: Acetonitrile (50:50) as mobile phase B and water: acetonitrile: H3PO4 (80:20:0.1) as diluent and the run considered as an isocratic elution. Flow rate was 0.5 ml/min with PDA detection at (λmax) 275 nm for paracetamol and caffeine and the injection volume was set at 2 μL with run time 7 min. The method was validated by using various validation parameters like accuracy, precision, linearity and specificity. These results show the method could find practical application as a quality control tool for analysis of the drug in its capsule dosage forms in pharmaceutical industries. The developed validated method and stability testing of new dosage forms as per ICH-Q2 (R1) and ICH-Q1C guidelines applicable for the analysis of bulk drug and in its capsules dosage form.

Keywords

UPLC; Validation; Precision; System suitability

Introduction

Ultra Performance Liquid Chromatography (UPLC) is a relatively new modern technique which gives a new direction for liquid chromatography and it is applicable for particle having less than 2 μm in diameter to acquire better resolution, speed and sensitivity as compared with High-Performance Liquid Chromatography (HPLC). It uses fine particles and saves time and reduces solvent consumption. The UPLC system reduces analysis time up to nine times comparing to the conventional system using 5 μm particle packed analytical columns. In UPLC the separation is performed under tremendous pressures (up to 100 MPa is possible), but it has no negative impact on analytical column as well as other components of chromatographic system. Separation efficiency remains maintained and also it is even improved more [1-4]. Paracetamol or acetaminophen which is chemically N-(4-hydroxyphenyl) acetamide or an "aniline analgesics", non-steroidal anti-inflammatory drug by some sources, and not as an NSAID by others sources while most sources implicitly distinguish them in today’s scenario as shown in Figure 1 [5-10]. Paracetamol has few anti-inflammatory effects in comparison to other NSAIDs. The antipyretic, analgesic and anti-inflammatory effects of paracetamol are due to inhibiting prostaglandin synthesis such as cyclooxygenase-1(COX-1) and cyclooxygenase-2 (COX-2) [11-13]. The major therapeutic uses of paracetamol are headache, migraine, neuralgia, muscular aches, rheumatic pain, backache, joint pain, toothache and reduction of fever in bacterial or viral origin [14,15]. Caffeine (1,3,7-trimethyl xanthine) is the world’s most widely consumed psychoactive drug as depicted in Figure 1 [16]. Apart from that it is also used as diuretic, CNS stimulant and CVS stimulant [17]. The merits of combination of paracetamol along with caffeine as an analgesic and antipyretic are properly established in pharmaceutical formulations [18]. From the various literature findings, it has been seen that paracetamol in combination with caffeine were analytically validated by different spectroscopic and chromatographic techniques [19-23]. The major objectives of this research is to develop a new, convenient UPLC method for determination of paracetamol and caffeine in capsule dosage forms, in accordance with international conference harmonized guidelines [24].

pharmaceutical-regulatory-affairs-structure-paracetamol

Figure 1: Chemical structure of paracetamol (a) and caffeine (b).

Experimental

Materials and reagents

The paracetamol and caffeine capsule samples were contain 99.71% (w/w) and 99.62% (w/w) respectively, on dried basis. The samples of paracetamol and caffeine capsules [325 mg, 30 mg] were obtained from local market for analysis. Methanol, acetonitrile, ortho-phosphoric acid, milli-Q water is used as a solvent of UPLC grade. Acquity UPLC (WATERS) liquid chromatographic system consists of following components: A gradient pump, variable wavelength programmable PDA detector, a flow rate 0.5 ml/min. It’s injection facility having 2 μl. Chromatographic analysis was performed using Acquity UPLC ethylene bridge hybrid (BEH) C18, column of dimensions of particle size 1.7 μm, and standard diameter (2.1 × 50 mm). In addition, UV Spectrophotometer (Shimadzu 1800) an electronic balance (Meter Toledo), pH meter (Metter Toledo), Sonicator (Power sonic), Hot air oven (Serve well), Centrifuge (REMI), Vacuum oven (Centex), Humidity cum photo stability chamber (Thermo lab), and Shaking water bath (JEIO Tech) were used in this study.

Instrumentation and chromatographic condition

The analytical method development was performed with a chromatographic system (WATERS), Acquity UPLC consisting of binary solvent manager along with PDA detector. It is having acquity UPLC ethylene Bridge Hybrid (BEH) C18 column of particle size 1.7 μm and internal diameter 2.1 × 50 mm. The injection facility with 2 μl and flow rate of 0.5 ml/min is maintained. Samples were injected with rheodyne injector system with PDA detection at (λmax) 275 nm. The system data analysis was performed with the Empower 2 software and peak areas were integrated with 7 min run time.

Preparation of solutions

Standard preparation: Weigh accurately about 325 mg of paracetamol to a 250 ml volumetric flask. Add 50 ml of acetonitrile and dissolved it (Solution-A) to get 1300 μg/ml.

Caffeine stock solution: Weigh about 151 mg of caffeine and transfer in to 100 ml volumetric flask. Dissolve it properly and make up to the mark with water. Pipetted out 20 ml of caffeine stock in volumetric flask (solution-A) to get 120 μg/ml and make up the volume with the diluent. For preparation of buffer, dilute 0.5 ml H3PO4 to 500 ml with water, i.e., (0.1% H3PO4) and for mobile phase mix buffer and methanol in the ratio of 550:450 (v/v). The diluent contains acetonitrile and water in the ratio 50:50 (v/v). Filtration is performed by using 0.22 μm membrane filter.

Test preparation

Weigh accurately about 10 capsules and take the average weight, remove the capsule cell and then transfer the capsules content in to the powder form and accurately take weight equivalent to 380 mg of capsule and transfer it in to the 250 ml volumetric flask. After that add 50 ml acetonitrile and dissolved it properly. Then add the diluent and sonicate it for 5 min until it dissolves the material and make up to the volume with the diluent. Filter the above solution through 0.22 μm PVDF Filter and then inject the clear solution in to the UPLC system.

Procedure

Separately inject equal volumes of about 2 μl of diluent as blank, six replicate injections of standard preparation and one injection of two test preparations into the to the UPLC system and record the chromatograms.

System suitability

Tailing factor for the paracetamol and caffeine capsules is not more than 2.0 for the first chromatogram of the standard injection. The relative standard deviation of sum of the peak areas of paracetamol and caffeine from six replicate injections of standard preparation should not be more than 2.0%.

Calculation

The paracetamol and caffeine present in the capsule as % of labelled amount by the formula:

Paracetamol:

image (1)

Where, AT1 is the sum of the paracetamol peak areas obtained from the test preparation of paracetamol and caffeine capsules and AS1 is the sum of the Paracetamol peak areas obtained from the standard preparation of Paracetamol, caffeine. WS1 is the weight of paracetamol working standard taken in mg for standard preparation. WT is the weight of test sample taken in mg for test preparation, AW is the average weight of paracetamol, caffeine capsules, P is the % purity of paracetamol, caffeine capsules working standard calculated as on as such basis and LC is the label claim of paracetamol, caffeine capsules in mg (of respective strength). For second assay test preparation (T2) is calculated by applying same formula, the average assay results as given below (% & mg);

image (2)

image (3)

Caffeine:

image (4)

Where, AT2 is the sum of the caffeine peak areas obtained from the test preparation of paracetamol, caffeine capsules and AS2 is the sum of the caffeine peak areas obtained from the standard preparation of paracetamol, caffeine. WS2 is the weight of caffeine working standard taken in mg for standard preparation. WT is the weight of test sample taken in mg for test preparation. AW is the average weight of paracetamol, caffeine capsules, P is the % purity of paracetamol, caffeine capsules working standard calculated as on as such basis and LC is the label claim of paracetamol, caffeine capsules in mg (of respective strength). For second assay Test preparation (T2) is calculated by applying same formula, the average assay results as given below (% & mg).

image (5)

image (6)

Method Development

Determination of maximum wavelength for paracetamol using PDA detector

The maximum wavelength for the paracetamol and caffeine were observed at 275 nm using PDA detector in UPLC.

At 275 nm paracetamol and caffeine both were showing absorbance.

So for estimation of these combinations we have selected 275 nm for paracetamol and caffeine respectively (Figure 2).

pharmaceutical-regulatory-affairs-PDA-detector

Figure 2: Spectrum for paracetamol (a) and caffeine (b) by PDA detector in UPLC.

Optimization method

Preparation of mobile phase: Mix 0.5 ml of Ortho Phosphoric Acid (OPA) and diluted to 500 ml with water (0.1% H3PO4), (mobile phase A) and it is filtered through 0.22 μ membrane filter. Similarly mix methanol and acetonitrile in the ratio of 50:50 (mobile phase B). Filtration is done through 0.22 μ membrane filter and both the mobile phases were sonicated for 5 min. The column temperature has to be maintained at 50°C with gradient composition of pressure 92:8.

Chromatographic condition: By Acquity UPLC (BEH C18 1.7 μm, 2.1 × 50 mm) with isocratic elution and a flow rate of 0.5 ml/min, the detail analysis was performed. The column temperature was maintained at 40°C. All solvents were filtered through a 0.22 μ membrane filter before use and degassed in an ultrasonic bath. The prepared standard solutions were injected into the column. By using PDA detector quantification was measured at 275 nm. The chromatographic run time was observed in 7 min.

Observation and major findings: Resolution between paracetamol and caffeine was 10.3. In this trail paracetamol is having a good resolution from caffeine peaks. It’s all passes the system suitability parameters caffeine peak shape is also good. This trail is considered as the optimized method trail (Figure 3). The observed data has been shown in Table 1.

pharmaceutical-regulatory-affairs-paracetamol-caffeine

Figure 3: Chromatogram of standard for paracetamol and caffeine.

Chromatogram data for standard paracetamol and caffeine
S. No. Name Rt Area % Area Height USP Tailing USP Plate count USP Resolution
1 Paracetamol 0.68 2323336 74.53 1207177 1.4 2927  
2  Caffeine 1.78 785452 25.2 191242 1.4 4397.6 10.1
Chromatogram data for sample paracetamol and caffeine
1 Paracetamol 0.67 4856755 78.55 1207177 1.1 841.1  
2 Caffeine 1.77 1326508 21.45 201785 1 1940 8.8

Table 1: Chromatogram data for standard, samples of paracetamol and caffeine.

Preparation of sample: Weigh accurately about 10 capsules and take an average weight, than transfer the capsules content in to the powder form and accurately take weight equivalent to 380 mg of capsule and transfer it in to the 250 ml volumetric flask. After that add 50 ml acetonitrile and dissolved it properly. Then add the diluent and sonicate it for 5 min until it dissolves the material and make up to the volume with the diluent. Filter the above solution through 0.22 μm PVDF filter and then inject the clear solution in UPLC (Figure 4).

pharmaceutical-regulatory-affairs-sample-paracetamol

Figure 4: Chromatogram of sample for paracetamol and caffeine.

Observation and major findings: In case of sample also all the individual impurities are resolved from the main peak of paracetamol and caffeine (Figure 4). So there is a good resolution is observed between the main peaks of all the two drugs. No interferences are observed and the results are shown in Table 1. The final developed method is confirmed to be a new analytical method for the simultaneous estimation of paracetamol and caffeine capsules and this method is to be validated.

The samples in all formulations were in good agreement with their respective label claims and they suggested non-interference of formulation excipients in the estimation. Here at the Retention Time (Rt) of paracetamol and caffeine having no interference of the mixture of excipients peak. So no interference of mixture of excipients is observed with the retention time of paracetamol and caffeine (Figure 5).

pharmaceutical-regulatory-affairs-mixture-excipients

Figure 5: Chromatogram of mixture of excipients at 275 nm.

Method Validation

The method validation for the assay of paracetamol, caffeine in paracetamol and caffeine capsules (USP 325 mg, 30 mg) has been performed. This report shows and discusses the results obtained in the validation study for the parameters such as system suitability, precision, specificity, linearity, accuracy, robustness. The method of analysis was validated as per the recommendations of ICH and USP [25,26].

System suitability test

The sample reproducibility was observed of the chromatographic system to measurement of peak area and was carried out using three replicates of standard and sample of having same concentration.

Precision

System precision: As per the test method standard solution was prepared by using paracetamol and caffeine working standard and six replicate injections were injected into the UPLC system. The percentage relative standard deviation (% RSD) was estimated.

Method precision: As per the test procedures, the precision of test method was found out by performing six samples assay, prepared from paracetamol and caffeine capsules 325 mg and 30 mg. Relative standard deviation of assay results was calculated.

Intermediate precision: For the determination of intermediate precision of assay method, the analyst to analyst variation study was conducted. Here the same UPLC systems and same UPLC columns were used. The assay of six samples was calculated as per the test procedures. The percentage relative standard deviation (% RSD) of assay results was found out and results are reported.

Specificity

A study to establish the interference of mixture of excipients was conducted. It was determined with excipients of formulated capsules dosage form. Assay was performed on paracetamol and caffeine capsules (32 5 mg and 30 mg) mixture of excipients in duplicate, equivalent to the weight of mixture of excipient present in portion of test preparation as per the test method and the results are found out. It is concluded that the method was specific.

