Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Setting up a Human Motion Analysis Laboratory: Camera Positioning for Kinematic Recording of Gait

Marcelo Riberto1*, Rogério Ferreira Liporaci2,3, Fernando Vieira2 and José Batista Volpon1,3

1Faculty Member at School of Medicine in Ribeirão Preto, University of São Paulo Brazil

2Motion Analysis Laboratory – Center of Rehabilitation – Clinical Hospital of Ribeirão Preto - School of Medicine in Ribeirão Preto, University of São Paulo USP, Brazil

3Bioengineering laboratory - School of Medicine in Ribeirão Preto, University of São Paulo, Brazil

*Corresponding Author:
Marcelo Riberto
Faculty Member at School of Medicine in Ribeirão Preto
University of São Paulo, Brazil
Tel: +55-16-36022416
E-mail: [email protected]

Received Date: April 15, 2013; Accepted Date: May 18, 2013; Published Date: May 20, 2013

Citation: Riberto M, Liporaci RF, Vieira F, Volpon JB (2013) Setting up a Human Motion Analysis Laboratory: Camera Positioning for Kinematic Recording of Gait. Int J Phys Med Rehabil 1:131. doi: 10.4172/2329-9096.1000131

Copyright: © 2013 Riberto M, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at International Journal of Physical Medicine & Rehabilitation

Abstract

This article compares the techniques used for camera positioning in a study room, proposing optimized data collection for three-dimensional analysis. This study analyzed the usable recording volume of three different video camera layouts: six equidistant cameras (C6), 8 equidistant cameras (C8E) and 8 cameras using enhanced positioning techniques (C8EP). In the last configuration, we attempted to place the cameras in certain locations in the room so that they had the same capture volume. It was observed that the volume of the 8 equidistant camera layout remained practically the same as that using the 6 camera layout (C6=10.579 m³ C8E x=11.565 m³). Regarding the enhanced positioning layout (C8EP), a 37% gain in usable walkway length was obtained, along with a 14% gain in width, 34% in height, and a consequent increase of 110% in total volume (22.247 m³). Although most modern kinematic analysis software have tools that are able to fill in the markings trajectory, the enhanced positioning camera layout provides more accurate information and data interpretation, besides ensuring better usage of rooms with restricted dimensions. The considerable gain in vertical dimension allows for the evaluation of performed activities at a higher level, such as those involving the upper limbs or steps.

Keywords

Motion analysis; Three-dimensional analysis; Kinematics motion; Setting up a motion laboratory; Gait analysis

Introduction

The Kinematic analysis of gait using three-dimensional analysis laboratories utilizes a system of cameras in standard fixed positions that capture the movement of reflective markings attached to the skin of the individual, forwarding it to a computer in order to create three dimensional movement simulations, models and visualization of real movements in a virtual environment. At least three markings, along with the definition of joint centers and Euller angles in a calibrated system, are necessary to describe the movements of each body segment in one possible solution to this issue.

The arrangement of the cameras along the walkway should be able to capture the movement of the markings in the sagittal, frontal and transverse planes; the space interval in all this three dimensions defines the recording volume where the gait must occur. There should be a sufficient number of properly positioned cameras so that all markings may be seen at all times by at least two of the cameras. When this condition is not met even on short sections of the walkway, more modern movement analysis resources should be used that have an interpolation tool, which fills in the markings trajectory in those sections by way of simulation and animation based on mathematical algorithms [1].

The location where the cameras will be positioned and affixed depends on certain factors: the size of the analysis room where the study will take place, the type of study, and the number of available cameras as well as camera characteristics (lens, sensibility), sizes of the markings and also experimental factors, like luminosity. Depending on the room size, the cameras should be arranged as far as possible from the walkway, so as to increase the range of view that will capture the movement of the markings, given the markers’ size is large enough to be visualized [2,3]. The evaluation of human gait requires a walkway with a minimum of 10-12-steps (corresponding to at least eight meters), while the evaluation of movements like jumping and throwing tasks do not require as much space [4]. Although improvement of gait capture only requires adequate length and width, such motion analysis system may be used for vertical movements too, thus an improvement in height is also desirable.

One problem faced in this study occurred during the setting up of the motion analysis laboratory. The building where it was to be located was not originally designated for this purpose, thus, the area available to organize the laboratory measured 8m long by 7.2m wide, and the walkway for recording movement measured 4.50 m × 1.50 m. All evaluations had to be performed within these pre-defined dimensions.

Objective

The aim of this study was to compare a camera arrangement alternative to the usual position of 6 or 8 cameras in regular intervals for a motion analysis laboratory with the goal of increasing kinematic data collection.

Method

With the room set and the total number of cameras already acquired, all focus was put on the best location to affix them. Some important points must be taken into account to facilitate this choice, regardless of the type of study.

Using six cameras:

• The typical arrangement for 6 cameras requires that the walkway used for data collection (where the subject is required to walk) be located in the center of a rectangular room, equidistant from the walls in the longitudinal and transverse axes of the ground plane. The logical reasoning is that the cameras should occupy the midpoints of the walls at the beginning and end of the walkway, while the sidewalls are divided into 3 equal parts by the other two cameras (Figure 1a).

Using 8 cameras with 2 alternative camera arrangements:

• In the first alternative arrangement, the walls at the ends of the walkway have a centralized camera. The same positioning is used on the sidewalls. The four remaining cameras are positioned in the corners of the room (Figure 1b). With this arrangement, each camera covers only a portion of the walkway, not the full length of it, which is more evident with cameras C3 and C7.

• In the second option of alternative arrangements, four cameras remain positioned in the corners of the room and in the center of the walls near the end of the walkway, and the remaining two are placed on the sidewalls, with a distance of about 2 meters from the opposite vertex (Figure 1c). The choice for this form of placement is based on the idea that all cameras should display the greatest possible extent of the walkway. To achieve this, markings are put on the ground to allow for the adjustment of each camera (Figure 2).

In order to quantitatively compare the proposed camera arrangements, the dimension “volume” was used, which is enough space to collect images and enough space for a study that allows the simultaneous viewing of all cameras. The calculation of the volume was obtained from the maximum length, width, and height obtained by each arrangement.

Results

Table 1 show the linear measurements from the reference point of the walkway where measurements were initially taken. The change in positioning and number of cameras led to a 37% increase in the usable length of the walkway (from 3.46 m, figure 1a, to 4.77 m, figure 2a). There was a 14% gain with respect to the width of the walkway, and a 34% gain in vertical dimension. The capture volume increased 110%. Note that changing from six to eight cameras placed in an equidistant manner, left the useful area for capturing the markers virtually the same (Figures 1a, 2a and 2b).

  Longitudinal* Transversal* Height* Volume**
6 equidistant cameras 3.465 1.966 1.553 10.579
8 equidistant cameras 4.382 1.793 1.472 11.565
8 cameras with optimal positioning 4.775 2.241 2.079 22.247

Table 1: Linear measurements in three directions and capture volumes for each camera arrangement.

physical-medicine-study-room

Figure 1: 1a –Positioning of 6 cameras equidistant to each other. Note the length and width of the recording area. 1b – Length and width of the recording area with 8 cameras in the same positions as in 1a. 1c – Positioning of the cameras with the goal of achieving optimal visualization of the entire central area of the study room. Both useful length and width increased considerably.

physical-medicine-cameras-room

Figure 2: Markers placed along the walkway for kinematic recording. Each camera was positioned so as to offer optimal visualization of the markings on the floor, shaping the data collection walkway. 2a: Positioning of the markings on the floor. The dotted lines in the center of the walkway connect the vertex markings of the force platform. 2b: 3D imaging of the illuminated dots captured by the set of cameras.

Discussion

Since the dimensions of the study room were already fixed and there were eight available cameras, the challenge faced by this study was to optimize the capture environment to be able to view a 10.59m2 walkway. The rearrangement of the cameras attempted to allow all cameras to visualize the same volume, which resulted in a considerable increase in the length, width and height of the useable recording area. Increased videotaping accuracy is obtained with a larger number of cameras focusing on the same point or with the object being near to the camera [2].

Although the positioning of cameras was based on markers on the floor, this strategy proved useful to optimize assessment of human gait once this study’s usable volume was increased due to enhancement of linear measurements in all three dimensions: laterolateral, anteroposterior and in the height in the alternative 8-camera arrangement. While gait study only requires an increase in the length and width of the recording area, the increase in vertical dimension capture allowed us to use the laboratory for other studies such as the study of upper limb movements and the use of steps [1].

Bontrager [4,5] suggests another way to ensure the best location for positioning cameras in the study room is through the use of a laboratory floor plan to scale, by cutting isosceles triangles with an acute angle similar to that of the camera viewing angle, and positioning them on the laboratory drawing to determine the position with better coverage of the motion analysis area. The locations of the cameras are marked on the scale drawing to obtain the real position coordinates of the camera in the laboratory. Given the availability of electronic resources that allow for the construction of models of overlapping cones observed by each camera and the instant calculation of overlap, it seemed pointless to use a method that required manual calculation. This method proposed by Bontrager disregards that the increased distance between the object and the camera implies a reduction in recording accuracy, which can be compensated electronically by changing the collection parameters (exposure time and camera sensitivity). It could be argued that semicircular arrangement of 4 cameras at both sides of the walk way might produce a larger capture volume [3,6]. We did not consider this possibility because we did not use tripods to support the cameras, so they would have to be affixed to the ceiling, thus more subject to vibration and imprecise recordings.

In short, arranging the infrared cameras so that all have the best range of view of the walkway infers an increase in linear capture dimensions of length, width and height, resulting in increased usable volume, which enables the laboratory to be used not only for gait studies, but also for other analyses of motion. The disadvantage of this arrangement of cameras we describe here is the higher cost than the 6 camera setting, but this cannot be said in relation to the use of 8 cameras in regular intervals.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Article Usage

  • Total views: 12022
  • [From(publication date):
    July-2013 - Dec 14, 2017]
  • Breakdown by view type
  • HTML page views : 8216
  • PDF downloads : 3806
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri & Aquaculture Journals

Dr. Krish

[email protected]

1-702-714-7001Extn: 9040

Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001Extn: 9040

Clinical Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

Food & Nutrition Journals

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

General Science

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics & Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Materials Science Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Nursing & Health Care Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

Ann Jose

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001Extn: 9042

 
© 2008- 2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version