alexa The effect of hematocrit on assay bias when using DBS samples for the quantitative bioanalysis of drugs.
Chemical Engineering

Chemical Engineering

Journal of Analytical & Bioanalytical Techniques

Author(s): Denniff P, Spooner N

Abstract Share this page

Abstract BACKGROUND: As hematocrit levels are known to vary between individuals and with disease state, its effect on the physical characteristics of dried blood spot (DBS) samples and on the accurate quantification of analytes within these samples is examined. RESULTS: The area of DBS samples decreases with increasing hematocrit levels in a linear manner on the three cellulose paper substrates tested. Furthermore, a bias was observed in the concentrations of two analytes determined in DBS samples at different hematocrits, which in some cases exceeded acceptable values, particularly for hematocrits outside normal values. CONCLUSION: If it is expected that the hematocrit of study samples will vary from values considered normal, then its effect on the quantitative determination of an analyte in DBS samples should be investigated as part of the method development and validation. If an unacceptable effect is observed, then this will need to be addressed, by modification of the analytical method, or the inclusion of quality control samples at different hematocrit levels to show control of the assay. This article was published in Bioanalysis and referenced in Journal of Analytical & Bioanalytical Techniques

Relevant Expert PPTs

Relevant Speaker PPTs

Recommended Conferences

Relevant Topics

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords