alexa Physical characterization and comparison of two commercially available micro-MLCs.
Oncology

Oncology

Journal of Nuclear Medicine & Radiation Therapy

Author(s): Podder TK, Bednarz G, Yu Y, Galvin JM

Abstract Share this page

Abstract In this study, the physical characteristics (penumbra width variation with the source size and shape, interleaf leakage, transmission through the leaves, and the tongue-and-groove effect) of two linear accelerators (BrainLAB's Novalis and Elekta's Synergy-S Beam Modulator) have been investigated. For similar square fields (about 4.5 cm x 4.5 cm) with source-to-surface/skin-distance (SSD) ranging from 90 cm to 115 cm and measurements taken at the depth of D(max)=1.5 cm for 6 MV photon beam. The Novalis MLC has penumbra width of 2.4 ± 0.11 mm-2.8 ± 0.11 mm at the leaf-end and 2.2 ± 0.1 mm-2.7 ± 0.1 mm on the leaf-side; and those for the Synergy-S MLC are 4.4 ± 0.17 mm-5.2 ± 0.2 mm and 3.0 ± 0.12 mm-3.5 ± 0.12 mm. Upon rotating the Synergy-S collimator by 90 ° (i.e., shifting the leaf movement to the gun-target direction), significant reduction of the leaf-end penumbra width (17\%) and increase of leaf-side penumbra width (28\%) suggest an elliptical shape of the radiation source spot. Similar rotation of the collimator yielded reduction of the penumbras on both leaf-end (34\%) and leaf-side (28\%) for Novalis, indicating that the Novalis has a more symmetric source size. For all the field sizes and settings, BrainLAB's Novalis μMLC produce a smaller penumbra for simple square fields compared to the Elekta's Synergy-S. However, this difference became less pronounced for leaf-side penumbra and also for circular fields. The tongue-and-groove effect of the Novalis (23 ± 0.9\%) is slightly smaller than that of the Synergy-S (25 ± 1\%); while the interleaf leakage and leakage directly through leaves for Synergy-S (1.6 ± 0.07\% & 0.9 ± 0.04\%) are lower than that of Novalis (2 ± 0.08\% & 1.3 ± 0.05\%). © 2010 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved. This article was published in Phys Med and referenced in Journal of Nuclear Medicine & Radiation Therapy

Relevant Expert PPTs

Relevant Speaker PPTs

Recommended Conferences

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords