alexa Electrode impedance in adults and children using the Nucleus 24 cochlear implant system.
Medicine

Medicine

Otolaryngology: Open Access

Author(s): Busby PA, Plant KL, Whitford LA

Abstract Share this page

Abstract This study measured changes in electrode impedance over time in 19 adults and 29 children implanted with the Nucleus 24 cochlear implant system, using common ground and three monopolar modes of stimulation, over a series of time intervals. Impedances increased from the intraoperative to the initial fitting session, decreased during the initial fitting session and for the following week, and were then stable. Impedances were lowest for the common ground mode, while for the monopolar modes, impedances were related to the surface area of the return electrode. Impedances for children were higher than those for adults at the initial fitting session and for the following three weeks. The clinically measured impedance values increased in a basal-to-apical direction. When these data were corrected for differences in electrode surface area, impedances decreased in a basal-to-apical direction. Impedances were influenced by whether the electrodes were stimulated or unstimulated. Adults with higher electrode impedances tended to have lower hearing thresholds and comfortable listening levels, but this did not appear to clearly be the case for children. This article was published in Cochlear Implants Int and referenced in Otolaryngology: Open Access

Relevant Expert PPTs

Recommended Conferences

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords