alexa Flow-through versus sidestream capnometry for detection of end tidal carbon dioxide in the sedated patient.
Anesthesiology

Anesthesiology

Journal of Anesthesia & Clinical Research

Author(s): Sakata DJ, Matsubara I, Gopalakrishnan NA, Westenskow DR, White JL

Abstract Share this page

BACKGROUND: End tidal carbon dioxide (ETCO(2)) in non-intubated patients can be monitored using either sidestream or flow-through capnometry [Yamamori et al., J Clin Monit Comput 22(3):209-220, 2008]. The hypothesis of this validation study is that, flow-through capnometry will yield a more accurate estimate of ETCO(2) than sidestream capnometry when evaluated in a bench study during low tidal volumes and high oxygen administration via nasal cannula. Secondarily, when ETCO(2) from each is compared to arterial CO(2) (PaCO(2)) during a study in which healthy, non-intubated volunteers are tested under normocapnic, hypocapnic and hypercapnic conditions, the flow-through capnometer will resemble PaCO(2) more closely than the sidestream capnometer. This will be especially true during periods of lower minute ventilation and high oxygen flow rates via mask in non-intubated, remifentanil sedated, healthy volunteers whose physiologic deadspace is small.

METHODS: The performance of a flow-through (cap-ONE, Nihon Kohden, Tokyo, Japan) and a sidestream (Microcap Smart CapnoLine Plus, Oridion Inc., Needham, MA) capnometer were compared in a bench study and a volunteer trial. A bench study evaluated ETCO(2) accuracy using waveforms generated via mechanical lungs during low tidal volumes and high oxygen flow rates. A volunteer study compared the ETCO(2) for each capnometer against PaCO(2) during sedation in which 8 l O(2) was delivered via mask rather than the nasal cannula.

RESULTS: In the bench study, the flow-through capnometer gave slightly higher values of ETCO(2) during high-flow oxygen and no discernable differences during variable tidal volumes. Bland and Altman plots comparing ETCO(2) to PaCO(2) showed essentially equal performance between the two capnometers in the volunteers.

CONCLUSIONS: Within a wide limit of agreement between the volunteer and bench study, flow-through and sidestream capnometry performed equally well during bench testing and in non-intubated, sedated patients.

This article was published in J Clin Monit Comput and referenced in Journal of Anesthesia & Clinical Research

Relevant Expert PPTs

Relevant Speaker PPTs

Recommended Conferences

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords