alexa Anterior segment biometry with 2 imaging technologies: very-high-frequency ultrasound scanning versus optical coherence tomography.
Ophthalmology

Ophthalmology

Journal of Clinical & Experimental Ophthalmology

Author(s): Piero DP, Plaza AB, Ali JL

Abstract Share this page

Abstract PURPOSE: To determine the interchangeability of 2 anterior segment imaging systems: a very-high-frequency (VHF) ultrasound scanning system (Artemis 2, Ultralink LLC) and an optical coherence tomography (OCT) system (Visante, Zeiss). SETTING: Vissum Instituto Oftalmologico de Alicante, Alicante, Spain. METHODS: This study comprised 20 eyes without pathology or previous surgery. The anterior chamber depth (ACD), central corneal thickness (CCT), angle-to-angle distance (ATA), and the iridocorneal angle size (IAS) at the 0-degree and 180-degree positions were measured with 2 imaging techniques: VHF ultrasound scanning and OCT. Analysis of agreement and interchangeability was performed by the Bland and Altman method. In addition, each measurement was performed 3 times consecutively to determine intrasession repeatability by means of the coefficient of variation (CV) and the intraclass correlation coefficient (ICC). RESULTS: No statistically significant differences were found between imaging techniques in ACD, CCT, or ATA (P>.40). The ranges of agreement were 0.20 mm, 16.11 mum, and 0.80 mm for ACD, CCT, and ATA, respectively. Regarding IAS, no statistically significant differences were found in the nasal (P = .78) or temporal (P = .63) measurements between devices. However, the range of agreement for nasal (14.3 degrees) and temporal (14.90 degrees) values was relevant, indicating the 2 techniques cannot be used interchangeably for IAS measurement. Excellent intrasession repeatability scores were obtained (CV and ICC). CONCLUSION: The Artemis 2 and the Visante OCT systems provide equivalent and repeatable measurements of the ACD, CCT, and ATA and can be used interchangeably for these purposes. This article was published in J Cataract Refract Surg and referenced in Journal of Clinical & Experimental Ophthalmology

Relevant Expert PPTs

Relevant Speaker PPTs

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords