alexa Investigation of breast dose in five screening mammography centres in Greece.
Oncology

Oncology

OMICS Journal of Radiology

Author(s): Tsapaki V, Tsalafoutas IA, Poga V, Louizi A, Kottou S,

Abstract Share this page

Abstract The objectives of this study were to investigate the techniques currently used for screening mammography in Greece, to estimate the mean glandular dose (MGD) for establishing a baseline radiation dose database, to analyse the effects of various factors on MGD, and to compare the results with others in the literature. Five mammographic facilities and 250 women having as a routine screening mammogram one craniocaudal (CC) and one mediolateral oblique (MLO) projection in each breast were included in the study. The parameters recorded were age, weight, compressed breast thickness (CBT), tube potential (kV), tube loading (mA s) and MLO projection angle. Large differences were observed among the different mammography facilities, mainly in terms of the tube potential setting and the MLO angle used. The average MGD per exposure was 1.4 +/- 0.6 mGy while the respective averages separately for the CC and MLO projections were 1.2 +/- 0.5 and 1.5 +/- 0.7 mGy, respectively. The average MGD values recorded in this study were below the limit of 2 mGy established for the reference medium-sized breast of 4.5 cm CBT. However, the variety of techniques observed revealed the need for a nationwide survey concerning screening mammography in Greece. This article was published in J Radiol Prot and referenced in OMICS Journal of Radiology

Relevant Expert PPTs

Relevant Speaker PPTs

Recommended Conferences

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords