Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Quantifying Reduced Glutathione by Square-wave Voltammetry

Marcos Vergilio Corrêa-da-Silva1,2, Acácio Antonio Pigoso1, Beatriz Felicio Ribeiro1, Laís Oliveira Barbosa1, Claudio Aparecido Rosado Miloch1 and Armindo Antonio Alves1,2,3*

1Herminio Ometto University Center, Araras-SP, Brazil

2Graduate Program in Biomedical Sciences, Herminio Ometto University Center, Brazil

3Department of Biochemistry, Institute of Biology, Campinas State University, Campinas-SP, Brazil

*Corresponding Author:
Armindo Antonio Alves
Graduate Program in Biomedical Sciences
Herminio Ometto University Center
Herminio Ometto-Uniararas, Brazil
Tel: 55-1996270855
Fax: 55-1935431412
E-mail: [email protected]

Received Date: February 19, 2013; Accepted Date: March 27, 2013; Published Date: March 29, 2013

Citation: Corrêa-da-Silva MV, Pigoso AA, Ribeiro BF, Barbosa LO, Rosado Miloch CA, et al. (2013) Quantifying Reduced Glutathione by Square-wave Voltammetry. J Biosens Bioelectron 4:133. doi: 10.4172/2155-6210.1000133

Copyright: © 2013 Corrêa-da-Silva MV, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Biosensors & Bioelectronics

Abstract

This study examines the use of square wave voltammetry (SWV) to quantify reduced glutathione (GSH) dissolved in phosphate buffer (pH 7.5) using a static mercury drop electrode (working), Ag/AgCl electrode (standard) and a platinum wire (auxiliary). The applied voltage ranged from -0.7 to -0.2 V. Increasing concentrations of GSH (13-188μmol/L) correlated with the voltammogram peak area (R2=0.99) and with the current at peak potential (Ip) (R=0.99). The reaction of GSH with diamide was monitored for validation of the method. Addition of increasing concentrations
of diamide (13.3-50.8 μmol/L) to a fixed concentration of GSH (120 μmol/L) decreased the Ip, and the results obtained presented a relative deviation (RD) ≤ 14.5% (compared with expected concentrations by stoichiometry) for GSH concentrations above 33.8 μmol/L, whereas the spectrophotometric method (Elman’s reagent) presented RD ≤ 25.6%. These data indicate that SWV method is more accurate and presented equal precision (SD<8%) as compared to the commonly used spectrophotometric method. This method seems suitable for measuring GSH concentrations at room temperature and pH 7.5 (near biological conditions). Other advantages of this method that make it highly desirable for rapid diagnostic purposes include low cost, simplicity, sensitivity, rapid response and no prior sample preparation.

Keywords

Square-wave voltammetry; Static mercury drop electrode; Reduced glutathione; Oxidative stress

Introduction

The reduced glutathione/glutathione disulfide ratio (GSH/GSSG) is crucially important for maintaining an appropriate redox state necessary for cellular metabolic functions. Many proteins, including enzymes of cell metabolism and regulation [1-6], transcription factors and regulators of cellular cycle [7-9] are potentially influenced by the formation of glutathione adducts [6]. In addition to regulating these processes, GSH participates in the antioxidant system as a preferred target of radical species. It chelates metal ions and is a substrate for the enzyme glutathione peroxidase, which maintains adequate levels of hydroperoxides (H2O2 or ROOH) [10-12]. GSH also acts as a substrate for transferase enzymes, linking these compounds to electrophilic xenobiotics to form S-conjugates that are exported from the cell to the blood or from the body to the environment [12,13].

Even small changes in GSH concentrations are related to oxidative stress, which, depending on its intensity, has been associated with the onset or progression of many diseases such as atherosclerosis, cancer, psoriasis, Alzheimer's, hypertension, heart and liver disease and with important physiological processes such as aging and physical exercise [12,14-21].

Therefore, it is of great importance to develop or improve methods to detect and quantify GSH and other thiols under biological conditions. Several techniques for the quantification of these species are described in the literature. Among these techniques is high performance liquid chromatography coupled to various detection methods [22-32], capillary electrophoresis, mass spectrometry, nuclear magnetic resonance [33] and the spectrophotometric method based on the Elman’s reagent [34-36]. Furthermore, Wang et al. [37] have proposed an electrochemical method (using a gold electrode) to quantify GSH that uses the decrease in the voltammogram of piazselenole induced by the reaction with GSH dissolved in Britton-Robinson buffer at pH (2.0).

Because the redox pair GSH/GSSG can donate electrons in negative potentials (using the Ag/AgCl electrode as a reference), voltammetric techniques, especially those based on mercury electrodes, appear to be able to detect and quantify GSH and other thiols in biological samples [13,38].

The electrode reaction of GSH at the static mercury drop has been studied by means of square-wave voltammetry (SWV) [13]. At potentials more positive than -0.350 V (vs. Ag/AgCl (3 mol/L KCl)), the oxidation of the mercury electrode in the presence of GSH produces a sparingly soluble mercury-GSH complex that deposits onto the electrode surface. Under cathodic potential scan, the deposited complex acts as a reducible reactant resulting in a well-defined cathodic stripping reversible SW response. In this study, quantitative results were not obtained. SWV appears to be an appropriate method for this type of analysis; its main advantages are speed of acquisition and good signal definition [39].

The objective of this study was to quantify GSH by SWV under near-physiological conditions using a static mercury drop electrode as the working electrode.

Materials and Methods

All chemicals were of analytical reagent grade. GSH is a product of Sigma, whereas all of the other chemicals were purchased from Merck. Double distilled water was used in all experiments. A stock solution of GSH was prepared by dissolving GSH in double distilled water. Phosphate buffer (0.10 mol/L), pH 7.5, prepared from Na2HPO4and NaH2PO4, was used as supporting electrolytes. Pure nitrogen was used to purge the electrolyte solutions for 5 min followed by 2 min of rest prior to each measurement. Thereafter, a nitrogen blanket was maintained over the electrolyte solution.

All voltammograms were recorded using an Autolab multimode potentiostat/galvanostat (PGSTAT) 101 (Metrohm, Netherlands) connected to a static mercury drop electrode (SMDE), Model 663 VA (Metrohm, Netherlands). A platinum wire was used as an auxiliary electrode and Ag/AgCl (3 mol/L KCl) (MME-Metrohm, Netherlands) was used as the reference electrode. To minimize noise, the cell was kept inside a grounded Faraday cage. Origin 6.0 was used for statistical analyses and to generate the graphs.

The formation of thionitrobenzoic acid (TNB) by the reaction between GSH and Elman’s reagent [5,5'-dithiobis-(2-nitrobenzoic acid)-DTNB] was monitored at 412 nm [35]. The reaction between GSH and diamide was monitored by the change in absorbance at 298 nm [40]. Measurements were performed 7 min after the addition of reagents.

Results and Discussion

SW voltammetric responses to increasing concentrations of reduced glutathione

A typical SW voltammetric response of GSH (61.5 μmol/L) at a static mercury drop electrode recorded in phosphate buffer at pH 7.5 is shown in figure 1. The net SW component, calculated as the difference between the cathodic and anodic currents, provides information on both the reduction and oxidation half electrode reactions and can be observed as a single bell-shaped curve with a peak potential of Ep=- 0.506 V. This Ep is less than that reported by Mladenov et al. [13], possibly because those authors used slightly different experimental parameters, including different GSH concentrations and phosphate buffer at pH 7.0.

biosensors-bioelectronics-voltammogram

Figure 1: SW voltammogram (δ current) of GSH (62.5 μmol/L) dissolved in 0.1 mol/L phosphate buffer, pH 7.5. The experimental conditions were: SW frequency f=50 Hz; SW amplitude Esw=0.02 V; step potential dE=0.005 V; deposition potential Edep=1.1 V and deposition time tdep=5 s.

Mladenov et al. [13] showed that the electrochemical activity of -SH groups is related more to the complexation reaction between GSH and mercury than its redox transformation; however, because the objective of this study was to quantify GSH levels, only the changes in the analyte concentration (GSH) that are directly correlated with variations in the signal (voltammogram) are relevant.

Figure 2 shows the means of the SW voltammograms obtained with increasing concentrations of GSH (from 12.5 to 187.5 μmol/L) for five independent experiments (n=5).

biosensors-bioelectronics-concentrations

Figure 2: SW voltammograms (δ current) of increasing concentrations of GSH dissolved in 0.1 mmol/L phosphate buffer (pH 7.5). a=12.5; b=37.5; c=62.5; d=125; e=187.5 μmol/L. Each voltammogram represents the mean of five independent experiments. Other experimental conditions were the same as those described in figure. 1.

The concentration of GSH directly correlates with the peak area (R2=0.99) obtained by integrating the curves of the voltammograms and the results are presented in figure 3. The equation that fits the curve is:

(1)

biosensors-bioelectronics-denote

Figure 3: Plot of the value of the Integral of the SW-voltammograms curves (described in figure 2) as a function of the increasing concentrations of GSH (12.5-187.5 μmol/L) dissolved in 0.1 mol/L phosphate buffer (pH 7.5). The regression equation is described in the text. Error bars denote standard deviation from 5 independent experiments.

where:

x=[GSH] (μmol/L)

y=Int.I(E).dE (μA.V)

a=2.7.10-6 (± 5.4.10-7) (μA.V.L2/μmol2)

b=7.2.10-4 (± 1.0.10-4) (μA.V.L/μmol)

c=f(0)=0.02 (± 0.003) (μA.V)

The concentration of GSH, also correlates with the current intensity of the peak (Ip/μA) (R=0.99), and the data was described in figure 4. Ip was defined as the point of maximum current in the voltammograms. The equation that fits the curve is:

(2)

biosensors-bioelectronics-intensity

Figure 4: Plot of the value Ip (peak current intensity) of the SW-voltammograms curves (described in figure 2) as a function of the increasing concentrations of GSH (12.5-187.5 μmol/L) dissolved in 0.1 mol/L phosphate buffer (pH 7.5). The linear regression is described in the text. Error bars denote standard deviation from 5 independent experiments.

where:

x=[GSH] (μmol/L)

y=Ip (μA)

a=slope=0.005 (± 0.0003) (μmol/L/μA)

b= f(0)=0.13 (± 0.02) (μA)

It should be noted that at lowest concentrations of GSH, the standard deviation between experiments is smaller, and therefore, the measurement is more precise in these conditions. The concentrations greater than 200 μmol/L, the mercury drop could not be sustained long enough to perform the measurement. Because the concentrations of GSH in mammalian cells range from 0.5 to 10 mmol/L [12,41,42], such samples must be diluted to precisely quantify GSH by this method. The dilution of biological samples is advantageous because it enables the use of smaller quantities of biological material conveniently dissolved in a suitable buffer.

Reaction between GSH and diamide

To validate the proposed method, the reaction between GSH and diamide, a specific ligand, was studied. Diamide [diazenedicarboxylic acid bis(N,N-dimethylamide)] has been shown to produce a rapid decrease in GSH concentration in rat erythrocytes [40] both in vitro and in vivo [43]; this process can be reversed by treatment with glucose [40] and is reversed more quickly with reduced cysteine [22,43]. In mitochondria from the rat liver, diamide induced a rapid efflux of Ca2+ associated with increased respiratory state 4, collapse of the membrane potential and large mitochondrial swelling; these effects were related to the decrease in GSH and NADPH [44,45]. These studies indicate that diamide could be used to induce controlled decreases of GSH in biological samples in vitro and in vivo. The stoichiometric diamide/ GSH ratio is 0.5 (1 mol of diamide to 2 mol of GSH) [40,46].

GSH can be quantified by the Elman’s reagent (DTNB), which reacts with GSH to produce TNB, which absorbs light at 412 nm [35]. In the presence of excess of DTNB, the stoichiometric ratio between GSH and TNB is 1:1 [34-36]. Figure 5 shows the effect of increasing concentrations of diamide (8.25-77 μmol/L) on the absorbance of TNB (412 nm) produced from the remaining GSH. At the first point, the GSH concentration is 100 μmol/L; upon addition of diamide, the concentration of GSH decreases. The curve shows two distinct behaviors: for the first seven points, the absorbance decreases as a function of GSH concentration; for all other points, the absorbance is constant, tending toward zero. Each point represents the mean of five independent experiments (n=5) and the measurements were performed at least 7 minutes after the addition of GSH.

biosensors-bioelectronics-fixed

Figure 5: Effect of addition of increasing concentrations of diamide (8.25-57 μmol/L) to a fixed GSH concentration (100μmol/L). The TNB, that absorbs 412 nm was produced when the remaining GSH were reduced by DTNB. All reagents were dissolved in 0.1 mol/L phosphate buffer (pH 7.5). Error bars denote standard deviation from 5 independent experiments.

This reaction was also performed using the SWV, the voltammograms are described in figure 6 and the Ip values from these voltammograms are shown in figure 7. In this experiment, increasing concentrations of diamide (13.25-63.2 μmol/L) were added to a fixed concentration of GSH (120 μmol/L). Each point represents the mean of five independent experiments (n=5) and the measurements were performed at least 7 minutes after the addition of GSH as described for figure 5. The curve shows two distinct behaviors for the first seven points, the Ip decreases as a function of the diamide concentration and for the last two points, the Ip increases. Changes in the integral of the voltammograms area indicate a high standard deviation.

biosensors-bioelectronics-dissolved

Figure 6: Effect of increasing concentrations of diamide (13.25-63.2 μmol/L) on the SW voltammograms (δ current) of reduced glutathione (120 μmol/L) dissolved in 0.1 mol/L phosphate buffer (pH 7.5). Voltammograms that decreases as a function of the concentration of diamide are described with solid line and those that increase in dashed line. Each voltammogram represents the mean of five independent experiments. Other experimental conditions were the same as those described in figure 1.

biosensors-bioelectronics-diamide

Figure 7: Effect of increasing concentrations of diamide (13.25-63.2 μmol/L) on the SW voltammogram peak potential current (Ip) of reduced glutathione (120 μmol/L) dissolved in 0.1 mol/L phosphate buffer (pH 7.5). The voltammograms are presented in figure 6. Error bars denote standard deviation from 5 independent experiments.

It should be noted that for the last two points, both in figures 5 and 7, the diamide/GSH ratio exceeded the stoichiometric ratio (0.5), but the behaviors were different. The spectrophotometry results (Figure 5) behaved as predicted by the rate constant of the reaction, which is quite high (300 mol/L in aqueous buffer pH 7.3) [40].

To explain the increase in Ip above the stoichiometric ratio, several control experiments were performed, as shown in figure 8. First, the absorbance of 90 μmol/L of diamide (298 nm) [40] was monitored for 15 minutes (Figure 8, line a), and no change in absorbance was observed during this time. Next, the absorbance of GSH alone was monitored; this absorbance remained near zero (Figure 8, line b). Upon addition of a fixed concentration of GSH (100 μmol/L) and varying concentrations of diamide (Figure 8, lines c, d, e, f), the diamide absorbance remained constant. When the diamide/GSH ratio was lower than 0.5, the diamide was completely consumed, and the absorbance decreased to a level similar to that observed with only GSH (Figure 8, lines c and d compared with line b). However, when the ratio was greater than 0.5, diamide became the excess reagent, and its absorbance could be detected (Figure 8, lines e, f), as predicted by the reaction mechanism. Notably, the absorbance of the diamide did not change during the period monitored, indicating that the reaction occurred before the start of measurements. This experiment demonstrates that the spectrophotometric (Figure 5) and voltammetric measurements (Figure 7), which were performed 7 minutes after initiating the reaction, represent the endpoint of the reaction, which is not reversible, as predicted by the rate constant of this reaction, which is high [40]. These findings rule out the possibility of increased Ip values, for points at which the diamide/GSH ratio is greater than 0.5 (Figure 7) due to the presence of GSH, because this compound would be completely consumed prior to signal acquisition and the reaction is not reversible.

biosensors-bioelectronics-phosphate

Figure 8: Variation of diamide induced absorbance during 15 min reacting with GSH (100 μmol/L) dissolved in 0.1 mol/L phosphate buffer (pH 7.5). (a) Only diamide (90 μmol/L); (b) Only GSH; (c) diamide (6 μmol/L) plus GSH; (d) diamide (30 μmol/L) plus GSH; (e) diamide (60 μmol/L) plus GSH; (f) diamide (90 μmol/L) plus GSH.

Another explanation for the increase in Ip is that diamide could be interacting with the electrodes to generate a voltammogram. Figure 9 depicts the SW voltammogram for 15 μmol.L-1 of diamide in phosphate buffer at pH 7.5. One current peak is observed near -0.1 V. Because diamide also interacts with the mercury electrode, it is possible that this compound may be responsible for the increase in Ip for the points at which the diamide/GSH ratio is greater than 0.5 (Figure 5). No experiments were performed to investigate whether hydrazine (other reaction product) interacts with the electrodes, although this reaction does not appear to occur at the first seven diamide concentrations.

biosensors-bioelectronics-dissolved

Figure 9: SW voltammogram (δ current) of 15 μmol/L diamide solution, dissolved in 0.1 mol/L phosphate buffer (pH 7.5), recorded in phosphate (pH 7.5). Other conditions as described in figure 1.

Precision and accuracy

To demonstrate the precision and accuracy of the proposed method, the GSH concentrations for the first seven points of figure 7 (diamide concentration versus Ip) were compared with the values expected by the reaction stoichiometry. These data are shown in table 1. The values for percent standard deviation (SD%) between the voltammograms are all below 8%, indicating an acceptable precision (low dispersal data). When the GSH concentrations obtained from the voltammograms (Obtained) were compared with those expected by the reaction stoichiometry, the percent relative deviation values (RD%) were equal or less than 14.5% for GSH concentrations greater than or equal to 33.8 μmol/L.

[Diamide] [GSH] µmol/L Precision (SD %)(d) Accuracy (RD %)(e)
µmol/L(a) Obtained (b) Expected (c)
0 114 120 4.50 6.50
13.3 96.2 93.5 6.00 6.70
19.5 78.2 81.0 2.10 4.40
25.8 60.3 68.5 2.10 14.5
38.2 38.1 43.6 4.40 14.5
44.4 33.8 31.2 7.90 13.6
50.8 25.9 18.4 5.20 *47.8

Table 1: Precision and accuracy of the voltammetric method proposed.
To a fixed concentration of GSH (120 umol/L) was added increasing concentrations of diamide(a). The final GSH concentrations obtained(b) presented acceptable precision (SD<7.9%)(d). When compared to the expected by the reaction stoichiometry(c), the accuracy was acceptable below 33.8 μmol/L of GSH(e) (RD% ≤ 14.5). The GSH concentrations(b) were obtained using the Ip of the first seven voltammograms described in figure 6 and the standard curve described in figure 4. *RD above 14.5%.

Similarly, GSH concentrations obtained for the first eight points of figure 7 (diamide concentration versus absorbance at 412 nm) were compared with the values provided by the reaction stoichiometry. The data in table 2 shows acceptable precision at all GSH concentrations used (SD ≤ 12.8). Accuracy was acceptable only at GSH concentrations greater than or equal to 57.5 μmol.L-1 GSH (RD ≤ 5.8%).

[Diamide] [GSH] µmol/L Precision (SD %)(d) Accuracy (RD %)(e)
µmol/L(a) Obtained (b) Expected (c)
0 104 100 3.50 5.80
8.25 86.4 83.5 1.10 4.10
14.5 70.2 71.0 0.80 2.70
20.8 57.5 58.5 0.50 2.10
27.0 36.1 46.0 8.10 *25.6
33.3 22.7 33.5 5.10 *37.3
39.5 4.29 21.0 12.8 *91.9
45.5 1.51 8.50 15.2 *94.9

Table 2: Precision and accuracy of the Elman’s method in this work conditions.
To a fixed concentration of GSH (100 μmol/L) was added increasing concentrations of diamide(a). The final GSH concentrations(b) presented acceptable precision (SD ≤ 12.8%)(d). When compared with to the expected by the reaction stoichiometry (1GSH:2diamide)(c) the accuracy(e) was acceptable below 57.5 μmol/L of GSH (RD% ≤ 5.80%) The GSH concentrations(b) were obtained with the absorbance of the first eight points described in figure 5 and a suitable standard curve. *RD above 5.80%.

The results in tables 1 and 2 indicate that the data obtained for the parameter Ip of the SW voltammograms, under the conditions described herein, are more accurate than and show precision equivalent to those obtained by the spectrophotometric method described by Favier [35]. It can be seen that the lowest concentration of GSH quantified with acceptable accuracy by the Elman’s method was 57.5 μmol/L comparing with 33.8 μmol/L obtained by the voltammetric method (58.8% less). Thus, the voltammetric method showed higher accuracy compared to the spectrophotometric method.

The experiments performed in the present study cannot determine whether reactions between diamide and GSH or between GSH and the electrode occur under ideal conditions, i.e., at 100% yield. However, the data were obtained at room temperature and pH 7.5, which is similar to biological conditions, thus indicating its use in biological samples.

Conclusion

This work demonstrates that it is possible to quantify GSH using both the area as the peak current (Ip) of the SWV voltammograms recorded using the static mercury drop electrode and the Ag/AgCl as standard electrode. These results were validated by monitoring the specific reaction with diamide that presented equal precision and better accuracy when compared to the results obtained by the spectrophotometric method that uses the Elman’s reagent.

Our results indicating the use of this method for use in samples containing cells, extracellular fluids, tissue homogenates or isolated proteins under physiological and pathological conditions, where monitoring alterations in the antioxidant capacity is a relevant factor

The main advantage of using the mercury drop electrode is that a new drop is used for each reading. Therefore, there much less interference from other molecules, particularly proteins that encrust solid electrodes and reduce the signal or even passivate the electrode, thereby decreasing its selectivity and sensitivity [47-49].

The concentration of GSH and other related thiol groups is very sensitive to levels of oxidative stress. Because SWV proved to be a rapid and precise technique, it may be useful for monitoring medical treatments, especially those based on chemotherapy, radiation and oxygen therapies [12,47]. It may also be useful for monitoring the oxidative stress induced by physical training related to both increased performance and injuries [50-52]. All of these situations require rapid intervention to address radical species or to increase the antioxidant capacity of the individual through proper supplementation [12,15,16].

The advantages of this method include the low cost of instrumentation, operational simplicity, speed of response, high sensitivity [37] and simple preparation (only suspension and dilution in buffer); these properties are desirable for rapid diagnostic methods. These properties indicate the importance of future applications of this method using biological samples.

Acknowledgements

This research was supported by the Brazilian Foundation-FAPESP (Fundação de Amparo à Pesquisa do Estado de São Paulo) (Proc. 2010/02791-7). The authors thank Dr. Denise Vaz de Macedo (Labex, IB-UNICAMP, Brazil) and Dr. Raquel Gianolla Miranda (UNIARARAS-Brazil), for help in writing the manuscript.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Article Usage

  • Total views: 11925
  • [From(publication date):
    May-2013 - May 23, 2018]
  • Breakdown by view type
  • HTML page views : 8145
  • PDF downloads : 3780
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2018-19
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2018 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
Leave Your Message 24x7