Visual Navigation System used to Coagulate Bleeding in a Semibiological Model

Petra Heinen, Athanasios K Petridis*, Homajoun Maslehaty and Martin Scholz
Department of Neurosurgery, Wedau Kliniken Duisburg, Germany
Corresponding Author : Athanasios K Petridis
Attending Neurosurgeon
Wedau Kliniken Duisburg
Department of Neurosurgery
Zu den Rehwiesen 947055 Duisburg, Germany
Tel: +49-203-7332425
Fax: +49-203-733-0
E-mail: [email protected]
Received August 19, 2013; Accepted October 16, 2013; Published October 19, 2013
Citation: Heinen P, Petridis AK, Maslehaty H, Scholz M (2013) Visual Navigation System used to Coagulate Bleeding in a Semibiological Model. J Neurol Disord 1:136. doi: 10.4172/2329-6895.1000136
Copyright: © 2013 Heinen P, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Related article at
DownloadPubmed DownloadScholar Google
 

Abstract

Objective: Endoscopical procedures in the ventricles for the treatment of hydrocephalus or removal of solid tumors are commonly used in neurosurgery. Even a minimal blood amount leads to a complete loss of visibility (“red out”) endangering the patient. To enable coagulation of a bleeding vessel in “red-out” we developed and tested the VNSoftware- System.

Methods: An optical-positioning-system (OPS) receives signals from 3 infrared Light-Emitting-Diodes (LED) on the endoscope and digitalized endoscopic images are created reproducing the real position of the endoscope. An imaging software correlates the transmitted images (max 8/sec) with the positional data. The virtual images of the VN-Software are available next to the real image of the endoscope-camera. In cases of bleeding a “red-out-sensor” activates the “virtual mode” and the surgeon can navigate the endoscope based on the positional coordinates and the saved images of the software program. The bleeding vessel can be localized and coagulated.

Results: When the target region included a radius of 2 mm around the epicenter the success rate was 96.9%. Only in 3.1% of the experiments the target was missed of more than 2 mm. The mean precision in all experiments was 0.81 mm around the epicenter. The mean time of target finding and coagulation was 48.23 sec.

Conclusion: Localizing a bleeding vessel in a “red-out” condition with the VN-Software is reliable and will improve endoscopical safety in removal of highly vascularised tissue. The VN-software showed a high precision-rate in finding a given target in “red-out” conditions.

Keywords

Share This Page

Additional Info

Loading
Loading Please wait..
 
Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords