Abstract

New Ion Selective Sensitive Electrode of Pd (II) as Multisensor Based on IRA-410 via Low Cost Oxidation Reduction Process

AT Kassem, N El Said and HF Aly

A novel ion selective IRA-410 membrane disc sensor for Pd (II) ions has been prepared and studied. This electrode has a wide linear dynamic range from 10-1 to 2.5�10-6 mol-1 with a Nernstian slope of 16.5 � 0.2 mV decade-1 and low detection limit of 1.6�10-6 mol l-1. It has a fast response time (