alexa Serial angiographic follow-up after Palmaz-Schatz stent implantation: comparison with conventional balloon angioplasty.

Author(s): Kimura T, Nosaka H, Yokoi H, Iwabuchi M, Nobuyoshi M

Abstract Share this page

Abstract OBJECTIVES: Serial angiographic follow-up study was designed to evaluate the temporal mode of lumen diameter changes after Palmaz-Schatz stent implantation, and the results were compared with those from a cohort of patients undergoing balloon angioplasty. BACKGROUND: Restenosis remains a major limitation of balloon angioplasty. The Palmaz-Schatz balloon expandable coronary stent is now under clinical investigation to evaluate its efficacy in preventing restenosis. METHODS: Serial angiographic follow-up study was performed the day after stent implantation and at 1, 3 and 6 months after the procedure. The stent group consisted of 96 patients who had 97 lesions with a single stent. A cohort of 179 patients with 192 lesions were selected as the balloon group by the criteria of final balloon size > or = 3 mm and lesion length < 20 mm. RESULTS: A significantly larger lumen diameter was obtained immediately after stent implantation (2.9 +/- 0.4 mm [mean +/- SD] in the stent group vs. 2.1 +/- 0.5 mm in the balloon group, p < 0.001). At 3 to 6 months of follow-up, a significantly larger lumen diameter was maintained in the stent group (2.2 +/- 0.6 vs. 1.5 +/- 0.7 mm, p < 0.001). The late restenosis rate according to a binary definition was significantly lower in the stent group (13\% vs. 39\%, p < 0.001). Stenosis exacerbation, frequently observed within 24 h after balloon angioplasty, was not found after stenting. Between the next day and 1 month, regression was dominant in the balloon group, whereas progression of stenosis was observed in the stent group. The greatest tendency to restenosis was observed in both groups between 1 and 3 months after the procedure. Between 3 and 6 months, significantly greater diameter loss was found in the stent group. CONCLUSIONS: The Palmaz-Schatz stent was effective in reducing the restenosis rate in this highly selected cohort of patients. Reduction in restenosis rate was dependent on a larger lumen obtained immediately. Late loss of diameter was significantly greater after stenting. The restenosis rate after stenting should be evaluated by follow-up angiography at 6 months rather than at 3 months, which is adequate after conventional balloon angioplasty.
This article was published in J Am Coll Cardiol and referenced in

Relevant Expert PPTs

Relevant Speaker PPTs

Recommended Conferences

Recommended Journals

Relevant Topics

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords