Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Zetasizer Technique in Biochemistry

Nagham Mahmood Aljamali*

Department of Chemistry, College of Education, Kufa University, Iraq

*Corresponding Author:
Nagham Mahmood Aljamali
Department of Chemistry
College of Education
Kufa University, Iraq
Tel: +96433 340952
E-mail: [email protected]

Received Date: March 20, 2015; Accepted Date: March 28, 2015; Published Date: April 06, 2015

Citation: Aljamali NM (2015) Zetasizer Technique in Biochemistry. Biochem Anal Biochem 4:168. doi:10.4172/2161-1009-1000168

Copyright: © 2015 Aljamali NM. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Biochemistry & Analytical Biochemistry

Abstract

In this review study of zetasizer technique and, using in molecular weight of compounds and large molecules, applications in most of fields like bio molecules, determination of molecular weight of organic chemistry

Keywords

Flexible; Nano; Coefficient; Physical

Introduction

Overview of Zetasizer: The SZ-100 nano partical series instruments are flexible analytical tools for characterizing the physical properties of small particles. Depending on the configuration and application the system can be used as a particle size analyzer, or also used to measure zeta potential, molecular weight, (MW) and second virial coefficient (A2). Typical applications for the SZ-100 include nano particles, colloids, emulsions, and submicron suspensions (Figure 1).

biochemistry-analytical-biochemistry-Zetasizer

Figure 1: Zetasizer instrument.

Particle size analysis is performed by dynamic light scattering (DLS). Depending on the physical properties of the sample, the dynamic range is 0.3 nm-8 μm. The lower limit is influenced by concentration, how strongly the sample scatters light, and the presence of large, unwanted particles. The upper limit is influenced by the density of the sample since DLS is modeled on all motion coming from Brownian motion, not gravitational settling.

The charge on the surface of particles is characterized by the SZ-100 by measuring the zeta potential of a suspension. The sample is injected into a disposable cell and a measurement of the particle electrophoretic mobility results in the calculated zeta potential. The zeta potential of the sample is most often used as an indicator of dispersion stability. Large magnitude zeta potential values indicate that an electrostatically stabilized suspension will remain stable. The zeta potential is often measured as a function of pH or other change in the chemistry to help formulators create new products with a long shelf life. Conversely identifying conditions at which the zeta potential is zero (that is, the sample is at the isoelectric point) allows one to choose optimum conditions for flocculating and separating particles (Figure 2).

biochemistry-analytical-biochemistry-Zetasizer

Figure 2: Zetasizer Nano.

The same instrument can also be used to measure the molecular weight and second virial coefficient of proteins, polymers, and other molecules. The user prepares several solutions with known concentrations and then uses the system in a static light scattering mode to create a Debye plot, which results in a calculation of both MW and A2.

Zetasizer

What does the Zetasizer Nano do?

The Zetasizer Nano range of instruments provides the ability to measure three characteristics of particles or molecules in a liquid medium. These three fundamental parameters are particle size, zeta potential and molecular weight. By using the unique technology within the Zetasizer system, these parameters can be measured over a wide range of concentrations (Table 1).

  Particle size and molecular size Zeta potential (particle size diameter) Molecular weight
Measurement range 0.3nm-10.0 microns (diameter) 3.8nm-100 microns 980Da-20M Da
Measurement principle Dynamic Light Scattering Electrophoretic Light Scattering Static Light ScatteringUsing Debye Plot
Sample volume range 12-20 µL 150µL 12-20 µL
Sample concentration Up to 40%w/v    
Temperature Range 0-120°C

Table 1: Measurements of molecules.

Features

• Particle size, zeta potential, molecular weight, and second virial coefficient all in one instrument

• Wide range of particle sizes and concentrations

• Particles size measurements at both 90° and 173°

• Multiple particle size measurement modes for working with small particles and weak scatterers

• Small volume cells for both particle size and zeta potential

• Zeta potential is a scientific term for electrokineticpotential [1] in colloidaldispersions. In the colloidal chemistry literature, it is usually denoted using the Greek letter zeta (ζ), hence ζ-potential. From a theoretical viewpoint, the zeta potential is the electric potential in the interfacial double layer (DL) at the location of the slipping plane relative to a point in the bulk fluid away from the interface. In other words, zeta potential is the potential difference between the dispersion medium and the stationary layer of fluid attached to the dispersed particle.

• The zeta potential is caused by the net electrical charge contained within the region bounded by the slipping plane, and also depends on the location of that plane. Thus it is widely used for quantification of the magnitude of the charge. However, zeta potential is not equal to the Stern potential or electric surface potential in the double layer [2], because these are defined at different locations. Such assumptions of equality should be applied with caution. Nevertheless, zeta potential is often the only available path for characterization of double-layer properties.

• The zeta potential is a key indicator of the stability of colloidal dispersions. The magnitude of the zeta potential indicates the degree of electrostatic repulsion between adjacent, similarly charged particles in a dispersion. For molecules and particles that are small enough, a high zeta potential will confer stability, i.e., the solution or dispersion will resist aggregation. When the potential is small, attractive forces may exceed this repulsion and the dispersion may break and flocculate. So, colloids with high zeta potential (negative or positive) are electrically stabilized while colloids with low zeta potentials [3].

Methods for Determining the Zeta Potential

Electrokinetic phenomena

There are four electrochemical phenomena that can be used for the determination of zeta-potential, electrophoresis, electroosmosis, streaming and sedimentation potential. Electrophoresis and streaming potential are the most important commercial effects that have been exploited to determine zeta potential. The Malvern Zetasizer uses electrophoresis, however electroosmosis is an interfering effect that must be considered. There are a number of ways of accounting for this effect (Figures 3 and 4). The use of measurements at the stationary layer is one method that is discussed in the next section [4].

biochemistry-analytical-biochemistry-electrophoresis

Figure 3: Basic principles of electrophoresis and electroosmosis.

biochemistry-analytical-biochemistry-constitution

Figure 4: Potential trends at interfaces with differences in the chemical constitution, influence of specific ionic adsorption.

• Positively charged particles on the surface, moderate

Adsorption of anions, moderate ionic strength positively charged particles with no specific adsorption, low concentration of electrolyte

• Negatively charged particles, strong specific cation adsorption, moderate electrolyte concentration [5].

Micro-Electrophoresis

The conventional method for measuring zeta potential is to observe the migration of particles in a dispersion with a microscope and measure their velocity using a stopwatch. The current technique of laser Doppler velocimetry (LDV) has mostly replaced this laborious measurement procedure.

Both techniques are based on measurement of the migration rate of dispersed particles under the influence of an electric field. The observed velocity (v), divided by the strength of the applied electrical field (E), is a direct measure of the electrophoretic mobility (μE) of the particles examined v/E=μE.

The zeta potential can be calculated from the mobility using the Henry equation.

Henry equation:

μE = ε * ξ * f(κ a)/6πη

ε − dielectric constant of the medium, η − viscosity of the medium, f(κa) - correction factor which takes into account the thickness of the double layer and particle diameter. The unit κ is a reciprocal length. 1/κ is frequently described as the thickness of the double layer. In practice an approximation can be made for f(κ a), this is f(κa)=1.0 for non-polar media,f(κ a)=1.5 for particle dispersions in polar media.

f(κa) = 1.5 is a good approximation for particles >100 nm in aqueous solutions with an ionic strength >10-3 M. This is known as Smoluchowski approximation [6].

Smoluchowski approximation

μE = ε * ξ /4 πη

at 25% C it is simplified to ξ= 12.85 μE mV.

The second limiting case f(κ a)=1.0 applies for very small particles in media with a low permittivity. This is known as the Hückel approximation.

• If the conventional microscopic observation technique is compared with modern techniques, the advantages of the LDV methods are many. Some important advantages are:

• Statistically better measurements

• Seconds measurement time rather than 10-30 minutes

• Measurement of smaller particles, 5-10 nm rather than a minimum of 200 nm

• Measurement of zeta potential distributions

• Improvement in measurement repeatability due to a reduction in the Joule heating effect

• Ability to automate complex measurements (e.g. pH or concentration dependences)

The electrophoresis cell

comprises a quartz glass capillary 50 mm long, cross section 5 mm by 2 mm, with electrodes at both ends. The capillary is sealed by valves to form a closed tube.