Linearity

Linearity was conducted by plotting a graph between concentration on X-axis and peak area on Y-axis and the correlation coefficient (R2) value was determined. Six different concentrations of paracetamol and caffeine working standard ranging from 25%, 50%, 75%, 100%, 150% and 200% (325-2600 ppm for paracetamol and 30-240 ppm for caffeine) to were prepared and analyzed as per test method. The calibration curves were obtained by plotting peak areas versus known concentrations in μg/mL.

Accuracy

The accuracy of assay method was found out by quantifying paracetamol and caffeine content at various concentrations ranging from 650 ppm, 1300 ppm, 1950 ppm for paracetamol and 60 ppm, 120 ppm, 180 ppm for caffeine of spike level 50%, 100%, 150% of the final test concentration (i.e., paracetamol and caffeine capsules 325 mg and 30 mg respectively) by mixing of mixture of excipients and drug substance. Assay was performed in triplicate for all levels and the results are reported and percentage of recovery was calculated.

Robustness

Robustness of the method was determined to ensure its capacity to remain unaffected by small deliberate variation in the method parameters such as mobile phase ratio, column oven temperature, flow rate of mobile phase, and pH. Robustness of the method was conducted by making slight changes in the chromatographic conditions, such as change in composition of mobile phase and flow rate.

Results and Discussion

System suitability

Standard solution of paracetamol and caffeine were prepared by using paracetamol, caffeine working standard as per the test method and made six replicate injections into the UPLC system (Table 2). Various parameters such as theoretical plates, tailing factor, retention time and resolution factor, were computed as reported by ICH and USP [25,26]. The relative standard deviation (RSD) for area response obtained from six replicate injections was found to be not more than 2.0. The tailing factor was found to be 1.8, 1.6 which are well within the acceptance criteria of NMT 2.0% respectively.

S. No Name Injection No Rt of paracetamol Rt of caffeine Paracetamol area Caffeine area
1 Standard 1 0.703 1.942 4318243 1399718
2 Standard 2 0.704 1.94 4323621 1402161
3 Standard 3 0.705 1.941 4315551 1399509
4 Standard 4 0.706 1.94 4310359 1398194
5 Standard 5 0.706 1.941 4325271 1402422
6 Standard 6 0.705 1.939 4326283 1403116
Mean 0.705 1.941 4319888 1400853
Standard deviation 0.001 0.001 6262.996 1972.63
%RSD 0.13 0.05 0.11 0.2

Table 2: Standard injections responses for system suitability test.

The number of theoretical plates is a measure of column efficiency which shows the high separation efficiency of the column used and was found to be 1824 and 4575.1 for paracetamol and caffeine capsules respectively which is in the acceptance criteria of NLT 1500. The typical chromatograms of blank, standard solution for system suitability test are shown in Figures 6 and 7.

pharmaceutical-regulatory-affairs-blank-solution

Figure 6: Chromatogram of blank solution (acetonitrile and water in the ratio 50:50 v/v) for system suitability test.

pharmaceutical-regulatory-affairs-system-suitability

Figure 7: Chromatogram of standard solution for system suitability test at (λmax) 275 nm.

It was observed from the results that the system suitability parameters meet the requirement of method validation. The results of all system suitability parameters were shown in Table 3.

System suitability parameters Observed value Acceptance criteria Rt Area %Area
Tailing factor for the paracetamol and caffeine capsules is not more than 2.0 for first injection of the standard solution 1.8, 1.6 Not more than 2.0 0.712 4229520 75.62
The % Relative standard deviation of sum of peak areas of paracetamol and caffeine from six replicate injections should not be more than 2.0 0.1, 0.1 Not more than 2.0 5.768 2880560 100
The USP plate count for paracetamol and caffeine 18244575 Not less than 1500 1.972 1363933 24.38

Table 3: System suitability test parameters for paracetamol and caffeine.

Precision

System precision was carried out by using preparation of standard to ensure that the analytical solution was working properly and giving optimum, precise results. The RSD obtained from six replicate injections were well within the acceptance limit of NMT 2.0%. Method precision was carried out by the same analyst on the same day with same instrument repeatedly for six times to ensure that the method gives consistent results. The RSD was calculated for standard and sample on six determinations and was found to be in the acceptance limit of NMT 2.0%. Intermediate precision was carried out on different days in different instruments to ensure that analytical results remain unaffected with change in instrument and day. The RSD was calculated for standard and sample on six determinations and was found to be less than 2.0%. The results of system, method and intermediate precision date are reported in the Table 4.

Injection No Peak response of system precision Peak response of method precision Peak response of intermediate precision
Analyst 1 Analyst 2
Paracetamol Caffeine Paracetamol Caffeine Paracetamol Caffeine Paracetamol Caffeine
1 4318243 1399718 4197528 1389898 101.7 98.4 99.7 99.2
2 4323621 1402161 4126330 1385496 99.9 98.4 101.3 99.3
3 4315551 1399509 4194117 1392549 101.6 98.8 99.9 98.3
4 4310359 1398194 4192119 1375869 101.5 98.3 99.1 98.8
5 4325271 1402422 4130056 1375269 100 99.1 101.5 100.3
6 4326283 1403116 4118578 1398965 99.8 100 99.4 100.2
Mean 4319888 1400853 4159788 1386341 100.8 98.8 100.2 99.4
Standard deviation 0.001 1972.63 38339.99 9418.311 0.9394 0.647 1.0075 0.7817
% Relative standard deviation 0.13 0.1 0.9 0.67 0.9 0.7 1 0.8

Table 4: Standard injection of paracetamol and caffeine peak response by system, method and intermediate precision test.

Linearity

This was performed to determine whether test results were proportional to the concentration of analyte in samples in a given range. It was determined using standard solutions of paracetamol and caffeine. The response was found to be linear over a concentration range of (25-200%) 325-2600 ppm for paracetamol and 30-240 ppm for caffeine. The correlation coefficient was found to be 0.999942 within the acceptance criteria limit of NLT 0.999. The results are mentioned in Table 5 and Figure 8.

pharmaceutical-regulatory-affairs-Linearity-plot

Figure 8: Linearity plot of paracetamol (a) and caffeine (b).

Level Stock weight taken (mg) Area response
Paracetamol 325 mg Concentration (ppm) Caffeine 30 mg Concentration ( ppm) Paracetamol Caffeine
25% 325 30 1087658 358436
50% 650 60 2154992 690650
75% 975 90 3186906 1081396
100% 1300 120 4318243 1399798
150% 1950 180 6425282 2006893
200% 2600 240 8614658 2718438
Slope Y = 03307 2x-1200.1 Y=1124x+24856
Correlation Coefficient (R2) 0.9999 0.9991

Table 5: Response of paracetamol and caffeine at various linearity levels.

Specificity

Interference from mixture of excipients: Specificity of the method was carried out and the peaks of diluent, excipients of capsules, mobile phase did not interfere with the peaks of two drugs. A study to establish the interference from mixture of excipients was conducted. The specificity of the method was assessed by comparing chromatograms obtained from drug standard with that obtained from capsule solutions. The Rt of drug standard and the drug from sample solution were same, so the method was specific. The method was also specific and selective because there was no interference of impurities with paracetamol and caffeine peaks. Thus, the method indicated a high degree of specificity (Figure 5).

Accuracy

Accuracy of the method was performed to ensure closeness of agreement between true value and reference value in three levels each. The mean % recoveries at concentrations ranging from (spike level) 50%, 100%, 150% were found to be 99.2 to 101.7 which were in the acceptance limit of 98.0 to 102.0% and the RSD was within the limit of NMT 2.0%. The results are shown in Table 6.

Sample No. Spike Level (%) Area “Mg added” paracetamol “Mg found” paracetamol %Recovery paracetamol Mean % Recovery paracetamol Area “Mg added” caffeine “Mg found” caffeine % Recovery caffeine Mean % Recovery caffeine
1 50 2137864 162.7 161.9 99.51 99.4 698893 15.1 14.9 100.08 99.65
2 50 2118864 162.7 161.2 99.3 696984 14.8 14.9 99.8
3 50 2139562 162.8 161.7 99.5 692092 15.3 15.2 99.07
4 100 4279243 325.3 325.8 100.2 99.8 1378789 30.2 30.1 100.52 99.8
5 100 4329481 325.1 324.9 99.4 1369843 30.4 30.2 99.85
6 100 4293364 324.9 323.1 99.8 1366962 29.9 30 99.64
7 150 6468797 487.3 486.5 99.7 99.9 2048762 45.2 45.1 100.17 99.9
8 150 6387653 487.2 485.9 100.1 2027764 45.4 45.2 99.13
9 150 6401765 487.6 488 99.29 2036587 44.8 44.9 99.57

Table 6: Percentage recovery of paracetamol and caffeine at different spike level.

Robustness

The system suitability parameters obtained from normal experimental conditions and that obtained from small and deliberately changed conditions for each of the system suitability parameters like RSD, tailing factor and theoretical plates passes for all the conditions. The results are reported in the Table 7.

Parameter Flow Rate (+10%) Flow Rate
 (-10%)
Column Temp. (+50°C) Column Temp.
(-50°C)
Wave length (+5 nm) Wave length
(-5 nm)
Flow Rate (+10%) Flow Rate
(-10%)
Column Temp. (+50°C) Column Temp.
(-50°C)
Wave length (+5nm) Wave length
(-5nm)
Peak Area (Inj.1) 4135056 4081671 4181519 4138536 4228698 4167561 1361581 1308261 1371621 1368267 1264894 1218894
Peak Area (Inj.2) 4160862 4060719 4172961 4151795 4209276 4149417 1359492 1311772 1374192 1366193 1270454 1213738
Peak Area (Inj.3) 4092617 4032781 4190178 4092578 4212851 4157158 1357235 1318784 1370784 1360173 1266282 1215932
Peak Area (Inj.4) 4151680 4050917 4157287 4117953 4201968 4098171 1369836 1309213 1366245 1358226 1273939 1213498
Peak Area (Inj.5) 4168076 4020761 4196151 4141395 4231317 4128643 1358737 1313287 1367279 1359239 1268384 1208584
Avg. Area 4141658 4049370 4179419 4128451 4216822 4140190 1361376 1312263 1370024 1362419 1268790 1214129
SD 30065.6 23812.9 15375.1 23512.1 12693.5 27482.5 4981.84 4157.27 3252.54 4505 3567.73 3783.32
% RSD 0.725 0.588 0.367 0.569 0.301 0.663 0.365 0.316 0.237 0.33 0.281 0.311

Table 7: Robustness data of paracetamol and caffeine.

Stability of analytical solution

This parameter was determined by injecting sample and standard solutions at regular intervals. The standard and sample solutions were found to be stable up to 48 h. The proposed method was validated and met the requirements as per the ICH guidelines. Hence, the method can be conveniently adopted for the routine analysis in quality control laboratories.

Conclusion

The present investigation represents a rapid, convenient, accurate and precise UPLC method was developed for the simultaneous estimation of paracetamol and caffeine in pharmaceutical formulation. The assay provides a linear response across a wide range of concentrations. This method can be said to be more economical as compared to other methods. The method is suitable for the determination of these drugs in capsules, and hence can be used for routine quality control of paracetamol and caffeine in this dosage form. From the results of linearity it was found that these drugs obey linearity within the level of 25-200 % for paracetamol and caffeine. From the results of precision it was found that % RSD is less than 2%; which indicates that the proposed methods have good reproducibility. From the results of accuracy, it was found that the percentage recovery values of pure drug from the pre-analyzed solutions of formulations were in between 99.65-102%, which indicates that the method was accurate and also reveals that the commonly used excipients and additives present in the formulations were not interfering in the proposed method. The method was validated by using various validation parameters like system suitability, precision, specificity, linearity, accuracy, robustness, stability in analytical solution. All the validation parameters were found to be well within the acceptance criteria. It is shown that the method was accurate, reproducible, repeatable, linear, precise, and selective, proving the reliability of the method. The run time is 7 min which enables proper separation and resolution between the peaks and quality control analysis of capsule formulations. The same solvent used throughout the experimental work and no interference from any excipient was observed. The proposed method is simple, economical, accurate, and precise and could be successfully employed in routine quality control for the simultaneous analysis of paracetamol and caffeine in pharmaceutical formulations.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Recommended Conferences

Article Usage

  • Total views: 216
  • [From(publication date):
    July-2017 - Sep 25, 2017]
  • Breakdown by view type
  • HTML page views : 183
  • PDF downloads :33
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri, Food, Aqua and Veterinary Science Journals

Dr. Krish

[email protected]

1-702-714-7001 Extn: 9040

Clinical and Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemical Engineering and Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001 Extn: 9040

Earth & Environmental Sciences

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

General Science and Health care Journals

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics and Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001 Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Informatics Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Material Sciences Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Mathematics and Physics Journals

Jim Willison

[email protected]

1-702-714-7001 Extn: 9042

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001 Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

John Behannon

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001 Extn: 9042

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords