Make the best use of Scientific Research and information from our 700+ peer reviewed, Open Access Journals that operates with the help of 50,000+ Editorial Board Members and esteemed reviewers and 1000+ Scientific associations in Medical, Clinical, Pharmaceutical, Engineering, Technology and Management Fields.
Meet Inspiring Speakers and Experts at our 3000+ Global Conferenceseries Events with over 600+ Conferences, 1200+ Symposiums and 1200+ Workshops on
Medical, Pharma, Engineering, Science, Technology and Business

Validation of SOS-lux Microbial Biosensors for Mutagenicity Assessment: Mitomycin-C as a Model Compound

Hani A Alhadrami1,2* and Graeme I Paton1,3

1Faculty of Applied Medical Sciences, King Abdulaziz University, P.O. Box 80324, Jeddah-21589, Kingdom of Saudi Arabia

2Institute of Biological and Environmental Sciences, Cruickshank Building, University of Aberdeen, Aberdeen AB24 3UU, United Kingdom

3Remedios Limited, Balgownie Technology Centre, Balgownie Drive, Aberdeen, AB22 8GW, United Kingdom

*Corresponding Author:
Hani A Alhadrami
Institute of Biological and Environmental Sciences
Cruickshank Building
University of Aberdeen
Aberdeen AB24 3UU, United Kingdom
Tel: +966505545275
E-mail: [email protected]

Received Date: August 13, 2013; Accepted Date: September 28, 2013; Published Date: October 06, 2013

Citation: Alhadrami HA, Paton GI (2013) Validation of SOS-lux Microbial Biosensors for Mutagenicity Assessment: Mitomycin-C as a Model Compound. J Biosens Bioelectron 4:142. doi: 10.4172/2155-6210.1000142

Copyright: © 2013 Alhadrami HA, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Biosensors & Bioelectronics

Keywords

SOS-lux biosensors; Salmonella assay; in vitro digestion; Mutagenicity; Bioluminescent bacteria; Mitomycin C

Introduction

Soil ingestion (both directly and through vegetable consumption) is a significant pathway by which pollutants enter the human body [1]. Permissible concentrations of ingested soil pollutants have been derived from animal studies (direct feeding or injection), and translated to humans. Such extrapolations often fail to consider the bioaccessible and bioavailable fractions of pollutants and their relative toxicity [2]. These extractions quantify the bioaccessible fraction of soil pollutants by simulating the oral exposure pathway for children (the most sensitive human receptors). The term, bioaccessible, in the context of human ingestion, is defined as the fraction of a substance that is soluble in the gastrointestinal environment and available to human biochemical processing [3].

Environmental mutagens are external agents that when activated, increase the rate of mutation in cells. Chemical analysis alone is unable to predict the biological impacts, antagonistic or synergistic effects in a mixture of mutagenic pollutants, nor to address the significance of bioavailability in terms of pollutant assimilation [4]. These shortcomings have required the development and application of rapid assays to screen large numbers of samples from environmental matrices.

Luminescence-based bacterial biosensors have been developed to detect a variety of mutagens [5]. Biosensors for mutagenicity assessments have been used due to their simplicity and sensitivity [6]. The SOS-lux based microbial biosensors have a promoter less luxoperon (luxCDABE) under control of the SOS-dependant col promoter, and thus, its synthesis is regulated by the SOS-system [7]. An exposure to a target analyte leads to an increase in the concentration of luciferase and bioluminescence. Subsequently, bioluminescence expression is proportional to the mutagenicity of the agent.

The Salmonella mutagenicity assay (Ames assay) is the most widely accepted bacterial assay for the screening and identification of mutagenic compounds [8]. Independent studies have shown a correlation between mutagenicity in the Salmonella assay and carcinogenicity in mammals [9]. The assay uses a number of Salmonella strains with pre-existing mutations that disable the cells from synthesising histidine, thus inhibiting growth. Fresh mutations at the site of these pre-existing mutations can restore the gene’s function and allow the cells to resynthesis histidine. Enumerations of these mutated colonies in the absence of histidine enable an assessment of mutagenicity [10]. The assay is reliable but laborious and requires working with Salmonella strains, which are classified as human pathogen.

In this study, MMC was extracted using a human digestion simulation procedure (the in vitro bioassay), and the mutagenicity of the extracted samples was measured using the Salmonella assay and SOS-lux based microbial biosensors. This enabled an overall evaluation of the optimised assays, before working with a complex environment like soil. MMC was then amended into soil which extracted using the in vitro bioassay, followed by comparative mutagenic assays. Furthermore, the effect of the simulated gastrointestinal constituents on the mutagenicity of MMC was assessed. The aim of this work was to validate the performance of microbial biosensors for mutagenicity assessment and consider the relevance of the in vitro bioassay in assessing pollutant exposure.

Materials and Methods

Stock solutions and soil preparation

Mutagenicity testing using the Salmonella assay and the SOSlux biosensors was performed against a range of doses of mitomycin C (MMC). MMC was purchased from Sigma (St.Louis, MO, USA), and dissolved in reverse osmosis water. Soil (sandy loam) of Insch Association/Insch Series (North-East Scotland) was collected to a depth of 50 cm, passed through a 2 mm stainless steel sieve, thoroughly mixed and amended with MMC to give final concentrations of 1, 10 and 50 μg MMC/g dry weight (dw) soil using standard procedures [11]. Soil amended with MMC was extracted by the in vitro digestion extraction and then assayed. The dose selection of MMC for soil amendment was based upon the dose response curves of MMC, which tested with the Salmonella assay and the SOS-lux biosensors. Un-amended soil (MMC free soil) was used as a negative control.

The in vitro bioassay

The sequential and the compartmental in vitro digestion extractions were carried out for soils amended with MMC. The sequential in vitro digestion assay was performed as described by Oomen et al. [1]. The compartmental approach was performed by extracting soil samples separately in each digestive compartment (i.e. saliva, gastric and duodenal). Synthetic gastrointestinal juices were simulated as described by Oomen et al. [1]. In summary, the saliva extraction was performed by adding 9 ml of synthetic saliva (pH 6.5 ± 0.2) to 0.6 g soil. The mixture was rotated for 5 min using an end-over-end shaker. The samples were centrifuged for 10 min at 1730 g and 37°C ± 2°C. The "digested soil" (the pellet) was discarded and the supernatant (the chyme) removed and stored at 4°C. Gastric extraction was performed by adding 13.5 ml of the synthetic gastric juice (pH 1.1 ± 0.1) to 0.6 g soil, and the mixture was mixed by end-over-end shaking for 2 h, centrifuged and stored at 4°C. The intestinal extraction was carried out by adding 27 ml of the synthetic duodenal juice (pH 7.8 ± 0.2) and 9 ml of the synthetic bile (pH 8.0 ± 0.2) to 0.6 g soil. The mixture was rotated for 2 h and centrifuged as above. The digestive juices were maintained at 37 ± 2°C in a water bath, prior to assay.

The Ames assay

The standard plate incorporation procedure described by Maron and Ames [12] was used for the Ames assay. In brief, Salmonella strains TA98, TA100 and TA102 were grown overnight in 150 ml Erlenmeyer flask containing 25 ml Oxoid nutrient broth at 37°C in an orbital shaker at 150 rpm with appropriate antibiotics (25 μg ml-1 ampicillin for TA98 and TA100, and 2 μg ml-1 tetracycline for TA102). The cultures were incubated until they reached an absorbance of 1.0 at 660 nm (corresponding to 1-2×109 CFU ml-1). Two ml of melted top agar supplemented with histidine and biotin solution was distributed into sterile glass tubes, and placed in a 45°C water bath. A hundred μl of MMC extracted by the in vitro bioassay and 100 μl of the tester strain was added, gently mixed by vortexing, and poured onto the surface of Minimal Glucose Agar plate. The plates were gently tilted and rotated to obtain an even distribution, placed onto a level surface to solidify and incubated at 37°C for 48 h. Following the incubation, the revertant colonies were enumerated on a Gallenkamp colony counter. Appropriate reagent and negative controls [12] were included to enumerate the spontaneous revertants.

The assay was conducted using triplicate of each sample and control. Salmonella typhimurium TA98, TA100 and TA102 were obtained from Molecular Toxicology Inc. (MD, USA). The strains were maintained and stored according to standard protocols [10].

SOS-lux biosensors

Biosensor strains E. coli K12C600 and E. coli DPD1718 were obtained from Rettberg et al. [7] and Vankemmelbeke et al. [13], respectively. Overnight cultures were grown on LB media at 37°C in an orbital shaker at 150 rpm in the presence of the appropriate antibiotics (50 μg ml-1 ampicillin for E. coli K12C600 and 30 μgml-1 chloramphenicol for E. coli DPD1718). Overnight cultures were diluted 1:50 in LB broth, and grown at 37°C, until they reached the appropriate pre-optimised optical density (0.3 at OD550 for E. coli K12C600 and 0.4 at OD492 for E. coli DPD1718). A negative control of 100 μl MilliQ water or 100 μl sample was mixed with 900 μl of overnight culture in 3 ml luminometer cuvettes. Bioluminescence was measured using a Jade bench-top luminometer (Labtech International, Uckfield, UK), over a period of 300 min, with readings taken every 30min.

Analytical techniques for MMC

MMC was analytically measured after the in vitro digestion extractions carried out for soils amended with MMC. The method adopted was modified from Metha et al. [14]. Soil samples (2 g ± 0.01) were grounded with 2 g of anhydrous sodium sulphide. Samples from the in vitro digestion assay were transferred to 30 ml Wheat on vials and 15 ml of methanol was added. The samples were sonicated, and for 6 h placed on an over and under shaker. Samples were centrifuged at 4°C for 10 min in glass centrifuge tubes. The sample was filtered through alumina and transferred to 1.5 ml HPLC sealed glass vials for analysis. Analysis was performed at 365 nm by HPLC (Thermoquest; Thermo Separation Products, San Jose, CA) on an octadecylsilicate column (15 cm 4.6 mm, i.e. particle diameter 5 ml), at a flow rate of 1.0 ml min-1.

The mobile phase was an acetonitrile: water mix with an elution gradientof 30:70 initially for 10 min; then increased to 60:40, over 10 min, and finally increased to 90:10 over 5 min.

Statistical analysis

Each sample was tested in triplicate. Statistical analysis was performed using Minitab 15 for Windows. A result with p ≤ 0.05 was considered significant. The two fold increase rule was applied for the Salmonella assay to evaluate the mutagenicity of the tested compounds [12]. For the biosensors, a compound was considered a mutagen if there was at least a two fold increase in the bioluminescence response relative to the negative control value [15,16]. If the bioluminescence values decreased during the incubation time, the sample was more likely to be cytotoxic [7]. If bioluminescence was not induced and the cell growth was comparable to that of the untreated control, the test sample was assumed to be neither mutagenic nor cytotoxic.

Results

Mutagenicity response of the Salmonella assay to MMC tested in the aqueous phase

Salmonella strain TA102 was the most sensitive strain to detect the mutagenicity of MMC (Table 1). Salmonella strains TA98 and TA100 were, by comparison, insensitive to MMC, as the numbers of histidine revertants were not twice the numbers of the spontaneous revertants (Table 1). A significant mutagenic response was detected for MMC tested with TA102, and extracted in the saliva and the duodenal compartments. There was no significant mutagenic response detected for MMC extracted in the gastric and in the sequential compartments (Table 1). To further investigate the reason for that, the pH of the two compartments were buffered, and the mutagenicity was re-evaluated with TA102 (Table 2). The pH values of samples extracted in the gastric juice were increased from 1.07 ± 0.10 to 7.4 ± 0.10. Similarly, the pH values were increased from 6.8 ± 0.20 to 7.5 ± 0.10 for samples extracted by the sequential in vitro bioassay. There was no significant mutagenic response for MMC extracted in the gastric, and in the sequential compartments for the buffered (pH values 7.4 ± 0.10 and 7.5 ± 0.10) and un-buffered (pH values 1.07 ± 0.10 and 6.8 ± 0.20) samples (Table 2).

Sample Dose (µg/Plate)B Number of revertants/plate (mean ± SE)D
    TA98 TA100 TA102
Saliva 0C 38 ± 4 E 151 ± 9 E 449 ± 29 E
  0.5 4 ± 1 0 ± 0 1836 ± 57 F
  1 1 ± 0 13 ± 1 2098 ± 155 F
Positive control A   1675 ± 0 F 484 ± 29 F 2151 ± 95 F
Gastric 0C 38 ± 4 E 151 ± 9 E 449 ± 29 E
  0.5 42 ± 3 61 ± 9 502 ± 21
  1 44 ± 5 84 ± 4 468 ± 15
Positive control A   1675 ± 0 F 484 ± 29 F 2151 ± 95 F
Duodenal + Bile 0C 38 ± 4 E 151 ± 9 E 449 ± 29 E
  0.5 9 ± 2 13 ± 2 1800 ± 74 F
  1 0 ± 0 54 ± 5 1699 ± 116 F
Positive control A   1675 ± 0 F 484 ± 29 F 2151 ± 95 F
Sequential 0C 38 ± 4 E 151 ± 9 E 449 ± 29 E
  0.5 37 ± 1 166 ± 5 524 ± 31
  1 38 ± 3 170 ± 10 473 ± 19
Positive control A   1675 ± 0 F 484 ± 29 F 2151 ± 95 F

Table 1: The number of reverse mutants of S. typhimurium TA98, TA100 and TA102 after exposure to MMC Samples were analysed in the aqueous phase after extraction by the in vitro procedure.

Sample Dose (µg/Plate) Number of revertants/plate (mean ± SE)C
Gastric (pH 1.07) 0B 467 ± 12 D
0.5 505 ± 20
1 424 ± 8
Positive control A   1223 ± 117E
Gastric (pH 7.4) 0B 467 ± 12 D
0.5 510 ± 11
1 493 ± 27
Positive control A   1223 ± 117 E
Sequential (pH 5.8) 0B 467 ± 12 D
0.5 507 ± 8
1 475 ± 27
Positive control A   1223 ± 117 E
Sequential (pH 7.5) 0B 467 ± 12 D
0.5 511 ± 12
1 480 ± 5
Positive control A   1223 ± 117 E

Table 2: The number of reverse mutants of S. typhimurium TA102 after exposure to MMC Samples were analysed in the aqueous phase after extraction by the in vitro procedure (buffered and un-buffered samples).

Mutagenicity response of SOS-lux biosensors toMMC tested in the aqueous phase

The biosensor (E. coli K12C600) dose response for the negative control (MilliQ water) was significantly different to the samples treated with MMC and extracted in the saliva and the duodenal compartments (Figure 1). The biosensor E. coli K12C600 was not significantly induced by samples treated with MMC and extracted by the gastric and the sequential in vitro bioassay (Figure 1). These results suggested that the biosensor E. coli K12C600 was induced by MMC extracted in the saliva and in the duodenal compartments, whereas MMC extracted by the gastric and the sequential in vitro bioassays was no longer a mutagen, as it was not able to induce E. coli K12C600 (Figure 1). The same results were observed for E. coli DPD1718 (Figure 2) (saliva and in the duodenal compartments, but not the gastric and in the sequential compartments). These findings were compatible with the results reported with the biosensor E. coli K12C600, suggesting that MMC was not a mutagen, once it passed through the gastric compartment of the simulated human gastrointestinal juices (Figures 1 and 2).

biosensors-bioelectronics-Mutagenicity

Figure 1: Mutagenicity response of the SOS-lux microbial biosensor Escherichia coli K12C600 to MMC in the aqueous phase and extracted by the in vitro extraction. The biosensor was only induced by samples treated with 0.5 and 1μg/ml MMC, and extracted in the saliva and the duodenal compartments.

biosensors-bioelectronics-microbial

Figure 2: Mutagenicity response of the SOS-lux microbial biosensor Escherichia coli DPD1718 to MMC tested in the aqueous phase and extracted by the in vitro extraction. The biosensor was only induced by samples treated with 0.5 and 1 μg/ml MMC, and extracted in the saliva and the duodenal compartments.

Mutagenicity response of the Salmonella assay to MMC amended soil

The Salmonella mutagenicity assay responded significantly to Insch soils amended with 10 and 50 μg MMC/g dw soil (Table 3). For Insch soil extracted with the gastric and the sequential in vitro extractions (Table 3), no significant mutagenic response was observed. The Salmonella results confirmed that MMC had a mutagenic effect at a concentration of 10 μg MMC/g dw soil and above, when extracted with synthetic saliva and duodenal juices. At the concentrations tested, MMC had no mutagenic impact on the Salmonella strain TA102 in the sequential and the gastric in vitro extractions due to the denaturing of MMC at the extraction pH value (Table 3).

Gastrointestinal Compartments Dose µg/g dry weight soil Number of revertants/plate (mean ± SE) D
Saliva 0B 314 ± 2E
0C 450 ± 39
1 477 ± 4
10 646 ± 11 F
50 950 ± 0 F
Positive ControlA   1073 ± 5 F
Gastric 0B 314 ± 2 E
0C 390 ± 7
1 394 ± 4
10 417 ± 2
50 381 ± 9
Positive ControlA   1073 ± 5 F
Duodenal + Bile 0B 314 ± 2 E
0C 356 ± 5
1 374 ± 60
10 687 ± 101 F
50 956 ± 0 F
Positive ControlA   1073 ± 5 F
Sequential 0B 314 ± 2 F
0C 393 ± 6
1 419 ± 4
10 428 ± 4
50 440 ± 15
Positive ControlA   1073 ± 5 F

Table 3: The number of reverse mutants of S. typhimurium TA102 after exposure to Insch soils amended with MMC and extracted by the sequential and compartmental (saliva, gastric and duodenal) in vitro extraction.

Mutagenicity response of SOS-lux biosensors to MMC amended soil: For soil amended with MMC and extracted in the oral compartment, a significant mutagenic response was measured at concentrations of 1, 10 and 50 μg MMC/g dw soil for E. coli DPD1718 (Figure 3) and E. coli K12C600 (Figure 4). Significant mutagenic responses to 10 and 50 μg MMC/g dw soil were reported for soil extracted in the intestinal compartment and assayed with E. coli DPD1718 (Figure 3) and E. coli K12C600 (Figure 4). The biosensors E. coli DPD1718 and E. coli K12C600 were not significantly induced by MMC in the gastric and the sequential in vitro extractions from Insch soils because MMC denatured at these pH values (Figures 3 and 4). MMC was more likely to be destroyed because of the acidic pH of the gastric compartment rather than the gastric enzymes. If the gastric enzymes should have an effect on MMC, this effect should be obvious with the saliva and the duodenal compartments, as both of them have enzymes such as amylase, lipase and pancreatin.

biosensors-bioelectronics-amended

Figure 3: Mutagenicity response of the SOS-lux microbial biosensor Escherichia coli DPD1718 to Insch soils amended with MMC and extracted by the sequential and the compartmental (saliva, gastric and duodenal) in vitro extraction. Error bars represent the standard errors of the mean of triplicate measurements.

biosensors-bioelectronics-Escherichia

Figure 4: Mutagenicity response of the SOS-lux microbial biosensor Escherichia coli K12C600 to Insch soils amended with MMC and extracted by the sequential, and the compartmental (saliva, gastric and duodenal) in vitro extraction. Error bars represent the standard errors of the mean of triplicate measurements.

HPLC results for soils amended with MMC: The pH values of the simulated gastrointestinal compartments had a significant influence (p<0.05) on the measured concentrations of MMC. The concentrations of MMC were 0.69, 8.50, and 49.01 μg MMC/g dw soil for soil amended with 1, 10 and 50 μg MMC/g dw soil, respectively, and extracted in the saliva. For the duodenal compartment, the measured concentrations of MMC were 0.90, 6.32 and 48.32 μg MMC/g dw soil. These results confirmed that the pH values of the saliva and the duodenal had no significant effect on the measured concentrations of MMC. The minute reduction in the measured concentrations of MMC was due to the low pH value (4.2 ± 0.0) of Insch soil [11]. For soil extracted in the gastric compartment, the measured concentrations of MMC were 0.35, 0.21, and 0.37 μg MMC/g dw soil for soils amended with 1, 10 and 50 μg MMC/g dw soil, respectively. Similar results were reported for soil extracted in the sequential compartment (the measured concentrations of MMC were 0.50, 0.34 and 0.18 μg MMC/g dw soil for soils amended with 1, 10 and 50 μg MMC/g dw soil, respectively). The acidic pH of the gastric compartment (1.1 ± 0.1), and Insch soil (4.2 ± 0.0) destroyed MMC and caused an enormous reduction in the measured concentrations of MMC.

Discussion

MMC was selected for this study because it is a highly water soluble chemical acknowledged to induce a response both to the Ames assay and the biosensor [17]. MMC is a natural occurring compound consisting of a pyrrolo (1,2-a) indole ring system with an aziridine ring.MMC is a potent DNA cross-linker, which has a strong ability to crosslink DNA with high efficiency and specificity for the sequence CpG [18]. To interact with DNA, MMC requires enzymatic activation by a one-electron pathway to a semiquinone, or by a two-electron reduction pathway to a hydroquinone [19]. MMC has several biological effects in mammalian cells, such as mutagenesis, stimulation of genetic recombination, selective inhibition of DNA synthesis, chromosome breakage and sister chromatid exchange and induction of the DNA repair system (SOS-response) [18]. The anti-cancer activity of MMC is based upon its covalent binding to DNA after chemical or enzymatic reductive activation [20].

Salmonella strain TA102 was the most sensitive among the Salmonella strains to detect the mutagenicity of MMC. Growth inhibition was observed with the Salmonella strains TA98 and TA100 when exposed to MMC (Table 1). The same results were also reported by Maron and Ames [12], who observed inhibition of growth with TA97, TA98 and TA100 when exposed to 2.5 μg/plate of MMC. Consequently, MMC was adopted as a positive control to assess the number of histidine revertants colonies for TA102 [10,12,21]. There is no doubt that the biosensor yielded rapid results when compared with the Ames assay, and this is widely acknowledged [22,23]. An acknowledged limitation of biosensors is that future developments will require the characterisation and adoption of strong promoters offering genuine relevance with key target receptors [24].

The selection of the particular assay procedures reported was because these were acknowledged to be responsive to MMC, but translation to other chemicals of concern (particularly hydrophobic compounds such as PAHs) may require a new suite of mutagenic responsive sensors and assays. In this study, the biosensor performance and sensitivity was similar to the Ames assay. This has been used to reinforce the merit of rapid screening assays to derive values for human protection. But there is a need to take a wider view. The main challenge is that the data that underpin the human protective strategy must be validated against relevant receptors. Comparing biosensor or Ames derived data with those from animal experiments where injection of the chemical into the animal has occurred may be of limited value. So, before the adoption of protective values or the use of routine screening approaches, users must be confident about the relevance of the adopted data.

The impact of soil borne mutagenic compounds on human receptors is of considerable interest when deriving threshold values that will prove protective [3]. There would be little relevance in simply using a solvent extraction procedure to collect the analyte and then measure the response, because this would not reflect the human bioaccessible fraction [1]. For MMC, there were problems with the stage of the assay procedure that mimicked the conditions of the stomach. The acidic nature of this step oxidised the test compound, and this is likely to be a problem with other chemicals of concern. A mutagenic response to MMC was recorded only for soil samples extracted in the oral and the intestinal compartments (pH 6.5 ± 0.2 and 7.8 ± 0.2, respectively) (Table 3 and Figures 3 and 4). There was no mutagenic response to MMC by the Salmonella assay or the SOS-lux biosensors for the gastric or the sequential in vitro digestion assay, because the compound had been destroyed under the acidic condition (pH 1.1 ± 0.1) of the gastric compartment. Even after a buffering step to mediate the pH value, there was no significant mutagenic response, which was confirmed by chemical analysis (Table 2). This concentration should not pose an unacceptable risk to a child ingesting up to 0.6 g of soil daily for six years contaminated with MMC.

The particular relevance and indeed the focus of this work was that it combined a human bioaccessibility assay with mutagenic testing. Such in vitro assays were developed for inorganic chemicals of concern, but there has been wider translation to other pollutants and mixed contaminants. This bioaccessibility in vitro procedure verifies a relevant environmental interface because this considers the potential pathways through which humans could become exposed to such mutagenic compounds. The question is not about the mutagenicity of soil, but is about the mutagenic nature of compounds that are bioaccessible in the soil. A constraint however, was that a specific stage of the in vitro assay oxidised the analyte. Further refinement of this procedure to adopt less harsh surrogate steps for the stomach and duodenal stage could mitigate this, but detailed correlation would be required against established data.

A key advantage of the application of biosensors is their ability to be deployed to genuine samples [24]. This means that while chromatographic analysis requires extraction, purification, clean-up, potentially derivatisation and then analysis, the biosensor can be used in crude and turbid solutions. This could be exclusively for a mutagenic sensor or for a wider suite of sensors for specific analytes and generic toxicity. Samples could be exposed to a suite of sensors simultaneously, allowing a real time response that could prove valuable for the protection of target receptors at a timescale commensurate with intervention [22]. The sensors could also be fabricated to permit compact analysis and the use of disposable electrodes. Such technologies could transform routine testing and interpretation.

Conclusion

This study enabled an empirical assessment of a model mutagen borne in soil and its potency on human receptors.

SOS-lux biosensors had numerous practical advantages over the traditional assays, including procedural simplicity, a rapid and unambiguous result, ease of measurement, rapid tabulation of exposure concentration and in vivo analysis without cell disruption. A critical insight for the bacterial SOS system and the regulatory pathway, which is responsible for inducible DNA repair system, is essential for the genetic engineering of these biosensors. This engineering must be placed in parallel with environmental applicability that enables the collection and analysis of relevant environmental samples and complementary chemical characterisation. As mutagenic-based biosensors evolve, the key traits of this study can be translated to ensure relevance and applicability.

Acknowledgements

The authors would like to thanks Dr. Petra Rettbergand Dr. Ying Zhang for providing the biosensors strains.

References

Select your language of interest to view the total content in your interested language
Post your comment

Share This Article

Relevant Topics

Recommended Conferences

Article Usage

  • Total views: 11560
  • [From(publication date):
    August-2013 - Sep 24, 2017]
  • Breakdown by view type
  • HTML page views : 7794
  • PDF downloads :3766
 

Post your comment

captcha   Reload  Can't read the image? click here to refresh

Peer Reviewed Journals
 
Make the best use of Scientific Research and information from our 700 + peer reviewed, Open Access Journals
International Conferences 2017-18
 
Meet Inspiring Speakers and Experts at our 3000+ Global Annual Meetings

Contact Us

Agri, Food, Aqua and Veterinary Science Journals

Dr. Krish

[email protected]

1-702-714-7001 Extn: 9040

Clinical and Biochemistry Journals

Datta A

[email protected]

1-702-714-7001Extn: 9037

Business & Management Journals

Ronald

[email protected]

1-702-714-7001Extn: 9042

Chemical Engineering and Chemistry Journals

Gabriel Shaw

[email protected]

1-702-714-7001 Extn: 9040

Earth & Environmental Sciences

Katie Wilson

[email protected]

1-702-714-7001Extn: 9042

Engineering Journals

James Franklin

[email protected]

1-702-714-7001Extn: 9042

General Science and Health care Journals

Andrea Jason

[email protected]

1-702-714-7001Extn: 9043

Genetics and Molecular Biology Journals

Anna Melissa

[email protected]

1-702-714-7001 Extn: 9006

Immunology & Microbiology Journals

David Gorantl

[email protected]

1-702-714-7001Extn: 9014

Informatics Journals

Stephanie Skinner

[email protected]

1-702-714-7001Extn: 9039

Material Sciences Journals

Rachle Green

[email protected]

1-702-714-7001Extn: 9039

Mathematics and Physics Journals

Jim Willison

[email protected]

1-702-714-7001 Extn: 9042

Medical Journals

Nimmi Anna

[email protected]

1-702-714-7001 Extn: 9038

Neuroscience & Psychology Journals

Nathan T

[email protected]

1-702-714-7001Extn: 9041

Pharmaceutical Sciences Journals

John Behannon

[email protected]

1-702-714-7001Extn: 9007

Social & Political Science Journals

Steve Harry

[email protected]

1-702-714-7001 Extn: 9042

 
© 2008-2017 OMICS International - Open Access Publisher. Best viewed in Mozilla Firefox | Google Chrome | Above IE 7.0 version
adwords