The stationary layer

In the same way that an electrical double layer is associated with charged particles, the same structure exists at the charged capillary wall. This leads to a liquid flow in the cell due to electroosmosis. This adds a velocity to the particle mobility at all points in the cell except at a special position called the stationary layer or stationary plane. This is created in a closed capillary because of the return flow in the center of the cell. At the stationary layer the fluid flow along the cell wall is exactly cancelled by the return flow. Measurements at this position give the true electrophoretic mobility (Figure 5) [7].

biochemistry-analytical-biochemistry-Electrokinetic

Figure 5: Electrokinetic effects in a closed electrophoresis cell.

The position of the stationary plane is dependent on the cell geometry and can be calculated for symmetrical shapes of capillary.

Zeta potential measurement technique

In contrast to the traditional microscope measuring technique, in the LDV method the particle image is replaced by the illumination of particles by intersecting laser beams, the observer's eye by the photomultiplier, and the stopwatch by the correlator [8].

The Malvern system is called the Zetasizer. This system is available with different combinations of measured parameters (Figure 6). These are zeta potential, size, pH and conductivity. The features of this system are its ease of use provided by pre-aligned optics and single button measurement software [9].

biochemistry-analytical-biochemistry-Electrokinetic

Figure 6: Measurement arrangement for the heterodyne measurement method.

Heterodyne optics

In this optical arrangement the scattered light from particles in the measurement zone is mixed with attenuated light from the illuminating laser that is directly focused on the detector. These two beams intersect at a low angle, 12.7 degrees. This effectively increases the resolution as the effect of Brownian motion is greatly reduced at forward scattering angles. At higher angles, Brownian motion causes a problem as it 'smears' out the velocity spectrum, making it difficult to resolve mixtures of particles with different mobilities [10].

Signal processing and data evaluation

The measurement technique used is often known as Laser Doppler Velocimetry (LDV).

Moving particles in the measurement zone shift the frequency of scattered light proportional to their velocity.

The scattered light is detected by a photon counting photomultiplier and the signal input to a digital correlator. This can be thought of as simply an efficient frequency analyzer using a Fourier transform algorithm to disentangling the mobilities of perhaps thousands or millions of particles into a number of mobility channels (Figure 7).

biochemistry-analytical-biochemistry-correlation

Figure 7: Stages of signal processing from the correlation function to a Fourier Transform analysis (FFT) to frequency, mobility and zeta potential spectrum.

As well as moving due to electrophoresis, the particles are moving due to Brownian motion. This makes the peak width broader than it should appear. To correct for this a measurement is done with no field present. The width of the peak produced will then be due only to the Brownian motion, and this width can be subtracted from the width produced by a measurement with the field applied [11].

Electrode polarization and determination of sign of the zeta potential

An alternating electrical field is used in order to exclude effects due to electrode polarization. The direction of movement of the particles is therefore periodically reversed during measurement.

In addition to this, one of the two laser beams is modulated by means of an oscillating mirror in order to determine the direction of movement and thus the sign of charge of the particle. The modulator adds a known frequency to the signal. The direction of particle movement, and therefore the sign of the charge, can be identified in this way because the frequency detected due to the particles either adds or subtracts from this modulator frequency. Figure 8 shows the time link between the direction of the electrical field and the phase of the modulated beam. With this scheme the maximum amount of data is collected with the minimum electrode polarization.

biochemistry-analytical-biochemistry-modulator

Figure 8: Link between the applied reversing field and the phase of the modulator.

A second advantage of the modulator is that particle mobilities at zero give a measured frequency, which is that of the modulator. This means that low mobility particles can be measured with the same high accuracy as particles with high mobility.

Applications - possibilities for automation

Properties of ultrafine disperse systems (dispersions, suspensions, and emulsions) can be examined or predicted using the zeta potential. To use this parameter, it is important to understand that the zeta potential is not a parameter of a particle alone. The zeta potential is dependent, as already discussed, on the particle and on its environment. Zeta potential data with no an exact description of the conditions under which it was measured are rarely meaningful. Controlled changes of the environment, such as pH, that lead to modification of the zeta potential, can give important information about the nature of the particle surface and the robustness of a dispersion to changes in the parameter [12].

A single measurement of the zeta potential of dispersion in a defined set of conditions is rarely useful for the purpose of characterization. Knowledge is required of how the zeta potential varies as parameters such as pH, conductivity and other factors change

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Article Usage

  • Total views: 11823
  • [From(publication date):
    April-2015 - Nov 23, 2017]
  • Breakdown by view type
  • HTML page views : 8040
  • PDF downloads : 3783
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords