alexa
Reach Us +441730450007
Visual Navigation System used to Coagulate Bleeding in a Semibiological Model | OMICS International | Abstract
ISSN: 2329-6895

Journal of Neurological Disorders
Open Access

Our Group organises 3000+ Global Conferenceseries Events every year across USA, Europe & Asia with support from 1000 more scientific Societies and Publishes 700+ Open Access Journals which contains over 50000 eminent personalities, reputed scientists as editorial board members.

Open Access Journals gaining more Readers and Citations
700 Journals and 15,000,000 Readers Each Journal is getting 25,000+ Readers

This Readership is 10 times more when compared to other Subscription Journals (Source: Google Analytics)

Research Article

Visual Navigation System used to Coagulate Bleeding in a Semibiological Model

Petra Heinen, Athanasios K Petridis*, Homajoun Maslehaty and Martin Scholz
Department of Neurosurgery, Wedau Kliniken Duisburg, Germany
Corresponding Author : Athanasios K Petridis
Attending Neurosurgeon
Wedau Kliniken Duisburg
Department of Neurosurgery
Zu den Rehwiesen 947055 Duisburg, Germany
Tel: +49-203-7332425
Fax: +49-203-733-0
E-mail: [email protected]
Received August 19, 2013; Accepted October 16, 2013; Published October 19, 2013
Citation: Heinen P, Petridis AK, Maslehaty H, Scholz M (2013) Visual Navigation System used to Coagulate Bleeding in a Semibiological Model. J Neurol Disord 1:136. doi: 10.4172/2329-6895.1000136
Copyright: © 2013 Heinen P, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Related article at
DownloadPubmed DownloadScholar Google

Abstract

Objective: Endoscopical procedures in the ventricles for the treatment of hydrocephalus or removal of solid tumors are commonly used in neurosurgery. Even a minimal blood amount leads to a complete loss of visibility (“red out”) endangering the patient. To enable coagulation of a bleeding vessel in “red-out” we developed and tested the VNSoftware- System.

Methods: An optical-positioning-system (OPS) receives signals from 3 infrared Light-Emitting-Diodes (LED) on the endoscope and digitalized endoscopic images are created reproducing the real position of the endoscope. An imaging software correlates the transmitted images (max 8/sec) with the positional data. The virtual images of the VN-Software are available next to the real image of the endoscope-camera. In cases of bleeding a “red-out-sensor” activates the “virtual mode” and the surgeon can navigate the endoscope based on the positional coordinates and the saved images of the software program. The bleeding vessel can be localized and coagulated.

Results: When the target region included a radius of 2 mm around the epicenter the success rate was 96.9%. Only in 3.1% of the experiments the target was missed of more than 2 mm. The mean precision in all experiments was 0.81 mm around the epicenter. The mean time of target finding and coagulation was 48.23 sec.

Conclusion: Localizing a bleeding vessel in a “red-out” condition with the VN-Software is reliable and will improve endoscopical safety in removal of highly vascularised tissue. The VN-software showed a high precision-rate in finding a given target in “red-out” conditions.

Keywords

Recommended Conferences
Share This Page
Top