alexa Ultra high performance liquid chromatography-tandem mass spectrometry method for cyclosporine a quantification in biological samples and lipid nanosystems.
Toxicology

Toxicology

Journal of Environmental & Analytical Toxicology

Author(s): Guada M, Imbuluzqueta E, EstellaHermoso de Mendoz, Lana H, DiosViitez MC, , Guada M, Imbuluzqueta E, EstellaHermoso de Mendoz, Lana H, DiosViitez MC,

Abstract Share this page

Abstract Cyclosporine A (CyA) is an immunosuppressant cyclic undecapeptide used for the prevention of organ transplant rejection and in the treatment of several autoimmune disorders. An ultra high performance liquid chromatography-tandem mass spectrometry method (UHPLC-MS/MS) to quantify CyA in lipid nanosystems and mouse biological matrices (whole blood, kidneys, lungs, spleen, liver, heart, brain, stomach and intestine) was developed and fully validated. Chromatographic separation was performed on an Acquity UPLC(®) BEH C18 column with a gradient elution consisting of methanol and 2mM ammonium acetate aqueous solution containing 0.1\% formic acid at a flow rate of 0.6mL/min. Amiodarone was used as internal standard (IS). Retention times of IS and CyA were 0.69min and 1.09min, respectively. Mass spectrometer operated in electrospray ionization positive mode (ESI+) and multiple reaction monitoring (MRM) transitions were detected, m/z 1220.69→1203.7 for CyA and m/z 646→58 for IS. The extraction method from biological samples consisted of a simple protein precipitation with 10\% trichloroacetic acid aqueous solution and acetonitrile and 5μL of supernatant were directly injected into the UHPLC-MS/MS system. Linearity was observed between 0.001μg/mL-2.5μg/mL (r≥0.99) in all matrices. The precision expressed in coefficient of variation (CV) was below 11.44\% and accuracy in bias ranged from -12.78\% to 7.99\% including methanol and biological matrices. Recovery in all cases was above 70.54\% and some matrix effect was observed. CyA was found to be stable in post-extraction whole blood and liver homogenate samples exposed for 6h at room temperature and 72h at 4°C. The present method was successfully applied for quality control of lipid nanocarriers as well as in vivo studies in BALB/c mice. Copyright © 2013 Elsevier B.V. All rights reserved. This article was published in J Chromatogr B Analyt Technol Biomed Life Sci and referenced in Journal of Environmental & Analytical Toxicology

Relevant Expert PPTs

Relevant Speaker PPTs

Recommended Conferences

Relevant Topics

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords