alexa Pelvic tilt makes acetabular cup navigation inaccurate.
Orthopaedics

Orthopaedics

Orthopedic & Muscular System: Current Research

Author(s): Lembeck B, Mueller O, Reize P, Wuelker N

Abstract Share this page

Abstract BACKGROUND: Modern navigation techniques allow precise positioning of the acetabular cup relative to the anterior pelvic plane. Variations in pelvic tilt will affect the resulting spatial orientation of the cup. METHODS: We measured pelvic tilt in 30 volunteers with an inclinometer combined with an ultrasonographic position measurement system. A mathematical algorithm was developed to calculate the resulting cup position measured on standard radiographs, depending on pelvic tilt. RESULTS: Average pelvic tilt at rest was -4 degrees in the lying position and -8 degrees in the standing position, and ranged from -27 degrees to +3 degrees. Pelvic reclination of 1 degree will lead to functional anteversion of the cup of approximately 0.7 degree. INTERPRETATION: Pelvic tilt makes navigation systems referring to the anterior plane inaccurate. This article was published in Acta Orthop and referenced in Orthopedic & Muscular System: Current Research